manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hioki
  6. •
  7. Measuring Instrument
  8. •
  9. Hioki LR8400-92 User manual

Hioki LR8400-92 User manual

About the Waveform/Values screens
PV Wave + Value screen
Estimate Power screen
Equations
w1: Power (kW) = Voltage (Ch. 1-1) × current (Ch. 1-2)
w2: Power integration (kWh) = Voltage (Ch. 1-1) × current (Ch. 1-2) × measurement time (h)
w3: Estimated power (kW) = Insolation strength (Ch. 1-3) / Gs (standard test condition of 1 kW/m2) × (1 + temperature loss*1) ×
(1 - coefficient for other losses) × photovoltaic cell rated output (kW)
w4: Estimated integration (kWh) = Insolation strength (Ch. 1-3) / Gs (standard test condition of 1 kW/m2) × (1 + temperature loss*1)
× (1 - coefficient for other losses) × photovoltaic cell rated output (kW) × measurement time (h)
w5: AC estimated power (kW) = Estimated power (w3) × power conditioner conversion efficiency (reference value)
w6: AC estimated integration (kWh) = Estimated energy (w4) × power conditioner conversion efficiency (reference value)
w7: Estimated power approximation rate (%) = Energy (w2) / estimated energy (w4)
* 1 Temperature loss = Photovoltaic cell maximum output temperature coefficient × (panel temperature (Ch. 1-4) - 25) / 100
* The estimated energy is calculated for reference purposes using JIS C8907:2005, “Estimation Method of Generating Electric Energy
by PV Power System.”
* In the state of "PV mode Set ON", the range of ch1-1 to ch1-4 and w1 to w7, scaling, an upper and lower limit, and a comment are
set up automatically. Moreover, the range of ch1-5 to ch1-11 set up on the "extended channel set" screen, scaling are also set up
automatically.
Note
The results of calculations performed by the instrument to yield estimated power, estimated integration, estimated power
approximation rate, AC estimated power (reference values), and AC estimated integration (reference values) are easily influenced by
the manner in which the target solar panels were installed as well as weather conditions and panel degradation. Calculated values
should be considered as guidelines for solar panel maintenance, rather than guaranteed values.
Measurement Procedure
1. Select “PV Mode Set ON” on the Measurement Settings screen.
-1. Press the “SET” key and select “PV Mode Set ON” on the
Measurement Settings screen.
2. Connect the sensors to the LR8400-23.
-1. Ch.1-1:
•Connect the BNC conversion cable(2.4m) to the BNC side of the 9322 Differential Probe and connect
the 9804-01 Magnet Adapter (red) and the 9804-02 Magnet Adapter (black) to the ends.
•Connect the BNC conversion cable to Ch.1-1.
•Connect the power cord to the power jack on the 9322 Differential Probe.
•Connect the power cord’s positive wire to the +12 V jack on the external I/O terminal block on the LR8400-23, and similarly
connect the power cord’s negative wire to the GND jack.
-2. Ch.1-2:
•Connect another BNC conversion cable(30cm) to the CT9692-90 (or CT9693-90) Clamp Sensor and connect it to Ch.1-2.
•Set the CT6590 Sensor Unit’s range (H or L).
•Check the AAbatteries in the CT6590 Sensor Unit.
-3. Ch.1-3:
•Connect the actinometer (EKO Instruments ML-01 or ML-020VM).
-4. Ch.1-4:
•Connect the thermocouple.
3. Place the sensors (see Figures 1 and 2).
-1. Ch.1-1:
•Connect the 9804-01 (red) and 9804-02 (black) magnet adapters that are connected to the tip of the 9322 Differential Probe to the
positive and negative terminals of the main DC line in the junction box or the electrical box (or when investigating a failed string,
the branch circuit) between the solar panels and the power conditioner. Cut the magnet sheet, pasted on the back of the 9322
Differential Probe.
-2. Ch.1-2:
•Clamp the CT9692-90 (or CT9693-90) Clamp Sensor to the positive side of the main DC line in the junction box or electrical box
(or when investigating a failed string, the branch circuit). At this time, orient the current direction mark on the clamp sensor so that
it indicates a flow from the solar panels to the power conditioner.
-3. Ch.1-3:
•Attach the actinometer (EKO Instruments ML-01 or ML-020VM) to the solar panel. So that it is parallel with the measurement
surface using the adjustment screw.
-4. Ch.1-4:
•Attach the thermocouple to the solar panel.
LR8400-92
LR8400-93
PV Power Verifier
Instruction Manual (PV Mode)
December 2013 Revised edition 1
Printed in Japan
LR8400E981-01 13-12H
Thank you for choosing the Hioki LR8400-92/
LR8400-93. This Measurement Guide introduces the
basic method for measuring the DC side of a
photovoltaic system. For more information about
standard methods for using the instrument in
applications other than the above and general handling
information, please see the LR8400-20 Instruction
Manual.
Turn on the instrument.
1. Select “PV Mode Set ON” on the Settin
g
s screen.
2. Connect the sensors to the LR8400-23.
3. Place the sensors (when investigating a failed string, change the measurement
c
ir
cu
i
t
in
s
i
de
t
h
e
ju
n
ct
i
o
n
bo
x
).
4. Confi
g
ure settin
g
s.
5. Check measured values on the Monito
r
screen.
6. Start time-series recordin
g
.End measurement.
End recordin
g
.
Check and anal
y
ze saved data.
Turn off the instrument. 7. Investi
g
ate the failed strin
g
.
Measurement preparations
-1. Inspect the instrument.
-2. Insert a CF card (or USB flash drive)
into the instrument as necessary.
-3. Connect the AC adapter.
-4. Turn on the instrument.
1
4
Figure 1
Figure 2
The instrument provides the following
six Waveform/Values screens:
•PV Wave + Value
•PV Wave + Crsr
•Gauge + PV Wave
•PV Value
•Estimate Power
•AC Estimate (reference values)
4. Configure settings.
-1. Select the “PV Mode Set ON” setting on the Measurement Settings screen.
Settings Settin
g
Descri
p
tio
n
Solar cell rated output
[kW] Enter the photovoltaic cell rated output or photovoltaic cell capacity for the solar panel being
measured. If measuring a branch circuit, for example in order to investigate a failed string, enter the
total rated out
p
ut for the tar
g
et
p
hotovoltaic cell onl
y
.
Cell temperature
coefficient
(maximum output)
Photovoltaic cells have negative temperature characteristics, meaning that their output decreases as
their surface temperature increases, and this coefficient expresses that relationship. This tendency is
particularly pronounced in crystalline silicon. Enter the appropriate value after checking the
specifications for the photovoltaic system. Following are reference values: Monocrystalline Si: -0.46;
p
ol
y
cr
y
stalline Si: -0.49; amor
p
hous Si: -0.26; CI
(
G
)
S: -0.46
Conditioner efficiency To display the AC estimated power (re
f
erence values) and AC estimated energy (reference values) on a
PV Waveform/Values screen, consult the power conditioner specifications and enter the device’s
conversion efficiency. The above value is subject to a larger error component in photovoltaic systems
that boost the voltage on the power conditioner side so that it is higher than the voltage at the
measurement
p
oint.
Other loss coefficient Coefficient for the effects of factors such as wiring resistance, dirt or snow buildup on the panel
surface, shadows, etc. A
g
eneral avera
g
e value of 0.07 is used b
y
default. Chan
g
e as necessar
y
.
Pyranomete
r
kind
Pyranometer output rate Choose the pyranometer kind. Output rate initial value of the ML-01 is 20.00. Output rate initial value
of the ML-020VM is 140.00. When Sensitivity indicated on the label of Pyranometer is A, the
calculated value of 1000/A is set to out
p
ut rate. Other
p
y
ranometer use Free Set.
Clamp sensor range Select the same range as set for the CT6590 Sensor Unit. Next, press the “Monitor” key on the LR8400
while the clamp is not applied to anything and adjust the 0ADJ knob on the CT6590 so that the “ch1-2
current” is 0.
PV mode initialize Resets the following settings to their default values: Solar cell rated output, Cell temperature
coefficient, Conditioner efficiency, Other loss coefficient Pyranometer kind, Pyranometer output rate,
and Clam
p
sensor ran
g
e.
(
Please
p
erform, when
y
ou initialize PV mode setu
p
.
)
Extended channel set An input can be added to ch1-5 to ch1-11.Voltage, Thermo couple, and a clamp (2000A, 200A, 100A,
20A, 10A) can be chosen from the input item of the channel which select check mark. (Please use it,
when
y
ou increase the measurement channel in PV mode.
)
Interval Set the time interval at which to record data. The longe
r
the time interval, the lower the cutoff
frequency, increasing noise rejection effectiveness. It is recommended to increase the recording
interval in environments that are
p
rone to noise from inverters or other devices.
Cont Set whether to perform measurement for the set time or continuously.
OFF: Normal measurement (measurement for the specified recording time)
ON: Continuous measurement
Time/DIV Set the time
p
er division.
Filte
r
Set to either 60 Hz or 50 Hz depending on the power supply frequency in your region. The longer the
recordin
g
interval, the
g
reater the noise re
j
ection effectiveness.
Timer Set when you wish to perform measurement regularly. This setting allows the instrument to perform
measurement at a re
g
ular interval from the set start time to the set sto
p
time.
Auto save Set when you wish to save data automatically during measurement. For more information, see “Saving
Data Automaticall
y
” in Section 6.2, “Savin
g
Data,” of the LR8400 User Manual.
Note
・For photovoltaic systems that boost the voltage on the power conditioner side so that it is higher than the value at the junction box
where the measurement is being taken, select channel number 1-1 on the Setting screen’s “CH” sheet and enter a voltage value as
appropriate for the “Offset” setting under “Scaling” so that the voltage shown on the LR8400-23’s Monitor screen matches the
photovoltaic system monitor’s DC voltage value.
・The waveform data measured in "PV mode Set ON" LR8400-23 can be displayed by attached Logger Utility. However, "PV mode"
cannot be set up by Logger Utility.
5. Check measured values on the Monitor screen.
-1. If you encounter any of the following displays on the Monitor screen, refer to the indicated solution.
Dis
p
la
y
Solutio
n
Erroneous voltage value •Check whether the 9322 Differential Probe switch is set to “DC.”
• Check whether the voltage range has been set to take into account the fact that the differential
probe’s output ratio is 1/1,000.
• For photovoltaic systems that boost the voltage on the power conditioner side so that it is higher
than the value at the junction box where the measurement is being taken, select channel number
1-1 on the Setting screen’s “CH” sheet and enter a voltage value as appropriate for the “Offset”
setting under “Scaling” so that the voltage shown on the LR8400’s Monitor screen matches the
photovoltaic system monitor’s DC voltage value.
• Check whether the solar panels are susceptible to the effects of factors such as shadow, dirt, or
snow buildup.
• There ma
y
be a wirin
g
break on a strin
g
or a module failure.
Erroneous current value •Check the CT6590 Sensor Unit’s range setting.
• Check the LR8400’s clamp range setting.
• Check the adjustment of the 0ADJ knob on the CT6590 Sensor Unit.
• Check the CT6590 Sensor Unit’s remaining battery life.
• Check for noise sources such as inverters in the vicinity of the system.
• Check the site to see if the solar panels are susceptible to the effects of factors such as shadow,
dirt, or snow buildup.
• There ma
y
be a wirin
g
break on a strin
g
or a module failure.
Erroneous insolation value •Check the wiring connections.
• Check for noise sources such as inverters in the vicinity of the system.
• Check the kind of
py
ranometer and out
p
ut rate to be used.
Erroneous
p
anel temperature
value •Check the installation of the thermocouple.
•Check for a broken thermocouple connection.
* Turning on the broken connection check on the “CH”sheet will cause “Burn Out”to be
displayed for the measurement data. However, the broken connection check function cannot be
used if the recordin
g
interval is 10 ms.
Significant difference between
power and estimated power •Check the solutions for erroneous voltage, current, insolation, and panel temperature values
above.
• Check whether the PV specification setup have been set to appropriate value. In particular,
double-check the photovoltaic cell rated output and clamp range settings as these are easy to
overlook.
• This error is more likely to occur in climates and time zones characterized by low insolation.
Perform measurement during a time of high insolation.
• There may be a wiring break on a string or a module failure. Pinpoint the location of the failure,
for exam
p
le b
y
usin
g
an I/V tracer.
6. Start time-series recording.
If performing time-series recording, start recording. If the auto-save setting has been activated, data will be recorded simultaneously
to the instrument’s internal memory and the specified CF card or USB flash drive.
-1. Start measurement with the Start key.
-2. Stop measurement with the Stop key.
Note
The continuous operating time for the AAalkaline batteries
in the CT6590 Sensor Unit that ships with the CT9692-90
(or CT9693-90) Clamp Sensor is approximately 25 hours.
When performing time-series recording, use the optional
9445-02 AC Adapter.
7. Investigate the failed string. (See Figure 3.)
If the power value is significantly less than the estimated
power, a solar panel may have failed. The instrument can be
used to look for a string with a wiring break or a module
failure by means of the following method.
-1. Return to Step 3, “Place the sensors,” and connect the
9322 Differential Probe and CT9692-90 (or CT9693-90) Clamp Sensor
to the branch circuit as shown in Figure 3. Next, change the
photovoltaic cell rated output on the Measurement Settings
screen to the output value for the string being measured and
check the voltage, current, and output values on the Monitor screen.
Perform similar measurements for the other strings. Circuits with
relatively lower values may indicate wiring breaks or module failures.
-2. Without changing the settings made in (-1) above, cover the solar cell that you expect may be damaged with a piece of cloth, board,
or other material and compare the resulting voltage, current, and energy values with the corresponding values before the panel was
covered. Strings whose measured values do not change when they are covered may have a module failure or wiring break.
-3. The instrument can also be used as a normal data logger by using the “PV Mode Set OFF” setting. Up to 30 channels of current
data can be recorded by purchasing additional CT9692-90 (or CT9693-90) Clamp Sensors and BNC conversion cables. Current data
for multiple strings can be recorded simultaneously, allowing current variations to be investigated.
Note
•The instrument cannot be used to search for failures or wiring breaks within individual modules. For such applications, please use a
thermal camera or dedicated checker.
32
Figure 3
Note: The version of the
software until the V9.12, it was
ML-020 only rather than
selection of pyranometer kind.
From V9.13, please use it to
select "ML-01", "ML-020", and
"Free Set" pyranometer kind
used as shown in the figure on
the left. (In the system screen,
you can check the version.)
4. Configure settings.
-1. Select the “PV Mode Set ON” setting on the Measurement Settings screen.
Settings Settin
g
Descri
p
tio
n
Solar cell rated output
[kW] Enter the photovoltaic cell rated output or photovoltaic cell capacity for the solar panel being
measured. If measuring a branch circuit, for example in order to investigate a failed string, enter the
total rated out
p
ut for the tar
g
et
p
hotovoltaic cell onl
y
.
Cell temperature
coefficient
(maximum output)
Photovoltaic cells have negative temperature characteristics, meaning that their output decreases as
their surface temperature increases, and this coefficient expresses that relationship. This tendency is
particularly pronounced in crystalline silicon. Enter the appropriate value after checking the
specifications for the photovoltaic system. Following are reference values: Monocrystalline Si: -0.46;
p
ol
y
cr
y
stalline Si: -0.49; amor
p
hous Si: -0.26; CI
(
G
)
S: -0.46
Conditioner efficiency To display the AC estimated power (re
f
erence values) and AC estimated energy (reference values) on a
PV Waveform/Values screen, consult the power conditioner specifications and enter the device’s
conversion efficiency. The above value is subject to a larger error component in photovoltaic systems
that boost the voltage on the power conditioner side so that it is higher than the voltage at the
measurement
p
oint.
Other loss coefficient Coefficient for the effects of factors such as wiring resistance, dirt or snow buildup on the panel
surface, shadows, etc. A
g
eneral avera
g
e value of 0.07 is used b
y
default. Chan
g
e as necessar
y
.
Pyranomete
r
kind
Pyranometer output rate Choose the pyranometer kind. Output rate initial value of the ML-01 is 20.00. Output rate initial value
of the ML-020VM is 140.00. When Sensitivity indicated on the label of Pyranometer is A, the
calculated value of 1000/A is set to out
p
ut rate. Other
p
y
ranometer use Free Set.
Clamp sensor range Select the same range as set for the CT6590 Sensor Unit. Next, press the “Monitor” key on the LR8400
while the clamp is not applied to anything and adjust the 0ADJ knob on the CT6590 so that the “ch1-2
current” is 0.
PV mode initialize Resets the following settings to their default values: Solar cell rated output, Cell temperature
coefficient, Conditioner efficiency, Other loss coefficient Pyranometer kind, Pyranometer output rate,
and Clam
p
sensor ran
g
e.
(
Please
p
erform, when
y
ou initialize PV mode setu
p
.
)
Extended channel set An input can be added to ch1-5 to ch1-11.Voltage, Thermo couple, and a clamp (2000A, 200A, 100A,
20A, 10A) can be chosen from the input item of the channel which select check mark. (Please use it,
when
y
ou increase the measurement channel in PV mode.
)
Interval Set the time interval at which to record data. The longe
r
the time interval, the lower the cutoff
frequency, increasing noise rejection effectiveness. It is recommended to increase the recording
interval in environments that are
p
rone to noise from inverters or other devices.
Cont Set whether to perform measurement for the set time or continuously.
OFF: Normal measurement (measurement for the specified recording time)
ON: Continuous measurement
Time/DIV Set the time
p
er division.
Filte
r
Set to either 60 Hz or 50 Hz depending on the power supply frequency in your region. The longer the
recordin
g
interval, the
g
reater the noise re
j
ection effectiveness.
Timer Set when you wish to perform measurement regularly. This setting allows the instrument to perform
measurement at a re
g
ular interval from the set start time to the set sto
p
time.
Auto save Set when you wish to save data automatically during measurement. For more information, see “Saving
Data Automaticall
y
” in Section 6.2, “Savin
g
Data,” of the LR8400 User Manual.
Note
・For photovoltaic systems that boost the voltage on the power conditioner side so that it is higher than the value at the junction box
where the measurement is being taken, select channel number 1-1 on the Setting screen’s “CH” sheet and enter a voltage value as
appropriate for the “Offset” setting under “Scaling” so that the voltage shown on the LR8400-23’s Monitor screen matches the
photovoltaic system monitor’s DC voltage value.
・The waveform data measured in "PV mode Set ON" LR8400-23 can be displayed by attached Logger Utility. However, "PV mode"
cannot be set up by Logger Utility.
5. Check measured values on the Monitor screen.
-1. If you encounter any of the following displays on the Monitor screen, refer to the indicated solution.
Dis
p
la
y
Solutio
n
Erroneous voltage value •Check whether the 9322 Differential Probe switch is set to “DC.”
• Check whether the voltage range has been set to take into account the fact that the differential
probe’s output ratio is 1/1,000.
• For photovoltaic systems that boost the voltage on the power conditioner side so that it is higher
than the value at the junction box where the measurement is being taken, select channel number
1-1 on the Setting screen’s “CH” sheet and enter a voltage value as appropriate for the “Offset”
setting under “Scaling” so that the voltage shown on the LR8400’s Monitor screen matches the
photovoltaic system monitor’s DC voltage value.
• Check whether the solar panels are susceptible to the effects of factors such as shadow, dirt, or
snow buildup.
• There ma
y
be a wirin
g
break on a strin
g
or a module failure.
Erroneous current value •Check the CT6590 Sensor Unit’s range setting.
• Check the LR8400’s clamp range setting.
• Check the adjustment of the 0ADJ knob on the CT6590 Sensor Unit.
• Check the CT6590 Sensor Unit’s remaining battery life.
• Check for noise sources such as inverters in the vicinity of the system.
• Check the site to see if the solar panels are susceptible to the effects of factors such as shadow,
dirt, or snow buildup.
• There ma
y
be a wirin
g
break on a strin
g
or a module failure.
Erroneous insolation value •Check the wiring connections.
• Check for noise sources such as inverters in the vicinity of the system.
• Check the kind of
py
ranometer and out
p
ut rate to be used.
Erroneous
p
anel temperature
value •Check the installation of the thermocouple.
•Check for a broken thermocouple connection.
* Turning on the broken connection check on the “CH”sheet will cause “Burn Out”to be
displayed for the measurement data. However, the broken connection check function cannot be
used if the recordin
g
interval is 10 ms.
Significant difference between
power and estimated power •Check the solutions for erroneous voltage, current, insolation, and panel temperature values
above.
• Check whether the PV specification setup have been set to appropriate value. In particular,
double-check the photovoltaic cell rated output and clamp range settings as these are easy to
overlook.
• This error is more likely to occur in climates and time zones characterized by low insolation.
Perform measurement during a time of high insolation.
• There may be a wiring break on a string or a module failure. Pinpoint the location of the failure,
for exam
p
le b
y
usin
g
an I/V tracer.
6. Start time-series recording.
If performing time-series recording, start recording. If the auto-save setting has been activated, data will be recorded simultaneously
to the instrument’s internal memory and the specified CF card or USB flash drive.
-1. Start measurement with the Start key.
-2. Stop measurement with the Stop key.
Note
The continuous operating time for the AAalkaline batteries
in the CT6590 Sensor Unit that ships with the CT9692-90
(or CT9693-90) Clamp Sensor is approximately 25 hours.
When performing time-series recording, use the optional
9445-02 AC Adapter.
7. Investigate the failed string. (See Figure 3.)
If the power value is significantly less than the estimated
power, a solar panel may have failed. The instrument can be
used to look for a string with a wiring break or a module
failure by means of the following method.
-1. Return to Step 3, “Place the sensors,” and connect the
9322 Differential Probe and CT9692-90 (or CT9693-90) Clamp Sensor
to the branch circuit as shown in Figure 3. Next, change the
photovoltaic cell rated output on the Measurement Settings
screen to the output value for the string being measured and
check the voltage, current, and output values on the Monitor screen.
Perform similar measurements for the other strings. Circuits with
relatively lower values may indicate wiring breaks or module failures.
-2. Without changing the settings made in (-1) above, cover the solar cell that you expect may be damaged with a piece of cloth, board,
or other material and compare the resulting voltage, current, and energy values with the corresponding values before the panel was
covered. Strings whose measured values do not change when they are covered may have a module failure or wiring break.
-3. The instrument can also be used as a normal data logger by using the “PV Mode Set OFF” setting. Up to 30 channels of current
data can be recorded by purchasing additional CT9692-90 (or CT9693-90) Clamp Sensors and BNC conversion cables. Current data
for multiple strings can be recorded simultaneously, allowing current variations to be investigated.
Note
•The instrument cannot be used to search for failures or wiring breaks within individual modules. For such applications, please use a
thermal camera or dedicated checker.
32
Figure 3
Note: The version of the
software until the V9.12, it was
ML-020 only rather than
selection of pyranometer kind.
From V9.13, please use it to
select "ML-01", "ML-020", and
"Free Set" pyranometer kind
used as shown in the figure on
the left. (In the system screen,
you can check the version.)
About the Waveform/Values screens
PV Wave + Value screen
Estimate Power screen
Equations
w1: Power (kW) = Voltage (Ch. 1-1) × current (Ch. 1-2)
w2: Power integration (kWh) = Voltage (Ch. 1-1) × current (Ch. 1-2) × measurement time (h)
w3: Estimated power (kW) = Insolation strength (Ch. 1-3) / Gs (standard test condition of 1 kW/m2) × (1 + temperature loss*1) ×
(1 - coefficient for other losses) × photovoltaic cell rated output (kW)
w4: Estimated integration (kWh) = Insolation strength (Ch. 1-3) / Gs (standard test condition of 1 kW/m2) × (1 + temperature loss*1)
× (1 - coefficient for other losses) × photovoltaic cell rated output (kW) × measurement time (h)
w5: AC estimated power (kW) = Estimated power (w3) × power conditioner conversion efficiency (reference value)
w6: AC estimated integration (kWh) = Estimated energy (w4) × power conditioner conversion efficiency (reference value)
w7: Estimated power approximation rate (%) = Energy (w2) / estimated energy (w4)
* 1 Temperature loss = Photovoltaic cell maximum output temperature coefficient × (panel temperature (Ch. 1-4) - 25) / 100
* The estimated energy is calculated for reference purposes using JIS C8907:2005, “Estimation Method of Generating Electric Energy
by PV Power System.”
* In the state of "PV mode Set ON", the range of ch1-1 to ch1-4 and w1 to w7, scaling, an upper and lower limit, and a comment are
set up automatically. Moreover, the range of ch1-5 to ch1-11 set up on the "extended channel set" screen, scaling are also set up
automatically.
Note
The results of calculations performed by the instrument to yield estimated power, estimated integration, estimated power
approximation rate, AC estimated power (reference values), and AC estimated integration (reference values) are easily influenced by
the manner in which the target solar panels were installed as well as weather conditions and panel degradation. Calculated values
should be considered as guidelines for solar panel maintenance, rather than guaranteed values.
Measurement Procedure
1. Select “PV Mode Set ON” on the Measurement Settings screen.
-1. Press the “SET” key and select “PV Mode Set ON” on the
Measurement Settings screen.
2. Connect the sensors to the LR8400-23.
-1. Ch.1-1:
•Connect the BNC conversion cable(2.4m) to the BNC side of the 9322 Differential Probe and connect
the 9804-01 Magnet Adapter (red) and the 9804-02 Magnet Adapter (black) to the ends.
•Connect the BNC conversion cable to Ch.1-1.
•Connect the power cord to the power jack on the 9322 Differential Probe.
•Connect the power cord’s positive wire to the +12 V jack on the external I/O terminal block on the LR8400-23, and similarly
connect the power cord’s negative wire to the GND jack.
-2. Ch.1-2:
•Connect another BNC conversion cable(30cm) to the CT9692-90 (or CT9693-90) Clamp Sensor and connect it to Ch.1-2.
•Set the CT6590 Sensor Unit’s range (H or L).
•Check the AAbatteries in the CT6590 Sensor Unit.
-3. Ch.1-3:
•Connect the actinometer (EKO Instruments ML-01 or ML-020VM).
-4. Ch.1-4:
•Connect the thermocouple.
3. Place the sensors (see Figures 1 and 2).
-1. Ch.1-1:
•Connect the 9804-01 (red) and 9804-02 (black) magnet adapters that are connected to the tip of the 9322 Differential Probe to the
positive and negative terminals of the main DC line in the junction box or the electrical box (or when investigating a failed string,
the branch circuit) between the solar panels and the power conditioner. Cut the magnet sheet, pasted on the back of the 9322
Differential Probe.
-2. Ch.1-2:
•Clamp the CT9692-90 (or CT9693-90) Clamp Sensor to the positive side of the main DC line in the junction box or electrical box
(or when investigating a failed string, the branch circuit). At this time, orient the current direction mark on the clamp sensor so that
it indicates a flow from the solar panels to the power conditioner.
-3. Ch.1-3:
•Attach the actinometer (EKO Instruments ML-01 or ML-020VM) to the solar panel. So that it is parallel with the measurement
surface using the adjustment screw.
-4. Ch.1-4:
•Attach the thermocouple to the solar panel.
LR8400-92
LR8400-93
PV Power Verifier
Instruction Manual (PV Mode)
December 2013 Revised edition 1
Printed in Japan
LR8400E981-01 13-12H
Thank you for choosing the Hioki LR8400-92/
LR8400-93. This Measurement Guide introduces the
basic method for measuring the DC side of a
photovoltaic system. For more information about
standard methods for using the instrument in
applications other than the above and general handling
information, please see the LR8400-20 Instruction
Manual.
Turn on the instrument.
1. Select “PV Mode Set ON” on the Settin
g
s screen.
2. Connect the sensors to the LR8400-23.
3. Place the sensors (when investigating a failed string, change the measurement
c
ir
cu
i
t
in
s
i
de
t
h
e
ju
n
ct
i
o
n
bo
x
).
4. Confi
g
ure settin
g
s.
5. Check measured values on the Monito
r
screen.
6. Start time-series recordin
g
.End measurement.
End recordin
g
.
Check and anal
y
ze saved data.
Turn off the instrument. 7. Investi
g
ate the failed strin
g
.
Measurement preparations
-1. Inspect the instrument.
-2. Insert a CF card (or USB flash drive)
into the instrument as necessary.
-3. Connect the AC adapter.
-4. Turn on the instrument.
1
4
Figure 1
Figure 2
The instrument provides the following
six Waveform/Values screens:
•PV Wave + Value
•PV Wave + Crsr
•Gauge + PV Wave
•PV Value
•Estimate Power
•AC Estimate (reference values)

This manual suits for next models

1

Other Hioki Measuring Instrument manuals

Hioki CM4141-90 User manual

Hioki

Hioki CM4141-90 User manual

Hioki 8846 User manual

Hioki

Hioki 8846 User manual

Hioki CM4001 User manual

Hioki

Hioki CM4001 User manual

Hioki 3270 User manual

Hioki

Hioki 3270 User manual

Hioki Memory HiCorder MR6000 User manual

Hioki

Hioki Memory HiCorder MR6000 User manual

Hioki 3532-50 User manual

Hioki

Hioki 3532-50 User manual

Hioki RM3545A-1 User guide

Hioki

Hioki RM3545A-1 User guide

Hioki LUX HiTESTER 3423 User manual

Hioki

Hioki LUX HiTESTER 3423 User manual

Hioki 3470 User manual

Hioki

Hioki 3470 User manual

Hioki 3661-20 User manual

Hioki

Hioki 3661-20 User manual

Hioki 8730-10 User manual

Hioki

Hioki 8730-10 User manual

Hioki CKT3548 User manual

Hioki

Hioki CKT3548 User manual

Hioki CM4371-50 User manual

Hioki

Hioki CM4371-50 User manual

Hioki IM7585 Series Use and care manual

Hioki

Hioki IM7585 Series Use and care manual

Hioki SAFETY HiTESTER 3258 Instruction manual

Hioki

Hioki SAFETY HiTESTER 3258 Instruction manual

Hioki CM4373-50 User manual

Hioki

Hioki CM4373-50 User manual

Hioki 3390 User manual

Hioki

Hioki 3390 User manual

Hioki IM3533-01 Use and care manual

Hioki

Hioki IM3533-01 Use and care manual

Hioki PR8111 User manual

Hioki

Hioki PR8111 User manual

Hioki PW3335 User manual

Hioki

Hioki PW3335 User manual

Hioki 3196 User manual

Hioki

Hioki 3196 User manual

Hioki CM4141-50 User manual

Hioki

Hioki CM4141-50 User manual

Hioki CM4141 User manual

Hioki

Hioki CM4141 User manual

Hioki 3274 User manual

Hioki

Hioki 3274 User manual

Popular Measuring Instrument manuals by other brands

PCB Piezotronics 3711E1150G Installation and operating manual

PCB Piezotronics

PCB Piezotronics 3711E1150G Installation and operating manual

Futech TERACROSS manual

Futech

Futech TERACROSS manual

Extech Instruments ET38 user manual

Extech Instruments

Extech Instruments ET38 user manual

BIRD SiteHawk SK-6000-TC quick start guide

BIRD

BIRD SiteHawk SK-6000-TC quick start guide

POSITECTOR Powder Checker instructions

POSITECTOR

POSITECTOR Powder Checker instructions

LOVATO ELECTRIC DMED301 instruction manual

LOVATO ELECTRIC

LOVATO ELECTRIC DMED301 instruction manual

Pembina BP1100 Series user manual

Pembina

Pembina BP1100 Series user manual

Velleman DVM1307 user manual

Velleman

Velleman DVM1307 user manual

Procentec ComBricks HE+ user manual

Procentec

Procentec ComBricks HE+ user manual

Deif MIB 8000 Designers handbook

Deif

Deif MIB 8000 Designers handbook

Trotec BE42 Original instructions

Trotec

Trotec BE42 Original instructions

Eaton Power Xpert PXM 4000 quick start guide

Eaton

Eaton Power Xpert PXM 4000 quick start guide

Siemens SITRANS FUP1010 quick start

Siemens

Siemens SITRANS FUP1010 quick start

Shodex OHpak SB-2000 Series Operation manual

Shodex

Shodex OHpak SB-2000 Series Operation manual

Spectrum 3401 product manual

Spectrum

Spectrum 3401 product manual

Renkforce 1404109 operating instructions

Renkforce

Renkforce 1404109 operating instructions

Weidmüller 2608C operating instructions

Weidmüller

Weidmüller 2608C operating instructions

PCE Instruments PCE-ECT 50 user manual

PCE Instruments

PCE Instruments PCE-ECT 50 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.