manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Riken Keiki
  6. •
  7. Gas Detector
  8. •
  9. Riken Keiki RI-415 User manual

Riken Keiki RI-415 User manual

PT3E-02!
RIKEN MODEL RI.415 CONDENSED USER GUIDE
CONDENSED INSTRUCTIONS
rO(-) PUMP/(+) POWERiENTER
Photo sensor window
\ Gasinre\
Calibration gas name plate
-'.' ..-,.",: :':1.in:.r'")"T:1tr5iff,.*, fffi'€ffi
1 . Press andhold "POWER' bufton for 5 seconds.
2. LCD displays as follows.
"Measuring gas fullscale" ) "Battery voltage" ) "PREPERATION" ) "PROBE/FILTER OK?"
press "ENTER" button. t "SENSOR CHECK ) ^WARMING UP 30seC'Countdown 30sec.
t "STAND BY OK' ) "Gas reading'
3. Zero adiustment
Press and hold "ZERO" button until LCD displays 'APPLY FRESH AlR". Press "ENTER" button.
LCD displays "ADJUSTING ZERO" t "ZERO SET OK'l "Gas reading".
Mode\ RI-415 is norw ready for norma\ operation.
4. Measurement
1)Lower the gas sampling hose to the sampling point (Not required for standard spiral sampling hose lm)
2) Check the flow monitor to veri$r proper flow.
3)Allow at least 3 minutes for a reading out the display.
5. Peak readinq
Press "PEAK" button to get maximum reading since the unit was turned on. Press again to retum to
normal operation.
6. "PUMP" button
Press and hold "PUMP" button until LCD displays "PUMP POWER OFF".
To return to normal operation, press and hold again until LCD displays "PUMP POWER ON'.
7 . Before tuming off the RI-415, let fresh air flow to purge the sampling system until the reading returns
to zero with sampling hose connected.
8. To turn offthe RI-415, press and hold'POWER" button until buzzer stops.
fiWAnrulNG : Do not replace battery in hazardous area.
Do not breathe exhaust gas from outlet.
Operate with carrying case in hazardous area.
Note : Regarding detail, refer to instruction manual. @ RIKEN KEII(
DAILY CHECK
1.
2.
3.
4.
5.
Check for damage on buttons, LED'S, or display.
Check pump's flow rate using flow monitor.
Check battery voltage.
Gheck that filter is clean.
Set zero reading.
R'f(Efr' 68ONEL R1415 COAIOEfr'SED UsER GUIDE
GNS ChL\BRN\\ON \ hHN\Nt : Do no\ ta\\bta\e \n \azat{ous atea.
1. Fill the gas sampling bag with the appropriate calibration gas. Then remove the filter tube and gas sampling
hose from the N-415.
HCversion:a)i-C4H10,50oloLEL(AirbalanceorN2balance) CH4version: a)CH4,7To/oLEL(NtbalanceorN2balance)
b)i-C4H10, l0vol%(N2 balance) b)CH4, 70vol%(N2 balance)
2. Turn on the RI-415 and set the zero reading.
3. Press both "PUMP- and"ZERO" buttons simultaneously until LCD displays "APPLY SPAN GAS".
4. Connect the sampling bagfilled with calibration gas a) to gas inlet and press "ENTER" button.
5. Once the gas reading becomes stable press "*- and "-" buttons to adjust the gn teading to match the
concentration of calibration gas, and then press *ENTER" button.
6. Calibration for YoLEL combustible range is complete.
'7. To calibrate %vol range press the "PUMP' and *ZERO" buttons simultaneously LCD displays "APPLY
SPA\{ GAS''.
8. Connect the sampling bag filled with calibration gas b) to *re gas inlet and press "ENTER" button.
g. Once the reading becomes stable press "+" and "-" buttons to adjust the gas reading to match the concentration
ofcalibration gas.
10. Press "ENTER" button to complete the calibration for the o/ovol combustible range.
Gas sampling bag re,..o1
Gas calibration cylinder
Note : Regarding detail, refer to instruction manual. @ RIKEN KEIKI
lr-trjl
I
TROUBLE SHOOTING
Trouble Ceute Trcalment
The LCD is blank Batedes are consumed.
Wrong polarity of batery.
Porrer buttttn not pressed long enougl
Replace batteries.
Put batteries in with correct polarity.
Press power button for approximately 5 seconds.
Pump is not running. No measurement and pump stops
running. (Disday cornment) 1. Tum pump back on. (Seemanual2€)
No pump suction, The tubing, filter, or fitting may be
clogged. Check the tubing, filter, or fitting (See manual $3 and 2-5)
Zero '0.0'%LEL of gas
display is dinking Zero point is deviated 1. Adjustzeroin agasfreearea(Seemanual2.4)
"100.Ovol%" blinking. Zero point or gas sensitivity is not
adjusted. Md<e zero adjusrnent (See manual 2-4)
V1/hen not adjus'ted to reading value, make gas calibration. (See manual $2)
lf not fixed by zero adjusrner( then do gas calibration. (see manual $2)

Other Riken Keiki Gas Detector manuals

Riken Keiki GD-D58-AC User manual

Riken Keiki

Riken Keiki GD-D58-AC User manual

Riken Keiki GD-D58-AC-GH User manual

Riken Keiki

Riken Keiki GD-D58-AC-GH User manual

Riken Keiki NP-1000 User manual

Riken Keiki

Riken Keiki NP-1000 User manual

Riken Keiki SD-1DRI User manual

Riken Keiki

Riken Keiki SD-1DRI User manual

Riken Keiki GD-F3A-SC-A User manual

Riken Keiki

Riken Keiki GD-F3A-SC-A User manual

Riken Keiki GD-A80D-AS Series User manual

Riken Keiki

Riken Keiki GD-A80D-AS Series User manual

Riken Keiki PT0E-1380 User manual

Riken Keiki

Riken Keiki PT0E-1380 User manual

Riken Keiki TP-70DGII User manual

Riken Keiki

Riken Keiki TP-70DGII User manual

Riken Keiki SD-1DRI-AS User manual

Riken Keiki

Riken Keiki SD-1DRI-AS User manual

Riken Keiki FP-30 User manual

Riken Keiki

Riken Keiki FP-30 User manual

Riken Keiki GD-F4A-SC-A User manual

Riken Keiki

Riken Keiki GD-F4A-SC-A User manual

Riken Keiki GD-F3A-A User manual

Riken Keiki

Riken Keiki GD-F3A-A User manual

Riken Keiki SD-1DOX User manual

Riken Keiki

Riken Keiki SD-1DOX User manual

Riken Keiki GP-147 User manual

Riken Keiki

Riken Keiki GP-147 User manual

Riken Keiki GD-1DOX-AS User manual

Riken Keiki

Riken Keiki GD-1DOX-AS User manual

Riken Keiki GD-K88Di User manual

Riken Keiki

Riken Keiki GD-K88Di User manual

Riken Keiki SC-8000 User manual

Riken Keiki

Riken Keiki SC-8000 User manual

Riken Keiki GD-K88Ai User manual

Riken Keiki

Riken Keiki GD-K88Ai User manual

Riken Keiki GD-F3A-1 User manual

Riken Keiki

Riken Keiki GD-F3A-1 User manual

Riken Keiki SP-220 Series User manual

Riken Keiki

Riken Keiki SP-220 Series User manual

Riken Keiki NC-1000 User manual

Riken Keiki

Riken Keiki NC-1000 User manual

Riken Keiki SD-1RI User manual

Riken Keiki

Riken Keiki SD-1RI User manual

Riken Keiki SD-3 Series User manual

Riken Keiki

Riken Keiki SD-3 Series User manual

Riken Keiki 04 Series User manual

Riken Keiki

Riken Keiki 04 Series User manual

Popular Gas Detector manuals by other brands

isweek AS8900 manual

isweek

isweek AS8900 manual

Cosmos XP-703D III instruction manual

Cosmos

Cosmos XP-703D III instruction manual

Simtronics GD10-P00 operating manual

Simtronics

Simtronics GD10-P00 operating manual

IMR EX660 user manual

IMR

IMR EX660 user manual

ION TIGER LT user manual

ION

ION TIGER LT user manual

ESP Safety SGOES operating manual

ESP Safety

ESP Safety SGOES operating manual

CGS TFT Installation operation & maintenance

CGS

CGS TFT Installation operation & maintenance

GASDNA DA-770 quick start guide

GASDNA

GASDNA DA-770 quick start guide

JB QUEST LD-3000 operating manual

JB

JB QUEST LD-3000 operating manual

ABB N9611-21 Installation and use manual

ABB

ABB N9611-21 Installation and use manual

PEMTECH Ultra 1000 Series Operator's manual

PEMTECH

PEMTECH Ultra 1000 Series Operator's manual

Carel GLD Series user manual

Carel

Carel GLD Series user manual

Honeywell 3000 MkIII Series Technical manual

Honeywell

Honeywell 3000 MkIII Series Technical manual

Seitron RGY 000 MBP4 Operation, installation, and maintenance manual

Seitron

Seitron RGY 000 MBP4 Operation, installation, and maintenance manual

Duomo BX150 manual

Duomo

Duomo BX150 manual

Dräger X-zone 5000 Instructions for use

Dräger

Dräger X-zone 5000 Instructions for use

Spectrex SafEye Quasar 900 manual

Spectrex

Spectrex SafEye Quasar 900 manual

Industrial Scientific Radius BZ1 product manual

Industrial Scientific

Industrial Scientific Radius BZ1 product manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.