PRÜFTECHNIK FLUKE VIBGUARD IIoT Instruction Manual

VIBGUARD®
IIoT
Operation
Edition:01.10.2018
Doc. no.:LIT 78.222.EN
Translation of the German manual
Type: VIB 7.800, VIB7.810, VIB 7.811, VIB 7.815, VIB7.820, VIB 7.825
Serial number and year of manufacture: see type plate
PRODUCER: Fluke Deutschland GmbH, Freisinger Str. 34, 85737
Ismaning, Germany, + 49 89 99616-0, www.pruftechnik.com
This manual is only intended for use with GL-
certified VIBGUARD CMS on wind turbines.

LEGAL NOTES
Protection notice
These instructions and the product described herein are protected by copyright. The authors
reserve all rights. These instructions must not be copied, reproduced translated or made availa-
ble to third parties in any other form - in part or in full - without prior approval.
Exclusion of liability
Any claims against the authors in relation to the product described in these instructions are
excluded. The authors do not provide any guarantee for the accuracy of the contents of these
instructions. Furthermore, the authors are not in any way liable for any direct or indirect loss ari-
sing from using the product or from using these instructions even if the authors have identified
the potential for such a loss.
The authors do not accept any liability for potential faults in the product. This exclusion of liability
also applies to dealers and distributors. Subject to errors and design changes, in particularly in
terms of further technical developments.
Trademark
The trademarks and registered trademarks referenced in these instructions are generally mar-
ked as such and are the property of their owners. The absence of a marking does not however
mean that names are not protected.
VIBGUARD is a registered trademark of PRÜFTECHNIK AG.
© Fluke Corporation. All rights reserved.
Fluke Deutschland GmbH
Freisinger Str. 34
85737 Ismaning, Germany
+ 49 89 99616-0
www.pruftechnik.com

Contents
1 Prior to starting 5
1.1 Notes regarding these instructions 5
1.2 Markup 5
1.3 Abbreviations 5
1.4 Service addresses 5
2 Safety 7
2.1 EU conformity 7
2.2 Intended use 7
2.3 Safety markings 8
2.4 Type plates 9
2.5 Information for the operator 10
2.6 Information for operating personnel 11
2.7 Residual hazards and protective measures 12
3 Control software 13
3.1 Hardware and software philosophy 13
3.2 Programming of the control software 14
4 Dialing into the CMS 15
4.1 Direct network connection (LAN) 15
4.2 Dial-up connection 16
4.3 Connection via a VPN tunnel 17
5 Operation via a web browser 18
5.1 Homepage 18
6 Alarm status 19
7 Displaying readings 20
8 Deleting measurements 22
8.1 Deleting signal data 22
8.2 Deleting the ring buffers 22
VIBGUARD IIoT│Operation 3

1 Prior to starting
1.1 Notes regarding these instructions
These instructions are a part of the product. As such, they must be kept throughout the product service life. These
instructions must be handed over to any subsequent owner or user of the product.
1.2 Markup
Texts are marked as follows in these instructions:
Action steps are indented and marked with a • bullet point.
List entries are indented and marked with a dash -.
Functional elements on the user interface, such as buttons, commands, links: <Element> in angle brackets.
Labels on the user interface, such as window titles, field names: '"Designation" in quotation marks.
Supplementary information / tips: Functional notes are introduced with a context-dependent keyword.
Safety notes:Warnings are introduced with the keyword CAUTION if there is a danger of personal injury.
Failure to observe these warnings may result in minor or moderate injury.
If there is a danger of damage to property, warning notices are introduced with the keyword Note.
1.3 Abbreviations
The following designations are considered equivalent in these instructions:
Condition Monitoring System = CMS
VIBGUARD IIoT Condition Monitoring System = VIBGUARD IIoT or System
VIBGUARD IIoT system module = system module
Sensors, cables, mountingadapters = measuring equipment.
VIB 7.800, VIB 7.810, VIB 7.811, VIB 7.815, VIB 7.820, VIB 7.825 = VIB 7.8xx
Current Linedrive = CLD
1.4 Service addresses
Hotline +49 89 99616-0
Please have the serial number of your system module ready when contacting our
hotline ("Type plates" on page 9).
Shipping address
Fluke Deutschland GmbH, Freisinger Str. 34, 85737 Ismaning, Germany
VIBGUARD IIoT 5
Operation

Empty page

2 Safety
VIBGUARD IIoT was designed and built following careful selection of the harmonized norms to be complied with as
well as other technical specifications. The system therefore corresponds to the state of the art and ensures the hig-
hest degree of safety.
Nevertheless, there are still risks relating to installation, commissioning and operation which are to be avoided.
Observe the general safety instructions in this section as well as the warnings in the instructions. Safety instructions
and warnings explain how you should act in order to protect yourself, others and objects from harm.
2.1 EU conformity
PRUFTECHNIK hereby declares that VIBGUARD IIoT conforms to the relevant European directives. The complete
text of the EU conformity declaration is available at the following Internet address:
2.2 Intended use
VIBGUARD IIoT is a stationary condition monitoring system for monitoring the condition of machines mounted on
anti-friction bearings. The system records and processes the following signals and parameters:
lVibration parameters, broadband and narrow-band
lTime signals
lSpectra
lTemperature
lProcess parameters
The system works continuously and records the machine signals synchronously on all channels1. It is therefore sui-
table for machines with dynamic operational behavior for which the operating and process parameters need to be
recorded simultaneously and at short intervals.
The system may only be operated within the specifications given in these instructions. PRUFTECHNIK shall not be lia-
ble for any damage caused by misuse.
1maximum number of channels is type-specific
VIBGUARD IIoT 7
Operation
https://www.pruftechnik.com/com/About-us/Company/Certificates-and-Material-Safety/

2.3 Safety markings
Please refer to the following figure for the safety markings on the VIBGUARD IIoT. The safety markings must be
observed and must not be concealed or removed. For the variants that are installed in a control cabinet (VIB 7.xxx-
PS), the safety labels must be attached at a suitable point in the control cabinet.
1 Vorsicht! Spannung Attention Danger
VIB 7.8xx-SDH A yellow sticker each on the housing lock and below the power supply.
VIB 7.8xx-PS A yellow sticker on the top hat rail.
2 100 - 240V AC / 1,3 A / 50-60 Hz
VIB 7.8xx-SDH A yellow sticker on the outside of the cable gland for the mains supply.
VIB 7.8xx-PS A yellow sticker on the top hat rail.
8 Edition: 10.2018
Operation

2.4 Type plates
System module
The type plate on the system module contains the following information:
Type: Item number (VIB 7.820), hardware status (V1.3) of the System module.
S/N: Serial number of the system module, 7 digits, starting with 95..
HW: Status (V 1.31), number (239), variant (LD = LineDrive) of the built-in board.
Protective housing
The type plate on the protective housing, or on the power supply cabling contains the following information about
the entire system:
Type: VIB 7.8xx-SDH or VIB 7.8xx-PS - Item number of the entire system.
S/N: Production week (WWYY) of the entire system and serial number of the system module installed.
HW: Hardware status of the protective housing (e.g. V 2.0).
VIBGUARD IIoT 9
Operation

2.5 Information for the operator
Obligations of the operator
Maximum safety can only be achieved in practice if all measures required for this are adopted. As the operator, it is
part of your duty of care to plan these measures and monitor their implementation.
Ensure that the following requirements are met:
Qualified specialist personnel for installation, commissioning and operation is available.
Installation material and tools are provided.
Voltage supply and network connection are present in accordance with the specification.
Potential equalization is provided
Integration of the system into existing equipment
The safety of a plant in which the system is integrated is the responsibility of the installer of the plant.
CAUTION! Except for the mains power supply to the system, all other connected circuits must be SELV cir-
cuits.
Ensure that the following requirements are noted:
Use the system in accordance with the intended use.
Only operate the system in a technically flawless state.
Only use original accessories and original spare parts.
Observe all nationally applicable regulations, all safety, accident prevention and environmental protection
regulations as well as all recognized technical rules for safe and proper work.
Training
Provide ongoing training for operating personnel in relation to the application of all safety regulations in safety
instructions. Ensure that the following requirements are complied with:
The applicable legal and otherwise binding safety and accident prevention regulations as well as the general safety
instructions and warnings must be adhered to and followed.
Ensure that operating personnel work in a safety-conscious manner.
10 Edition: 10.2018
Operation

2.6 Information for operating personnel
Operating personnel qualifications
Installation and disassembly may only be performed by a qualified specialist electrician.
Commissioning and operation must only be performed by personnel who have been trained and authorized to do
so.
Personal protective equipment
No protective equipment is required for installation, commissioning, normal operation and disassembly of the sys-
tem.
Regulations for normal operation
The operating state of the system module is displayed via the SYSTEM LED. The LED lights up orange while the sys-
tem module is starting up and changes to green (normal operation).
lCheck the following points at regular intervals:
Is there any identifiable damage on the system components and on the measuring equipment?
Are the cables crushed or damaged?
lRectify any defects detected immediately or report them to the operator. The system and measuring equip-
ment must only be operated in a flawless state!
lIn the event of functional faults, disconnect the system from the supply and secure it against a restart.
Operation of a machine is not impaired if the system is out of service. The machine can therefore remain in ope-
ration.
VIBGUARD IIoT 11
Operation

2.7 Residual hazards and protective measures
VIBGUARD IIoT is verifiably safe assuming it is used as intended. The following damage may occur if operated incor-
rectly or used improperly:
Personal damage
Damage to the system or to the machine
Danger due to running machine!
During installation and maintenance work on the machine, there is a risk of injury from moving machine com-
ponents.
lShut down the entire machine system and secure it against unintentional restarting.
Danger due to electric shock!
There is a risk of injury due to low voltage when working on the switch cabinet when it is open (230 V).
lObserve the safety markings on live components.
lBefore carrying out any installation, repair or maintenance work, disconnect the system from the power sup-
ply, determine that there is no voltage and secure the power supply against being switched on again.
lDo not open, repair or modify with power supply VIB 5.965-2.5.
Danger due to improperly laid cables
Personnel may stumble over an improperly laid cable and injure themselves. The cable can be damaged due to
external influences.
lLay the cable in such a way that no one can stumble over it.
lUse cable ties or a Velcro fastener to fix the cable in place.
lLay the cable in a cable duct or protective tube.
Damage due to electrostatic discharge
During installation, repair and maintenance work on the system, the electronic components can be damaged if tou-
ched due to electrostatic discharge.
lUse an earthing strap if contact with the components cannot be ruled out.
Damage due to contamination
In a loaded industrial environment, the system components may have their function impaired or be damaged due
to contamination or moisture when the switch cabinet is open.
lKeep the switch cabinet closed as much as possible.
Incorrect measurements due to electromagnetic interference
High frequency rays or electrostatic discharge in the vicinity of the system and measuring equipment can lead to
incorrect measurements.
lDo not lay the sensor cables in the vicinity of heavy current lines.
lSelect an installation site with low electromagnetic radiation exposure.
12 Edition: 10.2018
Operation

3 Control software
Executable real time software for controlling inputs and outputs, sequence control and visualization is already pre-
installed on the VIBGUARD IIoT. The following software is required for further parameterization, data storage and
analysis. The software is installed on a powerful PC:
Email client
OMNITREND Center
3.1 Hardware and software philosophy
As open systems, the telediagnostic systems offered by PRUFTECHNIK provide a large number of programming and
parameterization possibilities. The following section will provide an overview of the software structure.
As the VIBGUARD IIoT systems are certified together with the OMNITREND Center PC software, they are supplied
with defined program sequence control to ensure the minimum scope of measurements required by the cer-
tification body.
The user is not permitted to change program sequences on the CMS. Three different user levels serve as a means
of protection. The user names and passwords for each level can be found in your password document which is
included in the scope of delivery.
Security level 1: Allows the user to view the overview page, the alarm states, and the display of the readings
on the CMS user interface.
Security level 2: Allows the user to additionally output the log files and to download data as well as query the
system status.
Security level 3: Allows the user to configure the system and to change passwords.
Note!
With Security level 3, system configuration files can be changed or deleted. This can result in the CMS no
longer functioning faultlessly. PRUFTECHNIK will support you should work need to be performed on the
system which requires a high security level.
VIBGUARD IIoT 13
Operation

3.2 Programming of the control software
A real time operating system is installed on the VIBGUARD IIoT system. This operating system contains all of the func-
tions for controlling the measuring components. A web server provides the HTML pages required for operation.
These pages feature a series of parameterization, control and visualization masks.
A measuring program runs in parallel in different sections on the CMS to control the measuring and analysis func-
tions as well as data handling. After the measuring program has been loaded and validated, the measurements start
immediately, in parallel and continuously, in accordance with the configuration. Four parallel data calculation paths
exist for each channel. These paths can be used. In addition, the monitoring of the sensors (SensorCheck) is done
independently of the other measurements that are configured. The recording of the RPM or the Keyphaser signal
(max. two independent RPM inputs) also takes place simultaneously.
Up to six overall readings (Characteristic Group), at least 1x per second for the ‘Always’ persistent setting, and two
spectra or time signals or an envelope (Trending Group), at least 1x per minute, can be generated and stored per
measuring channel. Adaptive storage is generally used to reduce the data volume with maximum benefit and mini-
mum information. If more than the aforementioned trending signals are configured on one channel, such signals
are processed successively and the minute interval may be exceeded under certain circumstances. A 1-minute
interval cannot be adhered to if, for example, a prolonged time signal measurement over 180 seconds is con-
figured.
The continuous storage of data in the minimum time units is only recommended for special cases when inves-
tigating a certain type of behavior over a limited period of time. Otherwise, the data volume becomes too large. It
should also be noted that the data is usually sent via the Internet and the bandwidth is limited in this regard. The
CMS offers intelligent mechanisms for data reduction, such as adaptive storage, in which data is only saved after a
set variable change. There is no data loss with this function.
The measurements and program steps are generally processed continuously. Exceptions may, for example, occur if
a special case for additional and longer diagnostic measurements is started from a 'Diagnostic Group' or is triggered
and processed by an event.
Note!
It is not permissible to make changes to the program sequence.
If, however, changes are to be made, such as the extension of additional operational measurement varia-
bles, this can only be performed by using OMNITREND Center programming software. This software is not
part of the scope of delivery for the CMS, which is why it is advisable for the programming to be done by
trained personnel or by PRUFTECHNIK experts. The software can be purchased from PRUFTECHNIK if
necessary.
The measurement configuration loaded upon delivery is monitored by means of a checksum. If changes have been
made by mistake or for other reasons, such changes are detected and the OMNITREND Center software reports
their occurrence immediately when processing the measurement data.
14 Edition: 10.2018
Operation

4 Dialing into the CMS
Each VIBGUARD IIoT system has an IP address through which you can dial in from a PC.
4.1 Direct network connection (LAN)
Use this option if the XP system or the PC are running in a common network or are connected to each other by
means of a crossover network cable.
Prerequisites:
Functioning network connection
Internet browser
VIBGUARD IIoT In a ready for operation state
With regard to a direct network connection, the CMS and PC need to have an IP address in the same IP
address range, i.e. the first three numerical blocks of the IP address are identical.
The IP address is set during commissioning.
lEnter the IP address of the CMS in the browser.
“http://< IP address of the CMS>”
lWait for the connection to be established.
VIBGUARD IIoT 15
Operation

4.2 Dial-up connection
If the CMS is connected via a router and dial-up connection, a PPP dial-up is required to establish the connection.
Procedure for Windows 7
lOpen the Windows Start menu and select <Control Panel> / <Network and Sharing Center>.
lSelect <Set up new connection ..>.
lSelect <Set up dial-up connection>.
lEnter the <Dial-up number> and the <Username> and the <Password>.
Setting up the dial-up connection (Windows 7)
To dial up an analog modem connection, an analog telephone connection also needs to be dialed from the
PC. The telephone number can be obtained from your telephone provider once the telephone line has
been commissioned.
The username and password are either included in the documentation for the CMS, to the extent that this
is part of the scope of delivery, or they are to be obtained from the wind park operator.
Where appropriate, the company firewall must be set up in such a way that access rights exist and the
misuse of data is prevented.
16 Edition: 10.2018
Operation

4.3 Connection via a VPN tunnel
A VPN connection is required if the CMS is connected via a VPN network by means of an Internet router.
There are a number of VPN software providers. Setup may vary depending on the software package used.
Please contact the supplier in this regard. PRUFTECHNIK generally uses the OpenVPN software for dialing
the router packages they supply.
Installing the OpenVPN-Client
lStart the installation file (install.exe).
Administrator rights are required to run the file.
lIf necessary, ignore the certification warnings that are displayed.
With regard to an access certificate supplied separately, which you have requested and received befo-
rehand from the relevant authority, such an access certificate still needs to be inserted. If it is already inclu-
ded in the installation file, this step is not required.
lIf installation was successful, the openVPN icon appears in the bottom right corner of the system tray.
Establishing a connection to the VPN network
lUsing the right mouse button, click on the openVPN icon and select “Connect”.
lEnter the username and the password.
lIf necessary, deactivate the Internet proxy.
lIf the connection is established successfully, a confirmation message appears,
e.g. "mdex fixedip now connected, assigned IP address: x.x.x.x.x".
Example for entering the login details in an OpenVPN client using Windows.
Dialing into the CMS
lStart the dial-up connection to the CMS (see the previous section).
lEnter the IP address of the CMS in the browser.
lWait for the connection to be established. This can take some time depending on the connection type as the
user interface needs to be loaded.
The VIBGUARD IIoT homepage appears.
VIBGUARD IIoT 17
Operation

5 Operation via a web browser
VIBGUARD IIoT features a web server on which, inter alia, the user interface is implemented in the form of HTML
pages. The HTML pages are only available in English.
The functions for user inputs are provided via input boxes and buttons. This makes it possible to visualize mea-
surement data, display alarm states and status messages, and archive data. The communication is purely http based
and only requires the standard web port 80 for data exchange.
The HTML pages are specifically customized to the system to be monitored and installed on the CMS prior to deli-
very.
5.1 Homepage
After entering the IP address, the local homepage is loaded in the browser. If there are several CMS in a network,
each system can be reached via its IP address using the browser.
CMS homepage in the web browser
The homepage is the central document from which all other functions for oversight and monitoring are activated.
The pages that have been stored with additional links can be accessed via the links in the horizontal navigation bar.
Depending on the access level, it is necessary to enter a username and password in order to either branch out into
further links or to execute a function.
Navigation bar (hyperlinks)
The horizontal navigation bar contains the following links:
Home: the home page
System: Save and create information, configurations, logs
Data Transfer: Send, receive and delete data
Online Visualisation: Monitoring and change individual values
The other links are explained in the following pages.
18 Edition: 10.2018
Operation

6 Alarm status
VIBGUARD IIoT provides an overview listing the current threshold violations and sensor errors.
Opening the alarm status page
lClick on <System> in the navigation bar.
lClick on <Alarm Status Information> in the “Administration” field. The Alarm status page is displayed:
The list contains all the measurement locations that exhibit a deviation from the normal condition in the current
measurement cycle. The list displays the following information:
Alarm Location: Storage location number of the measurement in question
Alarm: Error type; threshold violation as Alarm, Warning, Prewarning or SensorFail (sensor error)
Job name: Label for measurement location
Task name: Label for measurement task
Clear (All): Resets the error message
Press the F5 key to update the list (refresh browser). You can reset an error message using <Clear>. However, the
error message is displayed again if the status in question still exists at the next measurement cycle. To this end, the
list must always be updated manually (F5).
The alarm messages are archived in the OMNITREND Center software. They must also be acknowledged
accordingly there.
Note
In the event of a sensor error, you must check the measurement location in question:
• Visual inspection of the sensor and wiring;
• Check the voltage supply if the sensors are supplied externally.
VIBGUARD IIoT 19
Operation

7 Displaying readings
The current readings can be displayed in two views:
Online readings
Online readings and status
Opening the views for the readings
lClick on <Online Visualisation> in the navigation bar.
lClick on
<System Online ISL Table> in the "Input Storage Locations (ISLs)” field to open the Online readings
view or on
<Display Input Storage Location Values> to open the Online readings and status view.
Both views list the readings in tabular form:
“Online readings” view
The Online readings view updates the values every 10 seconds. The table contains four columns (a-d):
a: Number of the input memory (ISL)
B:: label
c: Reading
d: Unit
The input memories can be selectively shown and hidden as required:
lTo this end, enter the desired input memory (ISL = Input Storage Location) numbers in the input field and
click on <submit>.
Example: 1-9,15,24,34-38
20 Edition: 10.2018
Operation
Table of contents
Other PRÜFTECHNIK Measuring Instrument manuals

PRÜFTECHNIK
PRÜFTECHNIK VIBXPERT II User manual

PRÜFTECHNIK
PRÜFTECHNIK VIBXPERT II Manual

PRÜFTECHNIK
PRÜFTECHNIK VIB 6.195 Owner's manual

PRÜFTECHNIK
PRÜFTECHNIK FLUKE VIB 6.12 Series Programming manual

PRÜFTECHNIK
PRÜFTECHNIK VIBXPERT II User manual

PRÜFTECHNIK
PRÜFTECHNIK VIBSCANNER 2 User manual

PRÜFTECHNIK
PRÜFTECHNIK INCLINEO User manual

PRÜFTECHNIK
PRÜFTECHNIK VIBSCANNER VIB 5.400 User manual

PRÜFTECHNIK
PRÜFTECHNIK VIBXPERT Balancing User manual

PRÜFTECHNIK
PRÜFTECHNIK VIBXPERT User manual