Note: During the direct calibration of an instrument with reference-junction
temperaturecompensation,itiscommontopresstheRJ-ONbuttonso
that the instrument can start the function of automatic reference-junction
compensation.Thusprovidingtherequiredthermo-electromotiveforcefor
outputfollowedbydisplayingthesymbolRJ-ONwhere:
Output thermoelectric force = corresponding emf of set
temperature – emf of room temperature
• Ittakestwosecondsfortheinstrumenttostartitsinternalreference-
junction temperature. After this, each automatic compensation occurs
at10secondintervals.
• Ifthereisachangeintheoperatingambienttemperature,donotstart
the operation until the built-in compensating sensor has become stable
(about10minutes).
• IfthereisnoneedfortheSimulatortoperformthefunctionofauto-
maticreference-junctioncompensation,presstheRJ-ONbuttonand
thesymbolRJ-ONwillnolongerappearinthedisplay.
Simulating output from thermal resistance or RTD
Note: Theinstrumentproducesthesimulationresistanceupto400Ωatits
outputterminal(RTD/Ω).Themethodofsimulatingresistanceoutputisto
sendoutanappropriatevoltage‘Vx’accordingtotheexcitationcurrent
‘Ix’producedbythecalibratedinstrument.BecauseR(setresistance)=
Vx(outputvoltage)/Ix(excitationcurrent),thecalibrateddevicemust
provideanexcitationcurrenttothesimulator.Theexcitationcurrent
shouldrangefrom0.5mAto2mA.
Note: Whentestingtheresistanceoutputofa4-wiresystemwhile
employing a 2-wire connection, you should take into consideration the
error(approx.0.1Ω)arisingfromtheleadresistanceofthetestleads.If
the capacitance between the resistance output terminal of the simulator
andthetestedinstrumentismorethan0.1µf,thesimulatorwillproduce
improper resistance.