Rosemount 396 User manual

Models 396/396VP/397/398/398VP
pH/ORP Sensors
Instruction Manual
PN 5100396
February 2001
MODEL 396 and
MODEL 398
MODEL 396VP and
MODEL 398VP
MODEL 397

CAUTION
SENSOR/PROCESS
APPLICATION COMPATIBILITY
The wetted sensor materials may not be
compatible with process composition and
operating conditions. Application compat-
ibility is entirely the responsibility of the
user.
DANGER
HAZARDOUS AREA INSTALLATION
This sensor is not Intrinsically Safe. or Explosion
Proof. Installations near flammable liquids or in
hazardous area locations must be carefully
evaluated by qualified on site safety personnel.
To secure and maintain an intrinsically safe
installation, an appropriate transmitter/ safety
barrier/sensor combination must be used.
The installation system must be in accor-
dance with the governing approval agency
(FM, CSA or BASEEFA/CENELEC) hazardous
area classification requirements. Consult your
analyzer/transmitter instruction manual for
details.
Proper installation, operation and servicing of
this sensor in a Hazardous Area Installation is
entirely the responsibility of the user.
SS-SE
March 1995
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures, and tests
its products to meet many national and international stan-
dards. Because these instruments are sophisticated
technical products, you must properly install, use, and
maintain them to ensure they continue to operate within
their normal specifications. The following instructions
must be adhered to and integrated into your safety pro-
gram when installing, using, and maintaining Rosemount
Analytical products. Failure to follow the proper instruc-
tions may cause any one of the following situations to
occur: Loss of life; personal injury; property damage;
damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and
servicing the product. If this Instruction Manual is not
the correct manual, telephone 1-800-654-7768 and the
requested manual will be provided. Save this
Instruction Manual for future reference.
• If you do not understand any of the instructions, con-
tact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked
on and supplied with the product.
• Inform and educate your personnel in the proper
installation, operation, and maintenance of the prod-
uct.
• Install your equipment as specified in the Installation
Instructions of the appropriate Instruction Manual and
per applicable local and national codes. Connect all
products to the proper electrical and pressure
sources.
• To ensure proper performance, use qualified person-
nel to install, operate, update, program, and maintain
the product.
• When replacement parts are required, ensure that
qualified people use replacement parts specified by
Rosemount. Unauthorized parts and procedures can
affect the product’s performance and place the safe
operation of your process at risk. Look alike substitu-
tions may result in fire, electrical hazards, or improper
operation.
Rosemount Analytical Inc.
Uniloc Division
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 863-1181
http://www.RAuniloc.com

i
MODEL 396/396VP/397/398/398VP TABLE OF CONTENTS
TUpH MODELS 396/396VP/397/398/398VP
pH/ORP SENSORS
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS............................................................ 1
1.1 TUpH™ Features and Applications................................................................. 1
1.2 Performance and Physical Specifications....................................................... 2
1.3 Ordering Information........................................................................................ 3
2.0 INSTALLATION................................................................................................. 8
2.1 Unpacking and Inspection............................................................................... 8
2.2 Mounting.......................................................................................................... 8
2.3 Electrical Installation ....................................................................................... 14
3.0 START UP AND CALIBRATION........................................................................ 28
3.1 Sensor Preparation .......................................................................................... 28
3.2 Calibration Using Buffer Solutions or Grab Samples....................................... 28
4.0 MAINTENANCE................................................................................................ 29
4.1 Electrode Cleaning .......................................................................................... 29
4.2 Automatic Temperature Compensator............................................................. 29
5.0 TROUBLESHOOTING...................................................................................... 30
6.0 RETURN OF MATERIAL................................................................................... 31
6.1 General ............................................................................................................ 31
6.2 Warranty Repair ............................................................................................... 31
6.3 Non Warranty Repair........................................................................................ 31
LIST OF TABLES
Number Title Page
4-1 R0and R1Values for Temperature Compensation Elements.......................... 29
4-2 Temperature vs Resistance of Auto T.C. Elements.......................................... 29
5-1 Troubleshooting................................................................................................ 30

MODEL 396/396VP/397/398/398VP TABLE OF CONTENTS
ii
LIST OF FIGURES
Number Title Page
2-1 Recommended Flow Through/Insertion Installation for Models 396, 396VP, .....
398, & 398VP.................................................................................................... 9
2-2 Dimensional Drawing For Models 396 & 398................................................... 10
2-3 Dimensional Drawing For Models 396VP & 398VP........................................... 11
2-4 Submersion Installations For Models 396, 396VP, 398, & 398VP ..................... 12
2-5 Dimensional Drawings For Model 397 in the Quik-Loc Unit............................ 13
2-6 Recommended Flow Through and Insertion Installation For Model 397......... 14
2-7 Models 396/397/398 Standard Wiring option used with Models 1054A, ........
1054B, 2054, and Preamps PN 22698-02 & 22698-03.................................... 14
2-8 Models 396/397/398 Option -62 for wiring to Models 1054, 1181, 2081, 54, .
81, 3081, 4081, SoluComp, and 2700 Preamp (PN 23054-03) and ...............
remote J-box PN 23555-00.............................................................................. 14
2-9 Wiring Models 396-54 (50), 397-54 (50), & 398-54 (50) to J-Box ...................
(PN 23309-04 (03)) & Remote Preamplifier ..................................................... 15
2-10 Wiring Models 396-54, 397-54, & 398-54 (PT-100-RTD) to J-Box ..................
(PN 23555-00) & Remote Preamplifier............................................................. 16
2-11 Wiring Models 396/397/398 to Model 1181pH ................................................ 17
2-12 Wiring Models 396-54, 397-54, & 398-54 to Models 1054A pH, 1054B pH, ..
& 2054pH-54.................................................................................................... 18
2-13 Wiring Models 396-54, 397-54, & 398-54 to Models 2081pH-05 & 1054pH-54.19
2-14 Wiring Models 396-54, 397-54, & 398-54 to Models 54, 81, 3081, & 4081..... 20
2-15 Wiring Models 396-54, 397-54, & 398-54 to SoluComp Model SCL-P-014..... 21
2-16 Wiring Models 396-54, 397-54, & 398-54 to Model 2700 Preamp................... 22
2-17 Wiring Models 396-( )-54-62-(71), 397-02-10-54-62, & 398-( )-54-62 to ........
Model 1055-01-10-22-32 ................................................................................. 22
2-18 Wiring Models 396-54-62, 397-02-54-62, & 398-54-62 to Model 1055-22-32.22
2-19 Wire Functions and Pin Connections for Variopol Connector Cable ............... 23
2-20 Wiring Models 396VP/398VP to Model 81 thru a remote J-box....................... 23
2-21 Wiring Models 396VP/398VP to Model 81 ....................................................... 23
2-22 Wiring Models 396VP/398VP to Model 1181 ................................................... 23
2-23 Wiring Models 396VP/398VP to Model 1181, 1050/1060, 1003/1023 thru a ..
remote J-box.................................................................................................... 24
2-24 Wiring Models 396VP/398VP to Model 2081 ................................................... 24
2-25 Wiring Models 396VP/398VP to Model 3081 & 4081....................................... 24
2-26 Wiring Models 396VP/398VP to Model 2081 thru a remote J-box................... 24
2-27 Wiring Models 396VP/398VP to Model 3081 & 4081 thru a remote J-box....... 24
2-28 Wiring Models 396VP/398VP to Model 1054 ................................................... 25
2-29 Wiring Models 396VP/398VP to Model 1054A/B & 2054 ................................ 25
2-30 Wiring Models 396VP/398VP to Model 1054 thru a remote J-box................... 25
2-31 Wiring Models 396VP/398VP to Model 1054A/B & 2054 thru a remote J-box.25
2-32 Wiring Models 396VP/398VP to Model 54/54e thru a remote J-box................ 26
2-33 Wiring Models 396VP/398VP to Model 54/54e ................................................ 26
2-34 Wiring Models 396VP/398VP to Model SCL-(P/Q)........................................... 26
2-35 Wiring Models 396VP/398VP to Model 2700 ................................................... 26
2-36 Wiring Models 396VP/398VP to Model 1055-22-32......................................... 27

MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
1
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.1 TUpH™ FEATURES AND APPLICATIONS.
Rosemount Analytical has achieved a new industry
standard for the life expectancy of pH sensors with
the TUpH Sensors. The low maintenance, disposable
Models 396, 396VP, 397, 398, and 398VP TUpH
Sensors offer long life and high performance in the
measurement of pH in aqueous solutions in
pipelines, open tanks, or ponds. These TUpH sen-
sors feature a patented, large area reference junc-
tion1for maximum resistance to process coatings,
generally found in dirty, high solid applications. In
addition, the secondary helical reference pathway2
serves as added protection from poisoning ions. The
simplified construction, designed with user conven-
ience in mind, does not require electrolyte replenish-
ment or any component replacement. All models fea-
ture a large glass bulb for increased resistance to the
effects of aging for longer life. All models are avail-
able only without an integral preamplifier. The pream-
plifier must be in a remote location or integral to the
analyzer/transmitter. Models with a “VP” after the
model number indicate the variopol connector is
used on the sensor in place of the cable. In this case,
a separate cable with a mating VP connector must be
used. All models are compatible with all Rosemount
Analytical and most manufacturers instruments.
1.1.1 Model 396 and 396VP TUpH
Features and Applications.
The Models 396 and 396VP
TUpH sensors are construct-
ed of polypropylene and
stainless steel and is com-
pletely sealed by EP (ethyl-
ene propylene) to eliminate
process intrusion. It is specifi-
cally designed for improved life in harsh, dirty appli-
cations such as lime slurry waste treatment and paper
machine headbox and pigment/dye applications
where large quantities of suspended solids are pres-
ent. Installation is easily achieved through a wide vari-
ety of mounting configurations. The Model 396 fea-
tures an optional front or rear 1 in. MNPT process con-
nection for insertion, submersion, or flow through
applications.
1.1.2 Model 397 TUpH and
Quik-Loc Features and
Applications. The Model 397 is
housed in a highly chemical
resistant polypropylene body
and completely sealed with EP
to eliminate process intrusion.
The Model 397 body is specifically designed for use
with the Quik-Loc Kit which consists of an adapter and
coupler. The PEEK (polyetheretherketone) adapter
enables the Model 397 sensor to fit into a 1 in. MNPT
Twin-Kam®4 Kamloc®4 coupler for quick and easy
removal without sensor cable twisting. The 316 stain-
less steel Twin-Kam®Kamloc®coupler is sealed with
EP and features “vibration resistant”5arms. The Quik-
Loc Kit is not recommended for use in processes with
hazardous, corrosive, or strong oxidizing chemicals
due to a risk of spray and bodily hazards.
1.1.3 Model 398 and 398VP TUpH
Features and Applications.
The chemical-resistant con-
struction of Tefzel, titanium,
and the patented TUpH refer-
ence junction make the
Models 398 and 398VP the
ideal sensors for measuring
pH in harsh process liquids.
Use Model 398 or 398VP to measure pH in sour water
strippers, in pulp bleaching towers that use chlorine
dioxide, and in process streams containing a variety
of organic solvents.
Models 398 and 398VP use the highest quality mate-
rials to provide superior chemical resistance. The sen-
sors are housed in a titanium tube and features an
optional 1 inch MNPT process connector for insertion,
submersion, or flow-through applications. The molded
Tefzel TUpH construction is offered with a choice of
seals (Viton, EPDM, or Kalrez). Combining high qual-
ity materials with the TUpH reference technology and
ACCUGLASS pH bulb allows for ultimate chemical
resistance and makes Models 398 and 398VP the per-
fect choice for measuring pH in harsh, demanding
processes.
1 May be Protected by U.S. Patent No. 5, 152, 882, Foreign Patent Pending
2 Protected by U.S. Patent No. 6, 054, 031, Foreign Patent Pending
3 ACCUGLASS is a registered trademark of Rosemount Analytical
4Kamlok® and Twin-Kam® are registered trademarks of CIVACON
5CIVACON catalog, Kam-92 (REV. May, 1992)

MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
2
1.2 PERFORMANCE & PHYSICAL SPECIFICATIONS
SPECIFICATIONS
MODELS 396 &396VP MODELS 398 & 398VP MODEL 397 QUIK-LOC KIT
Measurements pH: 0-14 pH: 0-14 pH: 0-14 —
and Ranges ORP: -1500 to 1500 mv
Available pH GPHT hemi bulb or GPHT hemi bulb or GPHT hemi bulb —
ACCUGLASSTypes GPLR flat bulb GPLR flat bulb
Wetted Materials 316 SST, Polypropylene, Titanium, Tefzel, glass, Polypropylene, EP, glass 316 SST, EP, PEEK
EPDM, glass choice of Kalrez, Viton, or
EPDM (platinum: ORP only)
Process Connection None, use 1 in. NPT None, use 1 in. NPT None, must use Quik-Loc 1-in. MNPT
process connector, process connector, kit which includes 1 in.
PN 23166-00 or 23166-01 PN 23166-00 or 23166-01 MNPT process connection
(sold separately) (sold separately)
Temperature Range 0-100°C (32-212°F) 0-100°C (32-212°F) 0-80°C (32-176°F) —
Pressure Range- 100-1136 kPa (0-150 psig) 100-1825 kPa (0-250 psig) 100-790 kPa (0-100 psig) —
Hemi bulb
Pressure Range- 100-790 kPa (0-100 psig) 100-790 kPa (0-100 psig) — —
Flat bulb
Minimum Conductivity 75 µS/cm, nominal 75 µS/cm, nominal 75 µS/cm, nominal —
Preamplifier Options Remote Remote Remote —
Weight/Shipping Weight 0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg (1 lb/2 lb)
PERCENT LINEARITY
396, 396VP, 397, 398, 398VP 396, 396VP, 397, 398, 398VP
pH Range GPHT Hemi GPLR Hemi
0-2 pH 94% 93%
2-12 pH 99% 98%
12-13 pH 97% 95%
13-14 pH 92% —

MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
3
1.3 ORDERING INFORMATION.
MODEL
396 TUpH INSERTION/SUBMERSION STAINLESS STEEL pH SENSOR
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For use with Models 1181, 1050, 1003 (3 K TC)
54 For use with Models 54, 1054, 1054A, 1054B, 2054, 2081, 2700, 81, SCL-P, 3081, and 4081 (PT-100 RTD)
CODE OPTIONAL ITEMS
62 Cable without BNC for wiring to 4081/3081/81/54 analyzers
71 Flat bulb, GPLR
396 - 54 EXAMPLE
The Model 396 pH Sensor features your selection of glass electrode type, either the
standard hemi bulb or the optional flat glass electrode combined with the coating
resistant, polypropylene reference with gel filled electrolyte. The sensor is housed
in a stainless steel body and is used with a 1 in. MNPT threaded process connector
suitable for insertion, submersion, or flow through installations. The Model 396 is
available without an integral preamplifier only and 15 ft of cable. Automatic
temperature compensation is standard 3 K Balco or PT-100 RTD.
MODEL
396VP TUpH INSERTION/SUBMERSION STAINLESS STEEL AND POLYPROPYLENE pH SENSOR (GPHT hemi bulb)
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For use with Models 1181, 1050, 1003 (3 K TC)
54 For use with Models 54, 1054, 1054A, 1054B, 2054, 2081, 2700, 81, 3081, 4081 and SCL-P (Pt 100 RTD)
CODE OPTIONAL OPTIONS
71 GPLR Flat bulb
396VP - 54 EXAMPLE
The Model 396VP insertion/submersion sensor is the same as Model 396 except,
instead of a cable on the back end of the sensor, it features a Variopol connector.
The watertight Variopol sensor-to-cable connector uses a mating connector cable
(ordered separately).

MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
The Model 398 Sensor, housed in a titanium tube and made with a patented Tefzel ref-
erence junction, can be used with a 1 inch MNPT process connector (purchased sepa-
rately). The sensor is available with a hemi bulb glass pH electrode or a platinum and
glass ORP electrode and with Pt100 or 3K temperature compensation.The 398 is pro-
vided with a standard 15 ft cable. A preamplifier must be used if the sensor is more than
15 ft from the analyzer/transmitter. Junction box kits with preamplifiers must be ordered
separately. Process connector must also be ordered separately.
CODE MEASURING ELECTRODETYPE (Required Selection)
10 GPHT hemi glass, General Purpose HighTemperature (0-14 pH)
12 ORP
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC) [no T.C. if ordered with option -12 (ORP)]
54 For Models 1054, 1054A/B, 2054, 2081,54, 4081, 3081, 81, SCL-(P/Q), Solu Cube (PT 100 RTD)
CODE O-RING MATERIAL (Required Selection)
30 EPDM
31 Viton
32 Kalrez (recommended for applications with Chlorine Dioxide)
CODE OPTIONAL SELECTIONS (see drawings on page 4)
62 15 ft Cable, no BNC (Not Valid w/Option 50) for wiring directly to 54, 81, 3081, and 4081Transmitter/Analyzers
MODEL
398 TUpH INSERTION/SUBMERSIONTITANIUM pH SENSOR
398 - 10 - 32 - 54 EXAMPLE
The Model 398VP insertion/submersion sensor is the same as Model 398, except
the cable is replaced with a Variopol connector. The watertight Variopol sensor-to-
cable connector uses the mating connector cable (ordered separately).
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC) [no T.C. if ordered with option -12 (ORP)]
54 For Models 1054, 1054A/B, 2054, 2081,54, 4081, 3081, 81, SCL-(P/Q), Solu Cube (PT 100 RTD)
CODE MEASURING ELECTRODETYPE (Required Selection)
10 GPHT hemi glass, General Purpose HighTemperature (0-14 pH)
12 ORP
13 Flat
CODE O-RING MATERIAL (Required Selection)
30 EPDM
31 Viton
32 Kalrez (recommended for applications with Chlorine Dioxide)
MODEL
398VP TUpH INSERTION/SUBMERSIONTITANIUM ANDTEFZEL pH/ORP SENSOR
398VP - 10 - 32 - 54 EXAMPLE
4

5
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
TheTUpH Model 397 pH Sensor is housed in a polypropylene body and is designed to be
used with the Quik-Loc Kit.The sensor includes a large general purpose (GPHT) glass pH
electrode and a large area polypropylene reference junction with gel filled reference elec-
trolyte.The Model 397 is available without an integral preamplifier only and 15 ft (4.6 m) of
cable.
MODEL
397 TUpH pH SENSOR
CODE PREAMPLIFIER/CABLE (Required Selection)
02 Without integral preamplifier, 15 ft (4.6 m) cable
CODE MEASURING ELECTRODETYPE (Required Selection)
10 GPHT, General Purpose glass (0-14 pH)
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181, 1050, 1003 (3 K Balco TC)
54 For Models 1054, 54, 81, 3081, 4081, 2054, 2081; 2700-Code 02 only (PT-100 RTD)
CODE OPTIONAL OPTIONS
62 Cable without BNC for wiring to 3081/4081/81/54 analyzers
1. Quik-Loc Mounting (required for all first time installations)
Choose one: PN 23757-00, Quik-Loc Kit: for use in 1 in. tees; insertion depth 1.4 in. (35 mm)
PN 23757-01, Quik-Loc Kit: for use in 1-1/2 in. and 2 in. tees; insertion depth 2 in. (50 mm)
2. Remote Junction Boxes (optional, recommended for sensor to analyzer distances of more than 15 ft)
Choose one: PN 23555-00 includes preamplifier for Models 54, 81, 3081, 4081
PN 23309-03 and PN 22698-02 plug-in preamplifier for Model 1181Analyzer
PN 23309-04 and PN 22698-03 plug-in preamplifier for Models 1054 series, 2054, and 2081 Analyzers
PN 23054-03 includes preamplifier for Solu Cube Model 2700
3. Extension cables (used with remote junction boxes)
Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped
4. Wiring Aids
Choose one: PN 9120531 BNC splitter, used in place of option -62 or BNC coax prepping for connections to
junction box (PN 23555-00); Models 1181, 1054, 2081, 54, 81, 3081, 4081; Solu Comp (SCL-P-014);
and Model 2700 Preamplifier (PN 23054-03)
NONE: No wiring aids required for connections to Models 1054A, 1054B, 2054, or preamplifiers
PN 22698-02 or 22698-03
FOR FIRST TIME 397/QUIK-LOC INSTALLATIONS, USINGTHE FOLLOWING GUIDE IS RECOMMENDED:

6
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1. Process Connector Accessories (required for all first time installations with 1-inch process connection threads)
Choose one: PN 23166-00, 316 SST, 1 in. x 1 in. NPT process connector, with EPDM o-ring
PN 23166-01, Titanium, 1 in. x 1 in. NPT process connector, with EPDM o-ring
PN 9510066, Nylon, 1 in. x 1 in. NPT process connector (submersion only)
Choose one (optional process connector o-rings)
PN 9550220, Kalrez o-ring, 2-214
PN 9550099, Viton o-ring, 2-214
2. Variopol Cable (required for all first time installations) of Models 396VP and 398VP
Choose one: PN 23645-06, 15 ft cable with mating VP connector, prepped with BNC on analyzer end
PN 23645-07, 15 ft cable with mating VP connector, prepped without BNC on analyzer end*
3. Mounting Accessories (optional)
Choose one: PN 915240-03 PVC flow through tee, 3/4 in. NPT process connection
PN 915240-04 PVC flow through tee, 1 in. NPT process connection
PN 915240-05 PVC flow through tee, 1 1/2 in. NPT process connection
PN 11275-01 Sensor handrail mounting assembly
PN 2002011 1-1/2 in. CPVC Tee with 1 in. FNPT connection
PN 23728-00 Low Flow Cell, acrylic
4. Remote Junction Boxes (optional, recommended for sensor to analyzer distances of more than 15 ft)
Choose one: PN 23555-00 includes preamplifier for Models 54, 81, 3081, 4081
PN 23309-03 and PN 22698-02 plug-in preamplifier for Model 1181Analyzer
PN 23309-04 and PN 22698-03 plug-in preamplifier for Models 1054 series, 2054, 2081 Analyzers
PN 23054-03 includes preamplifier for Solu Cube Model 2700
5. Extension cables (used with remote junction boxes)
Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped
6. Wiring Aids
Choose one: PN 9120531 BNC splitter, used in place of option -62 or BNC coax prepping for connections to junction box
(PN 23555-00); Models 1181, 1054, 2081, 54, 81, 3081, 4081; SoluComp (SCL-P-014);
and Model 2700 Preamplifier (PN 23054-03)
NONE: No wiring aids required for connections to preamplifier PN 22698-02, preamplifier PN 22698-03,
or Models 1054A, 1054B, and 2054
FOR FIRST TIME 396, 396VP, 398, AND 398VP INSTALLATIONS,
ROSEMOUNT ANALYTICAL RECOMMENDS USING THE FOLLOWING GUIDE:
*Used for connections to Models 1181, 1054, 2081, 54, 81, 3081, 4081, and remote junction box PN 23555-00.

7
MODEL 396/396VP/397/398/398VP SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
ADDITIONAL ACCESSORIES FOR MODEL 396, 396VP, 398 AND 398VPTUpH SENSORS
PART NUMBER DESCRIPTION
33046-00 Ferrule 1 in., split 316 SS
33211-00 Adapter retrofit for PN 915240-04
9550167 EPDM O-ring for process connector (PN 23166-00)
9310100 Ferrule, 1 in. Teflon1
9310096 Nut, Swage, 1 in. stainless steel
9550167 O-ring, 2-214 EPDM
1 Teflon is a registered trademark of E.I. du Pont de Nemours & Co.
ADDITIONAL ACCESSORIES FOR MODEL 397 TUpH SENSOR AND THE QUIK-LOC KIT
PART NUMBER DESCRIPTION
23753-00 PEEK replacement adapter for Quik-Loc kit (PN 23757-01)
23753-01 PEEK replacement adapter for Quik-Loc kit (PN 23757-00)
9160441 1 in. MNPT 316 stainless steel coupler
9160442 1 in. Plug, stainless steel
9160447 1 in. EP Gasket for coupler
2002011 1-1/2 in. CPVC Tee with 1 in. FNPT Connection
ADDITIONAL ACCESSORIES USED FOR MODELS 396, 396VP, 397, 398, AND 398VP TUpH SENSORS
PART NUMBER DESCRIPTION
22698-00 Preamplifier plug-in for J-box, 1003 compatible
22698-02 Preamplifier plug-in for J-box, 1181/1050 compatible
22698-03 Preamplifier plug-in for J-box, 1054A/B, 2054,2081 compatible
22743-01 Pt 100 preamplifier, 1181 compatible
22744-01 3K Preamplifier, 1181 compatible
23054-03 Remote J-box with preamplifier for Solu Cube Model 2700
23309-03 Remote Junction box, for preamplifier Model 1181
23309-04 Remote Junction box, for preamplifier Model 1054 Series, 2054, 2081
23555-00 Remote Junction box, with preamplifier, 54/3081/4081/81 compatible
23557-00 Preamplifier for Junction box for Models 54/3081/4081/81
23646-01 Extension cable 11 conductor, shielded, prepped
9200273 Extension cable11 conductor, shielded, unprepped
9210012 Buffer solution, 4.01 pH, 16 oz.
9210013 Buffer solution, 6.86 pH, 16 oz.
9210014 Buffer solution, 9.18 pH, 16 oz.
2001492 Stainless Steel Tag, Specify Marking
9200254 Cable, 4 cond., 22 AWG, 2 shielded pair

8
2.2.1 FlowThrough and Insertion Mounting for Models
396, 396VP, 398, and 398VP. The Models 396, 396VP,
398, and 398VP Sensors can be used with a 1 inch
MNPT process connector at the front of the sensor for
mounting into a 1-1/2 inch tee or the process. See
Figure 2-1 for installation configurations.
2.2.2 Submersion Mounting for Models 396, 396VP,
398, and 398VP. The Models 396, 396VP, 398, and
398VP Sensors also have a 1 inch MNPT process con-
nector available for use on the back of the sensor.
Tapered pipe threads in plastic fittings tend to loosen
after installation. It is therefore recommended that Teflon
tape be used on the threads and that the tightness of
the connection be checked frequently to assure that no
loosening has occurred. To prevent rain water or con-
densation from running into the sensor, a weatherproof
junction box is recommended (see Figure 2-4). The
sensor cable must be run through a protective conduit
for isolation from electrical interference or physical
abuse from the process. The sensor should be installed
within 80° of vertical, with the electrode facing down.
The sensor’s cable should not be run with power or con-
trol wiring.
2.2.3 Quik-Loc Mounting for Model 397. The Quik-
Loc mounting is used with the Model 397 TUpH sen-
sor only.
SAFETY WARNING
It is recommended that a thermome-
ter, drain valve to relieve pressure
and pres-sure gauge be inserted
near the Quik-Loc assembly (see
Figure 2-5).
WARNING
Once the Quik-Loc unit is installed the oper-
ator should wait for the process to cool to a
safe temperature, use the pressure drain
valve to relieve all process pressure and
observe the pressure on the pressure
gauge for proper removal of the sensor
without spray or bodily injury. The Quik-Loc
kit used with the 397 TUpH sensor is not
recommended for use with hazardous, cor-
rosive, or strong oxidizing chemicals due to
a risk of spray or bodily injury.
MODEL 396/396VP/397/398/398VP SECTION 2.0
INSTALLATION
2.1 UNPACKING AND INSPECTION. Inspect the out-
side of the carton for any damage. If damage is detect-
ed, contact the carrier immediately. Inspect the instru-
ment and hardware. Make sure all the items in the pack-
ing list are present and in good condition. Notify the fac-
tory if any part is missing. If the sensor appears to be in
satisfactory condition, proceed to Section 2.2,
Mounting.
NOTE
Save the original packing cartons and
materials as most carriers require proof of
damage due to mishandling, etc. Also, if it
is necessary to return the instrument to the
factory, you must pack the instrument in the
same manner as it was received. Refer to
Section 6 for return instructions. If the sen-
sor is to be stored, the vinyl boot should be
filled with pH buffer solution and replaced
on sensor tip until ready to use.
WARNING
Glass electrode must be wetted at all times
(in storage and in line) to maximize sensor
life.
2.2 MOUNTING. Each sensor has been designed to
be located in industrial process environments.
Temperature and pressure limitations must not
be exceeded at any time. A caution or warning label
regarding this matter is attached to each sensor. For
insertion, transfer the label as shown on label instruc-
tions. See Figure 2-2. For submersion applications, first
note limits then remove and discard label.
NOTE
Before mounting the sensor, shake down
the sensor to remove any air bubbles that
may be present at the tip of the pH glass
bulb. In most cases, the pH sensor can sim-
ply be installed as shipped and readings
with an accuracy of ±0.6 pH may be
obtained. To obtain greater accuracy or to
verify proper operation, the sensor must be
calibrated as a loop with its compatible ana-
lyzer or transmitter.
SECTION 2.0
INSTALLATION

9
MODEL 396/396VP/397/398/398VP SECTION 2.0
INSTALLATION
Wrap the pipe threads of the Twin-Kam Kamloc coupler
with Teflon tape before placing it into the process pipe.
The coupler can be connected to any 1 in. process con-
nection and must be mounted within 80° of vertical, with
the electrode facing down. Once the coupler is in place,
the adapter should be positioned onto the back end of the
sensor. (See Figure 2-5) Remove the parafilm wrapping
from the two 0-rings on the sensor, grease the o-rings with
the lube provided, and feed the sensor cable through the
adapter. Once the adapter is slipped over the sensor’s
back end, the retaining ring (which is included with every
397 sensor) should be installed on the black, grooved,
back end of the sensor body. The retaining ring secures
the sensor into the adapter. The adapter/sensor assembly
is now ready to be inserted into the coupler. With both
arms of the coupler in the released position, insert the
adapter/sensor assembly into the coupler.
NOTE
The adapter can not be inserted completely or
properly unless both arms are in the fully
released position (see Figure 2-5).
Once the adapter has been properly placed in the coupler,
both arms should be positioned in the locked position. A Sur-
Loc™ 1 spring arm has been provided on one arm so that the
“arm cannot be opened until the spring is released.”2
NOTE
The sensor may obstruct flow through smaller pipes.
Once the arms of the coupler have been locked in position,
use tamper-evident safety wire on the metal rings of the
Twin-Kam arms to prevent unauthorized and/or untrained
personnel from using the Quik-Loc unit.
SAFETY WARNING
It is the responsibility of each com-
pany using the Quik-Loc Kit/ 397
TUpH Sensor to train personnel of
the injury risks associated with
using a quick-release coupler that
is placed in a hot or pressurized
process.
The Quik-Loc unit should be used only within the pressure
and temperature limits stated for the Model 397 sensor in
Section 1.2.2.
FIGURE 2-1. Recommended FlowThrough and Insertion Installation
For Models 396, 396VP, 398 and 398VP
1-1/2 inch Pipe Tee (PN 2002011) with 1 inch threaded connections.
DWG. NO. REV.
40039609 B

10
MODEL 396/396VP/397/398/398VP SECTION 2.0
INSTALLATION
FIGURE 2-2. Dimensional Drawing For Models 396 and 398
DWG. NO. REV.
40039601 D
MILLIMETER
INCH
NOTE: PRESSURE RAT-
ING SHOWN IS FOR
MODEL 396 FLAT BULB.
SEE PAGE 2 FOR
MODEL 396 AND 398
PRESSURE RATINGS.

MODEL 396/396VP/397/398/398VP SECTION 2.0
INSTALLATION
INCH
MILLIMETER
Metal Process Connector PN 23166-xx
(xx = 00 for 316 SST and xx = 01 for tita-
nium) can be used for insertion or sub-
mersion mounting of Models 396VP or
398VP sensors in 1-inch fittings.
The metal process connector gives the sensor var-
ious insertions depths, depending on where the
user locates the compression fitting. Also the
threads can be switched to face the cable end of
the sensor for connection to submersion pipes.
FIGURE 2-3. Dimensional Drawings: Model 396VP and 398VP Shown with Process Connector
11

12
FIGURE 2-4. Submersion Installations For Models 396, 396VP, 398, and 398VP
DWG. NO. REV.
40039602 B
MILLIMETER
INCH
DWG. NO. REV.
40039603 B
DWG. NO. REV.
40039604 B
MODEL 396/396VP/397/398/398VP SECTION 2.0
INSTALLATION

13
MODEL 396/396VP/397/398/398VP SECTION 2.0
INSTALLATION
FIGURE 2-5. Dimensional Drawings For Model 397 in the Quik-Loc Unit
DWG. NO. REV.
40039701 A
MILLIMETER
INCH
DWG. NO. REV.
40039702 A
DWG. NO. REV.
40039704 A
NOTE: The EP gasket (see drawing #40039702)
provided with the coupler should be periodically
inspected. If gasket shows signs of corrosion,
replacement will be necessary to ensure a proper
and secure seal between the coupler and adapter.
QUIK-LOC A B
KIT SENSOR LENGTH INSERTION DEPTH
23757-00 6.15 IN / 156 MM 1.4 IN / 35 MM
23757-01 6.15 IN / 156 MM 2.0 IN / 50 MM

14
MODEL 396/396VP/397/398/398VP SECTION 2.0
INSTALLATION
2.3 ELECTRICAL INSTALLATION.
The Models 396, 396VP, 397, 398, and 398VP are for use
with a remote preamplifier. Each sensor comes with either a
special 15 ft low noise coax cable or a Variopol (VP) con-
nector, which is used with a mating Variopol cable. The
cable should be handled carefully, and kept dry and free of
corrosive chemicals at all times. Extreme care should be
used to prevent the cable from being twisted, damaged or
scraped by rough, sharp edges or surfaces. Please refer to
Figures 2-7 thru 2-18 for wiring Models 396, 397, and 398.
Please refer to Figures 2-19 thru 2-36 for wiring Models
396VP and 398VP.
DANGER
DO NOT CONNECT SENSOR CABLE TO POWER
LINES. SERIOUS INJURY MAY RESULT.
FIGURE 2-6. Recommended Flow Through and Insertion Installation For Model 397
DWG. NO. REV.
40039703 A
FIGURE 2-7. Models 396, 397, and 398 Standard Wiring
Option used with Models 1054A, 1054B, 2054, and
Preamplifiers PN 22698-02 and 22698-03.
FIGURE 2-8.Models 396,397, and 398 Option -62 for Wiring
to Models 1054, 1181, 2081, 54, 81, 3081, 4081, SoluComp,
and 2700 Preamplifier (PN 23054-03). Also for use with
remote Junction Box PN 23555-00.

MODEL 396/397 TABLE OF CONTENTS
15
MODEL 396/396VP/397/398/398VP SECTION 2.0
INSTALLATION
DWG. NO. REV.
40039706 C
DWG. NO. REV.
40039610 C
FIGURE 2-9. Wiring Details For Model 396-54 (50), 397-54 (50), and 398-54 (50).
For Use With Junction Box (PN 23309-04 (03)) and Remote Preamplifier

16
DWG. NO. REV.
40038920 E
MILLIMETER
INCH
MODEL 396/396VP/397/398/398VP SECTION 2.0
INSTALLATION
SENSOR
MODELS
54, 3081,
81, 4081
pH/ORP
FIGURE 2-10.Wiring Details for Model 396-54, 397-54, and 398-54 (Pt-100 RTD). For Use With
Junction Box (PN 23555-00) and Remote Preamplifier.
This manual suits for next models
4