NRG Systems 11174 User manual

Soiling Measurement Kit
NRG SRA System | Kit #11174
Authors:
Technical Services

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 2
1 November 2017
TABLE OF CONTENTS
INTRODUCTION 4
MATERIALS & TOOLS 4
Inside the Kit................................................................................................................. 4
Required Tools.............................................................................................................. 4
SECTION 1: ASSEMBLING THE SOILING MEASUREMENT KIT 5
1.1 Open and unpack all items. ..................................................................................... 5
1.2: Mount upper U-bolt to tower, ................................................................................ 6
1.3: Mount PV panel assembly to upper U-bolt.............................................................. 7
1.4: Attach the angle adjustment boom to PV assembly. ............................................... 8
1.5: Mount the angle boom to the tower....................................................................... 9
1.6: Adjust the PV panel angle..................................................................................... 10
1.7: Secure & tighten all hardware. ............................................................................. 11
SECTION 2: WIRING AND MOUNTING THE INTERFACE MODULE 12
2.1: Connect the 4C cable. ........................................................................................... 12
2.2: Mount DIN rail to shelter box back plate............................................................... 13
2.3: Wire the PV panels to the interface modules. ....................................................... 13
2.4: Mount the IMT Tm-V-4090 Temperature Amplifier Boxes. .................................... 14

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 3
1 November 2017
2.5: Connect PV panel to the SymphoniePRO logger & iPack........................................ 14
2.6: Affix Interface Module to DIN rail & wire snesor signals to the logger wiring panel.15
2.7: Reconnect wiring panel to the logger.................................................................... 16
SECTION 3: LOGGER PROGRAMMING & CONFIGURATION 17
3.1: Programming the SymphoniePRO logger. ............................................................. 17
Temperature Clean: 17
Temperature Soiled: 18
IscClean: 18
IscSoiled: 19
3.1.2 Scale Factors and Offsets Summary 19
3.1.3 Saving the Channel Configuration 20
SITE MAINTENANCE 20
DATA PROCESSING 20

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 4
1 November 2017
INTRODUCTION
This document explains how to integrate the Soiling Measurement Kit onto NRG Systems’ SRA tower.
The process includes assembling and installing hardware to the tower and shelter box, wiring the system
to the logger, and configuring the logger channels to read the signals. The Soiling Measurement Kit can
be installed on any NRG Systems Solar Resource Assessment (SRA) tower equipped with a
SymphoniePRO logger.
It is recommended that two people perform this installation.
MATERIALS & TOOLS
Inside the Kit
PV panel assembly
Angle adjustment boom
U-bolts & mounting brackets
Interface Module, DIN rail, & wiring cable
Required Tools
3/16” hex key
5/32” hex key
7/16” combination wrench
3/4” combination wrench
Small flathead screwdriver
Small diagonal cutters or snips
6ft stepladder
Elevation angle meter
Compass or GPS (Required for initial SRA tower orientation, optional for panel orientation)
Electrical tape
Documentation equipment (Camera, pen, & paper)
Windows PC with SymphoniePRO Desktop Software loaded, USB cable type-A to type-B
Gloves
Loctite 242 (optional)
Digital Voltmeter (optional)
Wire strippers (optional)

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 5
19 December 2018
SECTION 1: ASSEMBLING THE SOILING MEASUREMENT KIT
1.1 Open and unpack all items.
Verify that all contents are present and undamaged. Leave PV panel assembly inside the box for now to
prevent the panels from getting scratched.
Figure 1-1: Packaged Soiling Measurement Kit

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 6
19 December 2018
1.2: Mount upper U-bolt to tower,
Mount upper U-bolt (Figure 1-2, right) and bracket near the top of the NRG SRA tower using the 3/4”
wrench (Figure 1-3). Make sure that the 7/16” bolt heads are facing down and the slide nuts are facing
upwards. Do not overtighten the U-bolt nuts.
Note: The lower U-bolt will be mounted to the tower approximately 15” below the upper U-bolt later.
Verify that there is space on the tower tube at this location for the lower U-bolt.
Figure 1-2: U-bolts and brackets, lower (left) & upper (right)
Figure 1-3: Mounting the upper U-bolt

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 7
19 December 2018
1.3: Mount PV panel assembly to upper U-bolt.
If desired, remove the slide nuts and add Loctite to the threads. Allow to dry slightly, then reattach slide
nuts to the bolts, leaving them loose.
Otherwise, loosen the slide nuts on the upper U-bolt bracket so they are flush with the end of the bolts
(Figure 1-4).
Figure 1-4: Slide nuts (circled) on the upper U-bolt bracket.
Remove the PV panel assembly from the box and slide the short, hinged section of mounting rail onto
the slide nuts (Figure 1-5). Use 2 people if possible.
Figure 1-5: Mounting PV panel assembly to bracket using slide nuts.

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 8
19 December 2018
Center PV panel assembly and tighten mounting bolts with the 7/16” wrench (Figure 1-6). Do not
overtighten bolts.
Figure 1-6
1.4: Attach the angle adjustment boom to PV assembly.
Attach angle boom to the lower mounting rail of PV assembly approximately 4.5” from the left side of
the rail (Figure 1-7).
The bolts holding the hinge to the end of the angle boom must be loosened slightly in order to pivot the
hinge. Loosen the slide nut on the end of the hinge enough to mount it to the rail. The slide nut is
inserted into the channel opposite of the PV panels. Once in position, tighten the bolt inside the hinge
with a 3/16” hex key. Allow the angle boom to pivot freely for the time being.
Figure 1-7

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 9
19 December 2018
1.5: Mount the angle boom to the tower.
Disassemble the lower U-bolt assembly but leave slide nut/bolt attched through the outer bracket
(Figure 1-8).
Figure 1-8
Remove the rubber angle boom end cap and insert slide nut/bolt/bracket into channel opposite the SRA
tower. Tighten bolt slightly to keep bracket at approximately 3” from the end of the hinge (Figure 1-9).
Figure 1-9
Measure 15” down the tower from the upper U-bolt. Mark measurement with permanent marker if
desired (Figure 1-10).

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 10
19 December 2018
Attach lower U-bolt to the tower at this location and 90 degrees clockwise of the upper U-bolt. Insert
inner bracket and spacers onto the ends of the U-bolt. Place outer bracket and angle boom through the
ends of the U-bolt (Figure 1-9). Tighten nuts with 3/4” wrench. Do not overtighten.
Figure 1-10: Location of lower U-bolt measured from the upper U-bolt.
1.6: Adjust the PV panel angle.
Adjust the panels to your desired angle using the angle boom and verify with angle finder (Figure 1-11).
If the boom does not slide freely, then bolts in the system may be too tight. Ensure that the bolts are
loose enough to allow movement.
Figure 1-11

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 11
19 December 2018
1.7: Secure & tighten all hardware.
Tighten down all hardware sufficiently enough to prevent any unwanted movement (Figures 1-12, 1-13,
1-14).
Figures 1-12, 1-13, 1-14: Tighten the bolts in the system.
The system is now ready to be wired to your SymphoniePRO data logger.

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 12
19 December 2018
SECTION 2: WIRING AND MOUNTING THE INTERFACE MODULE
The Soiling Measurment Kit comes with an Interface Module used to create usable data for your
SymphoniePRO data logger and mounts inside the FRP shelter box on the included DIN rail section.
Figure 2-1: Interface Module, cabling, DIN rail, mounting bolts & IMT Temp Sensor Instructions
2.1: Connect the 4C cable.
Attach the 4C cable to the Interface Module according to Table 2-1. The cable ends are labeled for
guidance.
Table 2-1
Wire Color
Interface Module Location
26-Channel Wiring Panel
16-Channel Wiring Panel
Black
GND
16-26
16, 17, 20-22
Red
EXC
16-19, 20-26
16, 17, 20-22
Orange
ISC CLEAN
16-19, 20-26
16, 17, 20-22
White
ISC SOILED
16-19, 20-26
16, 17, 20-22

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 13
19 December 2018
2.2: Mount DIN rail to shelter box back plate.
Unplug and remove the SymphoniePRO wiring panel cable.
Attach DIN rail to the interior of the FRP shelter box approximately 6” from the top using the 3M VHB
tape (Figure 2-2).
Figure 2-2: DIN rail mounted to shelter box backing plate.
If desired, the DIN rail can be affixed to the back plate using the supplied self-tapping screws.
2.3: Wire the PV panels to the interface modules.
Unroll the PV panel and route them down the SRA tower as desired and into the holes on the bottom of
the FRP shelter box.
Route through the interior of the shelter box as desired, then attach the ends to the appropriate
terminals on the Interface Module (Figure 2-3).
Note: The [Temp +] and [Temp –] terminals in the PV Clean and PV Soiled sections are left empty.
Figure 2-3: Completed Interface Module wiring.

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 14
19 December 2018
2.4: Mount the IMT Tm-V-4090 Temperature Amplifier Boxes.
The Clean and Dirty PV panels have IMT Tm-V-4090 temperature sensors mounted on them. The
amplifier boxes need to be mounted to the back side of the shelter box with adhesive tape.
1. Double sided tape is adhered to the back of the measuring amplifier box.
2. Unpeel the backing and mount the boxes to back right of the shelter box (if looking at it
straight on), as shown in Figure 2-4 below.
3. Hold the box in place for at least a minute to properly bond the box to the shelter box.
Note: the amplifier boxes should be mounted horizontally with a drip loop on either end
of the cable to avoid water ingress.
Figure 2-4: IMT Amplifier Box Installation
2.5: Connect PV panel to the SymphoniePRO logger & iPack.
Wire the PV panel labeled “Logger Power” to the PV terminals on the NRG iPack (Figure 2-5). The cable
can be routed as desired from the panel, down the tower, and into the FRP shelter box.
Figure 2-5: iPack charging circuit wiring.

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 15
19 December 2018
2.6: Affix Interface Module to DIN rail & wire snesor signals to the logger wiring panel.
Attach the wired Interface Module onto the DIN rail and connect the other end of the 4C cable to the
wiring panel. Connect the two IMT back of the panel temprature sensors to the wiring panel.
The Soiling Measurement Kit consists of four inputs:
Two 0-30 V temperatures inputs - Temperature ‘CLEAN’ & Temperature ‘SOILED’:
oEach input requires a #9135 P-SCM cards.
oThe sensor signals come from the IMT amplifier boxes.
oBoth sensors require 12v DC excitation voltage.
Two 0-5 V inputs PV inputs - IscCLEAN and IscSOILED:
oEach input requires a #9132 P-SCM cards.
oThe sensor signals come from the Interface Module.
oThe Isc outputs require a single 12V excitation (EXC) and ground connection (GND).
Table 2-2 provides an example of how the sensor wires can be connected to the wirng panel (any
combination of P-SCM channels can be used ~ 20-26).
Channel
Sensor Output
Wire Color
Terminal
21
Temprature
‘CLEAN’
Power +
Red
EXC
Power/Signal –
Thin Black
GND
Signal Output
Brown
SIG +
Sheild
Thick Black
SHD
22
Temprature
‘SOILED’
Power +
Red
EXC
Power/Signal –
Thin Black
GND
Signal Output
Brown
SIG +
Sheild
Thick Black
SHD
24
IscCLEAN
Interface Module Power*
Red
EXC
Interface Module Ground*
Black
GND
IscCLEAN PV Signal
Orange
SIG +
26
IscSOILED
IscSOILED PV Signal
White
SIG+
*Interface Module Power and Ground can be placed on either IscClean or IscSoiled channel
Table 2-2: Example Configuration

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 16
19 December 2018
Figure 2-6
2.7: Reconnect wiring panel to the logger.
Reattach your SymphoniePRO wiring panel cable. The Soiling Measurement Kit is now completely wired
and ready to be programed into your SymphoniePRO logger.

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 17
19 December 2018
SECTION 3: LOGGER PROGRAMMING & CONFIGURATION
Open SymphoniePRO Desktop Application, connect to the logger and navigate to the Channels tab in the
sidebar. Expand the desired channels for the Soiling Measurement Kit.
For the purposes of these instructions, channels 21, 22, 24, 26 are used.
3.1: Programming the SymphoniePRO logger.
Temperature Clean:
Expand Channel 21 and choose “Custom Analog 0-1V” from the “Load From Defaults” drop-
down menu.
Choose “P-SCM #9135, (0 to 30) V, SE Input, Constant 12V EXC” from the SymphoniePRO Signal
Conditioning Module (P-SCM)” drop-down menu.
Edit the Description, Scale Factor, Offset and Units as show in the image below:

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 18
19 December 2018
Temperature Soiled:
Expand Channel 22 and choose “Custom Analog 0-1V” from the “Load From Defaults” drop-
down menu.
Choose “P-SCM #9135, (0 to 30) V, SE Input, Constant 12V EXC” from the SymphoniePRO Signal
Conditioning Module (P-SCM)” drop-down menu.
Edit the Description, Scale Factor, Offset and Units as show in the image below:
IscClean:
Expand Channel 24 and choose “Custom Analog 0-5V” from the “Load From Defaults” drop-
down menu.
This will auto select “P-SCM #9132, (0 to 5) V, SE Input, Constant 12V EXC” from the
SymphoniePRO Signal Conditioning Module (P-SCM)” drop-down menu.
Edit the Description, Scale Factor, Offset and Units as show in the image below:

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 19
19 December 2018
IscSoiled:
Expand Channel 26 and choose “Custom Analog 0-5V” from the “Load From Defaults” drop-
down menu.
This will auto select “P-SCM #9132, (0 to 5) V, SE Input, Constant 12V EXC” from the
SymphoniePRO Signal Conditioning Module (P-SCM)” drop-down menu.
Edit the Description, Scale Factor, Offset and Units as show in the image below:
3.1.2 Scale Factors and Offsets Summary
Table 3-1
Signal Input
Channel
Input Type
Scale Factor
Offset
Units
TempCLEAN
21
Analog 0-30 V
13
-40
Deg. C
IscCLEAN
24
Analog 0-5 V
0.3125
0
A
TempSOILED
22
Analog 0-30 V
13
-40
Deg. C
IscSOILED
26
Analog 0-5 V
0.3125
0
A

NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
support@nrgsystems.com | Page 20
19 December 2018
3.1.3 Saving the Channel Configuration
When all the channel configurations have been edited and are ready, click the “Save To Logger” button
at the top of the Channel Configuration page.
Note: It is a good idea to save the .LGR configuration file on your computer as well as to the logger. Go
to the Logger drop down menu in the upper right hand corner of the application screen and select
“Save Instrument Configuration File.”
SITE MAINTENANCE
Back-of-panel temperature and short-circuit current for each panel are continuously measured by the
system, and (typically) averaged every one minute.
The clean panel should be regularly cleaned thoroughly with deionized water and a soft non-abrasive
cloth (about once a week or as recommended by your analyst). Deionized water is free from calcium
and salt so it will not leave deposits on your PV panels.
Certain sites may have very abrasive mineral deposits, so be mindful of abrasion when cleaning the PV
panels.
Rainfall also has a natural cleaning effect and should also be recorded at the site with an instrument
such as a tipping rain bucket connected to the same data logger.
DATA PROCESSING
The measured values (Temp CLEAN, Isc CLEAN, Temp SOILED, Isc SOILED) can be used to determine a
Soiling Ratio which compares the soiled PV output to the clean PV output. By monitoring the soiling
ratio over time and using additional analysis techniques, the potential impact of soiling losses on a PV
farm’s energy production can be characterized.
For a quick check, the simplest of methods is to divide the soiled PV Isc by the clean PV Isc. Note that
other, more complex methods are often utilized.
Table of contents
Other NRG Systems Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Fluke
Fluke DSX-600 CableAnalyzer Troubleshooting handbook

HOKUYO AUTOMATIC
HOKUYO AUTOMATIC UHG-08LX instruction manual

ATAGO
ATAGO PAL-27S instruction manual

Lovibond
Lovibond Photometer-System MD100 instruction manual

Rlide
Rlide R2 Series manual

TSI Incorporated
TSI Incorporated Airmet VELOCICALC 5725 Operation and service manual