RMP60 installation guide
2.4
RMP60 basics
Enhanced trigger filter
Probes subjected to high levels of vibration or
shock loads may output probe trigger signals
without having contacted any surface. The
enhanced trigger filter improves the probe’s
resistance to these effects.
When the filter is enabled, a constant nominal 10
or 20 ms delay is introduced to the probe output.
It may be necessary to reduce the probe
approach speed to allow for the increased stylus
overtravel during the extended time delay.
Factory set to OFF.
Hibernation mode
Only applicable to ‘radio-on’ mode.
When RMP60 is in standby and RMI or RMI-Q
is powered off or out of range, the probe enters
hibernation; a low power mode designed to save
battery life. The probe ‘wakes’ from hibernation to
periodically check for it’s partnered RMI or RMI-Q.
The 'wake-up frequency can be set to 30 seconds,
5 seconds or set to off; the probe never goes into
hibernation.
Factory set to 30 seconds.
Multiple probe mode
The RMP60 can be configured, using Trigger
Logic™, to allow multiple radio probes in ‘spin on/
off’ or ‘shank on/off’ to be used with a single RMI
or RMI-Q.
Up to four RMP60s can be used with a single
RMI-Q in ‘radio on/off’ mode. For further details
of this functionality, please refer to the RMI-Q
installation guide.
NOTES:
Multiple probe mode is a function of the RMP60,
as such, the option will not appear when the 'radio
on' option has been selected.
RMP60 probes which are set to 'multiple probe
mode on' can coexist alongside any number of
RMP60 probes set to 'multiple probe mode off'.
To allow multiple radio probes to work in close
proximity, and with a single RMI or RMI-Q, 16
choices of 'mode on' colours are available, each
representing a different machine tool installation.
See 'Multiple probe settings' in Section 4 -
Trigger Logic™.
All probes operating with a single RMI or RMI-Q
must be set to the same 'mode on' colour choice;
any multiple probes located on adjacent machines
must all be set to an alternative 'mode on' colour
choice. Only one probe per 'mode on' colour
choice needs to be partnered with the RMI or
RMI-Q as, by configuring multiple probes to a
single 'mode on' colour choice, all probes using
this 'mode on' colour choice will have the same
identity. The probe to be partnered is partnered
after selecting the 'multiple probe mode' setting
and choosing the 'mode on' option. See 'Changing
the probe settings' in Section 4 - Trigger Logic™.
There is no limit to the number of probes that can
be used with a single RMI or RMI-Q so long as
they all have the same 'mode on' colour choice. All
RMP60 probes are factory set to 'mode off'.
The addition of any further probe(s) into a single
probe installation will require that all probes are
reconfigured to the same 'mode on' colour choice
and that one of the probes are then repartnered
with the RMI or RMI-Q.
The addition of any further probe(s), or
replacements, into a multi-probe installation can
be achieved simply through the reconfiguration of
the probe to the same 'mode on' colour choice.