Omron IntelliSense HEM-907 User manual

OMRON
IntelliSense™ Blood Pressure Monitor
ModelHEM-907
Table of Contents
Be Sure to Read This Section
Exemptions ........................................2
Notes on Safety..................................3
Know Your Unit
Features of the Product......................7
Components of the Product ...............8
Options...........................................9
Names and Functions of the Parts...10
Names of the Parts.......................10
Functions and setting
while the unit is in use ..................12
Preparations before Measurement
How to Apply the Cuff ......................14
How to Use the Power Source.........16
How to Use the AC Adapter............16
Installation and Replacement
of Battery Pack.............................17
How to Measure Blood Pressure
List of Measurement Modes.............19
SINGLE Mode ..............................20
AVG Mode....................................22
MANU Mode.................................24
How to Check Pressure Accuracy..26
How to Clean the Unit after Use ........27
List of Error Codes...........................28
Troubleshooting ...............................29
Specifications
Caution.............................................30
Five Year Limited Warranty..............31
Specifications...................................31
Thank you very much for purchasing the
OMRON IntelliSense™ Blood Pressure Monitor.
Please thoroughly read this Instruction Manual
before using this monitor to insure safe and
accurate use.
Please keep this manual near the monitor all
the time for future reference.
This Instruction Manual contains the Quality
Guarantee Card.
INSTRUCTION MANUAL

2
Exemptions
OMRON does not accept liability and warranty becomes void under the following
circumstances:
1. When persons, not authorized by OMRON, perform repairs or modifications of this
product.
2. When use and/or operation of this device is adversely effected by a device not
manufactured by OMRON.
3. When use and/or operation of this device is adversely effected by use of parts, not
authorized by OMRON, to repair or modify this product.
4. When Notes on Safety or Instructions for Use contained in this manual are not followed.
5. When use and/or operation of this device is caused by an act of God, such as fire,
earthquake, flood or other natural disasters.
1. The product and contents of this Instruction Manual may be changed without prior notice.
2. We have prepared the contents of this Instruction Manual thoroughly. However, if an
inadequate description or error is found, please let us know.
3. Reproducing or copying any or all of this Instruction Manual without OMRON’s written
consent is prohibited.

Specifications
3
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
Notes on Safety
• The warning signs and the sample icons shown here are listed to insure safe and accurate use.
• The icons and meanings are as follows.
Warning sign Contents
Indicates matters in which death or severe bodily damage may arise as a
result of incorrect handling.
Indicates matters in which bodily harm or material damage* may arise as a
result of incorrect handling.
* Material damage refers to a wide range of damage involving your house, household goods, domestic animals, and pets.
Examples of signs
The icon indicates caution (including warning and danger).
Matters involving actual caution are indicated by text or pictures in or near .
The pictured icon refers to "caution for flammability".
The icon indicates prohibitions (what you cannot do).
Matters involving actual prohibitions are indicated by text or pictures in or near .
The pictured icon refers to "prohibition to disassemble".
The icon indicates something that is compulsory (always follow).
Matters involving actual compulsory actions are indicated by text or pictures in or near .
The pictured icon refers to "unplugging the power source plug".
Do not wrap the cuff over an arm to which intravenous injection or transfusion
is being conducted, or when otherwise contraindicated.
Do not connect the air tube or the cuff to other equipment which is connected
to an intracorporeal organ. Air embolisms may result.
Self diagnosis of measured results or treatment is dangerous. Please follow
the instruction of the doctor or healthcare provider.
If cuff inflation does not stop, remove the cuff or pull out the air tube from the
main unit.
If battery fluid gets into your eye or comes in contact with skin, wash the
effected area with water repeatedly. Immediately consult a doctor for treatment.
Do not use this unit in the presence of flammable gas or anesthetics or in a
high pressure oxygen room or oxygen tent.
Do not use the battery pack (option) for devices other than for this unit.
Do not disassemble the battery pack.
Do not touch the AC adapter with wet hands.

4
Unplug the AC adapter from the electric outlet if this unit is unused for an
extended period of time.
Unplug the AC adapter from the electric outlet when installing, removing, or
cleaning the unit.
Confirm readings with a stethoscope when an irregular pulse wave is
displayed or when the measured value is questionable or erratic.
Use an AC adapter indicated for use with a power supply of 110 VAC.
Do not share an electric outlet with other unit or electric appliance.
After cleaning this unit, dry it well before plugging the AC adapter in the
electric outlet.
If this unit fails to perform as indicated, discontinue use, turn off the unit,
unplug the AC adapter from the electric outlet, and contact OMRON’s repair
department.
Do not disassemble or modify this unit.
Notes on Safety
Do not use any cuff other than the models exclusive for this unit.
Do not use this unit on infants.
Do not use this unit on patients using a pump oxygenator.
Do not use an AC adapter or battery pack not specified for this unit.

Specifications
5
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
Notes on Safety
Do not use a cellular phone near this unit.
Do not use this unit in a vehicle.
Do not install the parts and/or instruments not specified for this unit.
Do not use a broken power cord or AC adapter.
Do not install or store this unit where it may come in contact with water or
liquid medication.
General advice
Do not place or put anything on this unit.
Do not drop this unit.
Turn off power to the unit and unplug the AC adapter from the electric outlet before moving
the unit.
Read the instruction manual of the other devices to be used at the same time with this unit, to
understand and be aware of the interaction between the devices.
When using the unit
• Do not inflate the cuff without being wrapped over the arm.
• Do not use a damaged cuff.
• Be sure that patients do not touch the Buttons of this unit.
After using the unit
• Do not disinfect this unit by autoclave or gas sterilization (EtO, gluteraldehyde, or high
concentration ozone).
Do not install or store this unit in the following places.
• Under the direct sunlight,
• Dusty or salty environment,
• Places having slope, vibration, and/or shock,
• Storage of chemicals or where combustable gas may be generated,
• Under high temperature and high humidity.

6
Notes on Safety
Maintenance and inspection
1. Check the unit operation on regular basis.
2. If this unit has not been used for more than three months, be sure to check that this unit
operates normally and safely before use.
Troubleshooting
If device error 9 (Er9) occurs, take the following procedure promptly:
(1) Remove the cuff from the patient's arm.
(2) Turn off the power of the unit and unplug the AC adapter from the electric outlet.
(3) Display "Out of use" on this unit so that it cannot be used.
(4) Contact Omron repair service (1-877-216-1336).

7
Features of the Product
OMRON IntelliSense™ Blood Pressure unit, Model HEM-907 is developed to measure blood
pressure and pulse rate accurately and simply in a doctor's office, examination room, or patient
bedside.
• One-button operation
Simply wrap the cuff and push the START Button. Blood pressure and pulse rate are automatically
measured by the oscillometric method.
• Automatic pressure setting
When the P-SET (Pressure Setting) Knob is set to "AUTO," the unit will automatically inflate the
cuff to the optimal pressure according to each patient's blood pressure. Pre-setting inflation level is
not necessary.
• Noiseless operation
This unit operates so quietly that it can be used in the patient room at night.
• Average Mode (AVG Mode)
In the AVG Mode, this unit will automatically measure for two or three times. The average of
systolic and diastolic blood pressures and pulse rate are displayed. Each measurement can also
be shown individually. The number of measurements, waiting time before first measurement, and
the interval can be changed.
• Auscultation Mode (MANU Mode)
You can measure auscultatory blood pressure by using a stethoscope, with automatic cuff inflation
and deflation by this unit. Because the cuff pressures during deflation are displayed digitally and
synchronized with the heart beat, they can be read with accuracy. After taking systolic reading you
can accelerate cuff deflation to shorten measurement time.
• Large and easy to read display
Large and easy to read figures are displayed on the LCD display.
INTELLISENSE™ blood pressure monitor is a global brand name of blood pressure monitor
equipped with bio-information sensing and high performance fuzzy logic technology, which are
the key technologies of OMRON. With this excellent sensing technology, the monitor can
achieve “accurate” and “fast” measurement.

8
Components of the Product
Main unit
Accessories
Air tube 1.0m (39 1/3") *
Model: HEM-TUBE-100
AC adapter (80100H907S)
Model: HEM-AC-J Instruction Manual
Items identified with
an asterisk (*) are
consumables and not
covered by the
guarantee.
IntelliSense™ Blood Pressure
Monitor, Model HEM-907
(Included and also available separately)
Cuff / bladder set Large
(with built-in bladder) *
Model: HEM-907-CL19
Applicable arm circumference:
13
" to 17" (
32 to 42 cm)
Cuff / bladder set Small
(with built-in bladder) *
Model: HEM-907-CS19
Applicable arm circumference:
7” to 9
" (
17 to 22 cm)
Cuff / bladder set Medium
(with built-in bladder) *
Model: HEM-907-CR19
Applicable arm circumference:
9
" to 13"
(22 to 32 cm)

Specifications
9
Be Sure to Read This Section
Preparations before Measurement
How to Measure Blood Pressure
Know Your Unit
Components of the Product
Exclusive Stand
Model: HEM-907-STAND Wall-hanging kit
Model: HEM-907-WKIT
Bladders *
Pole-mounting kit
Model: HEM-907-PKIT
Cuffs without bladder *
Options
Large Cuff
Model:HEM-907-CUFFL
Medium Cuff
Model: HEM-907-CUFFM
Small Cuff
Model: HEM-907-CUFFS
Large Bladder
Model: HEM-907-BLDRL
Medium Bladder
Model: HEM-907-BLDRM
Small Bladder
Model: HEM-907-BLDRS
Battery pack (48H907N) *
Model: HEM-907-PBAT
Air tube 1.3 m *(51 1/6")
Model: HEM-TUBE-130
Items identified with
an asterisk (*) are
consumables and
not covered by the
guarantee.
Accessories and
options are subject
to change without
prior notice.
(not included)

10
Features and Functions
Main unit
1.Display: Displays blood pressure and, pulse rate readings, and oscillation pulse level.
2.HIDE Button: Switches display and non-display of measured results.
3.DC jack: Connects the AC adapter.
4.P-SET (pressure setting) Knob: In the AUTO position, inflation level is automatically set.
Otherwise, inflation level can optionally be set manually between 100 and 280 mmHg.
5.MODE Selector: Selects the operation mode.
• One-time Measurement Mode (SINGLE Mode): Measurement with automatic inflation.
• Average Mode (AVG Mode): Automatically measures two (or three) times consecutively.
• Auscultation Mode (MANU Mode): Automatic inflation, automatic deflation, and pressure display for auscultation (does
not measure blood pressure).
• Check Mode (CHECK Mode): Checks the accuracy of pressure display. Displays only pressure.
6.ON/OFF (power) Button: Turns on or off the unit.
7.START Button: Starts the measurement.
8. DEFLATION (deflation control) /Measurement Result Display Switch Button:
• In the MANU Mode, deflates the cuff by approximately 5 to 10 mmHg with each push during deflation.
• In the AVG Mode, switches the display of average values and the measurement results with each push.
9.Air Connector: Connects the air tube.
10.STOP Button: Stops the measurement and deflates air rapidly.
1. Display
2. HIDE (non-display) Button
3. DC jack
4. P-SET (pressure
setting) Knob
5. MODE Selector
6. ON/OFF (power)
Button
7. START Button
8. DEFLATION
(deflation control)
/Measurement
Result Display
Switch Button
9. Air Connector
10. STOP Button
Names of the Parts
(See Page 11 for detail)

Specifications
11
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
Features and Functions
Display
Systolic blood pressure
Diastolic blood pressure
Pulse rate
Contents when the AVG
Mode is selected
AVG.: Mean value
1st : First reading
2nd : Second reading
3rd : Third reading
Ready to Measure /
Pulse synchronization
Pulse level /
Number of irregular pulses
Number of irregular pulse waves: The pulse
wave having a difference of more than 25%
from the average interval of pulse waves is
determined irregular. Number of irregular
pulse waves is displayed up to three times.
Battery level
Displays for the operable level.
Blinks for a low charged
level
Displays for the inoperable
level.
Charging
Displays when the battery
pack is being charged.
External power source
Displays AC when the unit is connected
to the external power source via the AC
adapter.
Example of display
No. of irregular pulses
123

12
Features and Functions
Functions setting
(1) Inflation level setting
AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected.
The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure
(approximately 30-40 mmHg above the patient’s systolic pressure).
Manual level setting: Inflation level can be set manually between 100 and 280 mmHg. Set the
level to 30 to 40 mmHg higher than the expected systolic pressure.
• To set the P-SET to "AUTO," turn the P-SET Knob counterclockwise as far as it goes until you can hear a click.
• In the "AUTO" setting, inflation level may not be set automatically when the systolic blood pressure is more than
220 mmHg. Use the unit in the manual setting.
• If the cuff has not been inflated to the necessary level, it may be re-inflated automatically.
(2) Non-display function
Use to prohibit the display of measurement results. However,
the cuff pressure during measurement is displayed.
This function can be used in the SINGLE and AVG Modes.
By pushing the Button, display or non-display of status
is switched alternatingly.
(3) Manual deflation control
Accelerate deflation by pushing the DEFLATION (deflation
control) / Measurement Result Display Switch Button during
deflation in the measurement by the MANU Mode.
With each push of the Button, cuff is deflated rapidly
in increments of 5 to 10 mmHg.
Display Non-display

Specifications
13
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
Features and Functions
List of Functions for each Mode
Inflation level
setting
Non-display
function
Manual deflation
control
Single (SINGLE)
Mode Average
(AVG) Mode Auscultation
(MANU) Mode Check
(CHECK) Mode
Function
Measurement
Mode
(4) AVG Function setting
You can set the number of measurements, the waiting time until the 1st measurement, and the
measurement interval for the AVG Mode.
Items to set Set value
F1 Number of measurements 2 times or 3 times
F2
Waiting time until the start of 1st measurement
0 sec, 3 min, 5 min, or 10 min.
F3 Measurement interval 5 sec, 30 sec, 1 min, or 2 min.
Note: The bold letters represent the factory-set values.
Procedure to change the set values
1) When the power is OFF, push the ON/OFF (power) Button for
more than three seconds while holding the START Button;
F1 is displayed.
2) Push the START Button and select the function to set from
F1 to F3. Each time you push the START Button, the
functions change in the order of .
3) Push the DEFLATION (deflation control)/Measurement Result
Display Switch Button and change the set values.
4) When the setting is finished, push the ON/OFF (power)
Button to turn off the power. The setting is changed.
min
sec
Function #

14
How to Apply the Cuff
General advice
• Do not inflate the cuff without being wrapped over
the arm.
• Do not use a damaged cuff.
1. Select the cuff according to the arm size
Arm circumference Name of the cuff
(7” - 9”) 17-22 cm HEM-907-CS19(Small)
(9” - 13”) 22-32 cm HEM-907-CR19 (Medium)
(13” - 17”) 32-42 cm HEM-907-CL19 (Large)
• Check the following before applying the cuff:
1) The bladder is correctly installed in the cuff.
2) The bladder is not twisted inside the cuff.
3) The bladder tube is protruding from the cuff as shown in
the Figure on the right.
2. Connect the air tube securely.
• Connect the air tube to the main unit by securing
the air plug to the base of the air connector.
• Securely connect the air tube and the cuff/bladder
set by rotating the luer connection as shown in the
Figure on the right.
Do not wrap the cuff over on arm to which
intravenous injection or transfusion is being
conducted, or when otherwise contraindicated.
Do not connect the air tube or the cuff to other
equipment which is connected to intracorporeal
organ. Air embolisms may result.
The cuff of OMRON IntelliSense™ Blood Pressure Monitor HEM-907 plays an important role of
collecting the information on blood vessels. Please wrap the cuff according to the procedure, below.
Cuff
Air tube
Air connector
Air plug
Cuff side Air tube side
Extract the air tube from .
Enlarged figure of
the circled part

Specifications
15
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
How to Apply the Cuff
4. Align the Artery Position Mark
with the brachial artery.
5. Wrap the cuff snugly using both hands
and securely fasten it with the Velcro™
tape. At this time, the lower edge of
the cuff must be placed 1/2" to 1"
above the inner side of elbow joint.
• If the INDEX is positioned outside the RANGE, select the cuff suitable for the
patient's arm circumference and wrap it again.
• Wrap the cuff so that you can insert only one finger between the cuff and arm.
7. Keep the level of the cuff at the same
level as the heart during the
measurement.
3. Place the hand of the patient with the
palm of hand facing upward.
1/2"-1"

16
How to Use the Power Source
General advice
• Read the instruction manual of the other devices to
be used at the same time with this unit to
understand and be aware of the interaction between
the devices.
Do not use this unit in the presence of
flammable gas, or anesthetics, or in a high
pressure oxygen room or oxygen tent.
Do not touch the AC adapter with wet hands.
Be sure to use the AC adapter from the power
supply of 110 VAC.
Do not install or store this unit where it may
come in contact with water or liquid medication.
[1]
[2]
How to use the AC adapter
Connect the AC adapter to the DC jack of the main unit [1] and the electric outlet [2].
NOTE: When the AC adapter is connected and the unit is turned off, the AC adapter
charges the installed rechargeable battery.

Specifications
17
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
How to Use the Power Source
1. Remove both screws on the upper portion of
the battery cover of this unit, and remove
the cover.
2. Disconnect the old battery pack from the
connector and replace with a new one.
3. Install the battery cover and fasten it with both
screws. Be careful not to pinch the lead wire.
4. Connect the main unit to the AC adapter to
charge the new battery. The battery is not
charged when you purchase the monitor. When you use the battery for the first
time, charge it for more than twelve hours before use.
If battery fluid gets into your eye or comes in
contact with skin, wash the affected area with
water repeatedly. Immediately consult a doctor
for treatment.
Do not use the battery pack for devices other
than for this unit.
Do not disassemble the battery pack.
Installation and Replacement of Battery Pack (option)

18
Battery life
• You can use the unit for approximately three hundred measurements with one charge.
• Approximate life of battery is two years. However the battery life from each charging may be
shortened depending on the state of using. If the interval between charging becomes short and
the mark appears frequently, replace it.
Charging time
• At approximately five seconds after connecting the AC adapter, the unit will start battery
charging automatically.
• While the battery is being charged, the mark turns on.
• The battery can be completely charged in approximately twelve hours.
Battery low
• When the mark starts to blink, twenty to thirty measurements remain on the battery.
However if the Battery Low Mark starts to blink, charge it soon.
• If the mark is displayed, the battery is low and the unit cannot operate. Please charge
the battery.
Automatic Power Off
• When using the unit with the battery, the unit will turn off automatically after five minutes
of inactivity.
• While the AC adapter is connected, the Auto Power Off function does not work.
How to Use the Power Source

Specifications
19
Be Sure to Read This Section
Know Your Unit
Preparations before Measurement
How to Measure Blood Pressure
How to Measure Blood Pressure
General advice
• Patients should not touch the unit.
If cuff inflation does not stop, remove the cuff
or pull out the air tube from the main unit.
List of Measurement Modes
To measure
only once SINGLE Mode Refer to Page
19.
To measure two
times (or three times)
automatically and
consecutively AVG Mode Refer to Page
21.
To measure by
using a
stethoscope MANU Mode Refer to Page
23.
Confirm readings with the stethoscope when
an irregular pulse wave is displayed or when
the measured value is questionable or erratic.
Do not use a cellular phone near this unit.
Do not use this unit in a vehicle.

20
How to Measure Blood Pressure (in SINGLE Mode)
1. Push the ON/OFF (power) Button to
turn on the power.
2. Set the MODE Selector to "SINGLE".
3. Set the P-SET (inflation level) Knob
to "AUTO" or to the target pressure
value.
4. Measure the patient’s arm size, and
wrap appropriate cuff over the
patient's arm. (Refer to Pages 14
and 15.)
Other manuals for IntelliSense HEM-907
5
Table of contents
Other Omron Blood Pressure Monitor manuals

Omron
Omron BP4350 User manual

Omron
Omron M2 Compact User manual

Omron
Omron HEM-7155T-ESL User manual

Omron
Omron HEM-7156 User manual

Omron
Omron M6W User manual

Omron
Omron HEM-7203 User manual

Omron
Omron HEM-7361T-EBK User manual

Omron
Omron BP7450CAN User manual

Omron
Omron M2 Compact User manual

Omron
Omron T9P User manual

Omron
Omron 5 Series User manual

Omron
Omron X7 Smart User manual

Omron
Omron H-CR17 User manual

Omron
Omron BP769CAN User manual

Omron
Omron IntelliSense HEM-775 User manual

Omron
Omron M6 Comfort User manual

Omron
Omron HEM-9000AI User manual

Omron
Omron S1 User manual

Omron
Omron M2 Compact User manual

Omron
Omron M3 Intellisense User manual