manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. ABB
  6. •
  7. Measuring Instrument
  8. •
  9. ABB SM500F User manual

ABB SM500F User manual

Other manuals for SM500F

6

Other ABB Measuring Instrument manuals

ABB aquamaster User manual

ABB

ABB aquamaster User manual

ABB C1900 Series User manual

ABB

ABB C1900 Series User manual

ABB COPA-XM 3000 Series User manual

ABB

ABB COPA-XM 3000 Series User manual

ABB 8241 User manual

ABB

ABB 8241 User manual

ABB ScreenMaster RVG200 User manual

ABB

ABB ScreenMaster RVG200 User manual

ABB AX413 User manual

ABB

ABB AX413 User manual

ABB 2410 User manual

ABB

ABB 2410 User manual

ABB AX410 User manual

ABB

ABB AX410 User manual

ABB ProcessMaster Installation and operation manual

ABB

ABB ProcessMaster Installation and operation manual

ABB ReliaMod RTBCOVER User manual

ABB

ABB ReliaMod RTBCOVER User manual

ABB Sensyflow FMT400-VTS Training manual

ABB

ABB Sensyflow FMT400-VTS Training manual

ABB NETA-21 User manual

ABB

ABB NETA-21 User manual

ABB LS4000 User manual

ABB

ABB LS4000 User manual

ABB VortexMaster FSV430 Installation instructions

ABB

ABB VortexMaster FSV430 Installation instructions

ABB Relion RES670 User manual

ABB

ABB Relion RES670 User manual

ABB TRIO-WIRL ST4000 User manual

ABB

ABB TRIO-WIRL ST4000 User manual

ABB EQmatic User manual

ABB

ABB EQmatic User manual

ABB Navigator 550 User manual

ABB

ABB Navigator 550 User manual

ABB FV4000-VT4 User manual

ABB

ABB FV4000-VT4 User manual

ABB EXCOUNT-II User manual

ABB

ABB EXCOUNT-II User manual

ABB 10DX3111G User manual

ABB

ABB 10DX3111G User manual

ABB CoriolisMaster FCM2000 Training manual

ABB

ABB CoriolisMaster FCM2000 Training manual

ABB Sensyflow FMT400-VTS Training manual

ABB

ABB Sensyflow FMT400-VTS Training manual

ABB DELTAsingle User manual

ABB

ABB DELTAsingle User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

—
ABB MEASUREMENT & ANALYTICS | INSTRUCTION | IN/RANDC/007-EN REV. A
SM500F
Field mountable paperless recorder
Battery replacement
Measurement made easy
1 Introduction
This publication details replacement of the battery
(B12454) fitted to the processor board of the
SM500F Paperless recorder. The procedure must be
carried out by a trained technician.
Tools required
• T8 Torx driver
2 For more information
Further information is available from:
www.abb.com/recorders
or by scanning these codes:
Sales Service
3 Safety
IMPORTANT NOTE
Dispose of the old components as detailed in the
instruction manual (IM/SM500F).
SM500F
field mountable
paperless recorder
2 IN/RandC/007-EN Rev. A | SM500F | Paperless recorder | Battery replacement
4 Replacing the battery
1. Isolate the recorder from the power supply.
Referring to Fig. 4.1:
2. Unlock the door with the key (if required) A, press release
catch Band open the door.
Referring to Fig. 4.2:
3. Use a T8 Torx driver to remove 6 Torx screws Csecuring
the door assembly rear cover and remove the rear cover.
Referring to Fig. 4.3:
4. Disconnect processor board flexi-cable D, display
flexi-cable Ekeypad membrane cable Fand inverter
cable (if fitted) Gfrom the processor board. Lift out and
discard the processor board.
5. Turn the processor board over.
Referring to Fig. 4.4:
6. Remove battery Hand insert replacement.
Referring to Fig. 4.3:
7. Position the processor board in the door housing and
reconnect inverter cable G(color recorders only – if fitted)
keypad membrane cable F, display flexi-cable Eand
processor board flexi-cable D.
Referring to Fig. 4.2:
8. Refit the door assembly rear cover and use a T8 Torx driver
to secure 6 Torx screws C.
9. Close and lock the door.
10. Restore the recorder's power supply.
CAUTION – Property damage
The recorder is vulnerable to electrostatic damage.
Wear an anti-static strap or place the instrument on
an anti-static workbench.
Fig. 4.1 Opening the door
Fig. 4.2 Removing the rear cover
A
B
C
Fig. 4.3 Disconnecting / removing the processor board
Fig. 4.4 Replacing the battery
D
E
F
H
SM500F | Paperless recorder | Battery replacement | IN/RandC/007-EN Rev. A 3
5 Setting the time and date
1. Enter the Configuration level of the SM500F.
2. Enter password (if required) and select Edit current
configuration.
3. Select the Common configuration parameter:
The Common Configuration menu is displayed:
4. Use the arrows to highlight the Time tab:
5. Press the key to select the Date and time
parameter.
A warning dialog is displayed:
6. Press .
The Date and time menu is displayed:
7. Set the required time and date.
8. Once finished, press and exit and save the
configuration as normal.
Important note
Once the battery has been removed, the battery failed
icon is displayed on the front panel of the unit. To
restore normal operation, the correct time and date must
set.
Important note
Changing the time will potentially cause loss of
data. Ensure that all data archiving is up to date
before proceeding.
IN/RandC/007-EN Rev. A 12.2018
—
We reserve the right to make technical changes or modify the contents of this document
without prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.
We reserve all rights in this document and in the subject matter and illustrations contained
therein. Any reproduction, disclosure to third parties or utilization of its contents – in
whole or in parts – is forbidden without prior written consent of ABB.
© ABB 2018
—
ABB Limited
Measurement & Analytics
Howard Road, St. Neots
Cambridgeshire, PE19 8EU
UK
Tel: +44 (0)0870 600 6122
Fax: +44 (0)1480 213 339
Email: instrumen[email protected]om
ABB Inc.
Measurement & Analytics
125 E. County Line Road
Warminster
PA 18974
USA
Tel: +1 215 674 6000
Fax: +1 215 674 7183
abb.com/measurement