Penlon ESO 2 User manual



ESO 2 IU Ventlator User Manual
Important
Servicing and Repairs
In order to ensure the full operatonal lfe of ths ventlator,
servcng by an engneer traned by the manufacturer should
be undertaken perodcally
The ventlator must be servced to the followng schedule
1 Sx monthly servce - nspecton and functon testng
2 Annual / two year / four year servce - nspecton and
functon testng, and component replacement
Detals of these operatons are gven n the Servce Manual
for ths ventlator, avalable only for engneers traned by the
manufacturer
For any enqury regardng the servcng or repar of ths
product, contact Penlon Ltd
Techncal Support
Penlon Lmted
Abngdon Scence Park
Abngdon
OX14 3NB, UK
Tel +44 (0) 1235 547060
Fax +44 (0) 1235 547061
E-mal techsupportpenloncom
Always gve as much of the followng nformaton as possble
1 Type of equpment
2 Product name
3 Seral number (refer to the devce ID label)
4 Approxmate date of purchase
5 Apparent fault

ESO 2 IU Ventlator User Manual
Ths manual has been produced to provde authorsed
personnel wth nformaton on the functon, routne
performance and mantenance checks applcable to the ESO 2
Emergency Ventlator
Informaton contaned n ths manual s correct at the date of
publcaton The polcy of the manufacturer s one of contnued
mprovement to ts products Because of ths polcy, the
manufacturer reserves the rght to make any changes whch
may affect nstructons n ths manual, wthout gvng pror
notce
Personnel must make themselves famlar wth the contents
of ths manual and the machne’s functon before usng the
apparatus
The Importance of Patient Monitoring
WARNIN
Patent montors are very desrable ads but are not true
clncal montors as the condton of the patent s also
dependent on hs respraton and the functonng of hs
cardo-vascular system
IT IS ESSENTIAL THAT THESE ELEMENTS ARE MONITORED
FREQUENTLY AND REULARLY AND THAT ANY
OBSERVATIONS ARE IVEN PREEDENE OVER MAHINE
ONTROL PARAMETERS IN JUDIN THE STATE OF A
LINIAL PROEDURE
Before usng any montorng system or devce, the user must
check that t conforms to the relevant standards
Foreword

ESO 2 IU Ventlator User Manual
Contents
Servicing and Repairs ...............................................................................i
The Importance of Patient Monitoring.................................................................ii
1. Warnings and Cautions ......................................................................... 2
2. Purpose ...................................................................................... 5
3. Description ................................................................................... 6
3.1 ESO 2 System Description ...................................................................6
3.2 Ventilator .................................................................................7
3.3 Spirometry ...............................................................................19
3.4 Display Waveforms ........................................................................20
3.5 Alarms ..................................................................................21
3.6 Oxygen Monitor ...........................................................................22
4. Pre-operation Checks and Procedures ............................................................ 25
4.1 Ventilator Set-up ..........................................................................25
4.2 Pre-use Checklist .........................................................................30
4.3 Oxygen Monitor System Set-up ..............................................................32
5. Start-up and operating instructions .......................................................... 34
5.1 Pre-use checks and components.............................................................34
5.2 Start-up .................................................................................34
5.3 Suggested Protocol for closed endotracheal suction in Volume or Pressure Mode: .......................35
5.4 Oxygen monitor ...........................................................................36
6. Maintenance 37
6.1 Service Schedule ..........................................................................37
6.2 Cleaning and Disinfection...................................................................37
6.3 Sterilisation ..............................................................................39
6.4 Oxygen Sensor Replacement ................................................................40
6.5 Electrical Interface, Cables, Tubing...........................................................41
6.6 Bellows Assembly - Removal and Cleaning ....................................................41
6.7 Sterilisation and Disinfectant Treatment Table .................................................42
7. Appendix ..................................................................................... 43
APPENDIX 1 Care of Back-up Battery ...................................................................43
APPENDIX 2 Disposal at end of useful life - risk assessment ..............................................43
APPENDIX 3 Optional extras and approved accessories ....................................................43
APPENDIX 4 Operating system......................................................................44
APPENDIX 5 Spontaneous Mode and Patient Support Modes ............................................50
APPENDIX 6 Specification ................................................................................55
Application ...............................................................................55
Internal Compliance .......................................................................55
Physical .................................................................................55
Alarms ..................................................................................55
Functional................................................................................56
Advanced Spontaneous Breathing Modes (SIMV, SMMV, PSV)......................................57

ESO 2 IU Ventlator v User Manual
Contents
Page No
Disinfection and Sterilisation ................................................................57
Bacterial Filter............................................................................57
Fail Safe Mechanism.......................................................................57
Reliability ................................................................................57
Waveform Tests ...........................................................................57
Volume Tests .............................................................................57
Mobility and Mounting......................................................................57
Device Classification and Labelling ...........................................................58
Oxygen Monitor ...........................................................................59
APPENDIX 7 On-screen Menus .....................................................................60
APPENDIX 8 Symbols .............................................................................62
APPENDIX 9 References ..............................................................................63
APPENDIX 10. Electromagnetic compatibility (EMC) ....................................................63

ESO 2 IU Ventlator 1 User Manual
User Responsibility
Ths ventlator has been bult to conform wth the specfcaton
and operatng procedures stated n ths manual and/or
accompanyng labels and notces when checked, assembled,
operated, mantaned and servced n accordance wth these
nstructons
To ensure the safety of ths devce t must be checked and
servced to at least the mnmum standards lad out n ths
manual A defectve, or suspected defectve, product must not
under any crcumstances be used
The user must accept responsblty for any malfuncton whch
results from non-complance wth the servcng requrements
detaled n ths manual
Addtonally, the user must accept responsblty for any
malfuncton whch may result from msuse of any knd or non-
complance wth other requrements detaled n ths manual
Worn, broken, dstorted, contamnated or mssng components
must be replaced mmedately Should such a repar become
necessary t s recommended that a request for servce advce
be made to Penlon Ltd
Ths devce and any of ts consttuent parts must be repared
only n accordance wth wrtten nstructons ssued by the
manufacturer and must not be altered or modfed n any way
wthout the wrtten approval of the manufacturer The user
of ths equpment shall have the sole responsblty for any
malfuncton whch results from mproper use, mantenance,
repar, damage or alteraton by anyone other than the
manufacturer
USA and anada Federal Law restrcts the sale and use of
ths devce to, or on the order of, a lcensed practtoner
Statements n ths manual preceded by the followng words
are of specal sgnfcance
WARNIN
means there s a possblty of nury to the user or others
AUTION
means there s a possblty of damage to the apparatus or
other property
NOTE
ndcates ponts of partcular nterest for more effcent and
convenent operaton
Always take partcular notce of the warnngs, cautons and
notes provded throughout ths manual

ESO 2 IU Ventlator 2 User Manual
1. Warnings and Cautions
The followng WARNINS and AUTIONS must be read and
understood before usng ths ventlator
WARNINS
Important
Delverng hgh fresh gas flows
1 Hgh fresh gas flows when usng closed endotracheal
sucton Ths ventlator s confgured to delver flows of
oxygen n excess of the maxmum shown on the flowmeter
tube n order to fll the ventlator bellows durng closed
endotracheal sucton Up to approxmately 30 L/mn can be
delvered
Delverng O flows n excess of the maxmum calbrated
flow wll temporarly negate the possblty of accurate
volume delvery and measurement Once the ventlator
bellows s full the O flow should be reduced to that n range
of the calbrated flow tube, then accurate volume delvery
and measurement wll resume
See secton 53 for addtonal nformaton
eneral Informaton
2 Users must make themselves famlar wth the contents
of ths manual and the machne’s functon before usng
the ventlator
Before Usng the Ventlator
3 Before the ventlator s used clncally for the frst tme a
albraton heck and Output heck must be successfully
completed hecks must be carred out by a Penlon-
traned engneer
4 onnect the ventlator to a mans supply wth a protectve
earth Before frst clncal use, verfy that the hosptal
engneerng department has carred out an earth
contnuty test If the ntegrty of the protectve conductor
s n doubt, do not use the ventlator
5 Excessve electronc nose caused by other poorly
regulated devces, such as an electrocautery unt, may
adversely nterfere the functonng of the ventlator
Do not connect the ventlator’s power cord nto the same
electrcal outlet as an electrocautery unt
6 If used wth a mans extenson cord, the unt may be
subect to electro-magnetc nterference
7 Portable RF communcatons equpment (ncludng
perpherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12
nches) to any part of the ventlator, ncludng cables
specfed by the manufacturer Otherwse, degradaton of
the performance of ths equpment could result
8 The drvng gas supply must be clean and dry to prevent
ventlator malfuncton
9 Ths ventlator s desgned to be drven by medcal ar
only The drve gas s set durng manufacture and the
ventlator s calbrated for that gas
Before the ventlator s used clncally for the frst tme,
the commssonng engneer must confrm that the ar/
oxygen selecton s set correctly to Ar
The use of any other gas wll cause naccurate operaton
and may damage the ventlator, resultng n potental
nury to the patent
10 The drvng gas s dscharged through the exhaust
valve port at the back of the ventlator control unt
The dscharged gas may contamnate the envronment
onnect to an ASS system f avalable
11 The bellows can only support approxmately 1 kPa
(10cmHO) dfferental postve pressure, above whch
t may be dslodged from the mountng rng, resultng n
dangerous malfuncton of the ventlator Do not connect
a postve end expratory pressure (PEEP) valve or other
restrctve devce to the exhaust port on the bellows base
Ths would ncrease the pressure nsde the bellows and
the bellows could detach from the base, causng serous
malfuncton
12 The data output, VA, and prnter ports must only be
connected to equpment complyng wth IE safety
standards (eg IE 60950) whch are then confgured to
comply wth IE 60601-1 (Medcal Systems)
13 Breathng System
The breathng system whch conveys gases from the
ventlator to the patent, and dsposes of expred gases,
must conform to the requrements of ISO80601-2-13
Because breathng systems requre frequent cleanng
and dsnfecton they are not a permanent part of the
ventlator and therefore cannot be drectly under the
control of the manufacturer However, we strongly
recommend that only breathng systems whch have been
approved and authorsed by the manufacturer for use
wth the ventlator should be employed
14 Do not use conductve breathng system hoses
15 The sprometer sensors are mounted wthn the
absorber Do not ft a sprometer sensor to any other
locaton The devce wll not measure exhaled volumes n
any other poston
16 Verfy the operaton of each alarm functon daly
Perodcally check the alarms at clncally sutable
ntervals If the audble alarm or the vsual ndcator
of any alarm functon fals to actvate durng any alarm
condton or fals to reset after the alarm has been
cleared, refer the unt to an authorsed servce engneer
17 Before usng the ventlator check that all hose
connectons are correct, and verfy that there are no
leaks Patent crcut dsconnects are a hazard to the
patent Extreme care should be taken to prevent such
occurrences
Usng the Ventlator
18 Do not use Spontaneous (SPONT) mode
19 Before each clncal procedure, carry out pre-use
functon checks on the ventlator (see secton 4213)
20 Ths apparatus must not be used wth, or n close
proxmty to, flammable anaesthetc agents
There s a possble fre or exploson hazard

ESO 2 IU Ventlator 3 User Manual
Warnings and Cautions
21 Do not use the ventlator n an oxygen-rch envronment
22 Mans power supply solaton Dsconnect the mans
supply cable from the socket at the back of the ventlator
control unt (see 32) or at the mans power outlet
23 When the ventlator s connected to a patent, t
s recommended that a qualfed practtoner s n
attendance at all tmes to react to an alarm or other
ndcaton of a problem
24 An alternatve means of ventlaton must be avalable
whenever the ventlator s n use
25 It s recommended that the patent oxygen concentraton
should be montored contnuously
26 If the drve gas supply pressure drops below a nomnal
241 kPa (35 ps), the LOW DRIVE AS SUPPLY alarm wll
actvate both audbly and vsually Patent mnute volume
may be reduced due to lowered flow rates
27 An audble alarm ndcates an anomalous condton and
should never go unheeded
28 The characterstcs of the breathng crcut connected
between the ventlator and the patent can modfy or
change patent ventlaton To assst n the mantenance
of the delvered patent tdal volume, the ventlator
control system software ncludes
a) a complance compensaton algorthm,
b) a fresh gas compensaton algorthm
However, patent ventlaton must be montored
ndependently from the ventlator
It s the responsblty of the user to montor patent
ventlaton
29 heck that the flow sensors are connected correctly to
the nspratory and expratory ports of the breathng
module
30 The Vent Inop (ventlator noperatve) alarm ndcates that
one of the followng condtons has occurred
a) The drve gas solenod has faled
b) The flow control valve has faled
c) Internal electronc fault
d) Internal electrcal fault
e) Software error
31 Note that f a ventlator error s detected, ‘Ventlator
Inoperatve’ wll be dsplayed on the front control panel
dsplay
32 The Hgh and Low arway pressure alarms are mportant
for patent care It s mportant that the sensor s properly
located n the expratory lmb of the crcut - refer to
secton 418
33 The patent must be contnuously attended and
montored, even when advanced breathng modes are n
use
34 Do not touch any electrcal devce cable or connector at
the same tme as the patent
User Mantenance
35 User mantenance s restrcted to cleanng the
outsde surfaces of the ventlator, see secton 6 Other
procedures detaled n ths manual must be carred out
by traned engneers
Servce and repar operatons must only be carred out by
an engneer traned by the manufacturer
The warranty for ths product s vod f the product s
not mantaned n accordance wth the servce schedule
detaled n secton 61, and the procedures publshed n
the Servce Manual for ths product
36 The ventlator must not be modfed or dsassembled by
any unauthorsed person
ontrol Unt
37 Openng the control unt by unauthorsed personnel
automatcally vods all warrantes and specfcatons
Preventon of tamperng wth the control unt s
exclusvely the user’s responsblty If the control unt
seal s broken, the manufacturer assumes no lablty for
any malfuncton or falure of the ventlator
38 For contnued protecton aganst fre hazards, any
replacement fuses must be the dentcal type and
ratng as the orgnal components Replacement must be
carred out by a traned engneer
See secton 4 for fuse ratng
39 If the nternal battery s fully dscharged, the ventlator
wll not functon n the event of mans power falure The
battery must be recharged before the ventlator s used
clncally, otherwse backup cannot be guaranteed
See Appendx 1 for battery mantenance See auton 6
Used or defectve batteres must be dsposed of
accordng to hosptal, local, state, and federal
regulatons
40 No ol, grease or other flammable lubrcant or sealant
must be used on any part of the ventlator n close
proxmty to medcal gas dstrbuton components
There s a rsk of fre or exploson
41 Exteror panels must not be removed by unauthorsed
personnel and the apparatus must not be operated wth
such panels mssng There s a possble electrc shock
hazard
42 The ventlator has ppelne-only gas supply connectons
Note that a malfuncton of the central gas supply wthn
your faclty may cause mmedate cessaton of gas
delvery and total anaesthesa system falure
Bellows Assembly
43 The valve seat on the patent gas exhalaton daphragm
valve n the base of the bellows assembly must be
cleaned regularly
Falure to keep the valve seat clean could result n the
daphragm stckng, thus preventng exhalaton
reat care must be taken not to damage the precson
surface of the valve seat on the patent gas exhalaton
daphragm valve n the base of the bellows assembly
Never use any hard obect or abrasve detergent to clean

ESO 2 IU Ventlator 4 User Manual
Warnings and Cautions
the valve seat use only a soft cloth
If the valve seat s damaged, the valve wll leak and may
cause serous ventlator malfuncton
AUTIONS
1 Do not sterlse the ventlator control unt
Internal components (apart from the patent block
assembly) are not compatble wth sterlsaton
technques and damage may result
The patent block assembly must be removed from the
control unt before sterlsaton (see secton 627)
After cleanng and sterlsaton, the patent block
assembly must be reftted to the orgnal control unt
Refer to secton 6275
2 For ventlator components whch requre sterlsaton,
peak sterlsaton temperatures should not exceed 134o
(275oF) to prevent possble damage (See secton 6)
3 are must be taken not to let any lqud run nto the
control unt serous damage may result
4 The exhalaton valve located n the bellows base
assembly and the paedatrc bellows adaptor must be
cleaned and sterlsed separately Note that the bellows
assembly s bult nto the Breathng module
5 Always check for correct ftment, and carry out a full
functon test (4213 ) before clncal use, f the bellows
has been removed and reftted for any reason
6 Damage wll occur to the battery f t s allowed to reman
n a dscharged state heck the battery frequently f the
ventlator s n storage (see Appendx 1)
7 Fresh gas compensaton s dsabled f
a) The sprometry system s turned OFF through the
menu system, or
b) The sprometry system s not functonng correctly
8 Fresh gas mxture compensaton s dsabled f
a) The sprometry system s turned OFF through the
menu system, or the sprometry system s not
functonng correctly
b) The oxygen montor s swtched OFF
Refer to secton 417
9 rcut complance s not actvated untl Fresh as
ompensaton s swtched OFF Refer to secton 32
OXYEN MONITOR
WARNINS
1 We recommend a calbraton check of the oxygen
montor every tme the system s turned on, as a safety
precauton
2 Do not attempt to open the fuel cell
The sensor contans small quanttes of electrolyte,
classfed as a harmful rrtant whch s potentally
hazardous, and lead
Used or defectve cells must be dsposed of accordng to
hosptal, local, state, and federal regulatons
3 ALWAYS check the ntegrty of the sensor assembly
before use
4 Once exhausted, the sensor must be dsposed of
accordng to hosptal, local, state and federal regulatons
5 The sensor measures oxygen partal pressure, and ts
output wll rse and fall due to pressure change
An ncrease n pressure of 10% at the sensor nlet wll
produce a 10% ncrease n sensor output
6 The oxygen sensor s not sutable for sterlsaton
If contamnaton s suspected, ft a new sensor (see
secton 64) and dspose of the contamnated unt
accordng to hosptal, local, state and federal regulatons
AUTIONS
1 Do not sterlse any oxygen montor component
2 Do not autoclave or expose the sensor to hgh
temperatures
3 If the sensor shows sgns of beng affected by
condensaton, dry the sensor wth soft tssue
Do not use heat to dry the sensor
NOTES
1 The O SENSOR FAULT alarm ndcates that one of the
followng condtons has occurred
a) Internal electrcal fault
b) Software/electroncs fault
c) Oxygen sensor fault
2 The concentraton read-out may, n certan condtons of
excess pressure, show a value above 100%
To accommodate these condtons t s possble to set the
hgh alarm value up to 105% (see secton 5)
3 To mantan maxmum sensor lfe
a) Always swtch off the gas delvery swtch after use, to
ensure that the basal flow ceases
b) Dsconnect the breathng crcut after use
4 The accuracy of flow and volume measurements may be
reduced f the oxygen montor s not n use
5 Fresh gas mxture compensaton s dsabled f the oxygen
montor s swtched ‘off’ n the menu

ESO 2 IU Ventlator 5 User Manual
Restrcted devce
Only for use n OVID-19 pandemc
Only to be used for emergency ventlaton
All adverse ncdents must be reported to MHRA on
0800 731 6789
The devce s ntended to provde controlled concentratons
and flows of gases to the patent
The ventlator assembly s a pneumatcally drven, software
controlled, mult-mode ventlator, desgned for mechancal
ventlaton of adult and paedatrc patents
Envronment
The devce s sutable for use by a professonal operator n
a professonal healthcare faclty and must be contnuously
attended when n use
Do not use the devce near HF surgcal equpment, or outsde
the RF shelded room of a medcal system for magnetc
resonance magng where the ntensty of EM dsturbances s
hgh
Indcatons for use of the devce
The devce s ntended to provde contnuous mechancal
ventlatory support The devce s a restrcted medcal devce
ntended for use by qualfed traned personnel under the
drecton of a physcan Specfcally the ventlator s applcable
for adult and paedatrc patents
The devce s ntended for use by healthcare provders, e
Physcans and Nurses, n ntensve care applcatons
Oxygen Montor
The Oxygen Montor s ntended to contnuously measure and
dsplay the concentraton of oxygen n breathng gas mxtures,
and s ntended for adult and paedatrc patents The oxygen
montor s an ntegral part of the ventlator
The oxygen montor s ntended for use by health care
provders, e Physcans and Nurses, for use wth patents n
ntensve care applcatons
2. Purpose

ESO 2 IU Ventlator 6 User Manual
3. Description
3.1 ESO 2 System Description
WARNIN
Important Delverng hgh fresh gas flows when usng closed
endotracheal sucton
Ths ventlator s confgured to delver flows of oxygen n excess
of the maxmum shown on the flowmeter tube n order to fll the
ventlator bellows durng closed endotracheal sucton Up to
approxmately 30 L/mn can be delvered
Delverng O flows n excess of the maxmum calbrated flow
wll temporarly negate the possblty of accurate volume
delvery and measurement Once the ventlator bellows s full
the O flow should be reduced to that n range of the calbrated
flow tube, then accurate volume delvery and measurement wll
resume
See secton 53 for addtonal nformaton
3.1.1 Major sub-systems
1 Ventlator dsplay screen
210 mm (84 nch) hgh defnton, colour TFT screen, wth
sngle/dual waveform dsplay
2 Ventlator control unt
Pneumatcally drven, software controlled, mult-mode
ventlator
Drve gas supply Ar at 280 to 600 kPa (41 to 87 ps)
See also secton 32
3 as flowmeter system
4 Frame assembly
Frame constructon Alumnum
WARNIN KEEP UPRIHT
The devce must be level and secure
If the unt s not mounted on a trolley, do not stand on the
floor Place on a flat, stable surface to mnmze the rsk
of mpact or accdental movement
5 Breathng crcut module
Transfers the gas mxture to and from the patent
6 Bellows
7 Inspratory and expratory hose connectors
8 Inspratory and Expratory non-return valves (NRV)
The valves a control the drecton of the gas flow through
the system
Each valve conssts of a dsc located over a valve seat
The dscs operate by gravty and are retaned by gudes to
prevent lateral movement
9 Oxygen montor sensor
10 as delvery On/Off swtch
11 Oxygen supply falure ndcator
reen ndcates OK
Red ndcates oxygen supply falure
5
4
3
1
6
8
7
10
11
9
2

ESO 2 IU Ventlator 7 User Manual
Description
3.2 Ventilator
3.2.1 Description
• The AV-S Ventlator s a pneumatcally drven, software
controlled, mult-mode ventlator
• The ventlator s tme-cycled, volume/pressure controlled,
and pressure lmted
• The ventlator has complance compensaton and fresh
gas compensaton
• User-selectable gas mxture compensaton s a standard
feature, plus a user-selectable varable nspratory pause
and sgh opton
Ventlaton Modes
• Volume Mode contnuous mandatory ventlaton
• Pressure Mode pressure controlled mandatory ventlaton
• PEEP avalable n volume and pressure modes
• Spontaneous and advanced patent support SIMV, SMMV,
PSV
See Appendx
Patent Montorng
• Arway pressure, measured from the expratory lmb of the
breathng crcut
• Tdal Volume and Mnute Volume dsplay s provded by a
dual sprometry system
• An ntegral oxygen montor system measures oxygen
concentraton n the breathng crcut nspratory lmb
• The prnt functon provdes a permanent record of
ventlator functon actvty for up to eght hours procedure,
or can be used to record waveforms
Drve gas supply
WARNIN
1 The ventlator s desgned to be drven by medcal ar
only
The drve gas s set durng manufacture and the
ventlator s calbrated for that gas
Before the ventlator s used clncally for the frst tme,
the commssonng engneer must confrm that the drve
gas settng s set correctly to Ar
The use of any other gas wll cause naccurate operaton
and may damage the ventlator, resultng n potental
nury to the patent
1 The drvng gas supply must be clean and dry to prevent
ventlator malfuncton
• Supply pressure Ar at 280 to 600 kPa (41 to 87 ps)

ESO 2 IU Ventlator 8 User Manual
Description
3.2.2 Electrical System
The ventlator has a sngle connecton (1) to a mans power
supply, and an ntegral battery backup
WARNIN
1 Always use a mans power outlet socket that can be
easly accessed
2 To solate the ventlator from the mans power supply
Dsconnect the mans cable from the hosptal power
outlet
Mans Supply
The mans supply nlet s desgned for connecton to the
followng mans voltage supples
200 to 240 VA, 50 to 60 Hz, 06 A max
110 to 120 VA, 50 to 60 Hz, 12 A max
Note that the ventlator adusts automatcally to the supply
voltage range
The connector s a standard IE type
Back-up Battery
In the event of mans electrcal falure, the back-up battery
cuts n automatcally The back-up battery wll power the
ventlator for the perod lsted below f the battery has been
mantaned n a fully charged condton
NOTE
Manual Delvery of Fresh as s not possble
Battery back-up perod
A fully charged battery wll power the ventlator for
approxmately 60 mnutes (dependng on ventlator settngs)
hargng the battery
Battery chargng takes place automatcally when the ventlator
mans lead s connected to a ‘lve’ mans supply
The power on/off ndcator on the ventlator front control panel
wll show an amber lght durng chargng (see secton 323)
NOTE
1 The stated battery back-up perod wll only be avalable f
the battery s kept fully charged
2 If the battery has been allowed to dscharge below the
LOW BATTERY condton, the ventlator wll not functon
correctly untl the voltage rases above the LOW BATTERY
level
3 A fourteen hours recharge wll be necessary to brng the
battery to full charge
Battery care
See Appendx 1 for battery care procedures
1

ESO 2 IU Ventlator 9 User Manual
Description
323 User Interface
ontrol Panel and Touchscreen
ontrol Panel
1 On/Off control
Swtch On Short nternal test sequence
Swtch Off Power down sequence wth progress ndcator
2 Status ndcators for electrcal power (mans/battery
supply)
Amber ndcator - llumnated whenever power s appled
to the unt and nternal battery s beng charged
reen ndcator - llumnates when the unt s swtched on
3 Menu swtch
Press to gan access to the man menu
All avalable menus can be accessed through the man
menu system when the ventlator s n Standby mode
Selecton and modfcaton of menu optons s through
the navgator wheel (see 324) Menu optons are context
senstve
Lmted access s provded n other modes
4 Alarm mute swtch
30 second or 20 second alarm slence, dependng on alarm
status Note also that some alarms are not mutable
5 Navgator Wheel and Press Button
Turn the wheel Select a functon or parameter, alter the
value of an actve parameter Press to confrm the settng
Touchscreen
6 Actve Tabs
Touch the screen at the approprate tab area to actvate
the requred functon/parameter
7 Fast-access Menus
Use the touchscreen to access the menus n the actve
touch-selectable areas (A to E)
Sub menu Touch the waveform area (A)
Waveform Touch the waveform area (A)
Alarms Touch the alarm dsplay area (B)
as Mxture Touch the gas mxture dsplay ()
Oxygen alarm Touch the oxygen montor area (D)
Note the montor must be ‘on’
Breathng modes Touch the mode con area (E)
VTSET TPS INT
VTSET TPS INT SET SET PEEP LIMIT
STANDBY
SPONT
VOLUME
Area A Menu and sub-menu wndow
Waveform dsplay, pause and prnt symbols
Area B Alarm values wndow
Area as mxture selecton wndow
Area D Oxygen montor values wndow
Area E Breathng mode symbols
6
PRESS
PSV
SIMV
SMMV
Area A
Area
Area D
Area B Area E
7
SET SET PEEP LIMIT
STANDBY
SPONT
VOLUME
PRESS
PSV
SIMV
SMMV
5
3214

ESO 2 IU Ventlator 10 User Manual
Description
Parameter Dsplay Identfcaton
1 Actve Mode
Black text on rey
(see, for example, 1A)
Note n Standby Mode the last used mode s shown wth an
underlne
(see, for example, 1B)
2 Actve Parameters
Actve parameters that can be set for use n the current mode are
dsplayed as
Whte text on Blue
3 Inactve Parameters
Inactve parameters that can be set for any non-current mode are
dsplayed as
Whte text on Blue Label (3A)
Whte values on Black (3B)
4 Measured Parameters
Yellow values on Black
5 Breathng mode nformaton symbols
Paedatrc mode Inspratory pause
Ventlator mode
Adult mode Sgh
Bag mode
1B
5
2
1A
3A
3B
4
4
Parameters

ESO 2 IU Ventlator 11 User Manual
Description
3.2.4 On-Screen Menus
To Access
Press the menu swtch (1) on the front panel to access the
followng functons and parameters va drop-down menus
Man Menu
EXIT MENUS
O MONITOR & SPIROMETRY
LEAK TEST
FRESH AS OMPENSATIONON
MODES
WAVEFORM
ALARM SETTINS
AS MIXTURE O+AIR
SERVIE MENU
To Ext
Press the menu swtch (1) on the front panel, or, select EXIT
MENUS and press the navgator wheel (2)
NOTE
The menu wndow wll not be dsplayed f
1 ontrol parameters (VTMEAS, BPM, IE, PEEP, or LIMIT)
are enabled but not confrmed
2 A dsplay wndow s actve
To Operate
1 Rotate the navgator wheel (2) clockwse to scroll through
the menu optons - the cursor ( > ) algns wth each
parameter n turn
2 Press the wheel to enter the requred sub-menu
3 Rotate the navgator wheel to change any dsplayed values,
and press to confrm
4 To ext the menu dsplay
a) Press the menu swtch on the front panel
b) Scroll to EXIT MENUS, and press the navgator wheel
NOTE
1 If confrmaton does not take place wthn 8 seconds, the
parameter reverts to ts prevous value
2 If another parameter s selected usng the touchscreen,
the menu s de-selected
3 Whle any menu s selected
a) the alarms are actve
b) the ventlator can be swtched off
See Appendx for a further nformaton on Menu structure
2
1

ESO 2 IU Ventlator 12 User Manual
Description
3.2.5 Start-up Screens
1 Start-up
a) At start-up, the software verson
s dsplayed
b) The ntroducton screen allows
the user to select one of three
default settngs
ADULT DEFAULTS
PAEDIATRI DEFAULTS
SITE DEFAULTS
NOTE
a) The user must select one of the
above default groups before the
ventlator wll swtch to standby
n that default mode
b) SITE DEFAULT s edtable n
standby mode (see below)
c) Settngs can be saved va the
servce menu to create a new
ste default
2 Default Settngs Selecton
The user can select ADULT, or
PAEDIATRI, or SITE, and vew the
default parameter settngs
The optons wll reman, even after
the ventlator s turned off
Ste Default Settngs
(NOTE The values lsted are
examples )
Adust the parameter values from
wthn the Servce menu (SITE
DEFAULTS) Press to confrm the
new settngs for ste defaults
3 Touch Screen albraton
Reconfgure the touch screen
operatng system
<Adult Defaults<
VTSet 600 mL
VMSet 60 Ltres
T+PS INT 10 cmHO
Set BPM 10
I E 1 20
PEEP OFF
Lmt 38 cmHO
Trgger 10 L/mn
Apnoea Alarm
Lmt 15 secs
Volume Type Tdal
SITE DEFAULTS
VIEW
<Paedatrc Defaults<
VTSet 150 mL
VMSet 22 Ltres
T+PS INT 10 cmHO
Set BPM 15
I E 1 20
PEEP OFF
Lmt 38 cmHO
Trgger 10 L/mn
Apnoea Alarm Lmt 15 secs
Volume Type Tdal
ALIBRATE TOUH SREEN
ALIBRATE TOUH SREEN
<Ste Defaults<
VTSet 800 mL
VMSet 60 Ltres
T+PS INT 10 cmHO
Set BPM 10
I E 1 20
PEEP OFF
Lmt 38 cmHO
Trgger 10 L/mn
Apnoea Alarm Lmt 15 secs
Volume Type Tdal
ALIBRATE TOUH SREEN
PAEDIATRI DEFAULTS
VIEW
ADULT DEFAULTS
VIEW
SITE DEFAULTS
PRESS TO ONFIRM
PAEDIATRI DEFAULTS
PRESS TO ONFIRM
ADULT DEFAULTS
VIEW
SITE DEFAULTS
VIEW
PAEDIATRI DEFAULTS
VIEW
ADULT DEFAULTS
PRESS TO ONFIRM

ESO 2 IU Ventlator 13 User Manual
Description
3.2.6 Selecting Functions and Parameters
Functons/parameters shown on screen can be actvated as follows
1 Touch the screen at the approprate tab area, or
Rotate the navgator wheel and press t when the ndcator arrow s on the requred parameter tab
Note that unless Ste Defaults are selected (see above) parameters default to factory-set values for Adult
or Paedatrc patents when the ventlator s swtched on from OFF, and no further user selecton s made
3.2.7 User Adjustable Parameters
To alter a varable parameter, rotate the navgator wheel
When the requred value s dsplayed, press the actve tab, or the wheel, to confrm the settng
• Tdal Volume Range 20 - 1600 ml
• Rate 4 -100 bpm
• IE Rato 102 to 18
• PEEP 4 - 20 cmHO
(opton 4 - 30 cmHO)
PEEP can be set to OFF
• Pressure lmt - Volume mode 10 - 80 cmHO
• Pressure range - Pressure mode 5 - 70 cmHO
• Alarm lmts (user adustable alarms only - see 312)
• T+PS INIT (target and pressure support ntal value)
The ntal pressure value can be changed so that when enterng ether PRESSURE mode or PSV
(pressure supported ventlaton) mode the TARET value or PSUPP value are pre-selected
Note that changng ether of these lmts n ther actve modes wll mantan the value when changng
between PSV, PRESSURE, and STANDBY modes
3.2.8 Output Compensation Functions
WARNIN
The ventlator automatcally compensates for fresh gas flow (sprometry On), fresh gas mxture
(sprometry and oxygen montor On), and alttude
However, the actual tdal volume delvered to the patent may be dfferent to the ventlaton parameters
set by the user, due to
1 omplance effect
2 A substantal system leak
3 Patent crcut resstance effects
4 Extreme fresh gas flows
In addton, hgh fresh gas flows wll lead to an ncreased VTbeng delvered to the patent
The patent must be montored ndependently from the ventlator It s the responsblty of the user to
montor the patent for adequate ventlaton
WARNIN
Important
Delverng hgh fresh gas flows
1 Hgh fresh gas flows when usng closed endotracheal sucton Ths ventlator s confgured to delver
flows of oxygen n excess of the maxmum shown on the flowmeter tube n order to fll the ventlator
bellows durng closed endotracheal sucton Up to approxmately 30 L/mn can be delvered
Delverng O flows n excess of the maxmum calbrated flow wll temporarly negate the possblty of
accurate volume delvery and measurement Once the ventlator bellows s full the O flow should be
reduced to that n range of the calbrated flow tube, then accurate volume delvery and measurement wll
resume
See secton 53 for addtonal nformaton

ESO 2 IU Ventlator 14 User Manual
Description
3.2.8.1 Fresh Gas Compensation
Adusts delvered volume up to 60%
An alarm s trggered f measured volume vares by 50% from
the set volume
Ths functon s user adustable (329)
NOTE
Fresh gas compensaton s dsabled f
The sprometry system s turned OFF through the menu
system, or the sprometry system s malfunctonng
3.2.8.2 Fresh Gas Mixture Compensation - models with
Spirometry
The sprometry system compensates for the fresh gas mxture
n use
The as Mxture wndow (Area ) s an actve touch-selectable
area (n any mode), wth a drop down menu
1 Touch the screen n Area
2 Use the wheel to select O + AIR
Do not select O + NO
3 Press to confrm or, use the menu structure, as shown
Select the requred mxture wth the scroll wheel
NOTE
1 Fresh gas mxture compensaton s dsabled f
The sprometry system s turned OFF through the menu
system, or the sprometry system s malfunctonng
2 If the oxygen montor s swtched OFF, a default mxture s
assumed
3.2.8.3 Compliance Compensation
The ventlator wll apply complance compensaton to account
for complance loss n the breathng system n cases where
Fresh gas compensaton s dsabled, or
Sprometry s unavalable or dsabled
NOTE
For correct operaton the value of the breathng system
complance must be establshed frst, by completng the
ventlator leak-test as part of the Pre-operaton Procedure
Refer to secton 4, notng that breathng system complance s
dsplayed as ‘Bsyscomp’
If the leak test s not carred out, the default value wll be used
(7 ml/cmHO)
NOTE
In complance compensaton mode any fresh gas used wll be
n addton to the set tdal volume
3.2.8.4 Altitude Compensation
Ths functon montors ambent pressure, and adusts the
delvered volume accordngly
Alttude compensaton s automatcally appled durng
calbraton of the oxygen montor - see secton 43
Area A
Area
Area D
Area B Area E
Table of contents
Other Penlon Medical Equipment manuals