Oakton Temp-16 User manual

Instruction Manual
Temp-16
Precision RTD Thermometer
68X481904 Rev 0 09
/
0
9
Part of Thermo Fisher Scientific


TABLE OF CONTENTS
Quick
Start
Guide............................................ 1
1. Introduction.............................................. 2
2. Safety Precautions................................... 3
3. Specifications........................................... 4
4. Battery Installation And Replacement ...... 6
5. Inserting And Removing Optional
Rubber Armour ........................................ 7
6.
Assembling Optional
Handsfree
Accesories
............................................... 8
7.
Connecting
An Rtd Probe......................... 9
8. Key Functions.......................................... 10
9. Display Overview ..................................... 11
10. Measurement Mode ................................. 12
11. Selecting Temperature Scale................... 13
12. Min And Max Function............................. 14
13. Auto Off Function..................................... 14
14. Calibration ............................................... 15
15. Clear Field Calibration............................. 16
16. Field Calibration Unlock........................... 16
17.
Field Calibration Lockout
.......................... 17
18. Maintenance ............................................ 18
19.
Cleaning
................................................... 18
20. Troubleshooting ....................................... 19
21. Service..................................................... 20
22. Replacement Meter & Accessories .......... 20
23. Warranty .................................................. 21
24.
Product Return
......................................... 21
25. Innocal® Calibration And Repair
Services (North America)........................ 22
26. Technical Assistance ............................... 25


1
QUICK
START
GUIDE
To ensure best results please read the
complete manual.
1. Connect probe.
2. Press the on/off key.
Press again to turn on backlight.
3. Select °C or °F display using °C/°F key.
4. Insert probe in sample.
Allow adequate time for reading to
stabilize.
Response time will vary depending on
probe.
5. Take reading.
Press hold key to freeze display.
Press min/max to display minimum or
maximum temperature.
6. Press and hold the on/off key for three
seconds to turn meter off.

2
1. INTRODUCTION
Thank you for choosing the
Temp-16 Precision RTD
Thermometer.This versatile
hand-held instrument
provides highly accurate
temperature measurements
and is designed for easy
operation which includes the
following features:
•Operator selection of Celsius or
Fahrenheit scale
•Resolution of 0.1° (auto ranging)
•Four-digit backlit LCD
•Hold feature to retain a reading
temporarily
•Field calibration capability
•Low battery warning
•Auto power off with Enable/Disable
function
•Min and max readings display
•3-pin circular connector input
•Operates with a wide selection of
probes

3
2. SAFETY PRECAUTIONS
Voltages present at the RTD may also be
present at the battery terminals. Always
disconnect the RTD when changing
batteries.
W
WARNIN
G
This instrument is designed to accept low
level signals supplied by standard 100 Ω
RTDs. Under no circumstances should the
input voltage exceed the specified 10 v rms.
1. Do not use or store this instrument in
microwave ovens or any abnormally hot
or cold areas.
2. Weak batteries should not be left in the
instrument. Dead batteries can leak and
cause damage to unit.

4
3. SPECIFICATIONS
Platinum RTD Probes
100 ohm platinum: alpha = 0.003850
–200 °C to 850 °C (–392 °F to 1562 °F)
Out of range display: “Ur”, “Or”, “OPEN”
Range Accuracy
–200.0 to –100.0 °C
(–392 to –148 °F)
±2.0 °C
(±4.0 °F)
–99.9 to 199.9 °C
(–148 to 392 °F)
±0.2 °C
(±0.4 °F)
200.0 to 850.0 °C
(392 to 1562 °F)
±2.0 °C
(±4.0 °F)
Resolution 0.1° (–200.0 to 850.0 °C);
0.1° (–392.0 to 999.9 °F);
1° (1000 to 1562 °F)
Display 4-digit LCD (¼ X ½" digits);
2 1/3X 1 ½" backlit viewing
area
Display Update
Rate 0.5 sec per update.
Input One probe with 3 pin
circular connector
(Switchcraft TA3F)
Input Protection 10V rms
Battery Three AA, 1.5V alkaline
Battery Life 700 hours continuous
Auto Shutoff 17 minutes after last
keypress.
Stated Accuracy 18 to 28°C
(64 to 82°F)
Useful Range 0 to 40°C
(32 to 104°F)
Storage -40 to 65°C
(-40 to 149°F)
Humidity 10% to 90%
(non-condensing)
Dimensions
Without armour
With armour
(L x W x H )
175 mm x 97 mm x 42 mm
180 mm x 102 mm x 52 mm

5
Weight
Without armour
With armour
267 g
362 g
Ingress Protection
(With Probe
Attached)
Meets IEC-529 IP-54 for dust
and water-resistant
enclosures.
CE Compliance EN61326-1/A1:
1998 (EU EMC Directive)

6
4. BATTERY INSTALLATION AND
REPLACEMENT
Voltages present at the RTD may also be
present at the battery terminals. Always
disconnect the RTD when changing
batteries.
Weak batteries should not be left in the
instrument. Dead batteries can leak and
cause damage to unit.
The typical battery life is about 700 hours.
Selected settings are stored in memory and
will remain in memory even after power
is turned off, or while batteries are being
replaced.
1. Before
changing
batter
y
, tur
n
instrume
nt off
and
disconnect probe
.
2.
Loosen
screw
and
lift batter
y
cover
off
the
back
of case.
3.
Remove
the t
hree
AA batteries.
4. Insert three
new
batteri
es
observing
pol
arit
y
.
5. Install
cover
and
tig
h
ten
screw
.

7
5. INSERTING AND REMOVING
OPTIONAL RUBBER ARMOUR
Figure 1: Removing Optional Rubber Armour
1. To insert thermometer into the optional
rubber armor, slide in from the top of
meter before pushing the bottom edges
of meter down to set it into position. Lift
up the stand at the back of meter for
bench top applications if necessary.
2. To remove thermometer from armor,
push out from the bottom edges of
meter until it is completely out of boot.

8
6. ASSEMBLING OPTIONAL
HANDSFREE
ACCESORIES
You can use the optional magnets and strap
in the Hands Free Kit accessories for hands-
free operations.
Figure 2: Hands Free Kit

9
7.
CONNECTING
AN RTD PROBE
Use the correct 100 ΩRTD (alpha =
0.003850) for your instrument. Using an
incorrect probe type will result in erroneous
readings. Insert the 3-pin plug into the
mating connector on the top of the
instrument.
Figure 3: Top view of RTD Thermometer
RTD Input

10
8. KEY FUNCTIONS
Figure 4: Keypad
°C/°F
Toggles between Celsius or
Fahrenheit in measurement mode
and user calibration mode.
cal
Press and hold this key for 3 seconds
to enter Calibration mode if field
calibration is unprotected.
hold
Activates/ Deactivates freezing of the
measured reading while in
measurement mode.
min
max
Toggles between minimum and
maximum readings (MIN and MAX
readings are calculated from the last
min/max keypress).
Press and hold this key for 3
seconds to clear the MIN/MAX
reading stored.
on/off
light
Powers on and shuts off the
thermometer by holding it for 3
seconds.
Press this key to activate/ deactivate
the backlight display. (Backlight will
automatically turns off within 30
seconds of activation).

11
9. DISPLAY OVERVIEW
Figure 5: Meter Display
°C/°F Celsius or Fahrenheit indicator.
MIN Minimum reading annunciator.
MAX Maximum reading annunciator.
3 Bar : 700 – 550 hours
2 Bar : 350 – 550 hours
1 Bar : 150 – 350 hours
Blinking: < 150 hours
CAL
Blinks during calibration mode.
Remains in display indicating field
calibration is active.
HOLD Remain in display during hold mode.
STABLE Displayed upon recognizing final
value.

12
10. MEASUREMENT MODE
Press the on/off key. The thermometer
performs a self-test and all display digits and
indicators should remain on for
approximately two seconds before the meter
enters measurement mode.
If a probe is not connected or if the probe is
defective, the display will indicate “OPEN”.
°c
Figure 6: Display when no probe is detected
For optimum instrument accuracy, allow one
minute for ambient temperature stabilization.
If the unit has been stored at an extreme
ambient condition, more time will be needed.
A STABLE indicator appears at the top right
corner of the display when reading is
continuously stable for 5 seconds.
To freeze reading, press hold. Press hold
key again to release the reading and resume
normal operation.
24.6
°c
HOLD STABLE
Figure 7: HOLD and STABLE function
The display will show “Ur” (under range) or
“Or” (over range) if the temperature reading
is out of range of the instrument.

13
11. SELECTING TEMPERATURE
SCALE
24.6
°c
75.4
°F
Figure 8: Press °C/°F key to toggle
between °C and °F
Select °C
or
°F
by
pressing
the °C/°F
key
.
Eac
h
tim
e
the
key
is
pressed,
th
e
temperatur
e
scale
will
sw
itch. S
w
itc
hing
bet
ween
°
C
and
°F
can
be
done
at
any
time
during
operatio
n.
Each time
you
turn the
in
strument on, it
will
power
up
wi
t
h
the
same
setting
s
th
a
t were
set when the unit was last
turned
off.
°C/°F

14
12. MIN AND MAX FUNCTION
Press the min/max key to toggle between
the minimum and maximum readings. The
minimum and maximum reading function is
ideal for monitoring unattended operations
while continually displaying every
temperature change that occurs. The
minimum and maximum values are sensed
and automatically stored. To exit and clear
the MIN/MAX reading stored, press and hold
the min/max key for 3 seconds.
13. AUTO OFF FUNCTION
The thermometer has a default auto off
function of 17 minutes.
To deactivate this function, press the
min/max key while turning the thermometer
on with the on/off key. “A.OFF no” flashes
to show indicate that the auto off function
has been disabled for that session.
A.off
Figure 9: Disabling the auto off function

15
14. CALIBRATION
The CAL function allows single point
calibration of the thermometer at 0°C (32°F)
to compensate for probe offset error. Use
the field calibration feature to improve
thermometer/probe accuracy.
The Temp-16 Precision RTD Thermometer
comes with a calibration lock function to
prevent accidental changes to calibration
settings. A field calibration can only be
performed when the function has been
unlocked. To unlock field calibration, follow
steps 1 through 4 in FIELD CALIBRATION
UNLOCK section.
To calibrate:
1. Pack sensing end of probe in a
container tightly packed with crushed ice
and filled with distilled water. Allow
temperature to stabilize.
2. Press and release the cal key for 3
seconds to enter the calibration mode.
The CAL indicator starts blinking.
Release the cal key.
3. If the measured temperature is from
-5°C to 5°C (23 to 41°F) when the
temperature reading is stable, press the
cal key.
The CAL indicator will stop blinking, and the
reading will be set to 0°C (32°F). The CAL
indicator will remain in display, indicating
that field calibration is now active.
If “Err” is displayed, the displayed reading is
outside the above limits.
NOTE: Press any key (except the cal key)
while in calibration mode to abort
calibration.

16
15. CLEAR FIELD CALIBRATION
To clear user calibration:
1. Turn the thermometer off.
2. Hold the cal key down while pressing
the on/off key.
3. The measurement mode window
appears without the CAL indicator to
indicate that the user calibration has
been successfully cleared.
16. FIELD CALIBRATION UNLOCK
The calibration unlock feature enables field
calibration operation. To unlock field
calibration:
1. Turn the thermometer off.
2. Hold the cal key and hold key down
simultaneously, then press the on/off
key.
3. The “CAL” indicator appears
momentarily with the version-model
window to indicate that the user-
calibration function has been unlocked
successfully (Figure 10).
CAL
Figure 10: Successful field calibration unlock
Table of contents
Other Oakton Thermometer manuals

Oakton
Oakton WD-35427-50 User manual

Oakton
Oakton Temp-14 User manual

Oakton
Oakton 35625-40 User manual

Oakton
Oakton Type T User manual

Oakton
Oakton Temp-100 User manual

Oakton
Oakton Temp 4/5/6 User manual

Oakton
Oakton InfraPro 3 User manual

Oakton
Oakton 35625-40 User manual

Oakton
Oakton 35629-00 User manual

Oakton
Oakton WD-35426-60 User manual

Oakton
Oakton Temp-340 Thermistor User manual

Oakton
Oakton Acorn Temp TC User manual

Oakton
Oakton Temp-360 User manual

Oakton
Oakton TempTestr IR User manual

Oakton
Oakton Temp-340 Thermistor User manual

Oakton
Oakton Temp-360 User manual

Oakton
Oakton WD-35426-61 User manual

Oakton
Oakton 35629-20 User manual

Oakton
Oakton TEMP TC User manual