myomo MyoPro Motion E Operating and maintenance instructions

PN: 27533 Rev: 001
Quick Start Guides

PN: 27533 Rev: 001
WARNING:
Always read and refer to the user manual
for more complete and detailed instruction
before using this guide.

MyoPro Quick Start Guide: How to Don Your MyoPro (Motion E & W)
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
2020 design
PREPARATION: Ensure that the MyoPro is POWERED OFF and open all straps. Sit on a stable chair/surface.
STEP 1: Insert hand into hand shell. Double check hand position.
Motion E
Insert hand into shell and pull it all
the way through
Secure palm strap
Check hand and wrist position:
Knuckles should be visible
Wrist and forearm will be
held in “neutral” (thumb
pointing towards ceiling)
Motion W
Insert hand into shell and pull it all the way through
Secure palm strap
Check hand and wrist position:
Knuckles should be visible
Pivot point of wrist joint should align with the anatomical wrist bend
Wrist flexion/extension
How to make adjustments to the 2020 wrist design:
Pronation/Supination

MyoPro Quick Start Guide: How to Don Your MyoPro (Motion E & W)
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
STEP 2 (Motion E & W): Lower wrist and forearm into the MyoPro. Secure forearm strap.
Lower the forearm into the MyoPro
Feed the forearm strap through the D-ring
Tighten the strap
STEP 3 (Motion E & W): Secure upper arm sensor cuff and strap.
BLUE sensor over front of arm (bicep)
RED sensor over back of arm (tricep)
Feed outer strap through D-ring
Cuff and straps must be very snug
WARNING: This quick reference guide does not replace the device user manual. Always refer to user manual for more complete and detailed instructions.

MyoPro Quick Start Guide: How to Don Your MyoPro (Motion G)
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
PREPARATION: Ensure that the MyoPro is POWERED OFF. Remove finger saddle and open all straps.
STEP 1: Insert hand into hand shell. Reattach finger saddle. Double check hand position.
Insert hand into shell and pull it all the way through.
Secure palm strap
Place forearm into wrist component and
forearm shell
Reattach finger saddle. Check hand
position. There should be NO SPACE
between palmar bar and base of thumb

MyoPro Quick Start Guide: How to Don Your MyoPro (Motion G)
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
STEP 2: Secure forearm strap and push forearm sensor cuff into position
Feed strap through D-ring
GREEN sensor on inside of forearm
sensor on outside of forearm
STEP 3: Secure upper arm sensor cuff and strap.
BLUE sensor over front of arm (bicep)
RED sensor over back of arm (tricep)
Feed outer strap through D-ring
Cuff and straps must be very snug
WARNING: This quick reference guide does not replace the device user manual. Always refer to user manual for more complete and detailed instructions.

MyoPro Quick Start Guide: How to Don Your Harness
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
PREPARATION: Don your MyoPro first. The quick
release tab on the harness ( ) should be
disconnected for donning. Do not disconnect the
harness from the MyoPro.
HOW TO DON YOUR HARNESS BY YOURSELF:
1. 2. 3. 4.
Raise (abduct) your arm to the
side until the harness saddle is
positioned in the middle of your
shoulder
Clamp the saddle in place with
your chin. Keep your shoulder
raised to the side. Reach around
your back and bring the long strap
around your body and under your
opposite armpit
Tip: place your arm on a table or
armrest if you are unable to lift it
yourself
Secure the quick release tab to
the saddle
Lower your arm
The straps of the harness should feel
taut and you should feel that the
harness is suspending the MyoPro on
your arm

MyoPro Quick Start Guide: How to Don Your Harness
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
HOW TO DON YOUR HARNESS WITH ASSISTANCE
1. 2. 3.
Caregiver elevates arm until the saddle is
positioned in the middle of the shoulder
Keeping the arm elevated and the saddle
stabilized, attach the quick release tab to the
saddle
Lower the arm
Check that the straps feel taut
OPTIONAL: ANTI ROTATION STRAP
An additional strap can be added to help keep the user’s arm away from
their body. This anti-rotation strap starts from the second chafe in the
front and attaches to the back strap.
WARNING: This quick reference guide does not replace the device user manual. Always refer to user manual for more complete and detailed instructions.

MyoPro Quick Start Guide: Control Panel
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
Power Button: Push to turn MyoPro on/off
Bluetooth indicator light: turns blue when MyoPro is connected to Myconfig
Standby Mode
Bicep indicator light
Tricep indicator light
Close indicator light
Open indicator light
Dual Mode
Tricep Mode
Bicep Mode
Standby Mode
Dual Mode Close Mode
Standby Mode:
Open Mode
This mode is when the MyoPro has been powered on but neither an elbow nor a hand
mode has been selected. The motor will NOT respond in Standby Mode.
Return to Standby Mode at any time by pushing the elbow and/or hand mode buttons
until indicator lights turn off. Alternatively, the MyoPro can be completely powered off at
any time during use.
WARNING: This quick reference guide does not replace the device user manual. Always refer to user manual for more complete and detailed instructions.
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
Hand Mode Button (Motion G only)
Elbow Mode Button

MyoPro Quick Start Guide: Troubleshooting
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
If the solutions presented here do not fix your problem, please call Myomo Support: (617) 651 - 0226 or (877) 736 - 9666
Problem
Solution
Issues with the fit of the MyoPro on the arm:
Fingers falling out of the saddle
MyoPro sliding down the arm
Discomfort/pressure points
Make sure the MyoPro has been donned correctly and all joints are
aligned and cuffs/straps are snug. Refer to Quick Start donning guides
and user manual for detailed instructions.
Hand and hand strap routing:
Elbow:
Elbow aligned with MyoPro axis
Harness:
Poor Good
Contact your O&P provider or your Myomo Regional Manager to
address discomfort and any necessary modifications
Jerky or erratic movements
Check the fit of the MyoPro on the arm. If the cuffs and straps
are too loose, the sensors will lose contact with the skin,
causing erratic motion.
Jerky motion is common at the start of the training program,
especially when practicing muscle relaxation. Consistent
practice of proficiency drills and working with a therapist can
improve this.
Sensitivity is set too high. Settings in MyConfig may need to
be adjusted by the therapist/Myomo Regional Manager/ O&P
provider.

MyoPro Quick Start Guide: Troubleshooting
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
EMG settling time. If jerkiness is noted when the MyoPro is
turned on, there may be noise from static. Return to stand-by
mode and stay relaxed for 1 – 2min, then try again.
MyoPro is not moving/responding
Check to see if the MyoPro is in Standby mode.
Sensitivity is set too low or threshold is set too high.
Settings
in MyConfig may need to be adjusted.
The user should be relaxed when selecting modes.
Make sure the user is
activating the right muscle group(s) for
the mode.
BICEP MODE: squeeze bicep to bend, relax to straighten
TRICEP MODE: squeeze tricep to straighten, relax to bend
DUAL MODE: squeeze bicep to bend, squeeze tricep to
straighten
CLOSE MODE: squeeze fingers or bend wrist to close, relax to
open
OPEN MODE: extend fingers or bend wrist back to open, relax
to close
DUAL MODE: squeeze fingers to close, extend fingers to open
Low battery: indicated by
100%
75%
50%
25%
Push here Color of power button
Sensors need to be wiped or cleaned
(clean with alcohol prep
pad)
User is perspiring (skin should be clean and dry)
If perspiration is a frequent issue, Certain Dri can
help. This is found in any drugstore in the deodorant
section. Apply lightly, the night before.
How to charge the battery
Push button to
open battery
compartment
Remove battery,
charge on cradle
WARNING: This quick reference guide does not replace the device user manual. Always refer to user manual for
more complete and detailed instructions.

MyoPro Quick Start Guide: How to Use Go-To-Meeting & Zoom
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
1. Click on “camera” to turn it 2. Select an audio source: computer vs phone
If phone, enter the phone number listed and access code
3. Click “ok I’m ready” to join the meeting
Go-To-Meeting
ON YOUR DESKTOP or TASK BAR (at the bottom of your screen):
Double click on this icon to open Go-To-Meeting
Enter the meeting ID (this ID will have been provided to you in an e-mail from Myomo) and click “JOIN”.
The meeting will start:

MyoPro Quick Start Guide: How to Use Go-To-Meeting & Zoom
Myomo Inc. | One Broadway 14th Floor | Cambridge, MA 02142
TEL: 877.736.9666
www.myomo.com PN: 27533 Rev: 001
email.
Zoom
ON YOUR DESKTOP or TASK BAR (at the bottom of your screen):
1) Double click on this icon to open Zoom
2) Click “JOIN MEETING” (Do not sign in)
3) Enter the code that you received on your Telehealth email OR click the link within your
Do not click the box for “Do not connect to audio”
Do not click the box “Turn off my video”
Click “Join”
4) While you wait for the host, test your computer audio and computer microphone to ensure that it is working.
5) Join Video and Audio
6) You should then be connected to the Myomo Representative and you will be added to the meeting.
WARNING: This quick reference guide does not replace the device user manual. Always refer to user manual for more complete and detailed instructions.
This manual suits for next models
2
Table of contents
Other myomo Medical Equipment manuals

myomo
myomo MyoPro 2 Motion E User manual

myomo
myomo myopro 2 Plus Motion W User manual

myomo
myomo MyoPro 2 Motion G User manual

myomo
myomo myopro 2 Plus Motion W User manual

myomo
myomo MyoPro 2 Motion E User manual

myomo
myomo myopro 2+ Motion G User manual

myomo
myomo MyoPro 2 Motion E User manual

myomo
myomo MARK User manual

myomo
myomo MyoPro 2 Motion G User manual