Ruck STELLA 3s User manual

8020702 I Instructions for use RUCK®Treatment Chair STELLA 3s | REV 06 |
Revision status 18.01.2023
GB
FÜR FUSS UND PFLEGE
Treatment Chair STELLA 3s
INSTRUCTIONS FOR USE

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
3
|2|Dear customer,
With the RUCK®treatment chair STELLA 3s you have cho-
sen a treatment concept from RUCK. In the development of the
RUCK®treatment chair STELLA 3s, the declared objective was
to create an ideal tool for professional users in the foot care area.
The RUCK®treatment chair STELLA 3s is based on the Stella
treatment chair, which has proven itself over many years in the foot care
market. A completely new product has been developed taking into
account the current European Regulation (EU) 2017/745 for medical
devices, a sensible, ergonomic way of working and the increased de-
mands on medical practice hygiene. What you have in front of you now
is the synthesis of innovative functions, high-quality and contemporary
design. With the RUCK®treatment chair STELLA 3s as a Class 1 medi-
cal device and more than 90 years of practical experience we underline
our high quality requirements for professional foot care and give you
safety in your day-to-day work.
These operating instructions are a component of the medical device
and must be kept in its immediate vicinity. Please read these operating
instructions carefully all the way through before commissioning the device.
Our treatment chairs are manufactured and assembled in Germany and
are subject to ongoing quality control. Nevertheless, if you are not sat-
isfied or if you have questions please contact our service department.
We sincerely thank you for the trust you have placed in our company
and extend our best wishes for satisfaction in your daily work with your
new treatment chair.
Kind regards, HELLMUT RUCK GmbH
Class I
medical device
1. www.hellmut-ruck.de
2. 10137 I 10138
3. DOWNLOADS
Copyright
© 2023 Hellmut Ruck GmbH. All rights reserved. This document must not be copied, edited, reproduced or translated
without the prior written permission of Hellmut Ruck GmbH.
HELLMUT RUCK GmbH | Daimlerstraße 23 | D - 75305 Neuenbürg, Germany
web www.hellmut-ruck.de | tel. +49 (0)7082. 944 20 | fax +49 (0)7082. 944 22 22
MADE IN
GERMANY
YEARS
WARRANTY
https://www.hellmut-ruck.de
10137
QUALITY
since more than
YEARS
Treatment Chair STELLA 3s
Downloads
Gebrauchsanweisung
Colour variants:
dichromatic
(01) white/turquoise (02) white/slate
(03) white/mocha (04) white/sand
(06) white/bamboo (07) white/mint
(08) white/baltic (09) white/orchid
(10) white/mango
monochromatic
(21) turquoise (22) slate
(23) mocha (24) sand
(25) white (26) bamboo
(27) mint (28) baltic
(29) orchid (30) mango
Equipment variants:
Item no. 10137xx | RUCK®STELLA 3s treatment chair
Item no. 10138xx | RUCK®STELLA 3s treatment chair with memory function

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
CONTENTS
PRODUCT DESCRIPTION 66
ABBREVIATIONS AND SYMBOLS 88
SAFETY NOTICES 99
GENERAL 1212
SCOPE OF DELIVERY 1313
COMMISSIONING | CARRYING INSTRUCTIONS 1414
IMPORTANT USE REGULATIONS 1818
OPERATION 1919
CONTENTS 5
|4|
MAINTENANCE, CLEANING AND DISINFECTION 3030
SPARE PARTS 3333
DISPOSAL 3333
TECHNICAL DATA | TREATMENT POSITIONS 3434
CHANGING CUSHIONS 3636
WARRANTY 3737
SAFETY NOTICES 3838
ENVIRONMENTAL CONDITIONS 39
ELECTROMAGNETIC COMPATIBILITY 40ELECTROMAGNETIC COMPATIBILITY 40
LOOSENING THE SWITCH 19LOOSENING THE SWITCH 19
HOW THE PATIENT GETS ON AND OFF 19HOW THE PATIENT GETS ON AND OFF 19
OPERATION OF FOOT SWITCH WITH MEMORY FUNCTION 20OPERATION OF FOOT SWITCH WITH MEMORY FUNCTION 20
OPERATION HAND SWITCHOPERATION HAND SWITCH 2222
OPERATION FOOT SWITCHOPERATION FOOT SWITCH 2424
TURNING THE SEAT SURFACETURNING THE SEAT SURFACE 2525
SWINGING DOWN THE CALF SUPPORT AND FOOT SUPPORT 26SWINGING DOWN THE CALF SUPPORT AND FOOT SUPPORT 26
PULLING OUT THE FOOT ROLL 27PULLING OUT THE FOOT ROLL 27
ADJUSTING THE CLAMP LEVER 27ADJUSTING THE CLAMP LEVER 27
PIVOTING THE FOOT ROLL 28PIVOTING THE FOOT ROLL 28
PIVOTING THE CALF SUPPORT 29PIVOTING THE CALF SUPPORT 29
VARIABLE ADJUSTMENT OF THE HEAD SECTION 29VARIABLE ADJUSTMENT OF THE HEAD SECTION 29
DECLARATION RELATING TO THESE OPERATING INSTRUCTIONS
Hellmut Ruck GmbH reserves all rights to modify and supplement these operating
instructions. Hellmut Ruck GmbH has taken great care to keep these instructions free of
inaccuracies and omissions. Please contact us if you find mistakes or inaccuracies. How-
ever, Hellmut Ruck GmbH assumes no liability for mistakes in these operating instructions
or resultant follow-on damage, resulting from the provision of these operating instructions
or their application when using the product.
These operating instructions must not be disseminated or supplemented without the
express approval of Hellmut Ruck GmbH.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
PRODUCT DESCRIPTION 7
|6|PRODUCT DESCRIPTION
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
1.10
1.11
1.12
1.13
1.14
1.15
1.16
1.1 HEAD SECTION CUSHION
including NECK ROLL
1.2 BACK CUSHION
1.3 ARM RESTS
1.4 SEAT CUSHION
1.5 CALF SUPPORT
1.6 HAND WHEEL FOR
ROTATION
(Present on both sides)
1.7 CLAMP LEVER – FOOT
ROLL
1.8 FOOT ROLL
1.9 HEAD SECTION COVER
1.10 HEAD SECTION ADJUST -
MENT
1.11 BACK SECTION COVER
1.12 ACTIVATION LEVER – LEG
SUPPORT
1.13 HEIGHT LIFT COVER
1.14 BASE FROM COVER
1.15 SOCKET FOR SWITCHES
1.16 FOOT PLATE
1.17 CONNECTOR FOR SWITCHES
1.18 FOOT SWITCH
1.19 HAND SWITCH
1.20 FOOT SWITCH WITH MEMORY
FUNCTION
1.17
1.18
1.19
1.20

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
SAFETY NOTICES 9
|8|ABBREVIATIONS AND SYMBOLS
THE MANUFACTURER CANNOT BE MADE RESPONSIBLE
FOR DAMAGE THAT IS CAUSED THROUGH USE THAT DE-
VIATES FROM THE INTENDED USE OR FAILURE TO COM-
PLY WITH THE SAFETY NOTICES AND WARNINGS.
GENERAL SAFETY NOTICES
The RUCK®treatment chair STELLA 3s has been built in ac-
cordance with the state of technology and acknowledged safety
rules and standards. Each chair is tested for function and safety
prior to delivery and the inspection certificate is affixed to the
chair. However, if operated incorrectly or misused, hazards can
occur for:
• the health of the user and/or of the patient
• the chair and possibly the owner's other assets
• the function of the chair.
All persons who have anything to do with set up, commission-
ing, operation, service and maintenance must:
• be appropriately qualified
• have these operating instructions available to them,
have understood these operating instructions and precisely
comply with them.
Unauthorized persons and non-instructed persons are not al-
lowed to operate the electric motors under any circumstances.
ABBREVIATIONS AND SYMBOLS
Protection class II device
(Protective insulation)
The disposal notice
must be complied with!
Application part of the
type B
Manufacturing date
CE conformity mark
Refer to instruction manual
This symbol identifies a potential point of danger for people or for the
device itself. It must be strictly complied with.
This symbol identifies useful advice. Additional information about the
product and how to use it is provided here.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
10|SAFETY NOTICES SAFETY NOTICES 11
|
Comply with the Medical Devices Ordinance (applies to Germa-
ny only) (MPBetrV).
The owner must instruct the users, make the operating instruc-
tions accessible and ensure (verifiably) that these operating in-
structions have been read and understood.
Danger due to electrical energy
Danger due to electrical energy can occur through: Perform-
ing the wrong maintenance tasks on the transformer, using
"NON-ORIGINAL PARTS" in the transformer area, failure to un-
plug the mains plug for maintenance tasks, running over, dam-
aging the mains cable. Maintenance tasks may only be executed
by authorized specialists.
If the chair is not in use, always pull the power plug. The chair
must not be electrically adjusted, under any circumstances, as
long as people are within the movement range of the chair.
Interference with other electrical devices can occur under cer-
tain circumstances.
Protective devices
Safeguards and cladding elements are installed for the protec-
tion of patient and user. They must not be removed, changed or
bypassed.
HAVE IDENTIFIED DEFECTS RECTIFIED IMMEDIATELY!
Contact RUCK SERVICE immediately:
Tel + 49 (0)7082 944 22 55
Maintenance and safety instructions
The prescribed maintenance tasks, as specified under point
"MAINTENANCE" on page 28 must always be executed.
IF THIS REQUIREMENT IS NOT SATISFIED, THE GUARAN-
TEE WILL LAPSE AND WE CANNOT ACCEPT LIABILITY!
Daily function checks
Prior to each treatment the checks listed below must be executed:
• All electrically adjuschair sections, such as back and height
adjustment must be checked for functionality via the foot
switch.
• General visual inspection
IF THIS REQUIREMENT IS NOT SATISFIED, THE GUARAN-
TEE WILL LAPSE AND WE CANNOT ACCEPT LIABILITY!
Initial check for faults
If operation is no longer ensured without restriction, or if you
should determine that the power cable or any other cable is
damaged, unplug the mains plug and arrange for the faulty ca-
ble to be replaced by a qualified electrician. Contact RUCK Ser-
vice immediately: Phone +49 (0)7082 944 22 55
Danger of explosion!
Operation of the RUCK®treatment chair STELLA 3s in areas
subject to explosion hazard can trigger an explosion.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
12|
Intended purpose
The RUCK®treatment chair STELLA 3s is used for supporting
customers in a sitting or reclining position for the duration of
treatment under supervision (maximum 6-7 hours). It can be
variably adjusted from a sitting position to a flat position.
It is designed for use in podiatry practices, foot care, nail design,
cosmetics, wellness, orthopaedics and/or diabetic outpatient
foot clinics.
The treatment chair must only be occupied by one person with
a maximum weight of 200 kg. Under no circumstances should
two or more persons use the treatment chair simultaneously.
ANY OTHER USE IS NON-INTENDED USE!
Conversions, modifications, attachment of supplemental equip-
ment that is not provided by the manufacturer are not permit-
ted. Exceptions can only be approved by the manufacturer in
writing.
Only use original spare parts for repairs.
The prescribed maintenance tasks must always be executed.
IF THIS REQUIREMENT IS NOT SATISFIED, THE GUARAN-
TEE WILL LAPSE AND WE CANNOT ACCEPT LIABILITY!
Intended use also includes compliance with all instructions,
safety notices and regulations specified in these operating in-
structions.
Implementation site:
The RUCK®treatment chair STELLA 3s may only be operated
in dry areas that are not subject to explosion hazard.
SCOPE OF DELIVERY 13
|
Check that the delivery is complete.
1. Treatment chair
2. Neck roll
3. Foot switch
4. The operating instructions
TRANSPORT | UNPACKING | STORAGE
Delivery and intermediate storage of the chair occurs in the transport box.
For further information about transport and storage conditions see page 37.
Prior to commissioning, please check whether the mains voltage agrees with
the rated voltage and frequency specified on the rating plate (on the base
frame). The RUCK®treatment chair STELLA 3s can be connected with the
provided power cable (cable length rolled 2 m) to any 2-pin earthed socket.
GENERAL
PACKAGING
The packaging has been designed to protect the product while in transit. The
materials used have been selected with due regard for environmental issues
and with recycling in mind.
SN: REF
HELLMUT RUCK GmbH
Daimlerstraße 23 www.hellmut-ruck.de
75305 Neuenbürg MADE IN GERMANY
10137/10138
100 - 240 V | 50/60 Hz +/-10%| 420 VA
SWL 200 kg |IP 20
10137-10128 TYPSCHILD | Rev. A

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
14|COMMISSIONING | CARRYING INSTRUCTIONS
Danger of crushing: mains cable
Do not run over the mains cable (danger of crushing, electrical hazard),
the foot switch cable or a hand switch cable which may be hanging down
(danger of crushing, malfunction).
Stumbling hazard
Lay out the mains cable and foot switch in such a manner that they do
not pose a stumbling hazard.
Electrical hazard
The mains cable and cable for the foot switch must not get into the mov-
ing parts; danger of crushing and/or danger of electrical shock.
Accessibility of the mains plug
The accessibility of the mains plug must always be ensured for discon-
necting from the mains network in an emergency.
Set-up Diagram | Space Requirements
Always ensure that the chair is moved far enough away from the wall that
the head section can be freely moved from the sitting position to the flat
position.
COMMISSIONING | CARRYING INSTRUCTIONS 15
|
AFTER UNPACKING THE DEVICE
HEAD SECTION ASSEMBLY | Sliding-in and Fixing in Place
HEAD SECTION ASSEMBLY | Fixing in Place via the Rotary Knob

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
A) Rohrrahmen Kopfteil B) Rohrrahmen Sitzteil
A)
B)
16|COMMISSIONING | CARRYING INSTRUCTIONS
HANDLE POSITIONS
FOR CARRYING
REMOVAL OF THE WEIGHT PLATES
123
4
5
To reduce the total weight when carrying, two weight plates can be removed from
the base. To do this, slightly lift the cladding element, remove three hexagon socket
screws, and take out the two weight plates.
6
Weight plates
The treatment chair must only be operated with the weight plates inserted
and screwed into the pedestal.
COMMISSIONING | CARRYING INSTRUCTIONS 17
|
A) Kopfteilabdeckung fixieren
CARRYING EXAMPLE
FIXATION OF THE HEAD SECTION COVER
COMPENSATING FOR FLOOR UNEVENNESS
To achieve maximum stability,
floor unevenness can be com-
pensated using the height adjus-
chair foot plates. To do this, the
foot plates are adjusted in height
by turning.
Then, bond the provided silicone
pads on the four foot plates.
Thus, the chair is securely held
in position and floor coverings of
different types are protected from
scratches. To attach the silicone
pads, first the chair is lifted at the
front, the silicone pads are bond-
ed on the front feet, and then the
same procedure is executed for
the rear feet.
(See handle positions p. 16)
A) Tubular frame head section B) Tubular frame seat section

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
18|IMPORTANT USE REGULATIONS
Positioning of the chair
Push the chair far enough away from the wall that when moving
the back section into a low position, the back section cannot be
stopped by the wall. (If the back section is stopped by the wall and
if the button for the back section is kept depressed, the spindles
of the motor will continue to run. Result – the chair will be pushed
away from the wall, the back section drops to the spindle stop. The
motor can be destroyed, the patient can be frightened, damage
can occur. If the back section is in a low position, move the chair far
enough away from the wall that, when lifting up the back section
into the seated position, the back section cannot get stuck on a
windowsill or other obstructions (back section will be bent).
Armrests, back section
Do not sit on the armrests or on the head end of the back section.
Only use the armrests for arm support. Any other type of use is
prohibited. (Danger of falling, danger of injury) The safe work load
is 20 kg.
Load
The chair may only be loaded with one person weighing a maxi-
mum of 200 kg (including cushion support). The chair must not be
occupied by 2 or more persons at the same time under any circum-
stances (danger of falling, danger of breaking and danger of injury).
Moving parts
Do not reach into moving parts (mechanical elements). Significant
injuries can occur. The cushion is a component of the treatment
chair.
General note
The daily function check includes checking all movement sequenc-
es as described under point "Operation" starting on page 20.
OPERATION 19
|
RELEASING THE FOOT OR HAND SWITCH
The foot or hand switch may only be disconnected from the treatment chair
by removing the plug connector for service purposes following contact with
the RUCK customer service. Before replacing the switch, the treatment chair
must be disconnected from the mains. To disconnect the switch eject the
locking ring by using a slot screwdriver. Then the plug can be pulled out
carefully. To connect the switch first insert the Plug. Then press in the lock-
ing ring.
HOW THE PATIENT GETS ON AND OFF
The treatment chair can be
mounted or exited from the
front or from the side. To
mount from the side the arm-
rest is swung upward.
It is recommended that the
patients get on and off from
the side. In this way, walking
over waste materials that
accumulate during the treat-
ment, can be prevented.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
Instructions for use for RUCK® treatment chair STELLA 3s | REV 02 Instructions for use for RUCK® treatment chair STELLA 3s | REV 02
21
|
20|OPERATION OF FOOT SWITCH WITH MEMORY FUNCTION
1
5
2
3
6
4
MEMORY BUTTONS
By activating the memory
buttons [1], [2], [3] or [4],
the electric drives move the
treatment chair into the posi-
tions stored by the user. The
treatment chair moves only if
the user keeps the memory
button depressed, otherwise
the treatment chair stops mo-
ving.
HEIGHT ADJUSTMENT
The user activates the button
and the height adjustment
moves upward. To move in
the opposite direction, the
user “double-clicks” the but-
ton and keeps the button
depressed. This moves the
height adjustment down-
ward.
BUTTON BLOCK
Hold down the memory button
[1] and the memory button [2]
simultaneously until two beeps
are heard. All LEDs light up –
the button lock is now active.
Deactivate the button lock by
pressing the memory keys [1]
and [2] simultaneously until
two beeps are heard. Button
lock is switched off.
71
7
2 3 15
42
8
OPERATION OF FOOT SWITCH WITH MEMORY FUNCTION
Do not apply your full body weight
to the foot switch otherwise the but-
tons may be damaged. The seven
dark grey pressure points respond
to slight pressure.
BACK AND SEAT INCLINATION ADJUST-
MENT
The user activates the button [6] and the seat
surface tilts while the backrest moves along
with it. To move in the opposite direction, the
user “double-clicks” the button and keeps the
button depressed. This brings the seat sur-
face back into home position, the backrest
moves along with it.
CLEANING
The podiatry chair’s foot switch is protected
against ingress of dust and moisture in ac-
cordance with IP 64. For cleaning and dis-
infecting, use the RUCK surface disinfection
wipes, medical product, alcohol-free (item
no. 2967002, item no. 2967102 or item no.
2967202). Please observe the exposure
times according to the labelling of the surface
disinfection wipes.
SAVING A MEMORY POSITION
1. Move to any position via the function buttons
[5] and [6].
2. Select the desired user [8], [9] or [10].
The selection is visible as the corresponding
LED lights up.
3. Press memory button [7], LED lights up. Proce-
dure is automatically aborted after 10 seconds.
4. User selects desired memory button [1], [2], [3]
or [4] and keeps it depressed for 3 seconds. Suc-
cessful saving is confirmed by a signal tone.
6 7 8
THE FOOT SWITCH MUST NOT BE
RINSED UNDER RUNNING WATER OR
IMMERSED IN WATER. OTHERWISE IT
MAY STOP WORKING!
INITIALISATION AFTER REPLACEMENT OR MAINTENANCE OF THE FOOT SWITCH
Press memory button [1] and memory button [4] simultaneously and keep them pressed. The
control system is initialised. Several short beeps sound during initialisation. Release the buttons
again when no sound is heard for at least 3 seconds. All electric motors must now be moved once
via the function buttons [5] and [6] into the respective maximum and minimum position. Thereafter,
the treatment chair is again ready for operation. Pressing buttons [2] and [3] simultaneously for at
least 10 seconds resets the memory positions to the factory settings.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
22|OPERATION – HAND SWITCH OPERATION – HAND SWITCH 23
|
HEIGHT ADJUSTMENT
The user activates the button [1] and the height adjust-
ment moves upward. For movement in the opposite
direction, button [2] is activated. This moves the height
adjustment downward.
BACK AND SEAT INCLINATION ADJUSTMENT
The user activates the button [3] and the back and
seat section moves to the rear.
To move in the opposite direction the user activates the
button [4]. The back and seat section will be brought
back into the upright position.
BUTTON LOCK
Simultaneously keep memory button [5] and memory
button [6] depressed until two beep tones sounds. All
LEDs are illuminated. Button lock is active. Deactivate
the button lock by depressing button [5] and button [6]
until two signal tones sounds. Button lock is inactive.
CLEANING
The hand switch is protected against ingress of dust
and moisture in accordance with IP X4. To clean and
disinfect, use the RUCK® disinfectioncloths– alcohol
free (item no. 29428).
1
1
3
3
5
75
2
2
4
4
6
8
9
6
MEMORY BUTTONS
By activating the memory button [5] [6] [7] or [8] the electric drives move the treatment
chair into the positions stored by the user. The treatment chair moves only if the user keeps
the memory button depressed, other wise the treatment chair does not move.
1. Move to any position via the function buttons [1], [2], [3], [4].
2. Select the desired user by pressing the profile/save button [9]: A, B or C.
The selection is indicated by illumination of the appropriate LED.
3. Activate the profile/save button [9] for 3 seconds until the selected LED of the
selected user profile is flashing. Procedure aborts after 10 seconds.
4. User selects desired memory button [5], [6], [7], [8] and keeps it depressed for
3 seconds. Successful saving of the memory position is confirmed by a signal tone.
5. The LED of the selected user profile is again constantly illuminated.
SAVING A MEMORY POSITION
5 7
5
68
8
9
THE FOOT SWITCH MUST NOT BE RINSED UNDER RUNNING WATER OR SUB-
MERGED IN WATER OTHERWISE IT MAY STOP WORKING!
INITIALISATION AFTER REPLACEMENT OR MAINTENANCE OF THE HAND SWITCH
Press button [5] and button [8] simultaneously and keep them pressed. The control sys-
tem is initialised. Several short beeps sound during initialisation. Release the buttons again
when no sound is heard for at least 3 seconds. All electric motors must now be moved once
to the respective maximum and minimum position using the buttons [1] to [4]. Thereafter,
the treatment chair is again ready for operation.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
24|OPERATION – FOOT SWITCH
TURNING THE SEAT SURFACE
GRUND-
POSITION
To turn the seat area of the chair, turn one of the hand wheels (located on both
sides of the seat area) and pivot the podiatry chair into the desired position. At
approximately 90° a stop can be felt.
RUCK®treatment chair STELLA 3s enables two modes of operation:
For this mode of operation,
the chair is merely pivoted
out of the home position
briefly for the lateral foot
bath. In this process the
hand wheel is turned back
after a short time approxi-
mately 1/4 turn. Thus the
chair locks in place when it
pivots back into the home
position. The practitioner
works around his patient
starting from the right.
If the hand wheel is turned
to the stop, then the chair
can be pivoted beyond the
home position as often as
desired without locking in
place. This mode of op-
eration is preferred if the
practitioner wants to bring
the feet of his patient into
different positions during
the treatment.
01. Working in the home
position
02. Freely pivoting
HEIGHT ADJUSTMENT
The user activates the button [1] and the
height adjustment moves upward. For move-
ment in the opposite direction, the button [2]
is activated. This moves the height adjust-
ment downwards.
BACK INCLINATION ADJUSTMENT
The user activates the button [3] and the back
and seat section moves to the rear. For move-
ment in the opposite direction, the button [4]
is activated. The back and seat section will be
brought back into the upright position.
Do not apply your full bodyweight
to the foot switch other wise
the buttons may be damaged!
THE FOOT SWITCH MUST NOT BE
RINSED UNDER RUNNING WATER
OR SUBMERGED IN WATER OTHER-
WISE IT MAY STOP WORKING!
71 2 3 4
CLEANING
The foot switch for the RUCK®treatment chair STELLA 3s is protected against ingress
of dust and moisture in accordance with IP X6. To clean and disinfect, use the RUCK®
disinfection cloths – alcohol-free (item no.29428).
OPERATION 25
|
71
2 3
4
72 3
INITIALISATION AFTER REPLACEMENT OR MAINTENANCE OF THE FOOT SWITCH
Press button [2] and button [3] simultaneously and keep them pressed. The control system is ini-
tialised. Several short beeps sound during initialisation. Release the buttons again when no sound
is heard for at least 3 seconds. All electric motors must now be moved once to the respective max-
imum and minimum position using the buttons [1] to [4]. Thereafter, the treatment chair is again
ready for operation.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
26|OPERATION OPERATION 27
|
Rotating the chair
For safety reasons, the chair may only be turned in the lowered seat
position; in the bed position the chair must not be turned.
11
1
2
ADJUSTMENT OF CALF SUPPORT AND FOOT SUPPORT
For an optimal ergonomic mode of operation, the patient's leg or foot is
brought into different positions. For this, the support for calf and foot offers
versatile adaptation.
SWINGING DOWN THE CALF SUPPORT AND FOOT SUPPORT
So that you do not need to lift the entire weight of your patient's leg yourself,
you are supported by powerful gas pressure springs. To adjust calf and foot
support, activate the release mechanism (1) and (2) select the desired posi-
tion. This position will be reliably held by the gas pressure springs.
1
2
(1) release clamp lever – (2) bring leg support into the desired position, then
tighten the clamp lever (1).
PULLING OUT THE FOOT ROLL
ADJUSTING THE CLAMP LEVER
The procedure for changing the clamp lever in its position, without loosening the
foot roll in the process, is shown below in points (1) to (4). This adjustment pos-
sibility is always very helpful if the clamp lever is fixed in place in a position that
could hinder you in your work. To do this (1) grasp clamp lever (2) press the button
with your thumb – lift clamp lever (3) bring clamp lever into the desired position
(4) release button.
1 42 3
To the right, you see two
different treatment situa-
tions. Trimmin nails and
removing calluses par-
ticularly in the heel area
(rhagades).

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
28|OPERATION OPERATION 29
|
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
If you need more space between the foot rolls, the RUCK®treatment chair
STELLA 3s offers you new free space. An intelligent pivot mechanism (pro-
PIVOTING THE FOOT ROLL
12
tected under the utility model law) provides space without losing the proven
force of the gas pressure springs in the process. For this, lift the desired
foot roll (1) and turn it 180° to the outside. The foot rolls will then securely
engage on the fulcrum again and guarantee precise work. Patients consider
the possibility of
In order to turn the foot roll, it must be pulled out as described under
point "PULLING OUT THE FOOT ROLL" on page 27, and then
re-secured by fixing the clamp lever in place.
The foot roll must always be engaged before start-up!
positioning the angled foot laterally, as very pleasant.
It imparts a feeling of safety and in addition, provides you with free space for
treatment.
1
The calf support can be adapted depending on the position of your patient's
leg, and thus increase sitting comfort.
PIVOTING THE CALF SUPPORT
VARIABLE ADJUSTMENT OF THE HEAD SECTION
1
2
For variable adjustment of the head support unscrew the hand wheel (1). Po-
sition the head support (2) and arrest the hand wheel in place (1). If desired,
the head support can also be taken off.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
30|31
|
MAINTENANCE | CLEANING | DISINFECTION
Maintenance
To ensure continued safe and correct operation of the medical de-
vice, the RUCK®tr
eatment chair STELLA 3s must be maintained by
RUCK SERVICE after 24 months at the latest. During maintenance,
an electrical safety inspection is carried out in accordance with DIN
EN 62353.
For this contact RUCK SERVICE.
Undertaking maintenance and repairs
Maintenance tasks and repairs may only be undertaken by specialists
authorized by the manufacturer.
If all listed points are not observed/fulfilled, any guarantee claim or
liability claim for material damage, as well as for personal injury is
rendered null and void.
Only original spare parts may be used.
RUCK SERVICE
HELLMUT RUCK GmbH
Daimlerstrasse 23
75305 Neuenbürg, Germany
Tel. +49 (0)7082 944 22 55
Fax +49 (0)7082 944 22 59
e-mail ser[email protected]
web www.hellmut-ruck.de
CLEANING AND DISINFECTION I 31
Danger of damage due to improper care of artificial leather and sur-
faces!
Disinfect the contact surfaces after each treatment with RUCK surface disinfection
wipes, medical product, alcohol-free (item no. 2967002, item no. 2967102 or item
no. 2967202).
Tip: Using disposable covers helps to avoid direct skin contact (body grease and
sweat are sometimes quite aggressive) with the artificial leather and thus to protect it.
Clean the artificial leather surfaces with a damp cloth and clean with RUCK artificial
leather cleaner and care product (item no. 2947001). Please observe the exposure
times of the disinfectants according to the product labelling.
Danger of damage due to improper care of artificial leather and wood surfaces. Do
not use any concentrated disinfectants, solvents, cleaning agents or liquids that con-
tain alcohol (> 50 % alcohol content). Do not use spray disinfection.
Take great care to ensure that the cleaning utensils used are clean. If mineral conta-
minants such as grains of sand or dust are embedded in the cleaning agents, this
can destroy the surface. Do not use any scouring agents, such as abrasive cleaners,
cleaning sponges or similar items that may contain abrasive particles. Residues of
leaked fluids must be removed from the surface immediately with a soft cloth.
For a gentle cleaning of the surfaces use a soft cotton cloth or the RUCK®microfibre
cloth, which is gentle on surfaces, (item no. 29616), and is also suitable for wooden
surfaces. Caution: Other cleaning cloths with plastic fibres can have an abrasive ef-
fect and consequently they can damage the surface.
Over time, traces of use can appear on the surfaces. These are more or less noticeab-
le depending on light incidence. They do not constitute a lessening of utility. As is the
case with all surfaces over the long term, due to the wear involved in the daily routine
of the medical practice, they cannot be avoided. Spots or stains caused by liquids that
may corrode the surface (such as propolis, tea tree oil, cleaners that contain chlorine
or similar substances) are categorically excluded from the warranty.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
32|CHAPTER HEADING
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
DISINFECTION, CLEANING AND CARE 33
|
Danger due to disease transmission!
Insufficient disinfection can lead to transmission of disease. Always comply with the
requirements of the responsible national body for hygiene and disinfection.
Disinfect the product on a regular basis. Please refer to the notes on page 31
Cushion discolouration due to articles of clothing
Clothing, such as jeans, can lead to discolouration of cushion surfaces. If possib-
le, you should try to remove the colour residues using the RUCK® microfibre cloth,
which is gentle on surfaces (item no. 29616) and the RUCK®artificial leather cleaner
and care product (item no. 2947001). RUCK recommends that you use disposable
covers wherever possible. After cleaning and disinfection, surface disinfectants must
be allowed to effectively evaporate before the next treatment; i. e. allow surface disin-
fectants to thoroughly dry as specified by the manufacturer, but do not wipe surfaces
dry.
32|CLEANING AND DESINFECTION
Environmental hazard!
Incorrect disposal endangers our environment.
>>> Please use available return and collection systems for disposing of
the RUCK®treatment chair STELLA 3s at the end of its life-cycle!
>>> Dampers, gas push and pull springs are pressurised. They must not be
opened or heated. They have an oil fill; this must be disposed of in line
with waste legislation.
Danger of explosion!
Operation of the RUCK®treatment chair STELLA 3s in areas subject
to explosion hazard can trigger an explosion.
Spare parts
Art. no. 10389 Complete set of cushions, two-tone
Art. no. 10395 Complete set of cushions, single-tone
Available in turquoise (01), slate (02), mocha (03),
sand (04), white (05), bamboo (06), mint (07), Baltic (08), orchid (09)
and mango (10).
Please contact RUCK SERVICE for more spare parts.
If you have any questions, please also contact our RUCK SERVICE
department
Telephone +49 (0)7082 944 22 55
Fax +49 (0)7082 944 22 59
E-mail service@hellmut-ruck.de
SPARE PARTS
DISPOSAL

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
34|CHAPTER HEADING TECHNICAL DATA I TREATMENT POSITIONS 35
|
FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
Mounting height 53 cm
Highest seat position 100 cm
Highest position – leg rest 138 cm
Seat width 55 cm
Seat depth 48 cm
Total length – foot section and head section pulled out 210 cm
Total length – foot section and head section pulled in 174 cm
Total width including 82 cm
Total weight 140 kg
Power supply 100 - 240 VAC, 50/60 Hz -15% / +10%
Controller type CO 61
Protection class IP 20
Duty cycle 10% (= 2 min. ON /18 min. OFF)
Maximum load 200 kg
Cushion colour single-tone/two-tone
Warranty 2-years
Electromagnetic influence
The possibility that, through the use of additional electronic
devices, electromagnetic interference/influence can occur
cannot be excluded.
34|TECHNICAL DATA | TREATMENT POSITIONS
82 cm 172 - 202 cm
89 cm
138 cm
202 cm
82 cm

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
36 ICHANGING CUSHIONS
FÜR FUSS UND PFLEGE
WARRANTY 37
|
FÜR FUSS UND PFLEGE
2x
2x
2x
2x
1x
1x
1
1
2
2
24
5
3
2x
2x
3
3
The cushions of the RUCK®treatment chair
STELLA 3s can be replaced individually: (1)
Head section cushion (2) back section cush-
ion and (3) seat section cushion are fixed on
the tube frame using wing bolts.
To mount the back section cushion (2), first
remove the cladding element (4) which is se-
cured with four screws + cover caps.
To mount the seat section cushion (3), the seat
section is placed in the position of maximum
inclination.
Likewise, the head section is secured with
wing screws after the cover (5) has been re-
moved by simply pulling it.
www.hellmut-ruck.de
Video tip
Changing
cushions
Recommendation for changing cushions
Because the cushions of the treatment chairs
are constantly subjected to significant
stress, we recommend replacement after approx. three years (with normal use).
WARRANTY
The product comes with a 2-year warranty. Normal wear of surfaces, castors, etc., and
other moving parts is excluded from the warranty. Traces of use and signs of wear
and tear due to age and use, which are based on normal use corresponding to the
performance values and the nature of the treatment chair, are in accordance with the
contract and are not defects in the legal sense.
The warranty runs from the date of purchase when the product was first bought. The
date of purchase must be confirmed by the sales receipt. Within the scope of the war-
ranty, HELLMUT RUCK GmbH will, at no charge, either repair or replace the defective
products – at RUCK‘s discretion. The prerequisite is carriage paid shipment of the
defective product together with the purchase receipt. The local agent is responsible
for claims made outside of Germany. In case of doubt, please check with RUCK be-
fore returning the device.
The warranty becomes null and void if we find that the problem was caused by impro-
per or incorrect installation or usage, by non-observance of the operating instructions,
by external intervention or as a result of a non-authorised repair or modification.
The guarantee is limited to the repair or replacement of the product. Further liability
(particularly loss of revenue or earnings) is excluded. However, if it transpires that it
is a problem that is not covered by the terms of the warranty or that the period of the
guarantee has elapsed, any costs for inspection and repair must be borne by the
customer.
We reserve the right to make changes to and deviations from the technical design.

FÜR FUSS UND PFLEGE FÜR FUSS UND PFLEGE
8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023 8020702 I Instructions for use RUCK® Treatment Chair STELLA 3s | REV 06 | Revision status 18.01.2023
38||
SAFETY NOTICES 39
SAFETY NOTICES
The interference and interference immunity of the podiatry chair Stella 3s
agrees with the requirements imposed on devices that are not intended for
life-support in the typical clinical environment.
Because of EN 60601-1-2 relating to the electromagnetic compati
bility of electrical medical devices, we must draw your attention to
the following:
• Medical electrical devices are subject to special measures in
terms
of electromagnetic compatibility and must be operated in accord
ance with the requirements laid out in these Operating instructions.
• Portable and mobile high-frequency communication devices (in
cluding accessories, such as antenna cable and external cables)
can influence medical electrical equipment. Such equipment
should be operated at a minimum distance of 30cm from the
RUCK®treatment chair STELLA 3s and its parts.
Otherwise the possibility of device performance impairments can
not be excluded. Please also refer to: 8.2 Specifications referring to
electromagnetic compatibility in accordance with EN 60601-1-2
Risk of strangulation from power cable. Keep children away from
the device.
When using the device in domestic premises, there can be a risk
resulting from power cables that have been damaged by animals,
for example. Please check the mains power connection regularly
or signs of damage and if any are found, disconnect the device
from the mains.
Protect the device against pest infestation through regular checks
and cleaning.
The instructions regarding operation and transport of the device
must be adhered to.
FÜR FUSS UND PFLEGE
ENVIRONMENTAL CONDITIONS
Ambient temperature and humidity (non-condensing)
in operation +5°C to +40°C and 30% to 75% hu-midity
(non-condensing))
in transit and storage -10 °C (without monitoring of relative air humi-
dity) up to +50 °C (at relative air humidity from
15 % up to 93 %, non-condensing)
Air pressure 700 hPa (≈ 3000 metres above sea level)
- 1060 hPa
ELECTROMAGNETIC COMPATIBILITY
Guidelines and manufacturer’s declaration – electromagnetic emissions
The RUCK®STELLA 3s treatment chair is designed for use in an environment as specified
below. The customer or user of the RUCK®STELLA 3s treatment chair must ensure that it is
used in such an environment.
Emitted interference measure-
ments
Conformance Guidelines for the electromagnetic
environment
HF emissions according to
CISPR (Comité International
Spécial des Perturbations
Radioélectriques) [Special
International Committee on
Radio Interference] 11
Group 1 The RUCK®STELLA 3s treatment chair
uses HF energy exclusively for its inter-
nal function. Therefore, its HF emissions
are very low and it is unlikely that any
interference will be caused to nearby
electronic devices.
HF emissions according to
CISPR (Comité International
Spécial des Perturbations
Radioélectriques) [Special
International Committee on
Radio Interference] 11
Class B The RUCK®STELLA 3s treatment chair
is suitable for use in all facilities, includ-
ing those in residential areas, and those
directly connected to the public power
supply that also supplies buildings used
for residential purposes.
Harmonic component
emissions according to IEC
61000-3-2
Class A
Voltage fluctuations / flicker
emissions according to IEC
61000-3-3
Complies
Table of contents
Other Ruck Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Aquolab
Aquolab AQBAS01 user manual

Delft Imaging
Delft Imaging CAD4TB instruction manual

Carestream
Carestream Kodak DirectView CR 500 SYSTEM Preventive maintenance

Humanware
Humanware SmartView Graduate Getting started

Drive
Drive 13085RN Assembly instructions

Coltene
Coltene SciCan BRAVO G4 Technical & service manual

Nasco
Nasco Life/form LF00855U instruction manual

Smiths Medical
Smiths Medical Pneupac babyPAC 100 user manual

SELVAS Healthcare
SELVAS Healthcare SensePlayer T90ET user manual

Avita
Avita NS1 instruction manual

Orliman
Orliman 3010-C INSTRUCTIONS FOR USE AND PRESERVATION

VELA Medical
VELA Medical Dental Chair user manual