MONI LOG EnDaL curve User manual

EnDaL curve/vi
Release 09/2017
Manual
for
EnDaL curve
and
EnDaL vi


EnDaL curve/vi ENVIRONMENTAL AND SAFETY INSTRUCTIONS
- 1 -
T A B L E O F C O N T E N T S
1. ENVIRONMENTAL AND SAFETY INSTRUCTIONS.........................................3
2. INTRODUCTION ......................................................................................4
3. TECHNICAL PARAMETERS.........................................................................5
4. DEVICE DESCRIPTION...............................................................................7
4.1 DEVICE SETUP............................................................................................................7
4.2 CONNECTING EXTERNAL SENSORS AND SWITCHES............................................................8
4.2.1 External Acceleration Sensors And Switches..................................................................... 8
4.2.2 Digital / Analogue Input ............................................................................................... 8
4.3 OPERATION ..............................................................................................................9
4.3.1 Switching On And Off .................................................................................................. 9
4.3.2 Status indication........................................................................................................... 9
4.4 CHANGING THE BATTERIES.........................................................................................10
4.5 MOUNTING INFORMATION ........................................................................................11
4.6 MEASUREMENT DATA ACQUISITION .............................................................................12
4.6.1 Measuring Shock Events .............................................................................................. 12
4.6.2 Storing Digital Events .................................................................................................. 12
4.6.3 Synchronous measurements ......................................................................................... 13
4.6.4 Measuring The Analogue Signal................................................................................... 13
5. PC EVALUATION SOFTWARE ENCUR .......................................................14
5.1 SYSTEM REQUIREMENTS .............................................................................................14
5.2 INSTALLATION..........................................................................................................14
5.3 OVERVIEW..............................................................................................................15
5.4 CONFIGURATIONS (OPTIONS).....................................................................................18
5.4.1 Basic Settings............................................................................................................. 18
5.4.2 Classification ............................................................................................................. 20
5.4.3 Limit Value................................................................................................................. 20
5.4.4 Spectrum................................................................................................................... 20
5.5 MANAGE MEASURING AND PROGRAM FILES ...................................................................21
5.6 DEVICE CONFIGURATION ...........................................................................................22
5.6.1 Measuring Program .................................................................................................... 22

ENVIRONMENTAL AND SAFETY INSTRUCTIONS EnDaL curve/vi
- 2 -
5.6.2 Sensor Configuration.................................................................................................. 24
5.6.3 Autoprogramming...................................................................................................... 25
5.6.4 Password .................................................................................................................. 25
5.7 EVALUATION OF MEASURING DATA .............................................................................. 26
5.7.1 Device Status and Acquisition Periods ........................................................................... 26
5.7.2 Overall Graphic (total)................................................................................................ 27
5.7.3 Acceleration Data ...................................................................................................... 28
5.7.3.1 A-table.............................................................................................................. 28
5.7.3.2 Curve Analysis................................................................................................... 29
5.7.3.3 Distribution........................................................................................................ 31
5.7.3.4 Classification..................................................................................................... 32
5.7.3.5 Limit Table......................................................................................................... 32
5.7.4 Continuously measured data (Synchronous data)............................................................ 32
5.7.5 Digital Events ............................................................................................................ 32
5.7.6 GPS-Data.................................................................................................................. 33
5.8 HELP.................................................................................................................... 34
6. SERVICE AND CALIBRATION................................................................... 34
7. WARRANTY CERTIFICATE ....................................................................... 35

EnDaL curve/vi ENVIRONMENTAL AND SAFETY INSTRUCTIONS
- 3 -
1. E N V I R O N M E N T A L A N D S A F E T Y I N S TR U C T I O N S
Data Transfer Cable
To ensure full functionality of the system, please use only the supplied cables and accessories for connection
to a personal computer.
Environment & disposal
The EnDaL curve/vi contains no corrosive or environmentally damaging substances. This
does not apply to batteries.
Used batteries must be removed immediately from the device. Please dispose of used batteries in an
ecologically sound manner and according to the local regulations. Lithium batteries are hazardous waste.
Battery operation
The EnDaL curve is typically supplied with two exchangeable alkaline or lithium batteries
(type C|R14). They can be bought in specialized stores according to the specification stated in the
instruction manual. The usage of secondary battery cells (NiMH) is also possible. Furthermore there is a
special device configuration for external power supply.
Please label the EnDaL curve as hazardous material of class 9, according to the Worldwide
UN classification, when using it with lithium batteries.
ATTENTION!
Use only intrinsically safe batteries.
In case of non-compliance with the national regulations/laws applicable in the handling of lithium
batteries, any liability and warranty claim becomes void!
Decommissioning
In case of decommissioning,the devices must be properly switched off. All batteries must be
removed from the device and be disposed of in an ecologically sound manner.

INTRODUCTION EnDaL curve/vi
- 4 -
2. I N T R OD U C T I O N
The EnDaL curve/vi is a universal data logger. It is designed for long-term measurements,so
it is suitable to monitor transport cargo in critical environments.
The EnDaL curve/vi is available in many different configurations. Therefore,it can be
modified to meet a lot of special requirements. With internal or external acceleration sensors as well as the
sensor input for temperature, relative humidity and air pressure, the EnDaL curve/vi is
suitable for many applications. The configuration with external sensors makes it possible to measure with
optimal coupling on positions which are difficult to access. The configurable analogue input allows you to
use analogue sensors from other manufacturers. A device version suitable for use in a clean room is also
available.
Additionally you can connect the EnDaL curve/vi with an external tracking system like the
data link TM. Thereby, it is possible to track the GPS positions and send the device status as
well as measurement data via email.
Because of its versatility of power supply and its degree of protection IP65,the
EnDaL curve/vi is suitable as an independent long-term transport data logging system. By the use of an
external power supply,it is also suitable for stationary applications.
The license-free software ENCUR for configuration and evaluation enables graphical und tabular analyses
of the measured data. The measuring data can be archived on the computer or exported for use with other
software programs.

EnDaL curve/vi TECHNICAL PARAMETERS
- 5 -
3. T E C H N I C A L P A R A M E TE R S
Case and operational conditions
Case material
Aluminium, powder-coated
Degree of protection
IP65
IP40 for version with external NiMH secondary cells and internal 9 V
alkaline primary cell
Weight
800 g (with batteries excluding additional equipment)
Dimensions (H/W/D)
180 mm / 106 mm / 37 mm
Mounting type
Screw mounting, individual mounting plates are possible
Operating and storage
conditions
-20 °C to +70 °C with alkaline primary batteries
-40 °C to +85 °C with lithium primary batteries
-20 °C to +60 °C with NiMH secondary batteries
Data security
Data retention
minimum 10 years (independent of battery condition)
System time
internal clock supported independently from the power supply by an
internal battery, exchange of this battery at SMT ELEKTRONIK after 6
years recommended,
information of the time zone can be stored in the device
Password
3 levels of protection: no protection, protection of configuration, protection
of readout and configuration
Power supply
internal
2x batteries of type C/R14 changeable:
Primary cells: alkaline, lithium
Secondary cells: nickel metal hydride(NiMH)
Alternative 1x 9 V alkaline block battery
external
Alkaline battery pack
NiMH secondary
Mains supply
External interfaces
RS232
RS232 to USB adapter optionally available
Display and control elements
LED
2x red (error, alarm)

TECHNICAL PARAMETERS EnDaL curve/vi
- 6 -
3x green (power, active, memory)
Key button
1x “control“ to display the operating state via LEDs
Shackle
magnetic on/off-switch
Bypass switch
toggle switch to bypass the safety bar (permanently switched on)
Sound module
Only for EnDaL vi
Device sensors
Measured variable
Measuring range
Tolerance
Data storage
Acceleration
internal
2, 5, 10, 20, 50,
100, 200 g, discrete
configurable
±2 % of measuring
range
±5 % of measuring
value
16384
thereof 10 or 20 with
waveform (depending
on curve duration)
Acceleration
external
100, 200, 500,
1000, 2000 pC
Acceleration
external 3D ICP®
50, 100, 250, 500,
1000 mV
Temperature
-40…+120 °C
(depends on the
model)
±1 K
16384
Relative Humidity
0…100 %rh
±2 %rh at
20…80 %rh
Analogue Input
0…2.5V
±0.1%
Digital Input
4 channels with
respectively high/low
Low: 0…0.8V
High: 2.4…12 V
4096

EnDaL curve/vi DEVICE DESCRIPTION
- 7 -
4. D E V I C E D E S C R I P T I O N
4.1 D E V I C E S E T U P
The following picture shows a possible configuration of a EnDaL vi. The available inputs and
outputs on the EnDaL curve/vi depend on the specific model. Please contact the distribution
of SMT ELEKTRONIK GmbH for further information.
1EnDaL curve/vi
8 Connection for RS232 interface (COM-Port)
2Cover for battery compartment and bypass
switch
9Connection for external TH or THP sensor
3Magnetic shackle (on/off-switch)
10 Connection for external power supply
4 LEDs for device status
11 External TH or THP sensor
5 Control key
12 External acceleration sensor
6 Mounting plate
13 External power supply
7 Connection for external acceleration sensor
13
2
3
4
1
5
6
7
8
9
10
12
11

DEVICE DESCRIPTION EnDaL curve/vi
- 8 -
4.2 C O N N E C T I N G E X T E R N AL S E N S O R S A N D S W I T C H E S
4 . 2 . 1 E x t e r n a l A c c e l e r a t i o n S e n s o r s A n d S w i t c h e s
Depending on the model,EnDaL curve/vi can be equipped with internal acceleration
sensors as well as with inputs for external acceleration sensors. Piezoelectric acceleration sensors with
integrated electronics (EnDaL vi) or without (EnDaL curve) can be connected to the 3 inputs designated
with X, Y, Z. However, internal and external acceleration sensors cannot be used simultaneously.
The measuring ranges of the internal charge amplifiers are 100, 200, 500, 1000 or 2000 pC for each
channel (EnDaL curve). Measuring ranges of 50, 100, 250, 500 or 1000 mV can be set for ICP sensors
(EnDaL vi). The transmission factors of each sensor connected to a channel can also be entered in the
measuring program so that the stored measured values correspond to the accelerations in g transmitted by
the sensors.
In order to obtain correct spatial shock values (R) derived from the 3 channels x, y and z (see evaluation
software) make sure to place these sensor channels orthogonal to one another.
Always ensure to switch off the EnDaL curve/vi before connecting external acceleration sensors.
Otherwise the device may record acceleration events during the connector insertion.
To prevent effects of cable movement on the measurement, it is recommended to fasten the sensor cables
with a cable holder. An adhesive tape may also be used for fastening (see also DIN ISO 5348).
4 . 2 . 2 D i g i t a l / A n a l o g u e I n p u t
The “In D/A”socket provides the configurable analogue input and up to 4 digital inputs. The switching
state of switches can be evaluated without additional circuits. The analogue sensor is temporarily supplied
by 5 V at the time of measurement.
1Ground
2Analogue input 0 ... 2,5 V
3Digital channel D0 (cannot be used with external ICP sensor)
4Digital channel D1
5Digital channel D2
6Digital channel D3
75 V / 20 mA supply of sensor, switched
An optional terminal box is available to ease connection to that “In D/A”socket.

EnDaL curve/vi DEVICE DESCRIPTION
- 9 -
4.3 O P E R A T I O N
4 . 3 . 1 S w i t c h i n g O n A n d O f f
The device offers a magnetic switch for turning on and off. This is implemented by the special safety
shackle. After power up of the device (removal of the safety shackle), the LEDs signalize the status of the
device for a few seconds. When the LEDs go out, the device is ready for measuring.
To avoid an intentional or unintentional interruption of the measurement, the magnetic off-switch can be
bypassed by a toggle switch. This switch is located inside the battery compartment.
Toggle the switch corresponding to the figure or the label
inside the cover of battery compartment to “ON“, if you
want to bypass the functionality of the safety shackle. With
the position “Shackle”,the device can be switched on and
off by the safety shackle.
You can check the on/off-state of the device by pressing the
control key.
4 . 3 . 2 S t a t u s i n d i c a t i o n
The status of the device is always signalized after power up. Generally the device is switched on by
removing the safety shackle. When the LEDs do not light up afterwards, check - using the control key -
whether the bypass switch is activated (see 4.3.2).
The LEDs indicate the following states:
LED
Colour
Visualization
State Description
Error
Red
Permanent
Hardware failure which is unrecoverable by user,
Synchronization of the baud rate
Power
Green
Permanent
Battery reserve 100…30 % or external power supply
Flashing
Battery reserve <30 %
Alarm
Red
Permanent
Exceedance of threshold value for X, Y or Z
Active
Green
Permanent
Active acquisition period, within a start/stop period
Flashing
Inactive, Beyond a start/stop period
Memory
Green
Permanent
Usage of all memory areas below 50 %
Flashing
Usage of at least one area above 50 %
Off
At least one memory area is full
An alternative to the indication by means of the integrated LEDs is to signalise the current operating status
with an external test connector. This test connector needs to be connected to the
EnDaL curve/vi via the RS-232 data cable. The function of this test connector is described in a separate
appendix.

DEVICE DESCRIPTION EnDaL curve/vi
- 10 -
4.4 C H A N G I N G T H E B A T T E R I ES
Before changing the batteries, switch off the instrument using the safety shackle.
(Dashed circle must be covered).
Remove the two screws securing the battery cover with a flat screwdriver (4x0.8). Ensure that the toggle
switch next to the battery compartment is set to "Shackle" and not to "ON", so that the safety shackle can
switch off the EnDaL curve. After unscrewing the battery holder (cross recessed head H size 2), the used
batteries can be removed. Insert the new batteries, making sure their polarity is correct. Tighten the battery
holder with a tightening torque of 75 Ncm. When fitting the battery-cover, make sure that the gasket has
the correct orientation (lips pointing into the device). Ensure that the sealing faces are clean. The maximum
tightening torque for the screws of the cover is 30 Ncm.
In order to change the 9V battery of the EnDaL vi (version with an external power supply),
open the battery compartment cover and the inner battery cover by carefully applying a flat screwdriver. In
this case no screws need to be removed.

EnDaL curve/vi DEVICE DESCRIPTION
- 11 -
4.5 M O U N T I N G I N F O R M A T I O N
Depending on the type of measurement you want to perform, attach the instrument to the goods
to be transported, inside the package or to the means of transport. To avoid unintentional shock
absorption and thus the measurement of insufficient acceleration values, you should attach the
EnDaL curve/vi as close as possible to the goods or package.
Shocks absorbing padding or instable assemblies distort the measurement results!
The available mounting accessories give you the following possibilities of fastening the instrument:
Direct mounting of the instrument (recommended method)
If the freight or its package is suited for direct mounting, there is no need to fasten a mounting plate. The
instrument is then fastened directly with 3 M5 cheese head screws with spring washer or similar fasteners.
This method ensures optimum transmission to the internal sensor.
Mounting with mounting plate
Fasten the mounting plate with the screws to the freight or
the means of transport. Then fasten the instrument to this
plate using the screws M5x50 and a locking element
(spring washer, lock washer).
Attachment with adhesive tape
This method of fastening is suited for smooth, non-porous surfaces and for short-term fastening. Apply two
double-sided adhesive strips onto the rear of the mounting plate of the EnDaL curve/vi. Now
pull off the protective liner from the other side and stick the EnDaL curve/vi onto a suitable
surface of the freight (e.g. inside a container). The location can only be corrected by using new adhesive
tapes.
Provide for clean, non-greasy surfaces!
We urgently recommend securing the instrument additionally by tightening or tie-down straps!
06,6 / 011x6,8
M5
48
70
60
68

DEVICE DESCRIPTION EnDaL curve/vi
- 12 -
4 . 6 M E A S U R E M E N T D A T A A C Q U I S I T I O N
4 . 6 . 1 M e a s u r i n g S h o c k E v e n t s
Every acceleration event that exceeds the set recording threshold and the minimum shock duration on at
least one of the active measuring channels (X, Y, Z) is recorded. In addition to date and time, the peak
values of acceleration, the shock values integrated over time (speed difference) for the active measuring
channels and the event duration are recorded with their amount and direction.
The duration of the event (t) is the time while the
permanently set hardware threshold (approx. 5% of
the measuring range) is exceeded in at least one
channel.
All shock events are always recorded with their curve.
Depending on the curve duration that has been set,
this curve is stored in the memory for the 10 or 20
greatest shock events. Knowing the shock curve is
important for the determination of weak points in
securing transported goods.
The measuring ranges, the frequency range of the
analogue filter, the minimum shock duration and the
recording threshold can be set and defined before every application. We recommend using the preset
measuring program files.
To detect transport damage as early as during the transport, an alarm threshold can be set for every
measuring channel. Pressing the Ctrl button indicates a possible excess of a limit value by a red LED (see
indication of operating states). For devices starting from production 11/2002 this signaling is reset
automatically after each PC data communication.
4 . 6 . 2 S t o r i n g D i g i t a l E v e n t s
Every voltage level change at one of the 4 available digital inputs causes the input states of all digital
channels to be stored with a resolution of seconds with date, time and optionally geographical position.
The storage of digital events can only be activated by active channels.
Both low-active and high-active voltage levels and normally closed or open switches can be connected. This
allows the monitoring of door contacts, tilt switches or run times of connected units.
When an external ICP acceleration sensor is selected in the measuring program, the digital channel D0 is
used for sensor monitoring and can then not be used for the monitoring of external switching signals. In
case of an a-channel X failure D0 indicates a low level, otherwise a high level.
If a Tracking Module or data link TM GPS is connected, it is possible to
determine global location coordinates for acceleration and digital events.

EnDaL curve/vi DEVICE DESCRIPTION
- 13 -
4 . 6 . 3 S y n c h r o n o u s m e a s u r e m e n t s
The measurement of temperature and humidity and the free analogue measuring channel and optionally the
recording of geographical position are time-controlled and occur at regular intervals. The interval is
transmitted by the PC software (can be set from 2 seconds to 60 minutes) and is the same for every
measurement mentioned in this paragraph.
If the sensor is not connected or defective, the measured value is defined as the absolute lower-range value
(-40 °C or 0 % relative humidity).
A maximum of 16,384 records are possible which results in the following durations of operation.
Interval in min
Duration of operation with 16,384
possible measured values
1
11 days
5
56 days
10
113 days
20
227 days
60
682 days
4 . 6 . 4 M e a s u r i n g T h e A n a l o g u e S i g n a l
The analogue input In D/A pin1+2 allows the measurement of a voltage signal in the 0...2.5 V range.
Higher signal levels need an external voltage divider connected in series and low signals an amplifier. The
measured values are indicated as figure * unit of measurement if you have entered the parameters Offset,
Factor and Unit of measurement correctly in the menu EnDaL / Program / Sensors button / Analogue
measuring channel filecard before the recording. You can enter up to 6 characters for the unit of
measurement. The figure that is shown in the evaluation results from:
Figure = (1638 * UM/volts - offset) * factor
where UMis the voltage in volts applied at the analogue input In D/A between pin 2 and pin 1.
Example:
For a physically correct indication of a voltage between 0 and 10 volts connected via a 4:1 voltage
divider, you have to enter the following:
Offset: “0.0“
Factor: “0.002442“
Unit of measurement: “volts“

PC EVALUATION SOFTWARE ENCUR EnDaL curve/vi
- 14 -
5. PC E V A L U A T IO N S O F T W A R E E N C U R
5 . 1 S Y S T E M R E Q U I R E M E N T S
The evaluation software ENCUR is exclusively designed for the usage with the data logger
EnDaL curve and EnDaL vi.
The software ENCUR requires the following minimum PC specifications:
-Operating system: Windows Vista / 7 / 8 / 10
-30 MB hard drive space
-USB interface or COM port
-Internet connection for installation and use of Google Earth
The installer of Google Earth will download the current program files from the internet, so they cannot be
terminated. Please obtain information about the special system requirements of Google Earth.
5 . 2 I N S T A L L A T I O N
Start the setup “Setup_Encur_3.x.exe“. Follow the instructions of the installation wizard. The setup installs
the software files, some example files and help files into the chosen program path. The driver for the Prolific
USB serial adapter can be installed optionally. The adapter is available at SMT ELEKTRONIK GmbH.
Please contact the technical support for further information.

EnDaL curve/vi PC EVALUATION SOFTWARE ENCUR
- 15 -
5 . 3 O V E R V I E W
Function
Symbol
Main menu
(context menu)
Shortcut
Opened file
Closed file
Menu File
Transfer device data
Data transfer
Ctrl+T
X
Open measuring file (*.edz)
or program file(*.epz)
Open
Ctrl+O
Drag&Drop
X
open last files
Save file (*.edz)
Save
Ctrl+S
X
Print
Print
Ctrl+P
X
Print preview
Print preview
X
Close current project
(measuring file)
Close
Ctrl+W
X
Close ENCUR
Exit
Alt+F4
X
X
Menu EnDaL
Open measuring program
(file)
Meas. program
Ctrl+M
X
X
Auto programming
(transfers a predefined
measuring program file)
Autoprogramming
Ctrl+A
X

PC EVALUATION SOFTWARE ENCUR EnDaL curve/vi
- 16 -
Function
Symbol
Main menu
(context menu)
Shortcut
Opened file
Closed file
Overall graphic total
(graphical overview with all
measured values for all
acquisition periods)
Overall graphic total
(Overall graphic total)
Ctrl+Alt+G
X
Overall graphic
(graphical overview with all
measured values for selected
acquisition period)
Overall graphic
(Overall graphic)
Ctrl+Alt+O
X
Acceleration table
(for selected acquisition
period)
Acceleration a-table
(Acceleration a-table)
Ctrl+Alt+A
X
Distribution of acceleration
(for selected acquisition
period)
Acceleration Distribution
Table/Graphic
(Acceleration Distribution
Table/Graphic)
X
Classification of acceleration
(for selected acquisition
period)
Acceleration
Classification Table/Graphic
(Acceleration Classification
Table/Graphic)
X
Limit table
(a-table with values above
the chosen limit for the
selected acquisition period)
Acceleration
Limit table
(Acceleration Limit table)
X
Synchronous table
(table with the cyclic
measured values)
Synchronous table
(Synchronous table)
Ctrl+Alt+T
X
Table of digital events
Digital table
(Digital table)
Ctrl+Alt+D
X
GPS values
(opens the configured
standard program to show
the tracked GPS values for
the selected acquisition
period)
GPS values
(GPS values)
Ctrl+Alt+I
X

EnDaL curve/vi PC EVALUATION SOFTWARE ENCUR
- 17 -
Function
Symbol
Main menu
(context menu)
Shortcut
Opened file
Closed file
Change password
(configuration of the device
protection)
Change password
X
Menu Options
Basic settings (configuration
of interfaces and display)
Basic settings
X
X
Classification (parameter for
classification)
Classification Shock
duration/ Acceleration/
Velocity
X
Limit value (sets the threshold
for limit table)
Limit value
X
Spectrum (parameter for
spectral analysis)
Spectrum
X
X
Language setting
Language English/
Deutsch
X
Menu Help
Help topics (manual)
Help topics
F1
X
X
Info about…
Info about…
X
X

PC EVALUATION SOFTWARE ENCUR EnDaL curve/vi
- 18 -
5 . 4 C O N F I G U R A T I O N S ( O P T I O N S )
5 . 4 . 1 B a s i c S e t t i n g s
In the menu Basic settings the file path, the communication interface as well as the graphical view can be
configured. The types of measurement data to be transferred from the EnDaL curve/vi to the PC can also
be configured. This menu is located in the tab Options.
Confirming the basic settings with the button “OK”saves the configuration for the next program start in the
selected working directory.
Path settings
The transferred data will be automatically saved in the “Working directory”. This directory can be changed
in the tab “Path settings”. Additionally the user settings will be stored in this file path for the next start of
ENCUR.
The “Measuring program for automatic programming” is an
individually configured measuring program file, which can be
chosen to configure more than one device with the same
program. The file can be transferred by the menu “EnDaL
Autoprogramming” (Strg + A).
The selected “GPS program”is used to display the measured GPS
data. GPS values are tracked by a Tracking
Module or a data link TM, which can be connected
with the EnDaL curve or EnDaL vi. If the file path
is invalid, the ENCUR program will assume the standard software
of the PC system.