PART 1 - VERIFY PROPER OPERATION OF THE EQUIPMENT, cont.
Press to power on the ZOOM 2.
Press on the ZOOM 2 several times until the wavelength is set to 1310nm.
The ZOOM 2 should show approximately -10.00 dBm. If the reading is below -11.00 dBm, the patch cable should be replaced.
Once proper operation has been verified, remove the first patch cable from both units and set it aside.
Connect the power meter and light source together with the second patch cable as shown in Figure 2.
The ZOOM 2 should show approximately -10.00 dBm. If the reading is below -11.00 dBm, the patch cable should be replaced.
Once proper operation has been verified, continue on to the next step.
PART 2 - SET REFERENCE
With the ZOOM 2 and WaveSource still set to 1310nm, press and hold button on the ZOOM 2 to set it the 1310nm
reference.
The ZOOM 2 display now also should change to show readings in ‘dB’, and will be approximately 0.00 dB.
Press so that the ZOOM 2 is set to ‘1550nm’.
Press on the WaveSource once to set it to ‘1550nm’.
Press and hold on the ZOOM 2 to set the reference for 1550nm. Again, the ZOOM 2 display shows readings in ‘dB’, and
will be approximately 0.00 dB.
PART 3 - TAKE READINGS
Leaving the patch cable attached to the WaveSource, remove the patch cable from the detector port on the ZOOM 2, and
bring both units to the opposite ends of the link under test.
UNITS
ZERO
UNITS
ZERO
QUICK REFERENCE GUIDES
Optical Power Meter: ZOOM 2
Fiber Optic Light Source: WaveSource Series Singlemode Source PAGE 2 of 3
2
FIGURE 2
Verify
Patch
Cable SC connectors shown here;
connector style may vary
LIGHT SOURCE
CONNECTOR PORTS
Do NOT insert APC connector
into either light source port