
Deployment (for carbon dioxide removal)
The number of cartridges required is determined primarily by the number of people present but also by the volume
of the enclosed sealed space.The number of people determines the required CO2removal capacity and the volume of
the enclosed space determines the minimum required removal rate.The total CO2removal rate needs to be greater
than the production rate or the CO2level will continually increase.The deployed CO2capacity will determine the time
interval that the cartridge(s) can run before they need to be changed for fresh cartridges. The CO2production rate will
change with respiration rate or may be set by the relevant legislation.
Table 1 shows typical data for a range of operational conditions for people relaxed and exhaling an average of 0.24 l/
minute (14.4 l/hr) CO2.Table 2 shows the same data for a higher CO2exhalation rate of 0.4 l/minute (24 l/hr) CO2.
The main body of the table (shaded in blue) shows the normal operating range.The red shaded areas show marginal
or non-recommended conditions.
Select the volume of the enclosure in the left column then select the column with the maximum number of people in the
top row.The intersection gives the minimum number of cartridges that should be used to provide an acceptable removal
rate.The value to the right of this gives the expected time the cartridge(s) will run between changes.More cartridges than
this minimum will maintain a lower CO2concentration for a longer time interval.The time interval between changes will
depend on the number of people present and their activity level.The best way of operating the cartridges to ensure
maximum usable capacity is to monitor the CO2level in the enclosed space and change the cartridges when the CO2
concentration reaches a predetermined level – typically 1 or 2 % CO2.
The supplementary information is provided to allow other conditions to be calculated if required.
Table 1 – typical data for a range of normal operating conditions – people at rest (1).
(Assumes: 0.24 l/min CO2per person; 1,500 l/CO2per cartridge; 400 l/min air flow rate per cartridge
and 1% max CO2inlet concentration)
H i C a p
T M
C O 2
Instructions for use:
Use in accordance with refuge
manufacturer’s instructions.
Disposal and the Environment:
Please dispose of responsibly complying
with local legislation.
Contains Soda Lime
Ma n u f a c t u r e r :
C a r b o n D i o x i d e
A b s o r b e r C a r t r i d g e
Volume No. of run time No. of run time No. of run time No. of run time No. cartridges
of people (hr) people (hr) people (hr) people (hr) for 1air
enclosure 1to 10 11 to 20 21to 30 31to 40 change per hour
min No. min No. min No. min No.
m
3 cartridges cartridges cartridges cartridges
10 110.42 15.2113.47 12.60 0.42
20 110.42 15.2113.47 12.60 0.83
30 2 20.83 2 10.42 2 6.94 2 5.211.25
40 2 20.83 2 10.42 2 6.94 2 5.211.67
50 3 31.25 3 15.63 3 10.42 3 7.812.08
60 3 31.25 3 15.63 3 10.42 3 7.812.50
70 4 41.67 4 20.83 4 13.89 4 10.42 2.92
80 4 41.67 4 20.83 4 13.89 4 10.42 3.33
90 4 41.67 4 20.83 4 13.89 4 10.42 3.75
100 5 52.08 5 26.04 5 17.36 5 13.02 4.17
max no. 10 20 30 40
of people
max 144 288 432 576
CO2l/hr
No. 0.096 0.192 0.288 0.384
cartridges
per hr
1
cartridge
10.42 5.213.47 2.60
lasts (hr)
Normal operating range
Marginal or non
recommended conditions
ref (1) Criteria for the
design of refuge stations
Drake, Fellow & Bates
Version 4, 29/10/10 MCL, JS
Page 5 of 9