SCT-2200
Quick Setup Guide
To enter the setup menu, turn the instrument on and briefly press the mode key ( ) when the software version displays.
When settings are complete press until the indicator displays SAVE?. Press to save set up and return to the weigh mode.
Pressing any other key exits the setup and discards changes.
Press and to set decimals and minimum division.
Press , and to set the total capacity of the load cells.
Press , and to set the mV/V sensitivity of the load cell.
Press , and to set the weight value present on the load cell.
Press , and to set the scale capacity.
Quick calibration of zero.
Press , and to set the weight value used for the calibration.
Check the load cell signal in millivolts.
Set the analog output on net or gross weight.
Set the maximum value of the analog output.
Set the analog output when the scale shows zero weight.
Set the minimum value of the analog output.
Set the key to be emulated by the input.
Set the output normal state.
Set the output switching condition.
Enable time to disable the output.
Enable time to enable the output.
Enable or disable the hysteresis.
INPUTS
0 NONE
1 GROS
2 NET
3 PCS
4 GRO.0
5 NET.0
6 MOTI
23 K.PR
25.K.MOD
26 K.C
27.K.ZER
28.K.TAR
29 ERR
DIV.DEC
CEL.CAP
CEL.SEN
DEAD.LD
RANGE
ZERO
WEIGHT
ADC.MVV
AN.OUT
AO ZER
MODE
AO MAX
INP.02
AO MIN
INP.01
OUT.02
OUTPUT Disabled.
Setpoint for gross weight.
Setpoint for net weight, even with no tare.
Setpoint for pieces.
Gross weight at zero.
Net weight at zero.
Active on instability.
Active when the Ckey is pressed.
Error indication.
Active when is pressed.
Active when is pressed.
Active when is pressed.
Active when is pressed.
Setpoint for net weight when tare exists.
NO/NC
ONSTAT
RL.IST
ENAB.TM
DELAY
OUT.01 FUNC
1
2
3
4
5
6
7
8
9
10
11
30.NET.T
Optional Analog
Model Parameter
Key Function
ZERO
In setup: scroll forward through parameters
In numeric input: decreases the digit to be modified
TARE
In setup: scroll back through the functions
In numeric input: increases the digit to be modified
MODE
In setup: quickly position the first step of a menu
In numeric input: selects the digit to be modified, from
left to right
PRINT
In setup: enter into a parameter or to confirm a setting
In numeric input: confirms the entry made
ON/OFF
Turn the instrument on and off
In setup: exit a step without confirming the setting
In numeric input: clears the present value
C
Default Factory Calibration
The instrument has a default calibration with the following
features:
Capacity = 10,000 kg
Load cell sensitivity = 2.000mV/V
Divisions = 1
Theoretical Calibration
Use parameters 1,2,3,4 to perform a theoretical calibration.
Calibration Using Known Weight
Use parameters 1,5,6,7 to perform a standard calibration
with a known weight. For parameters 6 & 7, press and
wait for ERNOT, STORE?. Then press to accept.
For parameter 7, apply test weight before pressing .
Setpoint Programming
In the weigh mode, press and hold to access output’s
setpoint programming.
Press the print key ( ) in the setup menu to view/select the parameter choices.
PN 184974 Rev A
November 7, 2018