Chapter 1 General Policy
EVIS EXERA II TJF TYPE Q180V REPROCESSING MANUAL
• Patient debris and reprocessing chemicals are hazardous.
To guard against contact with dangerous chemicals and
potentially infectious material, wear appropriate personal
protective equipment during cleaning, disinfection, and
sterilization. Such protective equipment should include
appropriate eyewear, face mask, cap, moisture-resistant
clothing, shoe covers, and chemical-resistant gloves that fit
properly and are long enough to prevent skin exposure.
• The reprocessing room must be adequately ventilated.
Adequate ventilation protects against the buildup of toxic
chemical fumes.
• Always remove contaminated personal protective equipment
before leaving the reprocessing area to prevent
contamination from spreading.
• Only Olympus-recommended or Olympus-endorsed
automated endoscope reprocessors (AERs) have been
validated by Olympus. When using an AER that is not
recommended by Olympus, the manufacturer of the AER is
responsible for validating compatibility of the AER with each
Olympus endoscope and accessory.
• Before using an AER, confirm that it is capable of
reprocessing the endoscope including all channels, the
forceps elevator recess, and accessories. Be sure to attach
all required connectors. Otherwise, insufficient reprocessing
may pose an infection control risk. If you are uncertain as to
the ability of your AER to reprocess the endoscope including
all channels, the forceps elevator recess, and accessories,
contact the manufacturer of the AER for specific instructions
and information on compatibility and required connectors.
When you use an AER which allows you skip some steps in
precleaning and manual cleaning of endoscopes, confirm
with the AER manufacturer that such skip is applicable to this
endoscope and establish detailed precleaning and manual
cleaning procedures of this endoscope according to both
instructions of this manual and the AER manufacturer.
• Put the forceps elevator in intermediate position of the range
of movement by turning the elevator control lever and set it in
your AER.