manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Pall
  6. •
  7. Water Filtration System
  8. •
  9. Pall 5017 User manual

Pall 5017 User manual

Multi-Well Plate Vacuum Manifold
Description
The Multi-well Plate vacuum manifold is an anodized aluminum manifold that has
been designed and optimized for the vacuum filtration of the AcroPrep™and
AcroWell™lines of multi-well filter plates. The vacuum manifold comes complete
with the necessary O-ring and gasket. The control block includes the vacuum
pressure gauge, vacuum metering valve, vacuum release valve and the 1/4 inch
hose barb for vacuum line attachment. Included with the vacuum manifold unit is a
Delrin spacer block designed to accommodate standard 350 µL receiver plates.
The spacer block has been optimized to reduce the space between the receiver
plate and the filter plate during vacuum filtration. A spacer block that has been
optimized to work with our 1 mL plate is available separately. An accessory kit that
contains an additional O-ring, gasket and allen wrench is also included with the
vacuum manifold unit.
Ordering Information
Prod. No. Description Packaging
5017 Multi-well Plate Vacuum Manifold 1/pkg
5014 1 mL Receiver Plate Spacer Block 1/pkg
5015 350 µL Receiver Plate Spacer Block 1/pkg
5016 Replacement Accessory Kit (Includes O-Ring, 1/pkg
gasket and allen wrench)
Note: the 350 µL Receiver Plate Spacer Block can be used to accomodate 384
well receiver plates as well.
Specifications
Materials of Construction
Vacuum Manifold: Anodized aluminum
Gasket: EDPM (Ethylene propylene)
O-ring: Silicone
Spacer blocks: Delrin
Dimensions
Length: 17.48 cm (6.88 in)
Width: 12.37 cm (4.87 in)
Height: 8.05 cm (3.17 in)
Weight: 6.27 lbs
Maximum Operating Vacuum
71.12 cm Hg (28 in Hg).
Note: The multi-well plate vacuum manifold can be used with multi-well filter plates
that meet the specifications set forth by the Society for Biomolecular Screening
(SBS).
Instructions for Use
To set the vacuum manifold to a desired negative pressure setting prior to filtration,
follow the steps listed below.
11. Switch the on/off valve to the off position (See Figure 3).
12. Attach the vacuum line to the 1/4 inch hose barb.
13. Place a receiver plate on top of the vacuum manifold.
14. Turn on the vacuum pressure. You may need to press lightly on the receiver
plate to engage the vacuum seal.
15. Adjust the negative pressure to the desired setting by adjusting the metering
valve.
16. Turn off the vacuum pressure and remove the receiver plate. The vacuum
manifold is now set at the desired setting.
To filter the AcroPrep or AcroWell filter plates follow the steps listed below.
11. If collecting the filtrate, remove the top chamber of the vacuum manifold. The
indentations on the sides of the manifold allow for the top portion of the
manifold to be easily removed. If collecting the retentate, skip to step 3.
12. Place the appropriate spacer block into the lower chamber of the vacuum
manifold. The vacuum manifold can be used with any size receiver plate.
350 µL receiver plates: To use a 350 µL 96 well or 384 well receiver plate,
place the large spacer block (PN 5015) into the lower chamber of the vacuum
manifold. Place the 350 µL 96 well or 384 well receiver plate on top of the
spacer block.
1 mL receiver plates: To use 1 mL receiver plates, place the small spacer block
(PN 5014) into the lower chamber of the vacuum manifold. Place the 1 mL
receiver plate on top of the spacer block. (You can also use a standard 350 µL
receiver plate in place of the 1 mL spacer block.)
2 mL receiver plates: To use 2 mL receiver plates, place the 2 mL receiver plate
directly into the lower chamber of the vacuum manifold.
Instructions for Use (cont.)
13. Before replacing the top chamber of the vacuum manifold, ensure that all
surfaces are free from dirt, debris and any particulate matter that may have
accumulated on the vacuum manifold, O-ring and gasket.
Note: If the O-ring and gasket are not clean, you will not obtain a proper seal.
14. Replace the top chamber of the vacuum manifold.
15. Place the filter plate on the gasket located on the top chamber of the vacuum
manifold. Ensure that the gasket is clean.
16. Ensure that the on/off switch is in the position (See Figure 3).
17. Connect the vacuum line to the 1/4 inch hose barb.
18. When ready to evacuate the filter plate, turn the on/off switch on the vacuum
manifold to the on position. You may need to press lightly on the filter plate to
engage the vacuum seal.
19. The wells will begin to evacuate/empty once the vacuum has been applied to
the chamber.
10. If you need to adjust the vacuum pressure up or down, adjust by moving the
metering valve located to the right of the vacuum pressure gauge.
11. When all of the wells have completely evacuated, turn the on/off valve to the off
position. To release the residual vacuum pressure that remains in the chamber,
push the release valve located to the left of the vacuum gauge. The release
valve will allow the pressure within the manifold chamber to return to
atmospheric pressure and reduce the potential for cross contamination and
spraying of the filtrate. Do not release the vacuum by pulling the corner of
the plate as it will degrade the manifold gasket. You can also tap the top of
the filter plate prior to removing it to release any hanging drops that may be
attached to the outlet tips.
12. Remove the filter plate and place it aside for further processing or dispose of
properly.
13. If collecting the filtrate, remove the top chamber of the vacuum manifold from
the lower portion of the vacuum manifold.
14. Remove the receiver plate from the lower chamber of the vacuum manifold and
utilize the filtrate for further processing.
To replace the O-ring follow the steps below.
11. Remove existing O-ring from the bottom of the upper chamber.
12. Ensure that the new O-ring and O-ring groove are free from dirt, debris and
particulate matter.
13. Place new O-ring into place.
To replace the gasket follow the steps below.
11. With the allen wrench provided, remove the 12 screws located on the bottom
portion of the upper chamber.
12. Separate the top ring from the bottom section of the upper chamber.
13. Remove the old gasket and clean the gasket pocket area of any dirt and
debris.
14. Place the new gasket into the gasket pocket. Place the top ring back onto the
bottom section of the upper chamber. Ensure that the alignment posts on the
bottom section fit into the alignment holes in the top ring.
15. Invert the upper chamber and replace the 12 screws.
16. Lightly tighten the four corner screws with the allen wrench, then tighten all 12
screws. Once this is completed recheck the screws with the allen wrench to
ensure all screws are tightly secured.
17. The manifold is now ready for use.
Figure 1
Control block Vacuum release valve
Vacuum pressure gauge
Vacuum metering valve
On/Off switch
1/4 inch hose barb
350 µL spacer block
O-ring
Gasket
Figure 1
11951T_88293A 7/18/06 10:58 AM Page 1
Instructions for Use (cont.)
Figure 2
Note: A 384 well receiver plate can be interchanged with the 350 µL receiver plate
if using a 384 well filter plate.
Figure 3
Pall Life Sciences
600 South Wagner Road
Ann Arbor, MI 48103-9019 USA
For ordering or technical information:
Tel: 734-665-0651
800-521-1520 (in USA)
Fax: 734-913-6114
Visit Pall Life Sciences on the
Internet: www.pall.com/Lab or
e-mail us at: [email protected]
Offices:
Australia, Lane Cove, NSW, 02-9428 2333
Austria, Wien, 043-1-49 192 0
Canada, Ontario, 905-542-0330
Canada, Québec, 514-332-7255
China, P.R., Beijing, 86-10-8458 4010
France, St. Germain-en-Laye, 01 30 61 39 92
Germany, Dreieich, 06103-307 333
India, Mumbai, 91-22-5956050
Italy, Milano, 02-47-79-61
Japan, Tokyo, 3-3495-8319
Korea, Seoul, 2-569-9161
Poland, Warszawa, 22-811 12 83
Russia, Moscow, 095 787-76-14
Singapore, 65-389-6500
Spain, Madrid, 91-657-9876
Sweden, Lund, 46-0-46 15 84 00
Switzerland, Basel, 061-638 39 00
Taiwan, Taipei, 2-2545-5991
United Kingdom, Portsmouth, 023 92 302600
Pall, , AcroPrep, AcroWell, and BioTrace are trademarks of Pall Corporation.
® indicates a registered trademark in the USA.
©2003, Pall Corporation, 09/03 PN 88293A
On position
Metering valve fully open when
knob is turned completely down.
Turn counter clockwise to lower
vacuum (minimum vacuum
setting is 3" Hg).
On/Off valve
Off position
Vacuum relief valve,
push down on button
to activate.
Complementary Products
AcroWell™96-well Filter Plates with BioTrace™NT and BioTrace PVDF
Membranes exhibit high binding capacities for proteins and nucleic acids.
AcroPrep™96 and 384-well Filter Plates are an excellent platform for a wide
variety of molecular biology, analytical and high throughput sample preparation and
detection applications.
WARNING
Employment of the products in applications not specified, or failure to follow
all instructions contained in this product information insert, may result in
improper functioning of the product, personal injury, or damage to property
or the product. See Statement of Warranty in our most recent catalog.
Upper chamber
350 µL receiver
plate
350 µL receiver plate
spacer block (PN 5015)
1 mL receiver plate
spacer block (PN 5014)
Lower chamber
1 mL receiver
plate
2 mL receiver
plate
gasket
Figure
2Figure 2
11951T_88293A 7/18/06 10:58 AM Page 2

This manual suits for next models

3

Other Pall Water Filtration System manuals

Pall UR610 Series Operating instructions

Pall

Pall UR610 Series Operating instructions

Pall UH231 Series Operating instructions

Pall

Pall UH231 Series Operating instructions

Pall Profile User manual

Pall

Pall Profile User manual

Pall UT319 Series Operating instructions

Pall

Pall UT319 Series Operating instructions

Pall UH239 Series Operating instructions

Pall

Pall UH239 Series Operating instructions

Pall Versalon F310 Series Operating instructions

Pall

Pall Versalon F310 Series Operating instructions

Pall T200 Series Operating instructions

Pall

Pall T200 Series Operating instructions

Pall UR229 Series Operating instructions

Pall

Pall UR229 Series Operating instructions

Pall UH211 Series Operating instructions

Pall

Pall UH211 Series Operating instructions

Pall UH319C Series Operating instructions

Pall

Pall UH319C Series Operating instructions

Pall FP01087 Rev A Instruction Manual

Pall

Pall FP01087 Rev A Instruction Manual

Pall UR619 Series Operating instructions

Pall

Pall UR619 Series Operating instructions

Pall Stax User manual

Pall

Pall Stax User manual

Pall GASKLEEN PG550 Guide

Pall

Pall GASKLEEN PG550 Guide

Pall UR610 Series Operating instructions

Pall

Pall UR610 Series Operating instructions

Pall UR310 Series Operating instructions

Pall

Pall UR310 Series Operating instructions

Pall UR219 Series Operating instructions

Pall

Pall UR219 Series Operating instructions

Pall UH310 Series Operating instructions

Pall

Pall UH310 Series Operating instructions

Pall QPoint Installation and operating instructions

Pall

Pall QPoint Installation and operating instructions

Pall UH210 Series Operating instructions

Pall

Pall UH210 Series Operating instructions

Pall UR210 Series Operating instructions

Pall

Pall UR210 Series Operating instructions

Pall Versalon G410 Series Operating instructions

Pall

Pall Versalon G410 Series Operating instructions

Pall UT310 Series Operating instructions

Pall

Pall UT310 Series Operating instructions

Pall Kleen-Change 100 User manual

Pall

Pall Kleen-Change 100 User manual

Popular Water Filtration System manuals by other brands

Graco 24R707 instructions

Graco

Graco 24R707 instructions

BWT Woda-Pure Clear S Installation and operating manual

BWT

BWT Woda-Pure Clear S Installation and operating manual

Acqua Brevetti BravoMAX A Installation and operating manual

Acqua Brevetti

Acqua Brevetti BravoMAX A Installation and operating manual

Emerson A-W instruction sheet

Emerson

Emerson A-W instruction sheet

Hydac OLSW 20 Series Installation and maintenance instructions

Hydac

Hydac OLSW 20 Series Installation and maintenance instructions

BWT MINISTIL 1iD Installation and operating manual

BWT

BWT MINISTIL 1iD Installation and operating manual

SE Electronics REFLEXION FILTER PRO Assembly & use instructions

SE Electronics

SE Electronics REFLEXION FILTER PRO Assembly & use instructions

Sartorius Sartolab RF 50 Instructions for use

Sartorius

Sartorius Sartolab RF 50 Instructions for use

ETS-Lindgren EMI Installation and maintenance manual

ETS-Lindgren

ETS-Lindgren EMI Installation and maintenance manual

flamco Flamcovent Solar Installation and operating instructions

flamco

flamco Flamcovent Solar Installation and operating instructions

Water Right Sanitizer Plus Series Installation instructions & owner's manual

Water Right

Water Right Sanitizer Plus Series Installation instructions & owner's manual

amiad Sigma Pro 4 Installation & operation manual

amiad

amiad Sigma Pro 4 Installation & operation manual

BWT Multipur M DN65 Installation and operating manual

BWT

BWT Multipur M DN65 Installation and operating manual

Aquavolta hydrogen booster user manual

Aquavolta

Aquavolta hydrogen booster user manual

McNeilus STREET SMART MFS quick start guide

McNeilus

McNeilus STREET SMART MFS quick start guide

Bestway FLOWCLEAR 58404 owner's manual

Bestway

Bestway FLOWCLEAR 58404 owner's manual

Grundfos PHT-N130 Installation and operating instructions

Grundfos

Grundfos PHT-N130 Installation and operating instructions

Grohe Blue Pure 31345DC2 manual

Grohe

Grohe Blue Pure 31345DC2 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.