MRC PHT-3109 User manual

HUMIDITY/TEMP.
CONTROLLER/MONITOR
Model : PHT-3109
Your purchase of this HUMIDITY/TEMP. CONTROLLER,
MONITOR marks a step forward for you into the field of
precision measurement. Although this METER is a
complex and delicate instrument, its durable structure
developed. Please read the following instructions carefully
and always keep this manual within easy reach.
OPERATION MANUAL
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 35594529
MRC. 11.19

Caution Symbol
Caution :
* Risk of electric shock !
Caution :
* Do not use fingers or any tool
to touch the Wire Terminals.
* Do not apply the relay contact load
current > 0.5 Amp.
* The instrument contains no user
serviceable parts and should not
be opened by the user.
* Repair or after service should be
done by a qualified technician only.
* Power supply should apply the
correct ACV power voltage
* Cleaning - Only use the dry cloth to
clean the plastic case !
* Equipment protectted throughout
by Double Insulation or Reinforced
Insulation.
Environmental Condition
* Comply with EN61010.
Transient overvoltage at Mains Supply 2500V.
* Pollution Degree 2.
* Altitude up to 2000 meters.
* Indoor use.
* Relative humidity 80% max.

TABLE OF CONTENTS
1.FEATURES................................................................1
2.SPECIFICATIONS......................................................
.
1
2-1 General Specifications..........................................1
2-2 Electrical Specifications........................................3
3.FRONT PANEL DESCRIPTION.....................................4
3-1 Display...............................................................4
3-2 PV ( process value ) indicator...............................4
3-3 SV ( set value ) indicator.....................................
.
4
3-4 Set Button..........................................................
.
4
3-5 Button............................................................▼4
3-6 Button............................................................▲4
3-7 RH/Temp Button.................................................4
3-8 %RH control relay indicator.................................
.
4
3-9 Temp. control relay indicator................................4
3-10 %RH ( humidity ) indicator.................................4
3-11 Temp. indicator.................................................4
3-12 Wire terminals ..................................................4
3-13 RS232 terminal..................................................4
3-14 Input socket .....................................................4
3-15 Case holder ......................................................4
3-16 Probe plug........................................................4
3-17 Probe head ( Humidity & Temperature )..............4
3-18 Probe handle.....................................................4
3-19 Probe fix holder.................................................4
4.MEASURING PROCEDURE..........................................5
4-1 Terminal connection............................................5
4-2 Humidity/Temp. measurement............................
.
5
4-3 1st layer setting procedures.................................6
4-4 2nd layer setting procedures................................8
5. RS232 PC SERIAL INTERFACE...................................
.
12
6. SYSTEM RESET........................................................ 14
7.THE ADDRESS OF AFTER SERVICE CENTER................15

1. FEATURES
* Professional Humidity measurement monitor and
controller.
* Build in humidity control relay and the Temp.
control relay.
* Relay will be make action ( On/Off ) when the reading
value reach high limit or low limit value.
* Offset value setting.
* Hysteresis value setting for high and low alarm.
* , temp. unit setting with default.℃℉
* High precision humidity sensor with fast response time
and high accuracy.
* Large red LED display, high brightness and easy to read.
* RS232 computer interface, send out the humidity and the
temperature data at the same time.
* Optional data acquisition software.
* Optional GSM controller.
* Microprocessor circuit ensures high accuracy and
provides special functions and features.
* Standard 96 X 48 mm DIN case.
2. SPECIFICATIONS
2-1 General Specifications
Display 4 digits red LED, digit size : 14 mm.
Unit Temp. ,℃℉
Humidity %RH
Circuit Custom chip of microprocessor LSI
circuit.
Sensor Humidity Semiconductor
Structure Temperature Semiconductor
1

Sampling Time Approx. 1 second.
Relay outputs Number 2 relays
Function
Relay 1 :
Humidity control relay.
Relay 2 :
Temperature control relay.
Max load 0.5 ACA/250 ACV
0.5 DCA/24 DCV
* Do not apply the relay
contact load current
> 0.5 A, other wise the
relay may be damaged
permanently without
warranty.
Setting value High limit value setting.
Low limit value setting.
Hysteresis value setting.
Offset value setting.
* Setting for Humidity and Temp.
External DC 12 V, 50 mA max.
Power Supply
Data Output RS 232 PC serial interface.
Operating 0 to 50 .℃
Temperature
* Meter
Operating Less than 80% R.H.
Humidity
* Meter
Power Supply 90 to 260 ACV, 50/60 Hz.
Power Approx. 4.7 VA/AC 110V.
Consumption Approx. 5.3 VA/AC 220V.
Weight 384 g/ 0.84 LB.
* Including probe.
Dimension DIN size : 96 x 48 mm.
Depth : 110 mm.
Accessories Instruction manual.....................
.
1 PC
Included Humidity/Temp. probe................1 PC
Case holder with screw...............2 PCs
Probe fix holder..........................1 PC
2

Optional * Data Acquisition software,
Accessories SW-U801-WIN.
* RS232 cable, UPCB-02.
* GSM controller, GSM-889.
* Interface cable ( cable between meter
to GSM-889 ), GMCB-89.
2-2 Electrical Specifications (23±5 )℃
Humidity
Range 10 % to 95 % R.H.
Resolution 0.1 % R.H.
Accuracy 70% RH≧± (3% reading + 1% RH).
< 70% RH - 3% RH.
±3%RH.
Temperature
Measuring Range 0 to 50 /32 to 122℃℃℉ ℉
Resolution 0.1 /0.1℃℉
Accuracy ± 0.8 /1.5℃℉
*
T
he above specifications is for the probe only.
*
A
bove specification tests under the environment RF Field Stren
g
th less than
3
V
/M & frequency less than 30 MHz only.
3

3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display
3-2 PV ( process value ) indicator
3-3 SV ( set value ) indicator
3-4 Set Button
3-5 Button▼
3-6 Button▲
3-7 RH/Temp Button
3-8 %RH control relay indicator
3-9 Temp. control relay indicator
3-10 %RH ( humidity ) indicator
3-11 Temp. indicator
3-12 Wire terminals
3-13 RS232 terminal
3-14 Input socket
3-15 Case holder
3-16 Probe plug
3-17 Probe head ( Humidity & Temperature )
3-18 Probe handle
3-19 Probe fix holder
4

4. MEASURING PROCEDURE
Terminal layout Fig. 2
4-1 Terminal connection
1)Input the ACV power ( 90 to 260 ACV ) to T1, T2.
Do not input the
over voltage to the
AC input terminals.
2)Connect the " Humidity Control Relay " output from T3, T4.
Connect the " Temp. Control Relay " output from T5, T6.
4-2 Humidity/Temp. measurement
1)Connect the " Probe plug " ( 3-16, Fig. 1 ) to
" Input socket " ( 3-14, Fig. 1 ).
2)Power on the meter, the " Display " ( 3-1, Fig. 1 ) will
show the humidity value, in the same time the " %RH
indicator " ( 3-10, Fig. 1 ) will light.
5

]
3)Press the " RH/Temp Button " ( 3-7, Fig. 1 ) once,
the " Temp. indicator " ( 3-11, Fig. 1 ) will light, the
" Display " ( 3-1, Fig. 1 ) will show the Temp. value.
* Press the " RH/Temp Button " ( 3-7, Fig. 1 ) once again,
the " Display " will return the " Humidity value ", in the
same time the " %RH indicator " ( 3-10, Fig. 1 ) will light again.
4-3 1st layer setting procedures
LoLt Low Limit
HILt High Limit
Low Limit Value Setting
1)Press the " Set Button " ( 3-4, Fig. 1 ) once, the "
Display " will show " LoLt ", now the meter is ready
for the Humidity " Low Limit " value setting.
Press the " RH/Temp Button " ( 3-7, Fig. 1 ) once,
the " Temp indicator " ( 3-11, Fig. 1 ) will light, now
the meter is ready for the Temperature " Low Limit "
value setting.
Remark :
* Under " Display " show " LoLt ", if " % RH indicator "
( 3-10, Fig. 1 ) is lit, meter is ready for " Humidity
Low Limit " setting.
* Under " Display " show " LoLt ", if " Temp. indicator "
( 3-11, Fig. 1 ) is lit, meter is ready for
" Temperature Low Limit " valuesetting.
* The function of " Low Limit value " setting, refer to
page 10, Fig. 2.
6

2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring " Low
Limit " value.
* When adjust the value, the " SV indicator " ( 3-3, Fig. 1 )
will light.
High Limit Value Setting
1)After set the " Low Limit " value, press the " Set
Button " ( 3-4, Fig. 1 ) twice, the " Display " will show
" HILt ", now the meter is ready for the humidity
" High Limit " value setting.
Press the " RH/Temp Button " ( 3-7, Fig. 1 ) once,
the " Temp indicator " ( 3-11, Fig. 1 ) will light, now
the meter is ready for the Temperature " High Limit "
value setting.
Remark :
* Under " Display " show " HILt ", if " % RH indicator "
( 3-10, Fig. 1 ) is lit , meter is ready for " Humidity
High Limit value " setting.
* Under " Display " show " HILt ", if " Temp. indicator "
( 3-11, Fig. 1 ) is lit, it meter is ready for
" Temperature High Limit value " setting.
* The function of " High Limit value " setting, refer to
page 10, Fig. 2.
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring " High
Limit " value.
* When adjust the value, the " SV indicator " ( 3-3, Fig. 1 )
will light.
7

After adjust the " High Limit " value, press the " Set
Button " ( 3-4, Fig. 1 ) twice, " Display " will return to
the normal measuring screen.
4-4 2nd layer setting procedures
tPty Temp. unit setting
HySt Hysteresis setting
oFSt Offset setting
Temperature unit setting
1)Press the " Set Button " ( 3-4, Fig. 1 ) continuously at
least two seconds, the " Display " will show " tPty ",
now the meter is ready for the Temperature unit ( ,℃
) setting.℉
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring
temperature unit to " C " or " F ".
* When adjust the Temp. unit, the " SV indicator " ( 3-3, Fig. 1 )
will light.
8

Hysteresis value setting
1)After select the temperature unit ( , ), press the℃℉
" Set Button " ( 3-4, Fig. 1 ) twice, the " Display "
will show HySt ", now the meter is ready for the
the Hysteresis value setting.
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring
Hysteresis setting value.
* When adjust the Hysteresis value, the " SV indicator "
( 3-3, Fig. 1 ) will light.
Press the " RH/Temp Button " ( 3-7, Fig. 1 ) once,
the " Temp indicator " ( 3-11, Fig. 1 ) will light, now
the meter is ready for the Temperature " Hysteresis "
value setting.
Remark :
* Under " Display " show " HySt ", if " % RH indicator "
( 3-10, Fig. 1 ) is lit, meter is ready for " Humidity
Hysteresis value " setting.
* Under " Display " show " HySt ", if " Temp. indicator "
( 3-11, Fig. 1 ) is lit, meter is ready for
" Hysteresis value " setting.
* The function of " Hysteresis value " setting, refer to
page 10, Fig. 2.
9

Fig. 2
High limit value : 100
Low limit value : 20
Hysteresis value : 5
*The control relay will On when measuring
value up to 100. The control relay will Off
again when measuring value down to 95.
*The control relay will On when measuring
value down to 20. The control relay will
Off when measuring value up to 25.
Offset value setting
1)After finish the Hysteresis setting, press the
" Set Button " ( 3-4, Fig. 1 ) twice, the " Display "
will show " oFSt ", now the meter is ready for the
the offset value setting.
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring
offset value.
* When adjust the Offset value, the " SV indicator "
( 3-3, Fig. 1 ) will light.
10

Press the " RH/Temp Button " ( 3-7, Fig. 1 ) once,
the " Temp indicator " ( 3-11, Fig. 1 ) will light, now
the meter is ready for the Temperature " Offset "
value setting.
Remark :
* Under " Display " show " oFSt ", if " % RH indicator "
( 3-10, Fig. 1 ) is lit, meter is ready for " Humidity
Offset value " setting.
* Under " Display " show " oFSt ", if " Temp. indicator "
( 3-11, Fig. 1 ) is lit, meter is ready for " Temperature
Offset value " setting.
* For example of " Offset value setting " :
The reading value is 102.
The offset value is 5.
The new display value will be 107 ( 102 + 5 ).
11

5. RS232 PC SERIAL INTERFACE
The instrument has RS232 PC serial interface via a 3.5
mm terminal ( 3-13, Fig. 1 ).
The data output is a 16 digit stream which can be
utilized for user's specific application.
A RS232 lead with the following connection will be
required to link the instrument with the PC serial port.
Meter PC
(3.5 mm jack plug) (9W 'D" Connector)
Center Pin..............................
.
Pin 4
Ground/shield........................Pin 2
Pin 5
The 16 digits data stream will be displayed in the
following format :
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
12

Each digit indicates the following status :
D15 Start Word
D14 4
D13 When send the upper display data = 1
When send the lower display data = 2
D12 & D11 Annunciator for Display
= 01℃= 02℉m/S = 08
km/h = 10 mph = 12 knot = 09
FPM = 11 %RH = 04 dB = 17
LUX = 15 Ft-cd = 16
D10 Polarity
0 = Positive 1 = Negative
D9 Decimal Point(DP), position from right to the
left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D8 to D1 Display reading, D8 = MSD, D1 = LSD.
For example :
If the display reading is 1234, then D8 to
D1 is : 00001234
D0 End Word
RS232 setting
Baud rate 9600
Parity No parity
Data bit no. 8 Data bits
Stop bit 1 Stop bit
13

6. SYSTEM RESET
Power on the meter, use the two fingers to press " Set
Button " ( 3-4, Fig. 1 ) and " RH/Temp. Button " ( 3-7,
Fig. 1 ) continuously more than 5 seconds until the
Display show the text " rSt ", release the buttons. After
" rSt " text flashing 2 times will return to the normal
screen. The meter system will be reset, all the
calibration data will be cleared, the meter's internal
function will return the default value.
14

7. THE ADDRESS OF AFTER SERVICE
CENTER
15