18|IMPORTANT USE REGULATIONS
Positioning of the chair
Push the chair far enough away from the wall that when moving
the back section into a low position, the back section cannot be
stopped by the wall. (If the back section is stopped by the wall and
if the button for the back section is kept depressed, the spindles
of the motor will continue to run. Result – the chair will be pushed
away from the wall, the back section drops to the spindle stop. The
motor can be destroyed, the patient can be frightened, damage
can occur. If the back section is in a low position, move the chair far
enough away from the wall that, when lifting up the back section
into the seated position, the back section cannot get stuck on a
windowsill or other obstructions (back section will be bent).
Armrests, back section
Do not sit on the armrests or on the head end of the back section.
Only use the armrests for arm support. Any other type of use is
prohibited. (Danger of falling, danger of injury) The safe work load
is 20 kg.
Load
The chair may only be loaded with one person weighing a maxi-
mum of 200 kg (including cushion support). The chair must not be
occupied by 2 or more persons at the same time under any circum-
stances (danger of falling, danger of breaking and danger of injury).
Moving parts
Do not reach into moving parts (mechanical elements). Significant
injuries can occur. The cushion is a component of the treatment
chair.
General note
The daily function check includes checking all movement sequenc-
es as described under point "Operation" starting on page 20.
OPERATION 19
|
RELEASING THE FOOT OR HAND SWITCH
The foot or hand switch may only be disconnected from the treatment chair
by removing the plug connector for service purposes following contact with
the RUCK customer service. Before replacing the switch, the treatment chair
must be disconnected from the mains. To disconnect the switch eject the
locking ring by using a slot screwdriver. Then the plug can be pulled out
carefully. To connect the switch first insert the Plug. Then press in the lock-
ing ring.
HOW THE PATIENT GETS ON AND OFF
The treatment chair can be
mounted or exited from the
front or from the side. To
mount from the side the arm-
rest is swung upward.
It is recommended that the
patients get on and off from
the side. In this way, walking
over waste materials that
accumulate during the treat-
ment, can be prevented.