
Instruction Manual TAP / Temperature And Pressure Gauge EnglishEnglish Instruction Manual TAP / Temperature And Pressure Gauge
Calibration
The accuracy of all measuring devices will degrade over time. Cali-
bration improves the accuracy of the TAP. On this device temperature
and pressure can be re-calibrated. Temperature calibration can be
done using two procedures: calibration using ice water as reference
or by using room temperature as reference.
Calibration of TAP-CLAMP
Instruction Manual TAP / Temperature And Pressure Gauge English
5
The TAP must not be used for anything other than the below-stated purposes.
The TAP should not be exposed to moisture or used in damp or wet environments.
Remove refrigerants from the TAP and the hoses after use.
Purpose and use
The TAP has been developed for measuring pressure and temperature in both mobile
and stationary refrigeration equipment.
TAP is only designed for temporary measurement, don`t use it as a stationary long term
measurement unit.
Scope of delivery
Information about the various models and variations of our products can be found in the
REFCO catalogue or at www.refco.ch. The REFCO App is available on App Store and
Play Store.
Technical Data
Temperature clamp
-40 °C to +125 °C / -40 °F to +257 °F
6 mm to 41 mm / ¼" to 1-5/8"
0 °C to +50 °C / +32 °F to +122 °F
-20 °C to + 60 °C / -4 °F to 140 °F
3 x 1.5 V AAA / batteries
Service life of approx. 50 hours when used continuously.
Battery life varies with the brand or age of battery.
Automatic power save
mode:
Settable to: 10 min / 20 min / Off
Risk of short circuit. If the electronic components come into
contact with water, short circuits are possible. Don`t dunk
theTAP-CLAMPbelowmax.waterlevel.(Seegurebelow)
Ice water calibration (0 °C / 32 °F)
Instruction Manual TAP / Temperature And Pressure Gauge English
5
The TAP must not be used for anything other than the below-stated purposes.
The TAP should not be exposed to moisture or used in damp or wet environments.
Remove refrigerants from the TAP and the hoses after use.
Purpose and use
The TAP has been developed for measuring pressure and temperature in both mobile
and stationary refrigeration equipment.
TAP is only designed for temporary measurement, don`t use it as a stationary long term
measurement unit.
Scope of delivery
Information about the various models and variations of our products can be found in the
REFCO catalogue or at www.refco.ch. The REFCO App is available on App Store and
Play Store.
Technical Data
Temperature clamp
-40 °C to +125 °C / -40 °F to +257 °F
6 mm to 41 mm / ¼" to 1-5/8"
0 °C to +50 °C / +32 °F to +122 °F
-20 °C to + 60 °C / -4 °F to 140 °F
3 x 1.5 V AAA / batteries
Service life of approx. 50 hours when used continuously.
Battery life varies with the brand or age of battery.
Automatic power save
mode:
Settable to: 10 min / 20 min / Off
Risk of short circuit. If the electronic components come into
contact with water, short circuits are possible. Dunk the TAP-
CLAMP below max. water level as shown.
1. Prepare ice water:
Take your time, use lots of ice and stir wa-
ter often. It can take up to 15 minutes for
the ice water temperature to settle exact-
ly at 0.0 °C / 32.0 °F Use separate accura-
te thermometer to verify water temp.
For best results use an insulated cup, do
not let the sensor tips touch any ice cu-
bes, only the ice water in the top 2-3 cm
of the ice bath after +/- 15 min. Do not
touch or hold the sensor wires.
2. TurnotheTAP-CLAMP.
3. While holding [UNIT], press [ON / OFF] until display shows
"CALL"
4. Release both buttons, the display shows "00"
5. Press [UNIT] button < 1 sec. to adjust the number to "06"
6. Press [ON / OFF] button < 1 sec., the display shows "00 °C"
7. Press [UNIT] button < 1 sec. to show a code of temperature
8. Put the K – thermocouple (plate) into 0 °C ice water.
9. Wait the temperature code reading to be steady, press [UNIT]
button < 1 sec.
10. The display shows the ambient temperature.
11. Press [UNIT] button < 1 sec. to quit the temperature calibration.
Instruction Manual TAP / Temperature And Pressure Gauge English
23
Calibration
The accuracy of all measuring devices will degrade over time. Calibration improves the
accuracy of the TAP. On this device temperature and pressure can be re-calibrated.
Temperature calibration can be done using two procedures: calibration using ice water
as reference or by using room temperature as reference.
Calibration of temperature clamp
Risk of short circuit. If the electronic components come into contact with
water, short circuits are possible. Don`t dunk the temperature clamp below
max. water level. (See figure below)
Ice water calibration (0 °C / 32 °F)
Risk of short circuit. If the electronic components come into contact with
water, short circuits are possible. Dunk the temperature clamp below max.
water level as shown.
Prepare ice water:
Take your time, use lots of ice and stir water often. It can take
up to 15 minutes for the ice water temperature to settle exactly
at 0.0 °C / 32.0 °F Use separate accurate thermometer to
verify water temp.
For best results use an insulated cup, do not let the sensor
tips touch any ice cubes, only the ice water in the top 2-3 cm
of the ice bath after +/- 15 min. Do not touch or hold the
sensor wires.
Instruction Manual TAP / Temperature And Pressure Gauge English
23
Calibration
The accuracy of all measuring devices will degrade over time. Calibration improves the
accuracy of the TAP. On this device temperature and pressure can be re-calibrated.
Temperature calibration can be done using two procedures: calibration using ice water
as reference or by using room temperature as reference.
Calibration of temperature clamp
Risk of short circuit. If the electronic components come into contact with
water, short circuits are possible. Don`t dunk the temperature clamp below
max. water level. (See figure below)
Ice water calibration (0 °C / 32 °F)
Risk of short circuit. If the electronic components come into contact with
water, short circuits are possible. Dunk the temperature clamp below max.
water level as shown.
Prepare ice water:
Take your time, use lots of ice and stir water often. It can take
up to 15 minutes for the ice water temperature to settle exactly
at 0.0 °C / 32.0 °F Use separate accurate thermometer to
verify water temp.
For best results use an insulated cup, do not let the sensor
tips touch any ice cubes, only the ice water in the top 2-3 cm
of the ice bath after +/- 15 min. Do not touch or hold the
sensor wires.
Room temperature calibration
1. TurnotheTAP-CLAMP.
2. Place the TAP-CLAMP at constant temperature at 25 ± 3 °C con-
dition for 2 hrs. to ensure the temperature of the K – thermocou-
ple to be even.
3. While holding [UNIT], press [ON / OFF] until display shows
"CALL"
4. Release both buttons, the display shows "00"
5. Press [UNIT] button < 1 sec. to adjust the number to "09"
6. Press [ON / OFF] button < 1 sec. to show a code of temperature.
7. Wait the temperature code reading to be steady, press [ON /
OFF] button < 1 sec.
8. The display shows the ambient temperature.
9. Press [ON / OFF] button < 1 sec. to quit the temperature calibra-
tion.
Calibration of pressure gauge
Pressure calibration
1. TurnotheTAP.
2. While holding [UP / MENU] button, press [ON / OFF] button, the
display shows "P-CALL"
3. Press [DOWN / SET] button until the display shows
"Password 00"
4. Press [UP / MENU] button shortly to adjust the number to
"Password 8"
5. Press [DOWN / SET] button until the display shows a reading
which is presenting the unload pressure. (i.e. 0.0 psi)
6. Press [DOWN / SET] awhile to enter calibration
Instruction Manual TAP / Temperature And Pressure Gauge English
5
The TAP must not be used for anything other than the below-stated purposes.
The TAP should not be exposed to moisture or used in damp or wet environments.
Remove refrigerants from the TAP and the hoses after use.
Purpose and use
The TAP has been developed for measuring pressure and temperature in both mobile
and stationary refrigeration equipment.
TAP is only designed for temporary measurement, don`t use it as a stationary long term
measurement unit.
Scope of delivery
Information about the various models and variations of our products can be found in the
REFCO catalogue or at www.refco.ch. The REFCO App is available on App Store and
Play Store.
Technical Data
Temperature clamp
-40 °C to +125 °C / -40 °F to +257 °F
6 mm to 41 mm / ¼" to 1-5/8"
0 °C to +50 °C / +32 °F to +122 °F
-20 °C to + 60 °C / -4 °F to 140 °F
3 x 1.5 V AAA / batteries
Service life of approx. 50 hours when used continuously.
Battery life varies with the brand or age of battery.
Automatic power save
mode:
Settable to: 10 min / 20 min / Off
Ensure all the refrigerant is exhausted from TAP valve or
connected hoses before performing calibration. The ac-
curacywillbeaectedduetotheremainingrefrigerant.
7. While entering calibration, the display shows "400.0 psi". The
default pressure of calibration standard is 400 psi.
8. Press [UP / MENU] button shortly to select the calibration stan-
dard. The inlet pressure must be equal to the calibration stan-
dard pressure.
9. Press[DOWN/SET]awhileforcalibrationstandardconrmati-
on; the display shows the reading of unloaded pressure.
10. Wait until the reading of unload pressure ist stable, pressuirze
400 psi (the inlet pressure must be equal to the calibration stan-
dard pressure). The display shows the corresponding pressure
value (i.e. 400.0 psi)
11. Whenthereadingisstable,press[DOWN/SET]forconrmati-
on. Now the display shows "END"
12. Press [DOWN / SET] awhile to quit the pressure calibration.