LIPOSURG INSTRUCTIONS FOR USE
MALFUNCTIONS AND TROUBLESHOOTING
MALFUNCTION CAUSE SOLUTION REFER TO INSTRUCTIONS FOR USE
Device is
not functional
(Indicator
light is off)
Control unit not switched on Set the power switch « I/O» to
« I »
S >
Power connection not
established
Connect the control unit to the
mains power supply
C >
Incorrect operating voltage Check the mains voltage C >
Faulty fuse Replace fuse R >
Processor error Set the power switch to « O»,
wait10seconds and switch
back to position « I»
S >
No stroke
movement at
the Conform
cannula
handpiece
« LIPO » function is not
selected
Select function « LIPO » O: C >
Motor to drive cannula not
activated
Activate motor to drive cannula
by pressing pedal
O VARIO >
Motor to drive cannula not
connected
Insert motor plug into the
socket at the devices front
D >
P H C
>
Conform cannula handpiece
is not connected properly to
the motor
Press the Conform cannula
handpiece firmly to the motor
until it engages
P H C
>
No delivery
of infiltration
solution at the
cannula
« INFIL » is not selected Select function « INFIL» O: C >
Peristaltic pump is not
switched on
Activate peristaltic pump by
pressing the pedal
O VARIO >
Tube set not properly
inserted
Insert tube set correctly P >
Tube set is blocked Replace the tubing set P >
Infiltration container is not
ventilated
Open air bleed valve at
tube set
P >
Tube set is dripping Replace tube set P >
Roller clamp is closed Open the roller clamp all the
way
P >
Pedal is
not functional
Pedal not connected Insert the pedal plug into the
socket at the devices back
P >
D >
Control unit is not switched
on
Use mains switch, in the back
of the device, to switch on
S >
Incorrect operation Check operating instructions O VARIO >
If the problem cannot be solved please contact your supplier or an authorized service center. Addresses can be found in the appendix of this instruc-
tions for use under S > .