- 17 -
After pH7 is successfully calibrated, you can choose acid calibration or alkali calibration for the second calibration point.
If you choose acid calibration, clean beaker 3x with deionized water. Fill the beaker with pH=4 buffer solution. Enter
the value 4 in the calibration value dialog box, and click "Acid Calibration", then a dialog box will pop up to confirm
whether to perform this operation. Click "OK" if the calibration operation is confirmed and the dialog box will show
"Calibration Successful" if the calibration is successful. Similarly, a pH=10 buffer solution can be selected for the second
point calibration if desired.
Three Point pH Calibration
Remove the ST-765SS-O3 sensor and rinse 3x with DI water. Submerge the sensor into a beaker with pH=7 buffer
solution. Click "pH7 calibration". A dialog box will pop up to confirm whether to perform this operation, click "OK" if
the calibration operation is confirmed, if the calibration is successful the dialog box will show "calibration success".
After pH7 is successfully calibrated, you can choose acid calibration or alkali calibration for the second calibration point.
If you choose acid calibration, clean the beaker 3x with deionized water. Fill the beaker with pH=4 buffer solution.
Enter the value 4 in the calibration value dialog box, and click "Acid Calibration", then a dialog box will pop up to
confirm whether to perform this operation. Click "OK" if the calibration operation is confirmed and the dialog box will
show "Calibration Successful" if the calibration is successful.
After successful acid calibration, select pH=10 for alkali calibration. Clean the beaker 3x with deionized water. Fill the
beaker with pH=10 buffer solution. Enter the value 10 in the calibration value dialog box, and click "Alkali Calibration",
then a dialog box will pop up to confirm whether to perform this operation. Click "OK" if the calibration operation is
confirmed and the dialog box will show "Calibration Successful" if the calibration is successful. The three-point
calibration is completed.
4.7.2. O3 Calibration
The measurement module of the ST-765SS-O3 sensor is thoroughly calibrated at the Pyxis Lab factory. To calibrate, the
user can perform a single point according to the requirements of the application. (USEPA-334.0 / ISO-7393 compliant
methodology)
Calibration of the ST-765SS-O3 sensor for ozone should be done with the sensor inline exposed to active flowing
sample water. Use a portable or laboratory colorimeter (i.e. Pyxis SP-200 / SP-800 / SP-910 or similar) to test the active
(flowing) water sample in the flow tee assembly of the IK-765SS-O3 panel. Once you have tested and confirmed the
concentration value in the active (flowing) flow tee assembly, enter the test result value of the portable or laboratory
colorimeter in Calibration Screen and click "High Cal". A dialog box will pop up to confirm whether to perform this
operation. If the calibration operation is confirmed, click "OK", and if the calibration is successful, the dialog box will
show "Calibration Success".
*NOTE* Click the restore button in the calibration interface of each sensor to restore the data of pH/residual O3 sensor. If a user
error is made during calibration and other operations, you may restore the factory settings of the sensor through the restore
function.