Planer DATAssure D3 Manual

User Manual, Set Up & Installaon Guide
Covering
DATAssure D3
Monitoring System Base Staons

Page No. 2
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
Systems covered by this manual;
UK, Europe & Zone 2 (For 2.4Ghz systems)
P1-DATAssure D3101-0433 DATAssure D3 LITE 433Mhz
P1-DATAssure D3102-0433 DATAssure D3 433Mhz
P1-DATAssure D3103-0433 DATAssure D3 MEDICAL 433Mhz
P1-DATAssure D3104-0433 DATAssure D3 BRIDGE 433Mhz
P1-DATAssure D3101-2400E DATAssure D3 LITE 2.4Ghz Europe
P1-DATAssure D3102-2400E DATAssure D3 2.4Ghz Europe
P1-DATAssure D3103-2400E DATAssure D3 MEDICAL 2.4Ghz Europe
P1-DATAssure D3104-2400E DATAssure D3 BRIDGE 2.4Ghz Europe
Zone 1 (these units are not CE marked an must not be used in Europe)
P1-DATAssure D3101-2400 DATAssure D3 LITE 2.4Ghz
P1-DATAssure D3102-2400 DATAssure D3 2.4Ghz
P1-DATAssure D3103-2400 DATAssure D3 MEDICAL 2.4Ghz
P1-DATAssure D3104-2400 DATAssure D3 BRIDGE 2.4Ghz
Zone 1 & Zone 2
Refer to the radio power secon in the Technical Specicaon at the rear of this document for the transmission power of each system. Ensure
the correct system has been ordered to meet the radio power requirements for the territory the systems will operate within. If you are unsure
please contact your supplier for assistance.
Approvals
CE
Hereby, Tek Troniks declares that the radio equipment type; P1-DATAssure D3101-0433, P1-DATAssure D3102-0433, P1-DATAssure D3103-
0433, P1-DATAssure D3104-0433, P1-DATAssure D3101-2400E, P1-DATAssure D3102-2400E, P1-DATAssure D3103-2400E & P1-DA-
TAssure D3104-2400E is in compliance with Direcve 2014/53/EU. The full text of the EU declaraon of conformity is available from the
manufacturer at the following address: Tek Troniks Ltd, Unit 21 Manvers Business Park, High Hazles Road, Cotgrave, Nonghamshire, United
Kingdom, NG12 3GZ
FCC & ISEDC
The following products comply with FCC & ISEDC; P1-DATAssure D3101-2400E, P1-DATAssure D3102-2400E, P1-DATAssure D3103-2400E,
P1-DATAssure D3104-2400E, P1-DATAssure D3101-2400, P1-DATAssure D3102-2400, P1-DATAssure D3103-2400 & P1-DATAssure
D3104-2400.
FCC
Contains FCC-ID: QOQMGM12P3
ISEDC
Contains Transmier Module IC: 5123A-MGM12P3

Page No. 3
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
1. SAFETY INFORMATION
Read all the instrucons carefully before using the base staon and keep a copy of all documents for future review.
1.1. Electrical Safety
Before mounng/dismounng the base staon and/or connecng the power supply ensure the units is powered down and switched o at
the mains. Any connected peripheral, hard wired devices with live feeds should also be powered down before mounng/dismounng the
base staon.
Before opening the base staon casing to change the baery back-up baery, isolate the electrical supply and also the live feed to any hard
wired peripheral connected equipment.
Before connecng peripheral, hard wired devices with live feeds, power down and disconnect the base staon from the main supply.
Cauon, live feeds may derive from more than one source.
If the power supply or supply cable is damaged then it must be replaced, switch o the unit at the mains and contact the system supplier.
1.2. Baery Back-Up² Baery
The baery in the baery back-up² must be disposed of in accordance with your local laws & regulaons, or ship to the supplier in the new
baery packaging.
Replace the baery with the same part number as the type originally installed. Replacement instrucons are contained with the new baery
shipped.
CAUTION RISK OF EXPLOSION IF BATTERY IS REPLACED BY INCORRECT TYPE
Do not dispose of the baery in a re, the baery may explode.
Do not open or mulate the baery. They contain an electrolyte which is toxic and harmful to the environment and to the skin and eyes.
1.3. Damage
Inspect the base staon & power supply for damage aer unpacking.
If the base staon is damaged during normal operaon and you are unsure of its operaonal performance, switch-o at the mains & unplug.
Contact the system supplier for assistance.
The unit contains no user serviceable parts, the baery back-up² baery should only be replaced by suitable trained personnel. Do not
aempt to disassemble the unit or aempt repair.
1.4. Locaon of the unit
The base staon should be located on a at, even wall on the mounng plate supplied. The unit slides from right to le onto the mounng
plate, at least 30cms needs to be free of obstacles to the right of the unit for future mounng/dismounng.
Avoid mounng the unit next too or directly beneath electrical conduit, this may impact the wireless range of the unit
Ensure the mains socket and network port are near to the mounng point and are easily accessible
1.5. Temperature & Moisture
The base staon should be mounted in a dry environment with an operaonal temperature range of between 5°C to 55°C.
To reduce the risk of damaging the unit, causing a re or personal electrical injury, do not expose the base staon to rain, moisture, dripping
or splashing.
Note - ² Baery back-up not available on LITE versions
1.6. Loss of Life & Personal Harm Applicaons
The DATAssure D3 family of monitoring systems must not be used in life support, crical care, medical, safety equipment, or similar applica-
ons where product failure could result in loss of life or personal or physical harm.. The systems can be used for environment monitoring to
gain performance data but must not be used as a primary alarm system.

Page No. 4
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
System Overview
The DATAssure D3 family of monitoring system base staons are designed to communicate with the DATAssure D3 range of environment probes.
Probes communicate with their assigned base staon each minute, the probes recorded data is; automacally reviewed, assigned alarm acons (if
necessary) and stored within the base staon for later access.
Funconality across the DATAssure D3 family of base staons varies, this manual covers all base staon systems. Dierences in funconality will be
noted against each secon within the document as necessary.
Refer to the Set-up & Installaon Guide secon of this document for powering up, power connecon, networking, mounng of the base staon &
Administrator level access acons.
DATAssure D3 Base Staon Assembly
Front
1 - Colour Touch Screen
2 - Indicaon LED (Green & Red)
3 - Aerial Enclosure
4 - Micro USB Port
5 - On/O Switch
Back
6 - Wall Plate Fixing Points
7 - Power Connector
8 - Mini Din Connector
9 - Duel Relay Output
10 - Ethernet Connector
11 - Micro MMC Slot¹
Notes
¹ - Not acvated with DATAssure D3-LITE Soware Conguraon

Page No. 5
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
Table of Content
Systems covered by this manual; 2
1. SAFETY INFORMATION 3
1.1. Electrical Safety 3
1.2. Baery Back-Up² Baery 3
1.3. Damage 3
1.4. Locaon of the unit 3
1.5. Temperature & Moisture 3
1.6. Loss of Life & Personal Harm Applicaons 3
D3 Base Staon Assembly 4
2. Base Staon 9
2.1. Indicator LED 9
2.2. Base Staon Touch Screen 9
2.3. Status 9
2.4. List 9
2.5. Alarms 9
2.5.1. Alarm Acknowledgement via Base Staon 10
2.6. SET UP 10
2.6.1. Edit IP Conguraon (at Base Staon) 10
2.6.2. Re-Calibrang the Touch Screen 11
2.6.3. Drive 11
2.6.4. Ulies 11
2.7. Informaon 11
3. Networked Interface 12
3.1. Accessing the D3 Base Staon 12
4. Site (Default Screen on Login) 12
4.1. Sensor Readings & Meaning of Status Messages 13
4.1.1. Sensor Readings 13
4.1.2. Status Messages 13
4.2. Sensor Details 13
4.2.1. Edit Alarm Set Points 14
4.2.2. Warning level alarm set points (single points) 14
4.2.3. Warning & Crical level alarm set points 14
4.3. Sensor Inhibit/Clear Inhibit 14
5. GRAPHS 14
5.1. Graphing Tool 14
5.2. Graph View 15
6. ALARMS 15
6.1. Alarm Details 16
6.1.1. Adding an Alarm Note 16
6.2. Unacknowledged Alarms 16
6.2.1. Acknowledging an Alarm 16
6.3. Historical Alarms 16
6.4. Audit Informaon 17
7. REPORTS 17
7.4.1. Select Date 17
7.4.2. Daily Overview Report 17
7.4.3. Daily Sensor Report 17

Page No. 6
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
8. Users Login Details 17
8.1. Updang/Changing User Login Details 17
8.2. Forgot Login Details 17
Set-Up Guide 18
9. Powering the D3 Base Staon 18
9.1. Powering Up 18
9.2. Powering Down 18
10. Networking the Base Staon - IP Connecvity 18
10.1. IP Setup via the Base Staon 19
10.1.1. DHCP Client Mode 19
10.1.2. DHCP Server Mode 19
10.1.3. Stac Mode 19
10.2. IP Setup via the Network Interface 19
10.2.1. DHCP Client Mode 19
10.2.2. DHCP Server Mode 19
10.2.3. Stac Mode 19
SETTINGS 19
11. Sensor Administraon 20
11.1. Manage Devices 20
11.2. Add a Device/Sensor 20
11.2.1. Adding the Device/Sensor 21
11.2.2. Linking a Device 21
11.2.3. Legacy Sensors - compable with 433Mhz base staons ONLY 21
11.2.4. Intelligent Sensors 22
11.3. Adding a Device Group 22
11.4. Deleng, Disabling & Decommissioning Devices & Groups 23
11.4.1. Changing the Device’s State 23
11.4.2. Deleng a Device 23
11.4.3. Disabling a Device 23
11.4.4. Decommission a Device 23
11.4.5. Deleng a Group 23
11.5. Eding a Device/Group 23
11.5.1. Changing a Device Reference and Name 23
11.5.2. Adding or removing a Device from a Device Group 23
11.5.3. Changing the Device Type 23
11.5.4. Changing the Device Display Resoluon 23
11.6. Custom Sensor Database 24
11.7. Adding a Custom Sensor Type 24
11.7.1. Creang Pulse Kwh Sensor Type 24
11.7.2. Creang a Pulse Flow Sensor Type 24
11.7.3. Creang a Digital Sensor Type 24
11.7.4. Creang a Voltage, Current & Resistance Sensor Type 24

Page No. 7
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
12. User Sengs 25
12.1. My Details 25
12.2. Seng Password Expiry 25
12.3. User Management 25
12.3.1. Default Administrator Passwords/Login Details 25
12.3.2. Adding Users 26
12.3.3. Eding an Exisng User 26
12.3.4. Enable/Disable Users 26
12.3.5. Reseng a User Passwords 26
12.3.6. Deleng a User 26
13. Alarms 27
13.1. General Alarm Sengs 27
13.2. Alarm Proles 27
13.2.1. Creang an Alarm Prole 28
13.2.2. Eding an Alarm Prole 28
13.2.3. Adding/Eding a Sensor/Devices Alarm Prole 28
14. Site Sengs 29
14.1. System Setup 29
14.1.1. Adding A Site Name 29
14.1.2. Dimming/Brightening & Display Sleep Mode 29
14.1.3. Daylight Saving 29
14.1.4. Change Displayed Temperature (°C or °F) 29
14.1.5. Change the Default Displayed Resoluon 29
14.2. Scheduler 29
14.2.1. Seng/Eding a Schedule 29
14.2.2. Example - Acvate Inside Set Hours Schedule 30
14.2.3. Example - Acvate Outside Set Hours Schedule 30
14.2.4. Adding a Auto Inhibit Schedule to a Sensor/Device 30
14.3. Seng/Eding Time & Date 30
15. Networking 30
15.1. Email Setup/Eding 30
15.2. Bureau Access Setup 31
15.3. IP Dial Out Setup 31
15.4. Upload SSL Cercate 31
16. Administrator 31
16.1. Soware Update 31
16.2. Backup Conguraon 31
16.3. Restore Conguraon 31
16.4. Restore Conguraon to a Replacement Base Staon 31
16.5. Maintenance 32
16.5.1. Cancel all queued alarms 32
16.5.2. Re-index archive les 32

Page No. 8
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
Installing the Base Staon 33
17. What’s in the Box 33
18. Mounng the Base Staon 33
18.1. Power Supply & Ethernet Connecon 33
18.1.1. Power Supply 33
18.1.2. Ethernet Connecon 33
18.2. Mounng Wall Plate 33
18.3. Mounng/Dismounng the Base Staon 34
18.3.1. Mounng the Base on Wall Plate 34
18.3.2. Dismounng the Base from Wall Plate 34
19. Connecng to the Internal Relays 34
19.1. Relay 1 - Digital Relay Output 34
19.2. Relay 2 - Solid State Relay Output 34
20. Changing the Micro MMC Card 34
21. Baery Replacement 35
21.1. Coin Cell 35
21.2. Baery Back-Up (if installed) 35
21.3. Wireless Transmier Baery 35
21.3.1. Lithium AA Cell - Legacy Sensor can only use this baery 35
21.3.2. Alkaline AA Cell - 35
Annex 1 - Technical Specicaon 36

Page No. 9
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
User Manual
2. Base Staon
2.1. Indicator LED
During normal operaon the Indicator LED will show a constant
Green, indicang power is on and no alarms are currently acve.
Should an alarm occur then the LED ashes Red unl the alarm
is acknowledged. Once acknowledged the LED will be a con-
stant Red unl the alarm status has cleared.
Should a new alarm occur the LED will return to ashing Red
unl acknowledgement is completed.
2.2. Base Staon Touch Screen
The touch screen allows for selecon of various parameters by
rmly pressing on the icon/opon required.
The system date and me are shown on all system versions. The baery charge indicator shows the current charge level on the internal
baery back-up².
The home screen (seen above) allows access to all grouped areas on the system. The Icons displayed on the home screen are dened below.
Note -
Do not use sharp or pointed object as a selecon tool as these may damage the screen surface.
Within each screen there are acon icons and/or text boxes which allow acons to be completed.
2.3. Status
HOME => STATUS
The default screen for viewing each sensors live data individ-
ually, the Status screen shows the sensors; ID, Name, Current
Reading, Current Status with colour nocaon;
Green – OK
Amber – alarm delay count down
Red – sensor in alarm
By pressing the Info icon on the sensor status screen further
informaon (set points & alarm delays) can be viewed.
Each sensors status screen will scroll automacally aer 7
seconds, this allows for a connuous view of the current sensor
readings to be displayed. Sensor readings can be scrolled
through using the up & down icons, when scrolling ends the
current sensors readings will be displayed for 30 seconds before the auto scrolling recommences.
2.4. List
HOME => LIST
An overview list of all of the paired sensors sorted by their ID reference (lowest to highest). Shown are the sensor ID, Name, Current Reading
(with status denoted by colour) and the sensor reading type – e.g. °C or %RH.
The list can be scrolled using the up & down acon icons.
2.5. Alarms
HOME => ALARMS
An overview list of all unacknowledged and current alarms. The
list can be scrolled using the up & down acon icons.
Unacknowledged alarms are shown in Red, newly acknowledged
(via the base staon) & current alarms are Black.
NOTE -Alarms can be acknowledged by a user with the appropriate access
level on the base staon, however alarms acknowledgement on the base
staon is completed on each individual alarm in turn.
Acknowledgement via the networked access allows for a user to choose
mulple alarms and single click acknowledgement – see secon 6.2 Unac-
knowledged Alarms

Page No. 10
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
2.5.1. Alarm Acknowledgement via Base Staon
HOME => ALARMS => SELECT => ACKNOWL-
EDGE ALARM =>ACK
From the alarm list select the alarm to be ac-
knowledged using the up & down acon icons.
Once selected (highlighted with the black se-
lecon bar) press the alarm acknowledge acon
icon. The alarm details will be displayed, select
ACK (acknowledge alarm) and the Authorise
screen will appear.
Press within the Username box, the list of
system users will be displayed. Press or select
the relevant USER ID then press the Enter icon.
Now press within the Passcode box, the Pass-
code numeric keypad will display, enter pass-
code then press the keypad enter key. On the
Authorise screen press the CONFIRM buon.
The alarm details screen will now show the
alarm as acknowledge. Press the back icon to
return to the alarm list.
Note - The above steps need to be repeated for each alarm that
requires acknowledgement.
Set-Up HOME => SET UP
These screens are passcode protected, only system Admin-
istrator access level users have access to the Set-up menus.
• Home
Pressing home icon returns the screen to the home screen
• IP
Opens the IP conguraon menu where the IP sengs for
the system can be set (from new DHCP Client & DHCP
Sever ONLY)
• Touch
Allows the touch screen to be recalibrated
• Drives
Allows for conguraon of the system drives
2.6. SET UP
2.6.1. Edit IP Conguraon (at Base Staon)
HOME => SET UP => IP => CHANGE
From manufacture the base staon is shipped in DHCP Cli-
ent conguraon, this allows the network to assign the IP
sengs for the system automacally. The IP conguraon
at the base staon only needs to be enacted if the unit is to
be set up via a PC/Laptop prior to network installaon.
Pressing the change buon will open the network mode
conguraon
• Stac
The Stac IP conguraon can only be entered/updated via
the systems network interface. Once set the base staon
retains the Stac IP informaon allowing the unit to be
switched to Client or Server mode during maintenance (if
required) and back to Stac via this menu.
• DHCP Client
Default - allows the network to assign all required IP informaon for instant access via networked PC’s/Laptops or mobile devices.
• DHCP Server
Allows the base staon to be directly connected to a PC or Laptop, the IP is set at 192.168.100.100 and does not change. Opening a
web browser will allow access to the base staon log-in screen when this IP is entered in the address bar.
Select the relevant IP mode required and Accept.
The base staon will now be operang in the selected mode.

Page No. 11
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
See secon 10. Networking the Base Staon for more informaon on IP conguraon.
2.6.2. Re-Calibrang the Touch Screen
HOME => SET UP => TOUCH => START
The touch screen is calibrated at manufacture, should re-calibraon be required this can be done via the touch menu.
Within the touch menu simply press Start icon and follow the on-screen instrucon.
2.6.3. Drive
HOME => SET UP => DRIVES
The drives menu details the current informaon on the base staon internal and external memory drives.
2.6.4. Ulies
HOME => SET UP => UTILITIES
Allows for higher level acons to be taken on the base staon via its inbuilt USB drive and restoraon of conguraon from the uSD -
micro SD card if installed, contact your system supplier for more informaon.
2.7. Informaon
HOME => INFO
This screen gives an at a glance overview of key factors on the system set-up. It shows;
Network mode - shows the current selected network connecon mode of the system
IP address - shows the assigned IP to be used in an internet browser for networked access to the system
IP Subnet, Gateway & Preferred DNS - IP sengs assigned (preferred DNS will be blank in DHCP server mode)
IP Mac Address – the base staons unique mac address
Hp Port – the IP connecon port assigned
Baery – shows if a baery back-up is installed on the system
Version - shows the soware version on the base staon and the radio soware version and frequency.

Page No. 12
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
3. Networked Interface
The DATAssure D3 family of base staons are all designed to be accessed over a networked interface, the interface used is a standard internet
browser on a PC/Laptop or mobile device with access to the same network as the base staon. For the purpose of this manual Google Chrome
has been used.
3.1. Accessing the DATAssure D3 Base Staon
INTERNET BROWSER => LOGIN SCREEN => SITE PAGE
Each user will be assigned access privileges by the systems Administrator(s). When accessing the DATAssure D3 base staon a user’s access
privileges will determine what is seen on each screen. When creang users the Administrator can; manually create or enact an email for the
users to create their username and passwords.
To access the DATAssure D3 Base Staon open any standard internet browser on PC/Laptop or mobile device.
Within the browsers address bar type the base staons assigned IP address – e.g. 192.168.100.100 and press enter. The browser will con-
nect to the base staon and open the Login Screen. Enter the Username and Password provided by the Administrator and select the Sign In
buon.
The Main Site Page will open.
Note - Before accessing the system for the rst me or aer having a password reset, a user will be directed to a web login screen where they must change and
set their own secure login details. New users will also need to set their Base Staon password, see secon 12.1. My Details for more informaon.
Hint - Save the IP address to the browser favourites to make future access simpler.
4. Site (Default Screen on Login)
1. Main menu bar giving access to all system menu
pages
SITE (default page) – overview of sensor
details
ALARMS – Overview & Access to current,
unacknowledged and historical alarms and
audited system events
REPORTS – User generated reports
SETTINGS – The user privilege level will
determine the sengs page display
2. Current date and me
3. User Status Indicator – Click to log out of the sys-
tem, inacvity on the system for more than 5 minutes
will change the log in status of the user from Green
(logged in) to Grey (guest). As a guest the user can
view data but an aempt to access any restricted part of the system will require a re-login
4. View Bar
All – shows the readings and status of all sensor, with numerous sensors this page can be scrolled to view all
In Alarm – shows only those sensors currently in an alarm status, this is a view only screen, alarms can only be acknowledged from the
ALARMS menu – see secon 6. ALARMS
In Warning – shows only those sensors that are currently in a warning status and on alarm delay countdown, if the warning status remains
beyond the alarm delay period then these sensors will alarm.
5. Export Buons - allows the current screen data to be exported in various output types
6. Sensor Type – this can be expanded and individual or mulple sensor types can be chosen to be viewed by clicking and selecng them
7. Expand all - when sensors are grouped this allows the group to be expanded to show all sensors – see secon 11.3 Adding a Device Group for
more informaon on sensor grouping
8. Search - allows a search of the current viewed data by a user dened parameter
9. Sensor Detail & Details Bar
Details Bar– allows for the sensors to be quickly sorted by ID and current status by clicking on the sort buon next to the various headed
elements.
Sensor Details - sensor details are shown for each sensor on the system, it includes (from le to right)
Graphing Buon – clicking the icon opens the Graphing Tool for data review, for more informaon on the graphing see secon 5. Graphs
Sensor Descripon – clicking the underlined sensor descripon opens a Sensor Details page, see secon 4.2 Sensor Details for more
informaon
High/Low Alarm 1 – the warning alarm set points for the sensor (or sensor group if set)
High/Low Alarm 2 – the crical alarm set points for the sensor (or sensor group) if set
Current Reading – the most recent reading from the sensors
Status – the current status of the sensor

Page No. 13
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
The Site page is the default system landing page on login, it shows all paired sensors, their current readings and their status. Both the current
readings and status are colour coded to show those sensors that are OK, in Alarm or Warning.
Hint -No system changes can be administered on the SITE page itself, users should familiarise themselves with this page by clicking on the various elements.
Re-entering or refreshing the page will reset its default sengs.
4.1. Sensor Readings & Meaning of Status Messages
4.1.1. Sensor Readings
Power Up – shown on base staon at power up and will display unl a reading is received from the connected sensor, if no reading is
received the base staon will trigger a communicaons alarm
Current Reading – shows the last current reading of the sensor received by the base staon, the decimal placing of the reading will be as
dened by the system Administrator when seng the system up
Sensor Fault – shows that a paired sensor is faulty or not connected correctly to its wireless sensor or input module
Inhibit – The alarm status of the sensor has been inhibited, the current reading will show but no alarms will occur for this sensor while in
an inhibited status
4.1.2. Status Messages
OK – The sensors reading is within its dened high & low set points
Alarm Pending – The reading is outside of the dened set points and the set alarm delay period has been aconed, the remaining me
unl the status changes to alarm is also shown
Alarm – The reading is outside of the dened set points, the alarm delay has been exhausted and the sensor has alarmed
No Communicaons – The base staon has not received a reading from the sensor and an alarm has been triggered
4.2. Sensor Details
SITE => SELECT SENSOR
The sensor detail page gives a view of the full
current informaon for the sensor selected.
1. Current Sensors Name and Reference ID –
clicking on the 3 bar menu before the ref and ID
opens the sensor list where other sensor can be
selected. Sensors can be scrolled in the ID order
(low to high) by pressing the back and forward
arrow icons at each end of the name tab.
2. Graphing Tool, Signal Strengths and Baery
Level Icons –
Graphing Tool - selecng this icon will open
the graphing tool – see secon 5. Graphs
Wireless Signal Strength Indicator – this is
a simple display for the wireless signal strength. Four bars is full strength and no bars/red is low strength. A detailed view of the signal
strength and quality are available to the system Administrator privilege level, see secon 11.1 Manage Devices
Baery Wear Level Indicator - when red the sensor baery needs to be replaced.
3. The current live reading
4. The current sensor status
5. Alarm 1 & 2 Set Points & Delays – individual high and low set points and alarm delays can be set, this allows for a warning alarm set point
and a crical alarm set point to be created. See secon 4.2.1 Edit Alarm Set Points
If Alarm 2 is shaded then a 2nd alarm set point has not been set and the sensor will only alarm based on Alarm 1
6. Alarm Edit Buon – allows alarm set points to be amended, also allows Alarm 2 to be acvated.
Note - Only users with the correct access privileges can see the edit and amend alarm set points. All changes made are recorded into the base staon audit
history
7. Inhibit/Clear Inhibit Buon – allows for the sensors alarms to be inhibited and re-acvated. In an inhibited status the base staon will not
alarm. See secon 4.3 Sensor Inhibit/Clear Inhibit
Note - Only users with the correct access privileges can see, inhibit and clear inhibited sensors. All changes made are recorded into the base staon audit
history
8. Alarm Prole & Auto Inhibit – congures what acons the base staon will take in the event of an alarm.
Note - these funcons can only be set by the System Administrator privilege access level, see Secons 13.2 Alarm Proles & 14.2 Scheduler for more informa-
on.
Clicking the red cross in the le hand score of the detail screen returns the view to the default Status menu.

Page No. 14
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
4.2.1. Edit Alarm Set Points
SITE => SELECT SENSOR => ALARM SET POINTS => EDIT
Note - Only Users with the correct access privileges can see the edit alarm set points buon,
access levels are set by the System Administrator/s.
• High Alarm – the maximum point the sensor reading should achieve
before the alarm pending countdown iniates
• Low Alarm – the minimum point the sensor reading should achieve
before the alarm pending countdown iniates
• Delay – the amount of me of the alarm pending countdown
4.2.2. Warning level alarm set points (single points)
Set the required high, low and delay in Alarm 1, click the Submit buon to set
alarm set points, the screen returns to the Sensor Detail page
4.2.3. Warning & Crical level alarm set points
Set the required warning level high, low and delay in Alarm 1, then
Click the black toggle buon at the end of Alarm 2 to make it acve and editable. Set the required crical level high, low and delay in
Alarm 2. Click the Submit buon, both alarm set points will be set and the screen returns to the Sensor Detail page
4.3. Sensor Inhibit/Clear Inhibit
SITE => SELECT SENSOR => INHIBIT/CLEAR INHIBIT
Note - Only Users with the access privileges can see the Inhibit/Clear Inhibit alarm buon, access privileges are set by the System Administrator/s.
This funcon is oen used when a locaon being monitored is out of service or known to be faulty, by Inhibing the alarm it stops false
alarms occurring during a known outage.
To Inhibit Alarm – Click the Inhibit Alarm Buon, the sensor will be inhibited. At the next screen refresh the sensor status will change to show
as Inhibit, the Inhibit buon will now show Clear Inhibit. No alarm nocaons will be produced for this sensor while in an Inhibited status.
The current readings will sll be displayed.
To Clear Inhibit – Click the Clear Inhibit buon and the inhibit will be removed. At the next screen refresh the sensor status will change to
either OK, Alarm Pending or Alarm depending on the actual status of the sensor. Alarm nocaons will be produced from the this point
going forwards.
Note - See more on Auto Inhibit in secon 14.2 Scheduler for scheduling sensors to inhibit alarms at set me periods daily/weekly.
5. GRAPHS
5.1. Graphing Tool
The graphing tool can be entered on any page displaying the graphing icon by
clicking the icon.
1. Select Device – the default is the sensor where the graphing icon was selected,
however using the drop down box another sensor can be chosen.
2. Select Traces –
Default - Sensor Reading
Default - High Alarm 1 set point
Default - Low Alarm 1 set point
Default – Status
User Dened – High Alarm 2 set point
User Dened – Low Alarm 2 set point
User Dened – Communicaons – this shows the quality of the signal being received from the sensor by the base staon
User Dened – Signal Strength – this shows the strength of the signal be received from the sensor by the base staon.
Note – the default traces are loaded automacally when a date/date range is selected, therefore the traces only need to be selected if more than the default is
required.
3. Select Data Range – 4 selections are available, these are;
Today – Displays the data from 00:01am to the latest reading
Week – Displays the last 7 days of data
Month – Displays the last month’s data
User Dened – Displays the user selected date range (up to 1 year)

Page No. 15
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
5.2. Graph View
1. New Graph Trace, Pan Le, Refresh, Pan Right, PDF & CSV
New Graph Trace – opens the Select Trace tool and allows
addional traces to be added to the current selecon.
Pan Le & Right – allows for panning of the graph when
zoomed in.
Refresh – reverts the graph to 100% zoom level, this can also
be completed by double clicking on a zoomed graph.
PDF – downloads the selected element (see 2 below) of the
graph in a pdf format for prinng or saving.
CSV - allows for data in the selected element (see 2 below) of
the graph to be output into CSV format, users can select to
output data in 1, 5, 15 or 30 minute intervals, data in the se-
lected interval can also be averaged if required (if greater the
1 minute intervals are selected), these averages the readings
from the me over the interval selected and outputs this informaon.
2. Selected Traces – bringing the mouse cursor into trace area will bring a cursor line into view, the trace data will be shown in the trace detail
box (3 below). This cursor line can be moved along the trace, the trace data details will change to reect the data recorded at each point the
cursor line covers.
Click and drag to zoom in on a desired selecon of trace, data can be zoomed to the 1 minute interval period at which it is stored. Double
click within the selected graph to zoom out
3. Trace Detail Box – gives the id and name of traced sensor, shows the me, date and readings for the trace period covered by the cursor
line. Selecng the cross in front of trace element will remove that trace from the graph, the eye icon can be used to hide/show a trace but
not remove it from the graph. The + at the base of the box allows for addional traces to be added to a current graph. Clicking the top right
corner cross removes this sensor from the current graph
4. Brings up the Select Trace tool to allow for comparison traces to be added to the exisng graph. A comparison could be made for the same
sensor against a dierent me period and/or another sensor over the same or dierent me period.
Up to 4 sensors traces can be compared on the same graph by repeater step 5
When mulple sensor are being compared, each will show the me, date and selected data when the cursor line is moved over the graph,
zooming one will zoom all.
Trace colours can be changed by clicking the colour box in the trace box and selecng the desired colour from the colour swatch that
opens.
Clicking the red cross in the top right corner closes the graph and opens the Status page.
Hint - No system changes or data changes can be administered within the graphing tool, only data can be viewed, users should familiarise themselves with this tool by clicking on the various elements.
6. ALARMS
SITE => ALARMS
1. Selecon Tabs – the Alarms page is broken down in 4 areas, these
are;
• Current (default)– this is all alarms that are currently
on the system, this includes unacknowledged alarms. There-
fore the alarm status may have cleared and the sensor may no
longer be in alarm, but unl the alarm is acknowledged it will remain in the current alarm list to alert of the occurrence.
• Unacknowledged – this is all alarm events that have occurred and have not yet been acknowledged by an authorised user. Once
a user acknowledges an alarm it will be removed from this tab, if the sensor is sll in alarm it will remain in the Current tab, if the sensor’s
alarm has cleared it will be in Historical
• Historical – this is all the alarms that have occurred and stored in the system memory
• Audit – this is a record all alarm and user system change events stored within the system memory
2. Allows the current viewed data to be exported or printed/saved in various formats
3. Allows a search of the current viewed data by a user dened parameter
4. Descripon Bar – allows for the sensors to be quickly sorted by Alarm/Event ID, Date, Time, Device, Sensor Type, Alarm/Event by clicking on
the sort buon next to the various headed elements.
Graphing Buon – clicking the icon opens the Graphing Tool showing the point of the alarm occurrence
ID – this is the system generated unique ID given to the alarm event
Date & Time – show the system date and me when the alarm event occurred
Device - gives the device Ref and Name
Sensor Type – type of sensor that has alarmed
Alarm Event – gives reason for alarm event (Reading, Communicaons, etc), clicking the underlined alarm message will open the Alarm Details
Box where more informaon can been seen and alarm notes can be added
Status – shows the current status of the sensors readings; Alarm shows the sensor is sll in an alarm status, Cleared shows that the alarm has
cleared but indicates that the alarm event has not yet been acknowledged.

Page No. 16
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
Note - All system users can view the Alarms pages, however only users with the appropriate access privileges can acknowledge and add notes to an alarm
event. System access privileges are assigned to users by the System Administrator/s.
6.1. Alarm Details
SITE => ALARMS => select ALARM/EVENT
This box contains the details of an alarm event and allows for
alarm notes to be added, it lists:
Name – The system ID and name of the sensor
Serial Number – The serial number of the sensor
Sensor Type – The type of sensor
Details – the causal fault for the alarm
Reference (not seen on communicaons or fault alarms) – the
reading that caused the alarm
Occurred – the date and me the alarm event was triggered
Acknowledged – the date and me the alarm was acknowledged, if blank then the alarm has not yet been acknowledged
Acknowledged by – the details of the user who acknowledged the alarm, if blank then the alarm has not yet been acknowledged
Cleared, if the sensor has returned to within its dened set points then the point at which this occurred and the alarm cleared is recorded,
if blank then the sensor is sll in alarm
Email – if emailing of this alarm event is enabled then this will conrm an emails was sent
Add – this allows users with the correct access privileges to enter notes against the alarm event, these notes are then visible to other
system users. Up to 20 notes can be added per alarm event. Each note is unique and stored with the users ID details, Notes cannot be
deleted nor amended.
Graphing Icon (top le corner) – opens the graphing tool to the instance of the alarm event
6.1.1. Adding an Alarm Note
SITE => ALARMS => select ALARM/EVENT => ALARM DETAILS => ADD
When the Add buon is selected a record entering panel is opened at the boom of the Alarm Details box, the user can type a message
of up to 255 alpha/numeric characters. Clicking the Submit buon records the alarm comments into the system memory.
Up to 20 records can be entered per alarm event.
An audit review of the alarm details contain any notes entered against that alarm.
6.2. Unacknowledged Alarms
SITE => ALARMS => UNACKNOWLEDGED
The Unacknowledged alarm page is similar to the Current Alarm
page, the 2 dierence being
1. Acknowledge Alarm buon is available
2. The graphing icon is replaced with a selecon buon.
Exporng, sorng and the search capabilies are the same.
6.2.1. Acknowledging an Alarm
SITE => ALARMS => UNACKNOWLEDGED => select Alarm/s => ACKNOWLEDGE ALARMS
On entering the Unacknowledged alarms tab, the Acknowledge Alarm buon is shaded and the Select Buons are a solid colour.
Acknowledging alarms individually - Alarms can be acknowledged individually by selecng the Select Buon in front of the desired
alarm and clicking the Acknowledge Buon
Acknowledging a number of alarms at once – Click the Select Buon for those alarms to be acknowledged and Click the Acknowl-
edge Buon
Acknowledging all alarms at once – Click on the Select Buon in the Descripon Bar, all alarms will be selected, Click the Acknowl-
edge Buon
When the Acknowledge Buon is clicked the user will be required to enter their password to conrm acknowledgement.
Note - Acknowledged alarms will be moved from the Unacknowledged tab, if sll a current alarm they will appear in the Current secon, otherwise they will
move to Historical if the alarm has cleared
6.3. Historical Alarms
SITE => ALARMS => HISTORICAL
The Historical Page allows for access to all alarm events in the base staons recorded memory. It will also allow for archived alarm events to
be viewed and searched from the internal back-up facility (only on systems with this feature enabled).
The layout of the Historical page is similar to that of the Current alarm page with 2 addions. These addions are;
1. User Dened Date Search – search based on user dened start and end period.

Page No. 17
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
2. Show per page tab – users can choose the number records per page, the next and previous buons then allow users to scan through alarms
by page.
6.4. Audit Informaon
SITE => ALARMS => AUDIT
The Audit Page is idencal to the Historical Alarm page in look and
funcon.
The dierence with the Audit Page is that the view contains all
alarms as well as user dened events recorded by the base staon.
User dened events are system changes made by a user, these
are recorded against that users details to give an audit trail of all
acvity users complete on the system. The parameters changed
are highlighted along with the date and me details
7. REPORTS
SITE => REPORTS
All reports generated by the system are in a PDF format and contain the date of creaon and the creang users details (including electronic
signature).
Note - For reports to be created the network browser MUST support pop-ups
and pop-ups must be enabled for the systems assigned IP address.
7.4.1. Select Date
Before creang a report the date for the report to generate
must be selected. To select the date click within the date
picking box and the date selecon tool will appear.
Once a date is selected the report generaon buons become
available.
Note – any date from the full recorded history within the base staon can be
selected for the report to be created.
7.4.2. Daily Overview Report
The daily overview report includes a list of all sensors on the system on the date selected. It shows the set points, max, min and average
recording for each sensor and the readings taken every 2 hours during that full day.
7.4.3. Daily Sensor Report
The daily sensor report creates a detailed report with a page for each sensor on the date selected. It shows the set points, max, min and
average readings by sensor. It has a graphical representaon of the data over that day and shows the sensors readings over the 24 hour
period at 15 minute intervals.
The end of the report shows any alarm events that occurred on the system (from those alarms sll in the acve list).
Note – to ensure all associated alarms are included on the daily report it is best pracce to produce and save a copy of the report each day for the previous
days data.
All alarm details are saved to the Historical Alarm log and can be viewed by date to compare to a daily report.
8. Users Login Details
8.1. Updang/Changing User Login Details
SITE => SETTINGS => MY DETAILS => EDIT
A user set their Username, Unit Passcode and Web Password;
Username – can be up to 25 alpha/numeric characters in length and may be upper & lower case
Unit Passcode – Used where required on the base staon – can be up to 8 numeric characters only
Web Password – can be up to 25 alpha/numeric characters in length and may be upper & lower case
Once the user has input their Username, Unit & Web Codes they will need to enter the Administrator provided password and click the submit
buon, this will acon the changes made and the new username and passcodes/words will be required for future logins.
8.2. Forgot Login Details
Should a user forget their login details then they should contact the system Administrator, the Administrator can reset a default login and the
user can then complete secon 8.1 Updang/Changing User Login Details.

Page No. 18
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
Set-Up Guide
Overview
The set-up guide is aimed at System Administrators & System Installers. Where this guide overlaps acons outlined in the User Manual it will high-
light the relevant secon for the Administrator/Installer to refer to.
Set-up & Installaon should be completed by competent persons, please read this manual in full before commencing any work on the base staon.
Quick Start Guides are also issued with new sensors, these guides explain how to congure the sensors to a base staon, and these should be used in
conjuncon with this document during sensor set-up.
Note (No.) A number in brackets refers to relevant part of the system outlined in the System Overview DATAssure D3 Base Staon Assembly of the User Guide
New System Set-up
Hint - Once set-up the base staon retains the conguraon of all sensors, users, etc. even when powered down. Seng a new system up over a laptop connecon and then connecng to the site network is by
far the simplest and quickest set-up method.
Aer reading Sengs secon of the manual, and prior to commencing sensor set-up, thought should be put into how sensors will be; given a unique ID, named, the set points needed, the alarm delay needed and if
the sensors will be grouped together. Acons required on alarm for each sensor will also assist in dening the alarm proles needed. Planning and documenng this will make the set-up process far simpler.
New system set-up work-ow best pracce;
• Network the base staon
• Check and set me & date
• Complete the default Site Sengs
• Create addional sensor types required in the Customer Sensor Database
• Add & pair all required sensors
• Create alarms proles
• Create required Schedules (if needed)
• Edit sensor set points, aach to alarm proles & auto inhibit schedule (if needed)
• Complete Email Set-up & send test email
• Create Users & access levels
• Aach any peripheral hardware & Mount the Base Staon
• Install the sensors & repeaters as necessary
• Issues Users with login details
If a site plan is available, marking the posions of installed sensors and repeaters along with their assigned ID’s will simplify nding these items at a later date if required.
9. Powering the DATAssure D3 Base Staon
9.1. Powering Up
Connect the power supply male jack to the female input power connector socket (7), securing the power cable behind the internal securing
lug built into the base staon.
Plug the power supply into a mains socket and switch on, press and hold the power buon (5) unl the unit starts to power on. The unit’s
inialisaon progress bar will be displayed. Once power up is complete the menu screen will display and the base staon is ready.
The LED on the base staon display will show solid colour to indicate the system is powered.
9.2. Powering Down
Press and hold the power buon (5) for 15 seconds, the base staon will then fully power down.
Internal Baery Back-Up² (operaon & tesng)
On connecon to mains power supply the internal baery back-up will commence charging, on mains power loss the baery back-up will
auto switch and power the base staon.
Note - It is recommended that the baery back-up operaon is tested regularly (manufacturer recommends at least monthly). To test; remove the mains power
to the base staon (switch o at the mains power socket), the base staon should connue to funcon.
Replacement of the baery for the baery back-up should only be completed by competent persons, refer to the system supplier for assistance should the
baery need to be replaced.
10. Networking the Base Staon - IP Connecvity
The DATAssure D3 is capable of operang in 3 dierent IP modes, they are;
• DHCP Client - (default from new) this allows the system to be connected directly to an available network port and for that network
to provide all the relevant IP sengs to access the networked interface - see secon 3. Network Interface.
Note - Though this makes for an easier installaon of the system the network can assign the base staon a new IP, meaning any favourite saved within web
browser would need to be updated.
• DHCP Server – This allows the base staon to communicate with a pc/laptop directly via an ethernet connecon. The base sta-
on is acng as a server and gives out an IP that can be easily located on the PC/Laptop being used. This opon is frequently used when
seng the system up on installaon, as all can be done on a laptop/pc beforehand.
Where the user wants to connect to a dedicated PC or Laptop this opon can also be used but it will limit the alarm capabilies of the
system and is not recommended.
Note – do not connect a base staon in DHCP Server mode directly into a network as a conict in IP sengs on that network could occur.
• Stac – This base staon is set to a stac IP address on the network. The benet of the stac IP is that it does not change (unlike
DHCP Client). This means that even if the network loses power on reboot the same IP will be assigned to the base staon. This stops the
users having to rediscover the IP from the base staon for access via their preferred web browser (the stac IP can be saved as a favourite
or desk top icon).

Page No. 19
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
10.1. IP Setup via the Base Staon
HOME => SET UP => Authorise => IP
Also refer to the DATAssure D3 Quick Start Guide
10.1.1.DHCP Client Mode
HOME => SET UP => Authorise => IP => Change => NETWORK MODE => Select DHCP Client
In DHCP client the network will automacally assign the IP conguraon one the system is connected to the network. The assigned
network IP address is visible in the Info view (HOME => INFO).
10.1.2.DHCP Server Mode
HOME => SET UP => Authorise => IP => Change => NETWORK MODE => Select DHCP Server
In Server mode the base staon is congured for direct connecon to a PC/Laptop using a network cable. The IP Address is xed and is
192.168.100.100
10.1.3.Stac Mode
HOME => SET UP => Authorise => IP => Change => NETWORK MODE => Select Stac
The base staon retains the last Stac conguraon set, the Stac IP details can only be entered/updated over the network interface.
10.2. IP Setup via the Network Interface
SITE => SETTINGS => NETWORKING => IP SETUP
10.2.1.DHCP Client Mode
SITE => SETTINGS => NETWORKING => IP SETUP => DHCP CLIENT
Host Name – this can be edited, the host name is visible on the network and makes idencaon assigned base staon IP easier for
IT departments.
The IP Address, Subnet, Gateway, Preferred & Alternate DNS are all auto supplied by the network and no acon needs to be taken.
The HTTP Port is also normally auto supplied, however should the network administrator require this can be edited.
10.2.2.DHCP Server Mode
SITE => SETTINGS => NETWORKING => IP SETUP => DHCP SERVER
The IP is set at 192.168.100.100 and does not change, in this mode the unit can be connected directly to a PC/Laptop. Entering this IP
address into the browser will access the login window.
10.2.3.Stac Mode
SITE => SETTINGS => NETWORKING => IP SETUP => DHCP STATIC
The IP Address, Subnet, Gateway, Preferred & Alternate DNS and HTTP Port must all be manually congured.
Note – the system retains the last stac IP sengs entered. Therefore once the stac IP is set-up other modes can be used, on return to the stac mode the
last set details will be displayed and can be set. Failure to set all items correctly in stac mode may lead to the access and email funconality being impaired.
SETTINGS
SITE => SETTINGS
The basic set up of the system and its operaonal parameters are completed through the Seng page. Remember, users will only see those
elements of the Sengs page that their user dened access levels allow, only Administrator level users see the full Sengs page as outlined
below.
Note - The default view when entering the Sengs page is Manage Devices as below. The sengs menu is on the le of the screen with the acons selectable
grouped into their relevant headings, the selected acon is displayed and updated on the right.
1. Sengs Menu Bar
• Sensor Administraon – Allows for the
management of sensors/devices on the system and
includes the ability to create user dened device
types for pulse, linear and volt free input sensors and
add these to the sensor database.
• User Sengs – Includes the elements for all
users to amend their own user’s details, allows the
administrator to set password expiry and the overall
management of system users.
• Alarms – Allows for the general alarm sengs
to be created as well as for the administrators to
create bespoke alarm proles – sensors are aached
to an alarm prole, the prole outlines the system
acons when a sensor goes into an alarm state.
• Site Sengs – Includes the basic system set
up, the scheduler sengs that sensors and alarms
can be congured against and the ability to amend the systems Time/Date.

Page No. 20
Doc: 106-10051-03 (Planer DATAssure D3 User, Set-up & Installaon Manual)
• Networking – Allows for the seng up of the IP connecon, the email server details and connecon to 3rd party alarm nocaon
bureaus.
• Administrator – Includes the elements for copying the site conguraon, updang soware, backing-up and restoring the site
conguraon.
2. Menu Content
When an area from the Sengs Menu is chosen the content will open in this space for eding/conguraon.
11. Sensor Administraon
Sensor administraon is broken down into two secons, these are:
• Manage Devices - this allows for sensors and sensor groups to be added, edited and deleted from the system
• Custom Sensor Database - this allows for custom sensor types to be created to allow the connecon of 3rd party sensors and
meters using the DATAssure D3’s range of Linear, Digital and Pulse probes.
• Network Diagram - 2.4Ghz SYSTEMS ONLY - allows for a pictorial view of sensor communicaon routes and status.
11.1. Manage Devices
SITE => SETTINGS => MANAGE DEVICES
1. Add Device/Group buons – These buons open the sensor and
group sengs creaon pages
Device groups can be used to group sensors together – e.g. all the sensor
in one room could be grouped together and the name of the group could
be the room name. Grouping also allows for an average of the readings
of a chosen sensor type within that group to be displayed, e.g. if you have a number of temperature sensors taking readings in the grouped
room, you can display the average of all the temperature sensor readings
2. View Selecon
Default View - view, add, remove and manage devices/groups
Communicaon Status - shows the communicaon details of each aached device
Calibraon Oset - allows a single point calibraon o-set to be entered for each device
Remote Address - Used where 3rd party system gather data from the base staon for use in other systems (refer to your system supplier
for more informaon on this funcon).
Note - Best pracce is to achieve a signal quality greater than 80% and strength greater than 40%. If below these levels a repeaters should be considered
3. Expand All – toggling this switch expands any groups created so that all details of devices in each group can be seen
4. The details of any devices added to the system will show in this secon.
Ref – this is the device reference assigned when creang/eding the device to the system
Descripon – name of the device assigned when creang/eding
Type – shows the type of sensor added, e.g. Temperature
Data Source – shows the communicaons method for the data received, e.g. Wireless
Link/Link Icon – shows the status of the sensor link with the base staon, the link icon will be coloured:
Grey – Device not yet paired
Amber – Linked but waing for data from device
Green – Device paired
State – shows the connected status of the device/group on the base staon, the circular status icon will be coloured:
Green – Device is Acve
Purple – Device is Disabled
Grey – Device is Decommissioned
Delete – allows for a group to be deleted from the system
11.2. Add a Device/Sensor
SITE => SETTINGS => MANAGE DEVICES => ADD DEVICE
Adding a device to the system is completed in a 2 stage process, the device is created on the base staon and then it is linked (paired) with
the base staon. The rst element is to create the device.
Note - Devices can be added to groups when being created – see secon 11.3 Adding a Device Group. Or they can be added to a group later by eding the
device – see secon 11.5 Eding a Device/Group.
Hint - If a number of devices are being added to a system it is quicker to create all of the devices at one me. Once all devices are created each can then be linked (paired).
Table of contents
Other Planer Accessories manuals
Popular Accessories manuals by other brands

Simrad
Simrad PI DEPTH - MANUAL 2 instruction manual

Alesis
Alesis PHOTON X 25 Driver installation guide

DutchWest
DutchWest 16 installation instructions

Dorner
Dorner SmartFlex 2200 Series Installation, maintenance & parts manual

IPVideo Corporation
IPVideo Corporation Halo Administrator's manual

AVCOM
AVCOM Galaxy Installation & operation manual