Manual Zero Setting
When you set the display to zero (no pressure
applied to the connector), press button
for 2 seconds, the meter displays “0.0.0.0”
from right to left (See Fig.M) and then
delete each “0” from left back to right, the
LCD display shows a normal mode (See Fig.N).
Troubleshooting
• Power on but no display. Checkthatthebattery
isproperlyinstalledensuringthatcontactwith
terminalsandpolarityarecorrect,replacewitha
newbatteryorattachoptionalACadapter.
•BAT indication. Replace with a new battery when
LCD displays BAT at the middle bottom of screen.
•No display. Make sure that the battery is not
drained, if the display disappears, verify if
sleep mode is active. Refer to the Disable
sleep mode function for long term measurement.
Check that the tubing is tightly connected to
the meter.
•Err.1. If the pressure value exceeds the range
maximum, “Err.1“ appears on the display (See
Fig. O), it is recommended to use a meter with a
higher range, otherwise, the meter will be
damaged in use.
•Err.2. If the pressure value is less than the range
minimum, “Err.2“ appears on the display (See
Fig. P), it is recommended to use a meter with
a lower range to ensure measurement accuracy.
• Err.3. While operating the DIF function, if
the differential pressure value is larger than
maximum display digit, Err.3 appears on the
display (See Fig.Q ).
continued ...