Reliance 520 User manual

OWNERS MANUAL
INSTALLATION AND
OPERATING INSTRUCTIONS
MODEL 520
LOW VOLTAGE CONTROL CHAIR

QUALITY MANAGEMENT
SYSTEM
REGISTERED TO
ISO 9001
ISO 13485
MADE IN AMERICA
Reliance® Medical Products, Inc. • 3535 Kings Mills Road • Mason, Ohio 45040-2303 • 1-800-735-0357 • www.reliance-medical.com
SERIAL NUMBER
For your future reference, mark the serial number
in the space provided.

TABLE OF CONTENTS
IMPORTANT INFORMATION.............................................4
1. INTRODUCTION.........................................................5
TECHNICAL SPECIFICATIONS.........................................6
2. INSTALLATION...........................................................7
2.1. Unpacking.......................................................7
Chair......................................................................7
2.2. Assembly.........................................................7
Headrest......................................................................7
Power Receptacle...................................................7
3. OPERATION..............................................................8
3.1. Chair Lift........................................................8
Raising & Lowering the Chair Lift from the
Footswitch...............................................................8
3.2. Chair Top........................................................8
Posistioning the Chair Back.....................................8
3.3. Revolving/Rotation..........................................9
3.4. Armrest Movement..........................................9
3.5. Footrest..........................................................9
3.6. Headrest.........................................................9
4. LOCATING AND REPLACING FUSES............................9
4.1. Fuse Replacement............................................9
TROUBLESHOOTING GUIDE..........................................11
CLEANING AND MAINTENANCE....................................12
PARTS LIST.................................................................14
Model 520 Chair Assembly....................................15
Chair Top Assembly...............................................16
Chair Base Assembly.............................................18
Lift Assembly.........................................................20
Headrest Assembly................................................22
Wiring Diagram - 520............................................24
LIMITED WARRANTY.....................................................27

IN-5204
TRANSPORTATION / STORAGE CONDITIONS
Temperature: Operating: 59 to 104º F
15 to 40º C
Storage: -22 to +140º F
-30 to +60º C
R e l a t i v e H u m i d i t y : O p e r a t i n g : 5 t o 9 5 % w i t h o u t
condensation
Storage: 10 to 100% including
condensation
Atmospheric Pressure:
Operating: 700hPa - 1060hPa
Storage: 500hPa - 1060hPa
IMPORTANT INFORMATION
SAFETY SYMBOLS
“NOTE”
Amplies a procedure, practice, or condition.
“DANGER”, “WARNING”, or “CAUTION”
The exclamation point within an equilateral
triangle is intended to alert the user to
the presence of important operating and
maintenance (servicing) instructions in this
Installation and Operating Instructions.
« REMARQUE »
Amplie une procédure, pratique ou condition.
« DANGER », « AVERTISSEMENT », ou
« PRÉCAUTION »
Le point d’exclamation à l’intérieur d’un
triangle équilatéral vise à alerter l’utilisateur
à la présence d’importantes instructions
de fonctionnement et de maintenance
(service) dans ce manuel d’installation et
d’instructions de fonctionnement.
Alternating Current-AC
Safe Working Load
“Attention, Consult Accompanying
Documents”
Refer to Instruction Manual/Booklet
!
Type B, Applied Part
Dangerous Voltage / Shock Hazard
Fuse Rating Specication
Protective Earth Ground
Burn Hazard/Hot Surface

“NOTE”- Note indicates points of particular or
additional information.
« REMARQUE » : Remarque indique des points
d’intérêt particulier ou des renseignements
additionnels.
1.3 Should your product not perform properly, or if
you have any questions concerning the use and care
of any Reliance® product, contact the Reliance®
Distributor, where you purchased this product or contact
the Technical Service Department, Reliance® Medical
Products, Inc., 3535 Kings Mills Road, Mason, Ohio
45040-2303, or call (800) 735-0358.
NOTE:
Always have the model number and
serial number available before contacting
Reliance® or your authorized Reliance®
Distributor.
REMARQUE : Ayez toujours le numéro de
modèle et le numéro de série à portée de la
main avant de contacter Reliance® ou votre
distributeur Reliance® autorisé.
“CLASSIFIED BY CANADIAN STANDARDS
ASSOCIATION® WITH RESPECT TO ELECTRIC
SHOCK, FIRE AND MECHANICAL HAZARDS
ONLY IN ACCORDANCE WITH UL 60601-1.”
According to Clause 5 in IEC 60601-1, sec 6.8.1, this
unit is classied by the following:
• The type of protection against electric shock:
EQUIPMENT energized from an external electrical
power source: CLASS I EQUIPMENT
• The degree of protection against electric shock:
TYPE B EQUIPMENT
• The degree of protection against harmful ingress of
water: ORDINARY DEGREE
• The degree of safety of application in the presence
of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR
or WITH OXYGEN OR NITROUS OXIDE: EQUIPMENT
not suitable for use in the presence of a FLAMMABLE
ANAESTHETIC MIXTURE WITH AIR or WITH OXYGEN
OR NITROUS OXIDE
• The mode of operation: CONTINUOUS OPERATION
1. INTRODUCTION
1.1. This manual contains information applicable only
to the Reliance® Model 520 Chair.
1.2. Whenever you see the symbols shown below,
heed their instructions! Always follow safe operating
and maintenance practices.
“DANGER”-
THE DANGER SYMBOL
IDENTIFIES SPECIAL INSTRUCTIONS OR
PROCEDURES WHICH, IF NOT CORRECTLY
FOLLOWED, COULD RESULT IN LOSS OF
LIFE OR PERSONAL INJURY.
« DANGER » : LE SYMBOLE DANGER
IDENTIFIE DES INSTRUCTIONS OU
PROCÉDURES SPÉCIALES QUI, SI ELLES
NE SONT PAS SUIVIES CORRECTEMENT,
POURRAIENT CAUSER UNE PERTE DE
VIE OU UNE BLESSURE.
“WARNING”-
THE WARNING SYMBOL
IDENTIFIES SPECIAL INSTRUCTIONS
OR PROCEDURES WHICH, IF NOT
CORRECTLY FOLLOWED, COULD RESULT
IN PERSONAL INJURY.
« AVERTISSEMENT » : LE SYMBOLE
AVERTISSEMENT IDENTIFIE DES
INSTRUCTIONS OU PROCÉDURES
SPÉCIALES QUI, SI ELLES NE SONT PAS
SUIVIES CORRECTEMENT, POURRAIENT
CAUSER UNE BLESSURE.
“CAUTION”-
This caution symbol identifies
special instructions or procedures which,
if not strictly observed, could result in
damage to or destruction of equipment.
« PRÉCAUTION » : Ce symbole de
précaution identifie des instructions ou
procédures spéciales qui, si elles ne
sont pas strictement suivies, pourraient
causer un dommage ou la destruction
de l’équipement.
IN-520 5

WARNING: TO PREVENT FIRE OR
ELECTRICAL SHOCK HAZARD, DO NOT
EXPOSE THIS EQUIPMENT TO RAIN OR
MOISTURE.
AVERTISSEMENT : POUR ÉVITER UN
INCENDIE OU UN RISQUE DE CHOC
ÉLECTRIQUE, N’EXPOSEZ PAS CET
ÉQUIPEMENT À LA PLUIE OU À L’HUMIDITÉ.
TECHNICAL SPECIFICATIONS
Rating for 120V Low Voltage Chair
This equipment is rated Class I, Type B
Equipment Model
Low Voltage Chair 520
Volts Hertz Amps
120V 50/60 6.3
Class I Type B
Rating for 220-240V Low Voltage Chair
This equipment is rated Class I, Type B
Equipment Model
Low Voltage Chair 520
Volts Hertz Amps
220-240V 50/60 4.0
Class I Type B
IMPORTANT USER WARNINGS
WARNING: EXPLOSION HAZARD. THIS
EQUIPMENT MUST NOT BE USED IN THE
PRESENCE OF FLAMMABLE ANESTHETICS.
AVERTISSEMENT : RISQUE D’EXPLOSION.
CET ÉQUIPEMENT NE DOIT PAS ÊTRE
UTILISÉ ENPRÉSENCED’ANESTHÉSIQUES
INFLAMMABLES.
POTENTIAL ELECTROMAGNETIC or OTHER
INTERFERENCE
This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices
in the vicinity. However, there is no guarantee that interference
will not occur in a particular installation.
If this equipment does cause harmful interference to other
devices, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference
by one or more of the following measures:
• Reorient or relocate the receiving device.
• Increase the separation between the equipment.
• Connect the equipment into an outlet on a circuit different
from that to which the other device(s) are connected.
• Consult the manufacturer or eld service technician for
help.
WARNING - STORAGE OF THIS EQUIPMENT
BELOW 40°F (4.4ºC) WILL CAUSE THE
HYDRAULIC OIL TO BECOME THICK. THIS
MAY CAUSE UNUSUAL NOISES WHEN
THE BASE MOTOR IS OPERATED. IT IS
RECOMMENDED THAT THE EQUIPMENT
STAY AT A TEMPERATURE OF 60°F (15.6ºC)
OR ABOVE FOR AT LEAST 8 HOURS BEFORE
OPERATING.
AVERTISSEMENT : L’ENTREPOSAGE DE
CET ÉQUIPEMENT À MOINS DE 4,4°C
(40°F) FERA EN SORTE QUE L’HUILE
HYDRAULIQUE DEVIENDRA ÉPAISSE.
CECI POURRAIT CAUSER DES BRUITS
INHABITUELS LORSQUE LE MOTEUR DE LA
BASE FONCTIONNE. ON RECOMMANDE
QUE L’ÉQUIPEMENT DEMEURE À UNE
TEMPÉRATURE DE 15,6°C (60°F) OU PLUS
DURANT AU MOINS 8 HEURES AVANT DE
L’UTILISER.
IN-5206

IN-520 7
2. INSTALLATION
2.1. Unpacking
2.1.1. Chair
2.1.1.1. Cut the shipping bands, open the top of the
shipping carton, and remove all packing material and boxes
that can be easily reached. Remove (4) nuts and remove
both base clamping boards.
2.1.1.2. It is recommended that the skid be positioned as
close as possible to the desired Chair location.
2.1.1.3. Remove at least two small shipping blocks stapled
to the top of the skid. The Chair can now be slid and/or
tipped from the skid onto the oor.
2.1.1.4. Do not lift the Chair by the upper structure.
However, the Chair may be tipped or slid into position by
pushing or pulling on the upper structure.
2.1.1.5. With the Chair in positon, remove all remaining
paper pads, plastic, tape, strings, etc.
2.2. Assembly
2.2.1. Headrest
2.2.1.1. Remove the Headrest from its carton. Remove
seat upholstery by pulling the top away from Chair structure.
Attach Headrest to Chair frame using two 5/16 x 5/8
hex head cap screws supplied with Chair.
Replace seat
upholstery. Insert the two locating screws on seat frame
into the two holes in bottom of seat upholstery. Press back
to frame so the hook and loop fabric fasteners engage.
2.2.2. Power Receptacle
2.2.2.1. Locate the Footswitch assembly, Chair Control
Cable, and the Power Cord Assembly. Refer to gure 1 for
receptacle location and attach all cables.
2.2.2.2. To disconnect power to Chair unplug Power Cord
from Instrument Stand or from wall receptacle.
FIGURE 1
BASE
Power Cord &
Fusedrawer
Footswitch
Cable
Chair Control
Cable

NOTE: If this Chair is purchased in conjunction with
a Reliance® Instrument Stand, the Chair Control
Cable supplied will provide proper operation. Refer
to the Instrument Stand Manual for connection to
the Stand.
REMARQUE : Si cette chaise est achetée avec
un support à instruments Reliance®, le câble
de commande de la chaise fourni permettra un
fonctionnement adéquat. Se référer au manuel du
support d’instrument pour savoir comment faire la
connexion au support.
3. OPERATION
CAUTION- Caution should be taken to ensure
there are no obstructions (instruments, trays,
tables) impeding movement of the Chair
PRÉCAUTION : On devrait faire attention
pour s’assurer qu’il n’y a pas d’obstructions
(instruments, plateaux, tables) empêchant le
mouvement de la chaise.
Each time main power is applied to the Chair, the circuit
board inside the Chair will respond with a one second beep.
The beep indicates the electronics are functioning properly.
3.1. Chair Lift
3.1.1. Raising & Lowering the Chair Lift from the
Footswitch
• For operation from the Footswitch (Figure 2), depress the
right side of the pedal to raise the Chair lift.
• Depress the left side of the pedal to lower the Chair lift.
NOTE: The RAISE and LOWER BASE switches
on the Footswitch are programmable. By design,
the motion of the Chair will STOP when either
of these switches is released. If you want the
motion of the Chair to CONTINUE after either of
these switches is released, then see Section 4
for programming instructions.
REMARQUE : Les interrupteurs RAISE (LEVER) et
LOWER (BAISSER) la base sur l’interrupteur au pied
sont programmables. Selon la conception initiale,
le mouvement de la chaise S’ARRÊTE lorsque l’un
ou l’autre de ces interrupteurs est relâché. Si vous
désirez que le mouvement de la chaise CONTINUE,
après que l’un ou l’autre de ces interrupteurs est
relâché, consultez la section 4 pour connaître les
instructions de programmation.
3.2. Chair Top
3.2.1. Positioning The Chair Top
• The Chair top may be manually tilted. To do so, release
the lock by squeezing the Release Lever and the Grab Bar
on the back of the Chair together. Push back to recline.
Release two bars to lock top in desired position. (Figure
3)
.
WARNING- FIRM CONTROL OF THE CHAIR
TOP MUST BE MAINTAINED DURING THE
RECLINING OPERATION. THIS IS ESPECIALLY
IMPORTANT WITH HEAVIER OR LIGHTER THAN
AVERAGE PATIENTS.
“AVERTISSANT” - LE CONTRÔLE FERME DU
SOMMET DE PRÉSIDENCE DOIT ÊTRE MIS
À JOUR PENDANT L’EXÉCUTION ÉTENDUE.
C’EST PARTICULIÈREMENT IMPORTANT
AVEC PLUS LOURD OU LE CHALAND QUE LES
PATIENTS MOYENS.
FIGURE 2
FOOTSWITCH OPERATION
LIFT
DOWN LIFT
UP
IN-5208

IN-520 9
FIGURE 3
RELEASE LEVER
3.3. Revolving/Rotation
3.3.1. The Chair upper structure may be revolved 350°,
with a built-in stop to prevent continuous rotation. The lock
for revolving is controlled by a hand lever with black knob
located below the seat at the center and on either side. Push
the lever backward to release the lock and pull upward to
apply the lock.
3.4. Armrest Movement
3.4.1. The Chair Armrests may be raised to provide better
access to the patient or to permit entrance to or exit from
the Chair.
3.4.2. The movements of both Armrest are identical and
completely independent of each other.
3.4.3. To raise the Armrest simply lift Armrest to the stop
parallel to Chair back.
CAUTION: This exam Chair is designed
for use in a specific position called the
Intended Use Position. Prior to reclining a
patient, ensure the Chair Back is aligned
with the Chair’s power receptacle and
that the Chair Swivel is locked. The safe
working load is reduced if the Chair is
swiveled away from this Intended Use
Position, and may result in insufficient
stability.
GRAB BAR
RELEASE LEVER PRÉCAUTION : Cette chaise d’examen
est conçue pour utilisation dans une
position spécifique appelée la position
d’utilisation prévue. Avant d’incliner un
patient, assurez-vous que le dossier de
la chaise est aligné avec le réceptacle
d’alimentation de la chaise et que le pivot
de la chaise est verrouillé. La charge
de travail sécuritaire est réduite si la
chaise est tournée hors de sa position
d’utilisation prévue et cela pourrait causer
une stabilité insuffisante.
CAUTION-This exam Chair is designed
for use with a specific maximum patient
weight as a static (non-moving) load.
Sudden weight shifts or impact loads
can exceed the safe static load rating of
this Chair, and may result in insufficient
stability.
PRÉCAUTION : Cette chaise d’examen
est conçue pour utilisation avec un poids
de patient maximum spécifique comme
charge statique (sans mouvement). Les
déplacements de poids soudains ou les
charges dynamiques peuvent excéder
la norme de charge statique sécuritaire
de la chaise, ce qui pourrait causer une
stabilité insuffisante.
CAUTION-This exam Chair is designed
for use with a patient properly positioned
in the Chair. Prior to reclining the Chair,
ensure that the patient is seated in the
Chair and that the patient position does
not move to an extreme position either
toward the head of the Chair or toward
the foot of the Chair, since this may result
in insufficient stability.
PRÉCAUTION : Cette chaise d’examen est
conçue pour utilisation avec un patient
correctement positionné dans la chaise.
Avant d’incliner la chaise, assurez-vous
que le patient est assis dans la chaise et
que la position du patient ne se déplace
pas vers une position extrême soit vers
la tête, soit vers le pied de la chaise,
puisque cela pourrait causer une stabilité
insuffisante.

4.1. Fuse Replacement (AC Input Module)
4.1.1 Unplug the power cord from the AC power Input Module.
Referring to Figure 4, locate tab. With a tool similar to a small
common screwdriver placed under tab, pry Fuse Drawer out.
Remove the Fuse Drawer by pulling it out with your ngers.
4.1.2 To remove fuses from Fuse Drawer, tilt fuses in direction
of arrows as shown in Figure 4. Examine the fuses, replace
as necessary.
FIGURE 4
AC INPUT MODULE
3.5. Footrest
3.5.1. The Footrest may be retracted by lifting near its
outer edge. When retracted, the Footrest stays in place.
3.6. Headrest
3.6.1. A xed, non-articulating Headrest is the only Headrest
available on the 520 Chair. This Headrest is ergonamically
designed for patients of a wide range of heights.
4.0 LOCATING AND REPLACING FUSES
WARNING: DISCONNECT EQUIPMENT FROM
MAIN INPUT POWER BEFORE PROCEEDING WITH
ELECTRICAL INSPECTIONS OR MAINTENANCE.
AVERTISSEMENT : DÉCONNECTEZ
L’ÉQUIPEMENT DE L’ALIMENTATION
ÉLECTRIQUE PRINCIPALE AVANT DE
PROCÉDER À DES INSPECTIONS ÉLECTRIQUES
OU À DE L’ENTRETIEN.
CAUTION - Replace fuse(s) as marked. All
fuses must be replaced with a fuse of the same
size and rating. Refer to the Wire Diagrams at
the end of this Manual.
PRÉCAUTION : Remplacez le(s) fusible(s)
comme indiqué. Tous les fusibles doivent
être remplacés par un fusible de la même
dimension et de la même valeur. Se référer
aux schémas de câblage à la fin de ce manuel.
The Chair contain two fuses located inside the AC Input Module
in the Base Assembly.
IN-52010

PROBLEM PROBLEM CAUSE SOLUTION
CHAIR WILL NOT OPERATE. 1) POWER CORD NOT PLUGGED IN.
2) NO AC POWER AT WALL RECEPTACLE.
3) BLOWN LINE FUSE.
1) CHECK ALL WIRING CONNECTIONS.
2) RESTORE OUTLET POWER.
3) REPLACE FUSE.
OUTLET POWER OK, BUT NO
BEEP IS HEARD WHEN POWER
IS APPLIED.
1) BLOWN LINE FUSE.
2) LOOSE WIRING IN BASE.
3) OTHER.
1) REPLACE FUSE.
2) CHECK ALL WIRING CONNECTIONS.
3) CONTACT FACTORY
FOOTSWITCH FUNCTIONS NOT
OPERATING.
1) FOOTSWITCH CORD NOT CONNECTED. 1) CHECK ALL INTERNAL AND
EXTERNAL FOOTSWITCH WIRING
CHAIR BASE WILL NOT STAY UP 1) SOLENOID VALVE NOT CLOSED.
(DIRT IN VALVE.)
2) CONSTANT POWER APPLIED TO
SOLENOID.
3) OTHER
1) REPAIR OR REPLACE SOLENOID
VALVE.
2) CHECK FOR SWITCHES STUCK
"CLOSED" .
3) CONTACT FACTORY.
CHAIR BASE WILL NOT LOWER. 1) NO POWER TO SOLENOID ASS'Y.
2) SOLENOID COIL BURNED-OUT
3) OTHER
1) CHECK FOR VOLTAGE.
2) REPLACE SOLENOID COIL.
3) CONTACT FACTORY.
CHAIR BASE WILL NOT RAISE.
PUMP MOTOR DOES NOT RUN.
1) NO POWER TO MOTOR PUMP ASS'Y
2) UP-LIMIT SWITCH STUCK "OPEN".
3) PUMP MOTOR BURNED-OUT.
4) OTHER
1) CHECK FOR VOLTAGE.
2) RE-ADJUST OR REPLACE SWITCH.
3) REPLACE MOTOR-PUMP ASS'Y.
4) CONTACT FACTORY.
CHAIR BASE WILL NOT RAISE.
PUMP MOTOR RUNS.
1) SOLENOID VALVE NOT CLOSED.
(DIRT IN VALVE.)
2) CONSTANT POWER APPLIED TO
SOLENOID.
3) OTHER.
1) REPAIR OR REPLACE SOLENOID
VALVE.
2) CHECK FOR SWITCHES STUCK
"CLOSED".
3) CONTACT FACTORY.
BASE "STUTTERS" DURING
ASCENT.
1) HYDRAULIC FLUID LEVEL IS LOW. 1) CONTACT DEALER OR FACTORY
FOR INFORMATION ON ADDING OIL.
CHAIR BASE WILL NOT RAISE
OR LOWER FROM FLOOR UNIT.
1) CHAIR CONTROL CABLE NOT PLUGGED
IN.
2) BLOWN FUSE IN CHAIR BASE.
1) CHECK CABLE CONNECTIONS AT
STAND CHAIR
2) REPLACE FUSE.
CHAIR TOP "WOBBLES" ON
CHAIR BASE
1) LOCK NUT IS LOOSE.
(UNDER SEAT BOARD.)
1) TIGHTEN LOCK NUT.
CHAIR BACK RECLINES FULLY
WITHOUT HANDLE BEING
ACTIVATED.
1) GAS SPRING IN NEED OF ADJUSTMENT 1) ADJUST GAS SPRING
2) REPLACE GAS SPRING
2) GAS SPRING FAILURE
TROUBLESHOOTING GUIDE
IN-520 11

IN-52012
CLEANING AND MAINTENANCE
Cleaning the Upholstery
When cleaning upholstery, it is suggested that you use all
recommended commercial cleaning agents in an inconspicu-
ous spot to check for potential damage to the upholstery
material before applying to the entire area to be cleaned.
Day to Day soil:
Remove ordinary dirt and smudges with a mild soap and
warm water solution. Dry with a soft, lint-free cloth or towel.
For more difcult stains, use of a stronger detergent is
recommended; however, follow the detergent manufacturer’s
instructions closely.
Disinfecting Material:
In order to disinfect, use a 1:9 solution of bleach and water (1
part bleach to 9 parts water). Wipe surface with damp cloth
then rinse with water immediately, using a damp cloth.
Special cleaning problems:
The following methods are recommended to clean stains
on PreFixx-protected vinyl upholstery. Most stains will be
removed when these cleaning agents are used in the fol-
lowing order.
Type 1 Cleaners: Nonabrasive household cleaners to be
used with water and a soft cloth.
• Formula 409®~ All-Purpose Spray Cleaner
• Fantastik® Spray Cleaner
• Other similar household cleaners and bleaches
Type 2 Cleaners: Solvent-type cleaner to be liberally
applied with a soft cloth. Dry area with another cloth, rinse
with clean water and dry.
• Rubbing alcohol (isopropyl alcohol)
• Lighter uid (naptha)
Type 3 Cleaners: Strong, active solvent cleaners such as
nail polish remover (acetone/water) to be applied with a soft
cloth. Stain should be removed with less than six (6) rubs; if
stain persists after six rubs, contact manufacturer. Dry area
with another cloth, rinse with clean water and dry.
CAUTION- These solvents are highly
flammable. Exercise proper care in cleaning
and notify personnel in area of danger. Wear
rubber gloves during all cleaning activity. Use
caution in cleaning around buttons, stitching
and wooden or other decorative trim, since
these solvents could seriously damage such
areas.
PRÉCAUTION : Ces solvants sont extrêmement
inflammables. Faites très attention lors du
nettoyage et avisez le personnel dans la
zone de tout danger. Portez des gants de
caoutchouc durant toute activité de nettoyage.
Faites attention en nettoyant autour des
boutons, des coutures et autres finitions de
bois ou décoratives, puisque ces solvants
pourraient endommager sérieusement ces
endroits.
Cleaning Painted Surfaces
The painted metal surfaces are covered in durable, powder-
coated paint which is resistant to scratching and scufng.
It may be cleaned with a clean cloth dampened with mild,
soapy water or equivalent household product that uses spray
application.
Recommended Infection Control Products
The disinfectant/cleaner products that we have tested and
recommend for our upholstery, painted items and plastic
covers are as follows.
• Cavicide® Hospital Disinfectant
• Precise® Hospital Foam Cleaner Disinfectant
CAUTION:Followmanufacturer’sdirectionsfor
concentration and application of disinfecting
cleaner. Avoid prolonged application of any
disinfectant/cleaner products, because
they may cause staining or discoloration of
material.
PRÉCAUTION : Suivez les instructions
du fabricant pour la concentration et
l’application de nettoyeurs désinfectants.
Évitez une application prolongée de tout
produit désinfectant/nettoyant, parce que
ceux-ci pourraient causer des taches ou une
décoloration du matériel.

IN-520 13
(NOTES)

PARTS LIST
RELIANCE® MODEL 520
EXAMINATION AND TREATMENT CHAIR
Note: When ordering parts, please:
1. Advise dealer or factory of model and serial number of unit. These numbers are on the
plate that is located at rear, near bottom, of Chair back.
2. Specify color of painted parts. Painted parts have an asterisk(*) behind the part description.
REMARQUE : Lorsque vous commandez des pièces, veuillez :
1. Informez le distributeur ou l’usine du modèle et du numéro de série de l’ensemble. Ces
numéros sont de la plaque qui est située à l’arrière, près du bas, de l’arriére de présidence.
2. Spécier la couleur des pièces peintes. Les pièces peintes ont un astérisque (*) derrière la
description de pièce.
Should your Chair not perform properly, or if replacement parts are
needed, contact the Reliance® Distributor, where you purchased this
product or contact the Technical Service Department, Reliance® Medical
Products, Inc., 3535 Kings Mills Road, Mason, Ohio 45040-2303, or call
(800) 735-0358.
IN-52014

57, 8
6
42
MODEL AND SERIAL
3
1
NUMBER PLATE
19.09"
30.94"
44.00" LOW
46.50" HIGH
BASE TRAVEL
8.75" LOW
11.25" HIGH
11.12"
38.96"
40°
BASE
21.00" LOW
23.50 HIGH
BASE TRAVEL
8.75" LOW
11.25 HIGH
28.88" LOW
31.38" HIGH
BASE TRAVEL
8.75" LOW
11.25 HIGH
7.55"
FIGURE 5
520 CHAIR ASSEMBLY
IN-520 15
ITEM PART NO. DESCRIPTION ITEM PART NO. DESCRIPTION
1Chair Top Assembly* (See Figure 6)
2Chair Base Assembly* (See Figure 7)
3Headrest Assembly* (See Figure 9)
42342203 Footrest
5
3070973 Upholstered Seat Assembly W/Hole**
3070673 Upholstered Seat Assembly W/O Hole**
72347373 Upholstered Armrest Assembly - RH**
82347473 Upholstered Armrest Assembly - LH**
63070573 Upholstered Apron Assembly**
** Please specify upholstery color when ordering.
To order complete Upholstery Set use P/N.
3085673 Upholstery Set - 520 & 522 with holes
3070873 Upholstery Set - 520 & 522 without holes

22
42
43
54
29
44
16
32
17
27
18
31
11
6
15
40
24
41
46
33
4
38
10
19
45 49
7
36
47
30
5
48
15
26
34
20
52
14
55 13
39
53
2
28
25 37 21
50
1
23
51
9
835
56
(2) (4)
(2)
(2)
(2)
(2)
(2)
(2)
(2)
(2)
(6)
(4)
(4)
(2)
(2)
(2)
(2)
(2)
(2)
(2)
(4)
15
15
15
15
12
3
FIGURE 6
520 CHAIR TOP ASSEMBLY
IN-52016

*Please specify paint color when ordering.
IN-520 17

20 10
11
18
19
7
5
9
22
24
23
27
25
15
30
8
12
6
28, 17
33
32
13
21
30
28, 17
29
1
31, 30
2
3, 4
16, 17
14
FIGURE 7
CHAIR BASE ASSEMBLY
IN-52018

ITEM PART NO. DESCRIPTION ITEM PART NO. DESCRIPTION
12081203 Base 16 0458999 Tubing-5-1/2 in
22065599 Base Cover 17 1081999 Tubing Insert
3
2065199 Base Housing - Low - RH 18 0559999 Thrust Bearing
2065399 Base Housing - High - RH 19 0560099 Thrust Washer
4
2065299 Base Housing - Low - LH 20 0123885 Lock Nut
2065499 Base Housing - High - LH 21 1127599 Union Tee
5Lift Assembly (See Figure 8) 22 1659299 Power Cord Assembly
6
1707792 Solenoid Valve Assembly-120V 23 1550599 Chair Control Cable-LV To LV
1707892 Solenoid Valve Assembly-230V 24 2346492 Footswitch Assembly
7
1702892 Keyshield Assembly-Low*
25
1634299 Fuse, 6.3 A / 250V-120V
1702992 Keyshield Assembly-High* 1634999 Fuse, 4 A / 250V-230V
8
2720092 Pump/Motor Assembly-120V 26 1976799 Power Entry Module W/ Filter
2720192 Pump/Motor Assembly-230V 27 Fusedrawer, 2 Pole (In Item 26)
91547899 Piston Rod Bolt 28 0458999 Tubing-5 in
10 1525899 Set Screw-#10-24 x 1/4
29
1597299 Capacitor, Starting-120V
11 0607299 Washer 2002199 Capacitor, Starting-230V
12 2721092 Hose Assembly 30 1720399 PH Phil w/Lock Washer-#10-24 x 1/2
13 1596292 Pressure Tube Assembly 31 2081099 Housing Mount Bracket
14 2025599 P.C. Board Assembly 32 0708599 HHCS 3/8-16 x 1.5
15 2720892 AC Input Assembly 33 1569999 Circuit Board Support
To order complete Base Assembly use P/N’s.
2724092 Chair Base Assembly - Low - 120V*
2724192 Chair Base Assembly - High - 120V*
2724292 Chair Base Assembly - Low - 220-240V*
2724392 Chair Base Assembly - High - 220-240V*
*Please specify paint color when ordering.
IN-520 19

FIGURE 8
LIFT ASSEMBLY
IN-52020
1
2
10
9
3
4
6
5
7
8
11
41
38
26
42
40
43
20
30
31
32
34
27
28
29
35
36
12
13
15
16
19
18
14
17
25
2
4
23
22
21
18
39
33 & 37
Table of contents