2021/10/12 04:32 1/9 Olympus BH-2 Microscope Motor Drive Installation
Applied Scientific Instrumentation - https://asiimaging.com/docs/
Olympus BH-2 Microscope Motor Drive
Installation
Motor Drive Installation
This procedure steps you through the installation and alignment of the ASI motor drive onto the
Olympus BH-2 microscope. The following tools are required for this procedure:
Small knife or screwdriver
2.5 mm Allen wrench ( Provided )
3.0 mm Allen wrench ( Provided )
1.5 mm Allen wrench ( Provided )
3/32 mm Allen wrench ( Provided )
5/64 mm Allen wrench ( Provided )
7/64 mm Allen wrench ( Provided )
9/64 mm Allen wrench ( Provided )
1/16 mm Allen wrench ( Provided )
Preparing The Microscope
In this part of the procedure the left fine focus knob will be removed from the microscope:
Prior to removing left fine focus knob, the microscope stage should be fully lowered. To do this,
release the lock by rotating fully counter-clockwise the lock ring behind the left coarse focus knob,
then rotate the coarse knob to lower the stage all the way down.
Note: the terms left and right refer to the sides of the
microscope as viewed from the front.
Last update: 2021/09/23 17:15 oly_bh2_zdrive_installation https://asiimaging.com/docs/oly_bh2_zdrive_installation
https://asiimaging.com/docs/ Printed on 2021/10/12 04:32
1) Carefully slip the knife point into the inside edge of the left fine focus knob end and pry out the
20mm black disk that covers the fine focus knob attachment screw.
2) Remove the left fine focus knob by first removing the attachment screw using the 2.5mm hex
wrench then pulling the knob off of the fine focus shaft. It will be necessary to hold the right fine focus
knob while removing the screw to prevent the shaft from turning.
3) After the knob has been removed insure that the wave washer is still located on the brass clutch
assembly on the focus shaft. If the washer has stuck to the fine focus knob that was removed place it
as shown on to the focus shaft.
Installing the Drive and Back plate assembly
Locate the motor drive, the back plate and the brass clutch plate.
2021/10/12 04:32 3/9 Olympus BH-2 Microscope Motor Drive Installation
Applied Scientific Instrumentation - https://asiimaging.com/docs/
Back Plate
Drive Assembly
In this step we will install the ASI backplate and motor drive assembly as a unit. The ASI backplate will
be installed and clamped to the back of the microscope as shown only after the motor drive is
clamped onto the fine focus shaft of the microscope. Before installing the backplate and drive
assembly loosen but do not remove the horizontal and vertical adjustment screws. When these
screws are loosened the motor drive should move freely about in the horizontal and vertical
adjustment axis.
Alternatively the drive and the back plate can be installed separately. In this installation the drive is
removed from the back plate by removing the horizontal adjustment screw.
Horizontal adjustment screw
Last update: 2021/09/23 17:15 oly_bh2_zdrive_installation https://asiimaging.com/docs/oly_bh2_zdrive_installation
https://asiimaging.com/docs/ Printed on 2021/10/12 04:32
Vertical adjustment screws
Installing the Drive & Brass Clutch Plate
Use the 7/64 Allen wrench to loosen the clamp that is located on the ASI drive shaft. This clamp is
located at the end of the drive shaft as it protrudes out of the black encoder cover. Once the clamp is
loose slide it back towards the encoder and slide the brass clutch plate onto the ASI drive shaft. The
clutch plate should slide onto the drive shaft with the milled out portion pointing away from the
encoder. While holding the right fine focus knob slide the motor drive and back plate assembly in
towards the microscope, when the ASI drive shaft is close enough, position the hollow end of the drive
shaft (protruding from the black encoder) over the fine focus shaft on the microscope. Orient the
motor assembly so that the two shafts are aligned, and then push the motor assembly onto the
microscope. If you do not hold the right fine focus knob the microscope’s fine focus shaft may be
pressed over towards the right of the microscope. If this happens simply push the right fine focus
knob in towards the microscope to push the shaft back.
Press the drive onto the microscope until the side of the back plate assembly slides against the side
of the microscope, or if installing the drive and back plate separately: push the drive on until the tab
on the adjustment bar seats into the grove on the back plate. While holding the drive in place with
one hand use a screw driver to press the clamp against the brass clutch plate and press the brass
plate against the wave washer and focus shaft. Hold this in place and use the 7/64 inch hex wrench to
securely tighten the clamp. Note The Clamp Must Be Securely Tightened or the drive and
clutch may slip.
2021/10/12 04:32 5/9 Olympus BH-2 Microscope Motor Drive Installation
Applied Scientific Instrumentation - https://asiimaging.com/docs/
Slide ASI drive over fine focus shaft
Press clamp & clutch plate firmly against microscope
Last update: 2021/09/23 17:15 oly_bh2_zdrive_installation https://asiimaging.com/docs/oly_bh2_zdrive_installation
https://asiimaging.com/docs/ Printed on 2021/10/12 04:32
Securely Tighten Clamp
Clutch & Drive Securely Tightened
Aligning the drive
After installing & securing the drive and brass clutch plate to the focus shaft of the microscope the
back plate can be positioned and secured. Position the back plate to a point midway along the slots
for the two vertical adjustment screws and secure it in place by tightening the set screw to depress
2021/10/12 04:32 7/9 Olympus BH-2 Microscope Motor Drive Installation
Applied Scientific Instrumentation - https://asiimaging.com/docs/
the clamp on side of the microscope.
Position back plate so that vertical adjustment screws are midway in slots
Use the 5/64” Allen Wrench to tighten the Base plate clamp at a point where the vertical adjustment
screws are midway in the slots
After securing the back plate check for proper drive alignment be rotating the right fine focus knob.
The drive is usually self aligning and when properly aligned no drag is felt, other then that caused by
the gears, while rotating the fine focus knob. When the dive is in a position of proper alignment use
the 5/32 Allen wrench to tighten the horizontal, and vertical adjustment screws at the position where
minimum drag is felt.
Recheck for minimum drag on the fine focus shaft by turning the right fine focus knob. If there is any
drag felt at any particular point while rotating the right fine focus knob repeat the alignment
procedure by loosening the horizontal, and vertical adjustment screws and repositioning the drive.
The drive is usually self aligning , and if realignment is necessary loosen the horizontal, and z vertical
adjustment screws and move the drive so that is slides into its natural position and then tighten the
Last update: 2021/09/23 17:15 oly_bh2_zdrive_installation https://asiimaging.com/docs/oly_bh2_zdrive_installation
https://asiimaging.com/docs/ Printed on 2021/10/12 04:32
two adjustment screws with the 5/32 allen wrench.
Step 6 - Installing the cover & left fine focus knob
Remove the two screws from the edges of the drive assembly. Locate the ASI cover plate. Place it in
position over the motor drive assembly and secure it with the screws just removed. Place the left fine
focus knob over the protruding end of the drive shaft and secure it in place with the original screw
using the 2.5 mm allen wrench. Install the 20mm black disk that covers the fine focus knob
attachment screw. This 20mm black disk was removed in step #1.
Connecting the motor drive to the controller and computer
Please refer to the wiring drawings and diagrams given in the manual
Tools Needed: Small slotted screw driver.
1) Locate the drive cable. Connect one end of the drive cable to the pigtail extending from the motor
drive near the microscope and the other end to mating connector on the back of the Microscope
Focus Controller, or MS-2000 unit labeled “Drive”. Secure the cable to the drive by tightening the
screws at each end of the connector.
2) Locate the computer interconnect cable. Connect one end to the mating connector on the back of
the controller labeled “Serial In”. Connect the other end to the appropriate serial output connector on
your computer. .
3) Plug in the power cord / power supply
This completes the steps necessary to interconnect the ASI motor drive, focus controller and
computer. Please refer to the operation section of this manual before using the unit
zdrive, olympus, bh-2
2021/10/12 04:32 9/9 Olympus BH-2 Microscope Motor Drive Installation
Applied Scientific Instrumentation - https://asiimaging.com/docs/
From:
https://asiimaging.com/docs/ - Applied Scientific Instrumentation
Permanent link:
https://asiimaging.com/docs/oly_bh2_zdrive_installation
Last update: 2021/09/23 17:15

Other manuals for BH2 Series

6

Other Olympus Microscope manuals

Olympus BX2 SERIES User manual

Olympus

Olympus BX2 SERIES User manual

Olympus BX43 User manual

Olympus

Olympus BX43 User manual

Olympus BX53M User manual

Olympus

Olympus BX53M User manual

Olympus bx40 User manual

Olympus

Olympus bx40 User manual

Olympus CX31-P User manual

Olympus

Olympus CX31-P User manual

Olympus SZ61 User manual

Olympus

Olympus SZ61 User manual

Olympus BX53 User manual

Olympus

Olympus BX53 User manual

Olympus U-D7RES User manual

Olympus

Olympus U-D7RES User manual

Olympus BX-FLA User manual

Olympus

Olympus BX-FLA User manual

Olympus CX21 User manual

Olympus

Olympus CX21 User manual

Olympus BX61 User manual

Olympus

Olympus BX61 User manual

Olympus IX2-RFSPOT User manual

Olympus

Olympus IX2-RFSPOT User manual

Olympus BXC-CBB User manual

Olympus

Olympus BXC-CBB User manual

Olympus CKX41 User manual

Olympus

Olympus CKX41 User manual

Olympus CX21LED User manual

Olympus

Olympus CX21LED User manual

Olympus U-AW User manual

Olympus

Olympus U-AW User manual

Olympus BX61WI User manual

Olympus

Olympus BX61WI User manual

Olympus CX41 User manual

Olympus

Olympus CX41 User manual

Olympus BH2 Series Operating instructions

Olympus

Olympus BH2 Series Operating instructions

Olympus ORBEYE Manual

Olympus

Olympus ORBEYE Manual

Olympus SZX16 Installation guide

Olympus

Olympus SZX16 Installation guide

Olympus BH2 Series Installation instructions

Olympus

Olympus BH2 Series Installation instructions

Olympus IX71 User manual

Olympus

Olympus IX71 User manual

Olympus BX-URA2 User manual

Olympus

Olympus BX-URA2 User manual

Popular Microscope manuals by other brands

Nikon coolscope quick start guide

Nikon

Nikon coolscope quick start guide

Bresser National Geographic 300-1200x microscope operating instructions

Bresser

Bresser National Geographic 300-1200x microscope operating instructions

OPTO-EDU A15.2601-T instruction manual

OPTO-EDU

OPTO-EDU A15.2601-T instruction manual

Pulch + Lorenz ZEISS Stemi 305 quick start

Pulch + Lorenz

Pulch + Lorenz ZEISS Stemi 305 quick start

Labomed Sigma user manual

Labomed

Labomed Sigma user manual

National DCX-205-RLED instructions

National

National DCX-205-RLED instructions

Vision Engineering EVO Cam user guide

Vision Engineering

Vision Engineering EVO Cam user guide

AMG EVOS user guide

AMG

AMG EVOS user guide

BestScope BS-3015 Series user manual

BestScope

BestScope BS-3015 Series user manual

Optika B-180 Operation manual

Optika

Optika B-180 Operation manual

BestScope BLM-205 instruction manual

BestScope

BestScope BLM-205 instruction manual

Zeiss LSM 510 Inverted Brief Operating Manual

Zeiss

Zeiss LSM 510 Inverted Brief Operating Manual

Gima 31001 Use and maintenance book

Gima

Gima 31001 Use and maintenance book

National DCX-213-RLED instructions

National

National DCX-213-RLED instructions

Denville Scientific M3100 Series Operation manual

Denville Scientific

Denville Scientific M3100 Series Operation manual

Optika OPTIGEM Operation manual

Optika

Optika OPTIGEM Operation manual

Ecleris MICROSTAR OM-100 user manual

Ecleris

Ecleris MICROSTAR OM-100 user manual

Zeiss Axiocam 105 color instruction manual

Zeiss

Zeiss Axiocam 105 color instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.