Nouvag THP 3000 User manual

THP 3000
Tattooing handpiece
OPERATING INSTRUCTIONS
No. 31959 - 05/12
Safety Instructions
•The instrument is delivered in non-sterile condition! Before initial use and immediately after every use the handpiece
must be cleaned, disinfected and sterilised!
•Do not clean handpiece with compressed air!
•Only operate handpiece with attached needle!
•Never work on handpiece with the motor running, risk of injury!
•The handpiece must be used by skilled and trained persons only!
•The handpiece must not be operated at a speed exceeding 10 000 rpm!
•The clamp must not be stored clamped for an extended period without a clamped needle!
•No liability for warranty claims or any other liability will be accepted for improper use of the handpiece and operation
without observing the operating instructions!
Specifications
Weight: ....................................................................................................................................................................................................... 86 g
Optimum speed:................................................................................................................................................................................... 8000 rpm
Maximum speed:............................................................................................................................................................................... 10 000 rpm
Micromotor coupling:..................................................................................................................................................................as per EN23964
Icons on device
1275: E U - c o n f o r m i t y s y m b o l :Autoclavableuptomax.135°C :suitableforthermaldisinfection
Spare parts
Art. no. Name Units
5021............. Tattooing head blue (for 1x and 3x needles)..........................................................................................................................1 unit
5022............. Tattooing head red (for 7x, 10x and 12x needles)..................................................................................................................1 unit
40451........... Clamping nut .........................................................................................................................................................................1 unit
5024............. 1x needle sterilisable.............................................................................................................................................................1 unit
5025............. 3x needle sterilisable.............................................................................................................................................................1 unit
5026............. 7x needle sterilisable.............................................................................................................................................................1 unit
5027............. 10x needle sterilisable ...........................................................................................................................................................1 unit
5028............. 12x needle sterilisable ...........................................................................................................................................................1 unit
1984............. Nou-Clean Spray...................................................................................................................................................................1 unit
1958............. Attachment for Nou-Clean Spray ...........................................................................................................................................1 unit
Operation
Replacing needle
1. Unscrew the tattooing head in an anticlockwise direction.
2. Unscrew the clamping nut two revolutions anticlockwise.
3. Insert or replace needle.
4. Hand-tighten clamping nut clockwise.
5. Screw tattooing head clockwise by hand to the last notch (the insertion depth is now set to 2 mm).
!
!!
!
•Please use the blue tattooing head for the 1x and 3x needles. Please use the red tattooing head for the 7x, 10x and
12x needles.
•The insertion depth is adjusted on the tattooing head, which can be set for a stroke length from 0.2 mm to 2.0 mm.
The increment from one notch to the next is 0.2 mm.
•To ensure optimum colour pigmentation on the skin note that the variable pressure of the tattooing handpiece on the
skin will influence the insertion depth.
•It is also important to ensure that the handpiece is applied perpendicular to the skin during tattooing.
GB
GB
GB
GB
GB

Preparation Instructions
Restrictions on repeated
preparation
Frequent preparation has little effect on the handpieces. The end of the service life is normally the result of wear and damage
caused by use.
INSTRUCTIONS
At the usage location: remove surface dirt with a disposable cloth or paper towel.
Storage and transport: No special requirements. Avoid long periods before preparation because of the danger of drying out and corrosion.
Preparation for cleaning: Unscrew tattooing head and clamping nut, prepare needles separately (see preparation instructions for tattooing needles).
Remove dirt on the handpiece, tattooing head and clamping nut with a disposable cloth or paper towel.
Automatic cleaning and
disinfection:
Equipment: cleaning/disinfection device with special holder that connects handpieces to the cleaning/disinfection device and
ensures that the channels are rinsed. Rinse handpiece from the back. Use only neutral cleaners as cleaning agents, otherwise
the anodised surface will be damaged.
1) Insert handpiece in holder (it is important to ensure that the channels are rinsed), place tattooing head and
clamping nut in fine-mesh basket and close basket.
2) Set cleaning cycle with adequate cleaning and rinsing. Run last rinse cycle with fully deionised water.
3) Run 10-minute rinse at 93 °C to ensure thermal disi nfection.
4) When removing the parts inspect them for any remaining visible dirt in intermediate spaces and grooves. Repeat
cycle or clean manually if necessary.
Manual cleaning:
Equipment: neutral cleaning agent, soft brush, running water
Procedure:
1) Rinse and brush off surface dirt from handpiece, tattooing head and clamping nut.
2) Apply cleaning agent to all surfaces and intermediate spaces with brush.
3) Rinse handpiece, tattooing head and clamping nut thoroughly under running water for 10 minutes.
Manual disinfection: To disinfect handpiece, tattooing head and clamping nut manually wipe them with a cloth soaked in an ethanol-based
disinfectant solution.
Drying: If the cleaning/disinfection device does not have a drying program, the handpiece, tattooing head and clamping nut must be
dried manually or in a drying cabinet.
Inspection and
maintenance:
Inspect visually for damage, corrosion and wear. Mount clamping nut and tattooing head on handpiece and spray with NOU-
CLEAN SPRAY and wipe briefly with a moist cloth (see instructions on spray can).
Packaging: Single unit: pack handpiece in individual sterile packaging. The bag must be large enough so the seal is not under tension.
Sets: arrange handpieces in trays or place on all-purpose sterilisation trays.
Sterilisation: Autoclave in a vacuum autoclave at max. 135 °C for at le ast 5 minutes*. When sterilising more than one instrument in one
sterilisation cycle do not overload the steriliser. When using autoclaves without a post-vacuum a drying phase is required.
Allow handpiece to dry in the bag for at least 1 hour at room temperature with the paper side up.
* The temperature holding times must comply with the national and local directives and standards.
Storage: No special requirements. If the sterilised handpiece is not used immediately after sterilisation, the packaging must be labelled
with a sterile indicator and the date of sterilisation.
The above instructions have been classified by NOUVAG AG as suitable for preparation for re-use. The person doing the preparation is responsible for ensuring that the actual preparation with the equipment, materials and
personnel achieves the desired results. This normally requires validation and routine monitoring of the process. Every deviation from the provided instructions by the preparer must be carefully evaluated for its efficacy and any
possible negative consequences.
Faults and troubleshooting
Fault: Cause: Remedy:
Motor is running but the needle
does not move.
Handpiece is not correctly coupled to
motor.
Press handpiece to motor until it snaps into place.
Check that it is seated by pulling it lightly.
Needle does not run regularly. Needle is not optimally tensioned. Tighten clamping nut.
Service Center
Switzerland: Germany:
Manufacturer European Mandatory
NOUVAG AG NOUVAG GmbH
St. Gallerstrasse 23-25 • CH-9403 Goldach • Switzerland Schulthaißstrasse 15 • D-78462 Konstanz • Germany
Phone: ++41 (0)71 846 66 00 • Fax: ++41 (0)71 845 35 36 Phone: ++49 (0)7531 1290-0 • Fax: ++49 (0)7531 1290-12
E-mail: info@nouvag.com • www.nouvag.com E-mail info-de@nouvag.com • www.nouvag.com
USA:
NOUVAG USA Inc.
18058 Albyn Court • Lake Hughes • CA 93532 • USA
Phone: ++1 (661) 724 0217 • Fax: ++1 (661) 724 1590
E-mail: paul@nouvagusa.com • Toll Free: (800) 673 7427
If you require service, repairs or spare parts, please contact the dealer or representative in your country.
Instructions for disposal:
On disposal of the device, device components and accessories the national regulations for disposal must be observed.
GB
GB
GB
All NOUVAG
-
Service Centers worldwide pls. see at
www.nouvag.com
Table of contents
Other Nouvag Medical Equipment manuals

Nouvag
Nouvag HighSurg 30 User manual

Nouvag
Nouvag Dispenser DP 30 User manual

Nouvag
Nouvag MD 30 User manual

Nouvag
Nouvag Mesher User manual

Nouvag
Nouvag TCM 3000 BL User manual

Nouvag
Nouvag tcm endo v User manual

Nouvag
Nouvag md 20 User manual

Nouvag
Nouvag HighSurg 30 User manual

Nouvag
Nouvag Vacuson 60 LP User manual

Nouvag
Nouvag DP 30 LipoPlus User manual

Nouvag
Nouvag MD11 User manual

Nouvag
Nouvag Vacuson 60 User manual

Nouvag
Nouvag tcm 3000 User manual

Nouvag
Nouvag HighSurg 30 User manual

Nouvag
Nouvag MOS 5000 User manual

Nouvag
Nouvag Lipo Surg User manual

Nouvag
Nouvag Lipo Surg User manual

Nouvag
Nouvag TCM 3000 BL User manual

Nouvag
Nouvag OMS 5000 User manual

Nouvag
Nouvag OFA-Drill User manual