Cole Parmer Jenway 76 Series User manual

Spectrophotometer
7410,7415 and 7615
Instruction Manual
JEN0001 / Version 1.2

4
Contents
Section 1 - Introduction ................................................................................. 6
1.1 General Description....................................................................................................... 6
1.2 Important Safety Advice................................................................................................ 6
1.3 SymbolsDened........................................................................................................... 6
1.4 Electrical Requirements ................................................................................................. 7
Section 2 - Installation................................................................................... 8
2.1 Unpacking .................................................................................................................... 8
2.2 Installation Conditions................................................................................................... 8
2.3 Overview...................................................................................................................... 9
Section 3 - Theory and Practice of Spectroscopy Measurement ....................
10
3.1 Theory of Spectroscopy Measurement ............................................................................ 10
3.2 Spectroscopy Measurement ........................................................................................... 11
3.3 Good Practice Guidelines ............................................................................................... 12
Section 4 - Instrument Set up........................................................................
14
4.1 Start up Screen............................................................................................................. 14
4.2 Navigation.................................................................................................................... 14
4.3 Methods....................................................................................................................... 16
4.4 Results......................................................................................................................... 16
4.5 Settings ....................................................................................................................... 16
4.5.1 Instrument Status ................................................................................................... 16
4.5.1.1 Instrument Language ........................................................................................ 16
4.5.1.2 Setting the Time ............................................................................................... 17
4.5.1.3 Setting the Date................................................................................................ 17
4.5.1.4 Setting the Time Zone= 18
4.5.2 Measurement Settings............................................................................................. 18
4.5.3 Network Connections .............................................................................................. 19
4.5.4 Storage .................................................................................................................. 19
4.5.5 Service Settings ...................................................................................................... 19
Section 5 - Photometrics ................................................................................
20
5.1 Method Set up.............................................................................................................. 20
5.1.1 Selecting a Wavelength ........................................................................................... 20
5.2 Calibration.................................................................................................................... 21
5.3 Sample Measurement.................................................................................................... 21
Section 6 - Concentration...............................................................................

5
22
6.1 Method Set up.............................................................................................................. 22
6.1.1 Selecting a Wavelength ........................................................................................... 22
6.1.2 Using a Factor......................................................................................................... 22
6.1.3 Using a Standard .................................................................................................... 23
6.1.4 Selecting Concentration Units................................................................................... 23
6.2 Calibration.................................................................................................................... 23
6.2.1 Calibrating to a Factor ............................................................................................. 24
6.2.2 Calibrating to a Standard......................................................................................... 24
6.3 Sample Measurement.................................................................................................... 25
6.3.1 Measuring a Sample After Calibrating to a Factor....................................................... 25
6.3.2 Measuring a Sample After Calibrating to a Standard .................................................. 25
Section 7 - Spectrum ......................................................................................
26
7.1 Method Set up.............................................................................................................. 26
7.1.1 Setting Start and End Wavelengths........................................................................... 26
7.1.2 Setting the Scan Interval ......................................................................................... 27
7.1.3 Selecting Absorbance or % Transmittance................................................................. 27
7.2 Calibration.................................................................................................................... 27
7.3 Sample Measurement.................................................................................................... 27
7.4 Data Analysis................................................................................................................ 28
7.4.1 Peaks and Valleys.................................................................................................... 28
7.4.2 Area Under Curve ................................................................................................... 29
7.4.2.1 Area Under Curve - Baseline Mode ..................................................................... 29
7.4.2.2 Area Under Curve - Tangent Mode...................................................................... 30
Section 8 - Quantitation .................................................................................
31
8.1 Method Set up.............................................................................................................. 31
8.1.1 Selecting a Wavelength ........................................................................................... 31
8.1.2 Selecting Number of Replicates................................................................................ 32
8.1.3 Selecting Concentration Units................................................................................... 32
8.2 Measuring Calibration Standards .................................................................................... 33
8.3 Standard Curve............................................................................................................. 34
8.4 Sample Measurement.................................................................................................... 35
Section 9 - Kinetics.........................................................................................
36
9.1 Method Set up.............................................................................................................. 36
9.1.1 Selecting a Wavelength ........................................................................................... 36
9.1.2 Setting the Kinetics Measurement Time .................................................................... 37

6
9.1.3 Setting the Measurement Time Interval .................................................................... 37
9.1.4 Setting Lag Time..................................................................................................... 37
9.1.5 Selecting Absorbance or % Transmittance................................................................. 37
9.1.6 End Point Concentration .......................................................................................... 37
9.2 Calibration.................................................................................................................... 38
9.3 Sample Measurement.................................................................................................... 38
9.4 Data Analysis................................................................................................................ 40
Section 10 - Multi-Wavelength.......................................................................
41
10.1 Method Set up.............................................................................................................. 41
10.1.1 Selecting a Wavelength ........................................................................................... 41
10.1.2 Equation Parameters ............................................................................................... 42
10.1.2.1 Entering a Factor............................................................................................... 43
10.1.2.2 Selecting Concentration Units............................................................................. 43
10.1.3 Selecting Absorbance or % Transmittance................................................................. 43
10.2 Calibration.................................................................................................................... 44
10.3 Sample Measurement.................................................................................................... 44
Section 11 - Saving, Loading, Deleting and Printing......................................
45
11.1 Saving Methods ............................................................................................................ 45
11.1.1 Saving Methods to Internal Memory ......................................................................... 45
11.1.2 Saving Methods to CPLive........................................................................................ 45
11.1.3 Saving Methods to USB Memory Stick....................................................................... 45
11.2 Loading Methods .......................................................................................................... 45
11.2.1 Loading Methods from Internal Memory.................................................................... 45
11.2.2 Loading Methods from USB Memory Stick ................................................................. 45
11.3 Deleting Methods.......................................................................................................... 45
11.4 Saving Results .............................................................................................................. 46
11.4.1 Saving Results to Internal Memory ........................................................................... 46
11.4.2 Saving Results to CPLive.......................................................................................... 46
11.4.3 Saving Results to USB Memory Stick......................................................................... 46
11.5 Loading Results ............................................................................................................ 46
11.5.1 Loading Results from Internal Memory ..................................................................... 46
11.5.2 Loading Results from USB Memory Stick................................................................... 46
11.6 Deleting Results............................................................................................................ 46
11.7 Printing ........................................................................................................................ 46
Section 12 - Accessories and Spare Parts ......................................................
47
12.1 Optional Accessories ..................................................................................................... 47

7
12.2 Installing the Accessories .............................................................................................. 47
12.2.1 Passive Accessories ................................................................................................. 47
12.2.2 Active Accessories................................................................................................... 48
12.2.2.1 Installing Automatic 8 Cell Turret........................................................................ 49
12.3 Using the Accessories.................................................................................................... 50
12.3.1 Automatic 8 Cell Turret ............................................................................................ 50
12.3.1.1 Automatic 8 Cell Turret - Manual Mode ............................................................... 50
12.3.1.2 Automatic 8 Cell Turret - Automatic Mode ........................................................... 52
12.3.1.3 Automatic 8 Cell Turret - Supporting Creation of a Standard Curve in Quantitation . 54
12.4 Spare Parts................................................................................................................... 54
Section 13 - Maintenance, Servicing and Cleaning ........................................
55
13.1 Routine Maintenance..................................................................................................... 55
13.2 Lamp Replacement ....................................................................................................... 57
13.2.1 Tungsten Halogen Lamp Replacement ...................................................................... 57
13.2.2 Xenon Lamp Module Replacement............................................................................ 58
13.3 Service, Repairs and Support ......................................................................................... 58
13.4 Warranty...................................................................................................................... 58
Section 14 - Environmental Protection ..........................................................
58
14.1 Packaging Material........................................................................................................ 58
14.2 Waste Electrical and Electronic Equipment Directive (WEEE) ............................................ 58
Section 15 - Technical Specication...............................................................
59
15.1 GeneralSpecication..................................................................................................... 59
15.2 Weights and Dimensions ............................................................................................... 60
Section 16 - Troubleshooting .........................................................................
61
Section 17 - Glossary of Icons........................................................................
63
Index ..............................................................................................................
64
Declaration of Conformity ..............................................................................
67

8
Section 1 - Introduction
Thank you for purchasing this Jenway product. To get the best performance from the equipment, and
for your own safety, please read these instructions carefully before use.
If the equipment is not used in the manner described in this manual and with accessories other than
those recommended by the manufacturer, the protection provided may be impaired.
1.1 General Description
The 7410, 7415 and 7615 spectrophotometers are suited to a wide range of applications in education, quality
control, environmental and clinical analysis. The 7410 is a visible spectrophotometer covering a wavelength
range from 320nm to 1000nm. The 7415 and 7615 are UV/Visible spectrophotometers with a wavelength
range from 198nm to 1000nm. All models have measurement modes for Photometrics, Concentration, Spec-
trum scanning, Multi-wavelength, Quantitation and Kinetics. All models are compatible with CPLive.
1.2 Important Safety Advice
Users should be aware of the following safety advice:
● SHOCKHAZARDSORUNSAFEPRACTICESAREDANGEROUSastheycancauseseverepersonalinjury,re
or death.
● DONOTusecombustiblesubstancesnearhotobjects.
● DONOTusetheequipmentinhazardousatmospheres.
● DONOToperateorhandleanypartoftheequipmentwithwethandsoruseonsurfacesthatmaybecomeooded.
● NEVERmovetheequipmentwhilestillconnectedtothepowersupply.
● HIGHTEMPERATURESAREDANGEROUSastheycancauseseriousburnstooperatorsandignitecombustible
material.
● USECAREANDWEARPROTECTIVEGLOVESTOPROTECTHANDS.
● NEVERliftorcarrytheequipmentduringoperation.
● DONOTpositiontheequipmentunitsothatitisdifculttodisconnectfromthemainssupplyusingthemains
plug.
● Themainsoutletsocketusedshouldbelocatedclosetotheequipmentandreadilyidentiableandaccessible
to users.
● DONOTleaveequipmentswitchedonanditisnotrecommendedtoleaveanyheatingapparatusunattended
during operation.
● Theequipmentshouldbecarriedusingbothhands.
1.3 Symbols Dened
WARNINGEARTH RISK OF
ELECTRIC
SHOCK
BIOHAZARD HOT
SURFACE
INFORMATION UV LIGHT
SOURCE

9
1.4 Electrical Requirements
THIS INSTRUMENT MUST BE GROUNDED
Before connection please ensure that the line supply corresponds to the power requirements below:
Power Supply requirements
65W 100V-230V~50/60Hz
The equipment is provided with a power supply unit and three power cables consisting of a UK 3-pin and a “Schuko”
2-pin plug for 230 V installations and a NEMA 5-15 plug for 120 V installations.
Choose the power cable appropriate for your electrical installation and discard the others. Should one of the power
cables be suitable for connecting to the power supply, replace the plug with a suitable alternative..
THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A QUALIFIED ELECTRICIAN.
NOTE: Refer to the equipment rating plate to ensure that the plug and fusing are suitable for the voltage and
wattage stated. The wires in the mains cable are as follows:
230 V a.c. 120 V a.c.
HOT/LIVE - BROWN BLACK - HOT/LIVE
NEUTRAL - BLUE WHITE - NEUTRAL
EARTH - GREEN/YELLOW GREEN - EARTH
The appropriate power cable and power adaptor combination should be connected to the equipment BEFORE
connection to the mains supply. Should the mains lead require replacement, a cable of 1.25mm² (AWG16) of
harmonised code H05VV-F connected to an IEC320 plug should be used.
IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN

10
Section 2 - Installation
2.1 Unpacking
Before discarding the packaging check
that all parts are present and correct.
1 7410 / 7415 / 7615
2 Instruction manual
3 UK power lead
4 EU power lead
5 US power lead
6 Power supply unit
2.2 Installation Conditions
Whentheequipmentisusedforthersttimeormovedtoadifferentenvironmentaltemperature,itis
important to allow the equipment to equalise to the ambient temperature. We recommend you allow
the equipment to stand for 2 hours before switching on.
This equipment is designed to operate safely under the following conditions:
● For indoor use only
● Use in a well ventilated area
● Ambient temperature range 5°C to 40°C (41°F to 104°F)*
● Altitude to 2000m (6500 ft)
● Relative humidity not exceeding 80% (temperature 31ºC) decreasing to 50% (temperature 40ºC) and
free from condensation
● Mainssupplyuctuationsnotexceeding10%ofnominal
● Overvoltage category
● Pollution degree 2
● Use with a minimum distance all round of 300mm (12in.) from walls or other items
Placetheequipmentonaclean,rm,levelsurfacewhichisfreefromdrafts.Avoidinstallationonaslipperysurface
oronasurfacepronetovibrationoronasurfacepronetoooding.
Select the power lead and attach to the power supply unit. Connect the power supply unit to the power inlet
socket on the rear panel of the instrument and connect to the mains socket. Ensure that the sample chamber is
empty before turning the power on at the mains and switching the instrument on using the power switch on the
side of the instrument.
The equipment will perform several power-on tests and wavelength calibration before displaying the main screen.
NOTE: 7410 must be switched on 30 minutes before any readings are performed to allow the tungsten lamp to
warm up
NOTE: Leaving cuvettes in the sample holder during power up will result in failure of the power on tests.
345
1 2
6

11
2.3 Overview
Main View
1 Lid
2 Open and close catch
3 Colour touchscreen
and user interface
4 Removable panel
for sipper accessory
5 2 x USB Type A ports
6 On/Off power switch
Rear View
7 Ethernet (RJ-45) port
8 USB Type B port
9 Power inlet socket
Underneath View
10 Lamp holder cover
Refer to Section 13.2 for instructions on how to remove and replace the lamp.
1
2
3
4
10
5
6
7
4
8
9

12
Section 3 - Theory and Practice of Spectroscopy Measurement
3.1 Theory of Spectroscopy Measurement
UV-visiblespectroscopyisthemeasurementoftheabsorbanceoflightataspecicwavelengthinasample.Thisis
used to identify the presence and concentration of molecular entities within the sample. The Beer-Lambert law is
used to relate the absorption of light to the properties of the sample through which the light is travelling through.
The Beer-Lambert law states that:
A = Ɛ l c
A is the absorbance
Ɛ isthemolarabsorptioncoefcient(lmol-1cm-1)
l is the path length (cm)
c is the concentration (mol I-1)
This law shows that absorbance is linear to concentration but this is only true for low concentrations. For
absorbance levels above 3 the concentration starts to move away from the linear relationship.
Transmittance is the proportion of the light which passes through the sample:
Where:
lois the incident light
l
tis the transmitted light
l is the path length
Therefore:
Absorbance is inversely related to transmittance:
Io
I
It
Io
It
T =
T
Log 1
A =
Io
I
It
Io
I
t
T =
T
Log 1
A =
Io
I
It
Io
It
T =
T
Log 1
A =

13
3.2 Spectroscopy Measurement
There are four main components of a spectrophotometer. These are a light source to emit a high and constant
amount of energy over the full wavelength range; a method for separating the light into discreet wavelengths; a
sample holder and a light detector.
The optical layout of the 7410 and 7415 spectrophotometers is shown below:
The light from the pre-focused tungsten halogen (7410) or pre-aligned xenon (7415) lamp is focused onto
the grating, with 1200 lines per millimetre, which separates the light into discrete wavelengths. The diffracted
spectrum of light then passes through a further slit and lens arrangement before passing through the sample in
the sample chamber from left to right. The light which is not absorbed by the sample is transmitted through a
collecting lens and onto the signal detector. The photo-diode detector used is mounted directly onto the detector
PCB and the output is used to calculate the % transmittance. The result is displayed either as % transmittance or
absorbance on the instrument display.
Note:
On the 7615 model an additional detector at the exit slit detects the incident light and noise through the
optical system
Exit Slit
Entrance Slit
Lamp
Grating
Collimator Mirror
Detector
Collecting Lens
Sample

14
3.3 Good Practice Guidelines
1. For optimum performance all spectrophotometers should be sited in a clean, dry, dust free atmosphere. When
in use ambient temperature and light levels should remain as constant as possible.
2. If required adherence to Standard Operating Procedures (SOP) and Good Laboratory Practice (GLP) should be
monitored with regular calibration checks and a suitable Quality Control (QC) programme.
3. The sample chamber lid must be fully closed during measurement and before any readings are recorded or
printed.
4. The correct selection of sample containers is imperative for accurate and reproducible results:
a) Check that the material of the sample container is compatible with the wavelengths to be used for
measurement. In general glass can only be used down to 360nm or 320nm depending on quality.
Standard plastic cuvettes can be used down to 320nm. Special UV versions can be used down to 260nm.
Belowthislevelquartzcuvettesmustbeused.
b) Plastic disposable cuvettes should only be used ONCE.
c) Glass cuvettes should be thoroughly cleaned after use. Discard when scratches become evident on optical
surfaces.
d) Care should be taken when selecting semi-micro or micro cuvettes. The cuvette window on the inner
chamber(thearealledwithsample)mustbewiderthantheapertureinthesampleholderorlightwill
reach the detector without passing through the sample. In this case, semi-micro or micro cuvettes with
self-screening black surrounds must be used or, alternative holders for these cuvettes should be used.
e) Glass test tubes and other sample tubes should be used with care. Where possible, matched tubes should
be used and any index mark set to the correct position before measurements are made.
f) Ensure any sample containers used are compatible with the constituents of both the samples and
standards they are to hold. Plastic cuvettes are not compatible with organic solvents.
g) All sample containers must be handled with care; by the top, bottom and non-optical surfaces only. Any
ngermarksevidentmustberemovedbyasuitablecleaningprocess.
h) Flow-through cuvettes must be selected with care and consideration for the sample type, sample volume,
pumping system, rinse, sample and waste handling to be used.
5. Samples and standards should not be stored in open cuvettes or sample containers as evaporation will change
the value and lead to staining of the walls which may be irreversible. If stored in stoppered and sealed
cuvettes,theyshouldbelledwithlittleornoairspaceandthevaluesregularlycheckedagainstareference
standard or quality control material.
6. Samples should be allowed to equilibrate to ambient temperature before measurement (unless a suitable
temperature controlled sample holder is in use). Temperature change during measurement may cause air
bubbles to form on the walls of the sample holder. This is a common cause of drift during measurement.
7. In the preparation of samples and standards high grade borosilicate glass and AR grade chemicals and
reagents must be used. Good quality deionised water or other suitable solvents must be used for dissolving or
diluting samples, chemicals and reagents.
8. All measurements require calibration to a blank, for maximum accuracy this should be prepared with care
using the same deionised water or solvent used for dissolving or diluting the sample. Where reagents are
added to the sample to produce a colour proportional to its concentration a ‘sample based’ blank should be
used. In this case the blank should consist of all reagents or chemicals to be used, except the sample which
will produce the colour to be measured.

15
9. Deviations from the Beer-Lambert Law may occur at high and low concentrations giving non-linear response
duringsampleconcentrationmeasurements.Forallnewmethodsalinearrangeshouldbedenedbythe
preparation of a calibration curve. The quantitation mode may be used to construct such a curve against
which sample results are automatically measured.
10.Cuvettesandsampleholdersmustbelledtoaminimumlevel whichcoversthelightpath.AllJenway
spectrophotometers have a beam height of 15mm.
11.Theinstrumentmustbecalibratedtozeroabsorbance/100%transmittancepriortotakingreadings.Inthe
spectrum measurement mode a baseline scan must be performed before performing a sample scan.

16
Section 4 - Instrument Set up
4.1 Start up Screen
The power up screen is shown below:
4.2 Navigation
The main menu is shown below:
These spectrophotometers are controlled solely through the touchscreen interface of the equipment and follow a
basic Android user interface. If the number of options available in a menu exceeds the number that can be
displayed on the screen, swipe to the left to view the other modes.

17
The main menu screen provides access to all Measurement modes, Methods, Results and the Settings menu.
Additional icons are displayed when the unit is connected to a network and if an active accessory is installed.
The settings menu enables access to Instrument status, Measurement settings, Network connections
Storage and Service settings.
Throughout, the software options can be turned ON and OFF using a switch:
Option is ON
Option is OFF
Ineachmeasurementmodethereisanoverowicon giving additional save and load method options.
Touching gives options to load a previously saved method, touch to
save the entered method parameters, touch to upload to CPLive* or touch
to clear recently used method parameters. *Refer to CPLive instructions for more informa-
tion.
When required to enter numbers, a keypad will pop up. Touch the required numbers and touch to apply. To
exit the keypad without changing the entered value touch ortheminimizeicon .
When required to enter letters, a keypad will pop up. Touch the required letters and touch Save to apply. To
exit the keypad without changing the entered value touch Save .
Save

18
4.3 Methods
Touch to access methods that have been saved. Touch the required method to view the details of the
method set up. You will then be able to delete, upload, export, edit or run the selected method. See section 11 for
more information.
4.4 Results
Touch to access results that have been saved. Touch the required results to view the details of the
result. You will then be able to delete, upload or export the selected result. See section 11 for more informa-
tion.
4.5 Settings
Touch to enable access to instrument status, measurement settings, network connections, storage
and service settings.
4.5.1 Instrument Status
Touch toviewthestatusofthespectrophotometer,checkconnections,ttedaccessories,
the serial number of the unit and the software version the instrument is using. The language and date and time
can also be set here.
4.5.1.1 Instrument Language
ThesoftwarecanbeviewedinvedifferentlanguageswithachoiceofEnglish,French,German,Spanishor
Italian. To select the required language touch and select from the menu. Touch next to the
required language to apply.

19
4.5.1.2 Setting the Time
To set the instrument time touch .Touch and move the clock hand to the correct hour position,
repeat the same process for minutes, select AM or PM and touch to apply. Touching will return to the
instrument settings screen without altering the time. The set time will be displayed in each measurement
mode and will be recorded against saved methods and results.
4.5.1.3 Setting the Date
To set the instrument date touch . Scroll up or down to change the month viewed. Touch the
required date and touch to apply. To set the year touch and scroll up or down and touch the
required date and touch to apply. Touching will return to the instrument settings screen
without altering the date. The set date will be displayed in each measurement mode and will be recorded
against

20
4.5.1.4 Setting the Time Zone
Tosettheinstrumenttimezonetouch . Scroll up or down to locate the required time
zoneandtouch nexttotherequiredtimezonetoapply.
4.5.2 Measurement Settings
Touch to select options for autosave results and the favourites panel. Results can be
saved automatically to the instrument’s internal memory by sliding the button to the
position. To view the favourites panel on the home screen slide the button to the position.
If auto-save is not selected, then results need to be saved manually. Refer to section 11.4 for more details.

21
4.5.3 Network Connections
Touch to view available network connections. Options include Ethernet (RJ-45), IP
conguration,connectionstatusandCPLive.
4.5.4 Storage
Touch to view the amount of available storage on the internal memory of the spectrophotometer. If a
USB memory stick is inserted the amount of free space on the USB stick will also be shown.
4.5.5 Service Settings
Service settings is for service engineers only
Touch to view the serial number and options to change serial number and to open windchime.

22
Section 5 - Photometrics
The photometrics measurement mode enables simple measurements of absorbance and % transmittance to be
performed. The sample is measured at one wavelength and at one point in time. There are no post measurement
calculations available in this measurement mode. Touch the Photometrics icon on the main menu to enter this
measurement mode.
5.1 Method Set up
Thismeasurementmodeisverysimpleandtheonlyparameterwhichcanbeadjustedisthewavelength.Oncethe
required wavelength has been entered a calibration can be performed.
5.1.1 Selecting a Wavelength
Toadjustthewavelength,touch and use the keypad to enter the required wavelength. Touch
to apply the entered wavelength and return to the method set up.
This manual suits for next models
8
Table of contents
Other Cole Parmer Measuring Instrument manuals

Cole Parmer
Cole Parmer 81150-48 User manual

Cole Parmer
Cole Parmer Melting Point S MP-800D Series User manual

Cole Parmer
Cole Parmer 1100 Series User manual

Cole Parmer
Cole Parmer Jenway 7415 NANO User manual

Cole Parmer
Cole Parmer SP-300 Series User manual

Cole Parmer
Cole Parmer P100 User manual

Cole Parmer
Cole Parmer 1200 Series User manual

Cole Parmer
Cole Parmer Techne FSC400D User manual

Cole Parmer
Cole Parmer Stuart SMP11 User manual

Cole Parmer
Cole Parmer Jenway 7415 User manual