corticosteroids for a long time, or is taking now.
4. Precautions for Use
(1) Clean the operating head with normal saline after use and keep it properly.
(2) Plug with ground pin must be used, and power socket that already factually
grounded must be ensured before using the equipment.
(3) To make sure equipment’s voltage is adaptive. If voltage of local power supply is
unstable, we suggest users add voltage regulator with matching power between
the mains supply and the equipment.
(4) To guarantee curative effect and normal service life of the equipment, please
uniformly use specified parts provided or suggested by original manufacturer.
(5) The equipment can’t be placed in damp places or near water, and also can’t be
exposed to direct sunlight.
(6) Don’t place the equipment near a strong heat source since this may affect its
service life and normal use.
(7) Please remove all the metal objects from body before treatment so as to avoid
unexpected situations which may affect curative effect.
(8) Please restrain from using the equipment aiming at eyes, thyroid, parathyroid,
testicles, pregnant woman’s abdomen, pacemaker, etc.
(9) People who are suffering from illness should use with caution unless he gets
permission from a doctor.
(10) Please turn off power switch of the equipment if no one uses it, and ensure main
power is off after someone used it and before he left, thus safety of the
equipment can be guaranteed.
(11) Using the equipment and training the operators in strictly accordance with the
instructions specified in the manual.
(12) It’s advised to stop taking other weight-loss medicines for 1 to 2 months before
treatment if you are taking them now. The course of treatment shall be extended
if you want to have it started immediately.
(13) Don't use the equipment with an empty stomach, and wait for at least 1 hour
after a full meal before starting the course for weight reduction.
(14) The equipment should contact skin fully so as to avoid uneven heating when in
use.
(15) Start with the lowest energy level when just operated, and increase the level
gradually after client got used to it.
(16) Matched freezing film must be applied to the treated part to avoid dry skin
treatment or treating with the device contacting directly.
(17) Clean the equipment with normal saline after operation so as to ensure its
cleanliness and hygiene, as well as extending its service life..
5. Troubleshooting & Solutions
(1) The equipment can’t be started, and the lights in its back are not working?