Sartorius mLINE User manual

User Manual
Bedienungsanleitung
Mode d'emploi
Manual Usuario
Instruzioni d'impiego
Инструкция пользователя
mLINE®pipette

User Manual ............................................................1
Bedienungsanleitung .......................................... 17
Mode d’ emploi ..................................................... 35
Manual Usario ....................................................... 53
Instruzioni d’impiego .......................................... 71
Инструкция пользователя
................................. 89
Specifications ......................................................109
Huom! Suomenkielinen käyttöohje ladattavissa
osoitteesta: www.sartorius.com
Obs: Manual på svenska kan du ladda ner
på adressen www.sartorius.com
mLINE®pipette

1
Contents
1. Intended use .....................................................................................................2
2. Product description........................................................................................2
3. Sartorius pipette tips.....................................................................................4
4. Unpacking.......................................................................................................... 4
5. Personal identification..................................................................................5
6. Pipette holder and carousel stand............................................................ 5
6.1. Installing the pipette holder........................................................5
6.2. Carousel stand ..................................................................................5
7. Calibration tool (also tube and ID window opener)........................... 5
8. Operating the pipette.................................................................................... 6
8.1. Setting the volume.......................................................................... 6
8.2. Sealing and ejecting tips...............................................................6
8.3. Protective tip cone filters............................................................. 6
9. Pipetting techniques...................................................................................... 7
9.1. Forward pipetting............................................................................7
9.2. Reverse pipetting............................................................................. 7
9.3. Repetitive reverse technique....................................................... 8
10. Recommendations for good pipetting....................................................8
11. Maintenance..................................................................................................... 9
11.1. Daily cleaning of the outer surface of the pipette ............. 9
11.2. Cleaning the lower part of the pipette....................................9
11.3. Sterilization of the pipette.........................................................11
12. Testing the performance and recalibration ........................................11
12.1. Testing the performance.............................................................12
12.2. Recalibration....................................................................................13
13. Troubleshooting.............................................................................................15
14. Warranty information.................................................................................16
15. Performance specifications ......................................................................16

2
1. Intended use
mLINE pipette is intended to be used in liquid
handling applications for dispensing liquids in
volume range of 0,1 µl to 10 ml. It is recommended
to use Optifit Tips or SafetySpace™ Filter Tips to
ensure the best compatibility and performance
with Sartorius pipettes.
This liquid handling device is designed and
manufactured to be used as general purpose
laboratory equipment. Before operating it is
recommended to read the user manual, which
contains useful information e.g. for good
pipetting practice.
2. Product description
Your new mLINE is an autoclavable air
displacement pipette. The attractive and
ergonomical design of the mLINE pipette together
with its very low pipetting forces reduce the risk
of repetitive strain injuries (RSI). The pipette is
designed for both right- and left-handed use.
NOTE: It is known that prolonged pipetting can
cause RSI. The manufacturer is not responsible for
RSI or any related diseases caused by prolonged
pipetting.
Cat. No. Colour
Code
Channels Volume
Range/µl
Increment
/µl
Tip
/µl
Safe-Cone Filters
50 pcs/bag
Standard Plus
725010 Grey 10.1-3 0.002 10 - -
725020 Grey 10.5-10 0.01 10 - -
725030 Yellow 1 2-20 0.02 200 721014 -
725050 Yellow 1 10-100 0.1 200, 350 721008 721018
725060 Yellow 1 20-200 0.2 200, 350 721007 721017
725070 Blue 1 100-1000 11000 721006 721016
725080 Green 1500-5000 10 5000 721005 721015
725090 Red 11-10 ml 20 10 ml 721005 721015
725120 Grey 8 0.5-10 0.01 10 - -
725130 Yellow 8 5-100 0.1 200, 350 721008 721018
725140 Orange 8 30-300 0.2 350 721007 721017
725220 Grey 12 0.5-10 0.01 10 - -
725230 Yellow 12 5 - 100 0.1 200, 350 721008 721018
725240 Orange 12 30 - 300 0.2 350 721007 721017
mLINE single- and multichannel pipettes

3
Pipette description
Cap .......................
Operating
button ...................
Tip
ejector .......
Display .............
Volume
lock ..................
Colour
coding...............
Tip ejector
collar ......................
Tip cone....................
ID window
and tag
Finger
support
........ Handle
... Calibration
nut
Safe-Cone
....... Filter
Tip cone ...............
........ Tip cone
housing
Tip
........ ejector bar
........ Safe-Cone
Filter

4
The special tip cone design of the pipette offers
the possibility to use replaceable Safe-Cone
Filters to help prevent the contamination and
damage of the pipette. The operating button of
the mLINE allows safe and easy filter ejection. The
pipette uses disposable tips.
3. Sartorius pipette tips
Sartorius pipette tips are recommended for
use with mLINE pipettes. The use of fully
compatible tips will ensure the maximal pipetting
accuracy and precision, and thereby guarantee
the performance specifications given for the
pipette. Sartorius tips are made of pure virgin
polypropylene and manufactured in protected
clean room conditions.
Sartorius offers a full range of standard Optifit,
and SafetySpace™ Filter tips. Sartorius non-
filter tips are available in trays, bulk packages
and space-saving refill systems. Sartorius non-
filtered tips and trays are autoclavable at 121°C
(252F), 20 min, 1 bar (15 p.s.i). All Single Tray
Racks and clean Refill Packs are certified RNase,
DNase and endotoxin-free.
For more information, please go to www.
sartorius.com or contact your local Sartorius
support.
4. Unpacking
The mLINE pipette package contains the following
items:
1. Pipette
2. Tip
3. Caps for colourcoding
4. Calibration tool (also tube and ID window
opener)
5. Pipette holder
6. Identification tags
7. Safe-Cone Filters (pipettes >10 µl)
8. Grease
9. Instruction manual
10. Performance certification in accordance
with ISO 8655-6
Please check that all items are included and that
no damage has occurred during shipment.

5
A
B
C
5. Personal identification
Each mLINE pipette can be identified with a tag
located under the ID window which on the finger
support.
1. Remove the ID window by using the
calibration tool.
2. Remove the tag and mark the identifying
information on it.
3. Position the tag in place and clip the window
back into place.
6. Pipette holder and
carousel stand
For convenience and safety, always keep the
pipette vertically on its own holder or carousel
stand when not in use.
Cat.No. Product
725600 Carousel Stand for 6 pipettes
725610 Pipette Holder for 1 pipette
6.1. Installing the pipette holder
1. Clean the shelf surface with ethanol.
2. Remove the protective paper from the
adhesive tape.
3. Install the holder by pressing it against the
edge of the shelf.
4. Place the pipette onto the holder.
6.2. Carousel stand
A convenient and ergonomical Carousel stand for
six mLINE pipettes is also available
7. Calibration tool
(also tube and ID window opener)
The calibration tool is designed for the following
purposes:
1. Tool for recalibration (A).
2. Opening tool for the ID window and lid of
the calibration nut (B).
3. Opening tool for the tubes (C).

6
8. Operating the pipette
8.1. Setting the volume
The volume of the pipette is clearly shown
through the display. The volume setting is carried
out with the following steps:
1. Press the volume lock and hold it down.
2. Set the volume by rotating the operating
button (clockwise to decrease the volume
and counterclockwise to increase).
3. Release the volume lock.
NOTE: Do not rotate the operating button
without pressing down the volume lock. Do not
rotate the operating button outside the volume
range of the pipette.
8.2. Sealing and ejecting tips
To ensure the maximum accuracy and precision,
Sartorius tips are recommended to be used with
mLINE pipettes. Before fitting a tip ensure that
the pipette tip cone is clean. Press the tip on the
tip cone of the pipette. The pipette is provided
with a spring loaded tip cone for optimal sealing.
Eject the tip by pressing the tip ejector with your
thumb. Make sure that the tip is disposed into a
suitable waste container.
NOTE: If the tip is pressed too hard the tip cone
yields and, as a result, the counter and operating
button moves.
8.3. Protective tip cone filters
The tip cone design of the mLINE pipettes (>10
µl) allows the use of Safe-Cone Filters in the tip
cones. These removable filters prevent liquids
and aerosols from entering the pipette and thus,
protects the pipette from contamination and
damage.
The Safe-Cone Filters are available as Standard
or Plus versions. It is recommended to use the
Standard filter for general applications and the
Plus filter for more demanding applications such
as cell culture, bacterial and virological work and
molecular biology. Filters need to be exchanged
regularly. The interval for filter exchanging is
application dependant but the recommendation is
Tip ejector
+
–
Volume
lock

7
to change the filter daily (after 50 - 250 pipetting
cycles) and always in case of over-aspiration. (See
Chapter 2 for ordering information.)
The Safe-Cone Filters can be removed with the
unique filter ejector without touching the filters
by hand. Remove the operating button cap and
eject the filter by pressing the operating button
down. Make sure that the filter is disposed of into
a suitable waste container. Clean the tip cone, if
needed, and put the new filter in place.
9. Pipetting techniques
Two basic pipetting techniques, forward and
reverse pipetting, are associated with mLINE
pipettes. Forward pipetting is the most common
used pipetting technique. The technique employs
the blow-out function ensuring complete delivery
of the liquid. Reverse pipetting is recommended
for highly viscous, biological or foaming liquids,
or very small volumes of liquid. A selected
volume plus an excess is aspirated into the tip.
The delivery is done without blow-out, and, thus,
the excess volume remains in the tip. The reverse
technique also facilitates the repeated delivery of
the same volume. If reverse pipetting technique is
used the pipette might need recalibration.
9.1. Forward pipetting
1. Fit the tip onto the pipette tip cone.
2. Press the operating button to the first stop.
3. Place the tip just under the surface of the
liquid (2-3 mm) and smoothly release the
operating button up to the starting position.
Wait one second. Carefully withdraw the tip
from the liquid, touching against the edge
of the container to remove excess from the
outside of the tip.
4. Dispense the liquid by pressing the operating
button to the first stop. After a short delay
press the operating button to the second
stop. This action will empty the tip.
9.2. Reverse pipetting
1. Fit the tip onto the pipette tip cone.
2. Press the operating button all the way to the
second stop.
3. Place the tip just under the surface of the
0
1
2
0
1
2

8
liquid (2-3 mm) and smoothly release the
operating button up to the starting position.
4. Withdraw the tip from the liquid touching
against the edge of the container to remove
excess.
5. Press the operating button smoothly to the
first stop to deliver the desired volume. Hold
the operating button at the first stop. The
liquid that remains in the tip should not be
included in the delivery.
6. Discard the remaining liquid by pressing the
operating button to the second stop.
9.3. Repetitive reverse technique
1. Follow the reverse technique steps 1 to 5.
2. Continue pipetting by repeating steps 3 to 5
as long as needed.
3. Finally discard the remaining liquid by
pressing the operating button to the second
stop.
10. Recommendations
for good pipetting
- Make sure that the tip is firmly attached to
the tip cone.
- Hold the pipette vertically when aspirating
the liquid and place the tip only a few
millimeters into the liquid.
- Always control the operating button slowly
and smoothly.
- Pre-rinse the tip before aspirating the liquid
by filling and emptying the tip for three to
five times. This is important especially when
pipetting liquids with a viscosity and density
greater than water or liquids with high vapor
pressure (e.g. ethanol).
- Check that the pipette, tip and liquid are at
the same temperature.
- When pipetting liquids with temperatures
different to the ambient temperature change
the tip after each pipetting. Do not pre-rinse
the tip.
- To avoid contamination, do not rest the
pipette on its side especially with the tip
attached.
- Change the tip cone filter regularly
(recommendation after 50 - 250 pipetting
cycles).
- Never strike the tip cone against the tip tray

9
when mounting the tip as this can damage
the pipette.
- Avoid rough handling and do not drop the
pipette.
- Avoid exposing the unit to extreme
temperature changes, humidity and dust
(operating temperature from 15°C to 40°C).
11. Maintenance
mLINE pipettes have been designed for easy
in-house service. If the pipette is used daily it
is recommended to clean/decontaminate and
check the performance of the pipette every
three months. Sartorius also provides complete
repair and recalibration service for your pipette
including a service report and performance
certificate.
NOTE: Please make sure that the pipette has
been decontaminated before you send it to us
or your local representative. Please advise if any
hazardous material has been used with your
pipette.
NOTE: The use of the tip cone filters may prolong
the service interval. Change the filters regularly.
NOTE: It is recommended to always use gloves
when cleaning the pipette.
11.1. Daily cleaning of the outer
surface of the pipette
Your mLINE pipette should be checked every day
for cleanliness. To clean and decontaminate the
outer surface of your pipette use ethanol (70%),
isopropanol (60%) or a mild detergent and a soft
lint-free cloth.
Gently clean the surface of the pipette with
moistened cloth and wipe dry. Pay special
attention to the tip cone. Change the tip cone
filter if needed (Chapter 8.3.).
11.2. Cleaning the lower
part of the pipette
If your pipette is in daily use it is recommended
to clean/decontaminate and grease the pipette
every three months. It is recommended to send

10
the multichannel pipettes to your local Sartorius
service representative for cleaning and greasing.
To clean and decontaminate the lower parts of
the single channel pipette, follow these steps:
Disassembling and cleaning
(see the pictures under the cover page):
1. Eject the tip cone filter (if fitted, Chapter 8.3.)
2. Unscrew the tip ejector collar (1)
counterclockwise and remove it.
3. Pipettes m3, m10, m20, m100, m200, m1000:
Unscrew the tip cone holder (2) counter-
clockwise and carefully remove it with the
tip cone (3).
Pipettes 5000 µl and 10 ml: Please note
different opening method between these
models.
Pipette 5000 µl:
Hold the tip cone holder firmly with your
fingers and turn the tip cone cylinder counter
clockwise with the other hand.
Pipette 10 ml:
Hold the tip cone cylinder firmly and turn the
tip cone holder clockwise with your fingers.
NOTE: Don't turn the tip cone cylinder +
holder system together. That will break the
pipette!
4. Clean the tip ejector collar, the tip cone
holder, the tip cone (cylinder) and the piston
(5) with ethanol (70%), isopropanol (60%) or
mild detergent and soft lint-free cloth.
5. Clean the interior of the tip ejector collar and
the tip cone (cylinder) with a cotton swab. Be
careful with the pipettes m3, m10, m20 and
m100 so that the seal inside the tip cone will
not damage.
6. Rinse the parts with distilled water if needed
and let the parts dry.
7. Pipettes m3, m10, m20 and m100:
Put a thin layer of grease on the piston (5).
Pipettes m200 and m1000:
Put a thin layer of grease around the seal (6).
Pipettes m5000 and m10 ml:
Put a thin layer of grease on the seal (6) and
the interior of the tip cone cylinder (4).
NOTE:Avoid excess grease. Use only the grease
provided with the pipette.
NOTE: Before reassembling check that no lint or
particles are on the surface of the piston.
Hold tip
cone
holder
tightly
Rotate
tip cone
cylinder
Rotate tip
cone
cylinder
Hold tip
cone
holder
tightly

11
Reassembling:
1. Pipettes m3, m10, m20, m100, m200, m1000:
Carefully place the tip cone (3) on the piston
and attach it by screwing the tip cone holder
(2) clockwise.
Pipettes m5000 and m10 ml:
Carefully place the tip cone cylinder (4)
on the piston and screw clockwise. Make
sure that the tip cone (cylinder) is properly
tightened. Avoid over-tightening.
2. Attach the tip ejector collar (1) by screwing it
clockwise.
3. Place the new tip cone filter on its place.
4. Press the operating button several times to
ensure that the grease has spread evenly.
5. Check the performance of the pipette.
NOTE: It is always necessary to check the
performance of the pipette after in-house service
or maintenance.
11.3. Sterilization of the pipette
The entire mLINE pipette can be sterilized by
steam, autoclaving at 121°C, (252°F), 1 atm for
20 minutes. Remove the tip cone filter (if fitted,
Chapter 8.3). The single channel pipettes can be
autoclaved without special preparations. Unscrew
the lower part of the multichannel pipettes by
holding the connecting collar and turning the
lower part 360° counterclockwise. Put the pipette
into the sterilization bag and place it into the
autoclave. After autoclaving, the pipette must
be cooled down and left to dry overnight before
use. Screw the lower part of the multichannel
pipettes by holding the connecting collar and
turning the lower part 360° clockwise. Make
sure that the lower part is properly tightened.
It is recommended to check the performance
of the pipette after every autoclaving. It is also
recommended to grease the piston/seal of the
pipette after every 10 autoclaving.
12. Testing the
performance and
recalibration
It is recommended to check the performance
of your mLINE pipettes regularly (e.g. every 3
months) and always after in-house maintenance.
However, the user should establish a regular
360°
Connecting
collar

12
testing routine for their pipettes with regard
to accuracy requirements of the application,
frequency of use, number of operators using
the pipette, nature of the liquid dispensed and
the acceptable maximum permissible errors
established by the user. (ISO 8655-1.)
12.1. Testing the performance
Performance testing should take place in a
draught-free room at 15 - 30°C, constant to
+/- 0.5°C and humidity above 50%. The pipette,
tips and the test water should have stood in the
test room a sufficient time (at least 2 hours) to
reach equilibrium with the room conditions. Use
distilled or deionised water (grade 3, ISO 3696).
Use an analytical balance with a readability of
0.01 mgs. (ISO 8655-6.)
Weighing
1. Adjust the desired test volume Vs.
2. Carefully fit the tip onto the tip cone.
3. Fill the tip with test water and expel to waste
five times to reach a humidity equilibrium in
the dead air volume.
4. Replace the tip. Pre-wet the tip by filling it
once with test water and expel to waste.
5. Aspirate the test water, immersing the tip
only 2-3 mm below the surface of the water.
Keep the pipette vertical.
6. Withdraw the pipette vertically and touch
the tip against the inside wall of the test
water container.
7. Pipette the water into the weighing vessel,
touching the tip against the inside wall of
the vessel just above the liquid surface at an
angle of 30° to 45°. Withdraw the pipette by
drawing the tip 8-10 mm along the inner wall
of the weighing vessel.
8. Read the weight in mgs (mi).
9. Repeat the test cycle until 10 measurements
have been recorded.
10. Convert the recorded masses (mi) to volumes
(Vi)
Vi= miZ Z = correction factor (Table 1)
11. Calculate the mean volume (⊽) delivered:
⊽= (∑Vi)/10
12. For conformity evaluation calculate the
systematic error esof the measurement:
in µl: es= ⊽- VsVs= selected test volume
or in %: es= 100 (⊽- Vs)/Vs

13
13. For conformity evaluation calculate the
random error of the measurement:
as standard deviation s = ∑(Vi- ⊽)2
n - 1
n = number of measurement (10)
or as coefficient of variation CV = 100s/⊽
14. Compare the systematic error (inaccuracy)
and random error (imprecision) with the
values in the performance specifications
- (p. 109) or the specifications of your
own laboratory. If the results fall within
the specifications, the pipette is ready for
use. Otherwise check both systematic and
random errors and, when necessary, proceed
to the recalibration procedure (Chapter 12.2).
NOTE: Systemic error (inaccuracy) is the
difference between the dispensed volume and the
selected test volume. Random error (imprecision)
is the scatter of the dispensed volumes around
the mean of the dispensed volume. (ISO 8655-1.)
NOTE: Sartorius specifications are achieved in
strictly controlled conditions (ISO 8655-6). The
user should establish own specifications based on
the field of use and the accuracy requirements
placed on the pipette (ISO8655-1).
Table 1
Z-values (µl/mg):
Temp. (°C) Air Pressure (kPa)
95 100 101.3 105
20.0
20.5
21.0
21.5
22.0
22.5
23.0
23.5
1.0028
1.0029
1.0030
1.0031
1.0032
1.0033
1.0034
1.0036
1.0028
1.0029
1.0031
1.0032
1.0033
1.0034
1.0035
1.0036
1.0029
1.0030
1.0031
1.0032
1.0033
1.0034
1.0035
1.0036
1.0029
1.0030
1.0031
1.0032
1.0033
1.0034
1.0036
1.0037
NOTE: This method is based on ISO 8655.
12.2. Recalibration
The calibration of your mLINE pipette has been
factory-checked by using forward pipetting
technique and certified at 22°C using grade
3 distilled water according to ISO 3696. The

14
calibration is based on ISO 8655-6: Gravimetric
test method for volumetric instruments. The
pipette's specifications are guaranteed only with
genuine Sartorius tips. If you find the pipette to
be inaccurate after performance testing, please
follow the instructions below:
1. Remove the lid of the calibration nut, located
on the back of the handle, with the aid of the
calibration tool.
2. Place the hexagonal head of the calibration
tool into the hole of the calibration nut.
3. Turn the adjustment lock counterclockwise
to decrease and clockwise to increase the
volume.
4. Repeat testing the performance procedure
(Chapter 12.1.). Continue until the results
are correct.
NOTE: Recalibration is recommended if reverse
pipetting technique is used.
NOTE: Sartorius offers accredited calibration
service. Please contact your local Sartorius
representative for further information.
+
–

15
13. Troubleshooting
Symptom Possible cause Solution
Droplets left
inside the tip
Unsuitable tip Use original
Sartorius tips
Leakage or pipetted
volume too small
Non uniform wetting
of the plastic
Attach new tip
Tip incorrectly
attached
Attach firmly
Unsuitable tip Use original
Sartorius tips
Foreign particles
between tip and cone
Clean the tip cone,
attach new tip
Tip cone holder
incorrectly tightened
Tighten the tip cone
holder
Pipette damaged Return to your
Sartorius service
representative for
servicing
Pipette outside
the established
specifications
Incorrect operation Follow instructions
Unsuitable tip Use original
Sartorius tips
Calibration altered Reclibrate
Operating button
jammed or moves
erratically
Liquid has penetrated
tip cone and dried
Clean and grease
piston and seal
Clean the tip cone
Safe-Cone Filter has
been contaminated
Change filter
Insufficient amount
of grease on a piston
and seal
Grease accordingly
Tip ejector jammed
or moves erratically
Tip ejector collar has
been contaminated
Remove and clean
ejector collar and
tip cone

16
14. Warranty information
The mLINE pipettes are covered by warranty
for 2 years against defects in materials and
workmanship. Should your mLINE pipette fail to
function at any time, please contact your local
Sartorius representative.
ANY WARRANTY WILL, HOWEVER, BE DEEMED AS
VOID IF FAULT IS FOUND TO HAVE BEEN CAUSED
BY MALTREATMENT, MISUSE, UNAUTHORIZED
MAINTENANCE OF SERVICE OR NEGLIGENCE
OF REGULAR MAINTENANCE AND SERVICE,
ACCIDENTAL DAMAGE, INCORRECT STORAGE
OR USE OF THE PRODUCTS FOR OPERATIONS
OUTSIDE THEIR SPECIFIED LIMITATIONS, OUTSIDE
THEIR SPECIFICATIONS, CONTRARY TO THE
INSTRUCTIONS GIVEN IN THIS MANUAL OR WITH
OTHER THAN THE MANUFACTURER’S ORIGINAL
TIPS.
Each mLINE pipette is tested before shipping
by the manufacturer. The Sartorius Quality
Assurance Procedure guarantees that the mLINE
pipette you have purchased is ready for use.
15. Performance
specifications
The manufacturer's specifications (p. 109) are
guaranteed only when the manufacturer's
original tips are used. The manufacturer's
specifications should be used as guidelines when
establishing your own performance specifications
in accordance with ISO 8655.

Specifications
mLINE single-, 8- and 12-channel pipettes
Cat. No Channels Volume
Range µl
Test
Volume µl
Inacc. (%) Impr. (%)
725010 10.1-3 3
1.5
0.3
1.30
2.40
10.0
0.80
1.60
6.00
725020 10.5-10 10
5
1
1.00
1.50
2.50
0.60
1.00
1.50
725030 12-20 20
10
2
0.90
1.20
3.00
0.40
1.00
2.00
725050 110-100 100
50
10
0.80
1.00
2.00
0.15
0.40
1.00
725060 120-200 200
100
20
0.60
0.80
2.00
0.15
0.30
0.80
725070 1100-1000 1000
500
100
0.60
0.60
1.00
0.20
0.20
0.40
725080 1500-5000 5000
2500
500
0.50
0.60
2.00
0.20
0.30
0.60
725090 11-10 ml 10 000
5000
1000
0.60
1.20
3.00
0.20
0.30
0.60
725120 80.5-10 10
5
1
1.50
2.50
4.00
1.00
2.50
4.00
725130 85-100 100
50
10
0.70
1.00
3.00
0.25
0.70
1.50
725140 830-300 300
150
30
0.60
1.00
2.00
0.25
0.50
1.00
725220 12 0.5-10 10
5
1
1.50
2.50
4.00
1.00
2.50
4.00
725230 12 5-100 100
50
10
0.70
1.00
3.00
0.25
0.70
1.50
725240 12 30-300 300
150
30
0.60
1.00
2.00
0.25
0.50
1.00

4002080 • 03/2013 • © Sartorius Biohit Liquid Handling Oy
Sartorius Biohit Liquid Handling Oy
Laippatie 1
FI-00880 Helsinki
Finland
Phone +358.755.951
www.sartorius.com
Headquarter
Sartorius Corporate
Administration GmbH
Weender Landstrasse 94-108
37075 Goettingen, Germany
Phone +49.551.308.0
Fax +49.551.308.3289
www.sartorius.com
Copyright by Sartorius, Helsinki,
Finland.
All rights reserved. No part of this
publication may be reprinted or
translated in any form or by any
means without the prior written
permission of Sartorius.
The status of the information,
specifications and illustrations in this
manual is indicated by the date given
aside.
Sartorius reserves the right to make
changes to the technology, features,
specifications and design of the
equipment without notice.
All trademarks are Sartorius property
unless otherwise stated. Patents
granted or pending.
Table of contents
Other Sartorius Test Equipment manuals
Popular Test Equipment manuals by other brands

Redtech
Redtech TRAILERteck T05 user manual

Venmar
Venmar AVS Constructo 1.0 HRV user guide

Test Instrument Solutions
Test Instrument Solutions SafetyPAT operating manual

Hanna Instruments
Hanna Instruments HI 38078 instruction manual

Kistler
Kistler 5495C Series instruction manual

Waygate Technologies
Waygate Technologies DM5E Basic quick start guide

StoneL
StoneL DeviceNet CK464002A manual

Seica
Seica RAPID 220 Site preparation guide

Kingfisher
Kingfisher KI7400 Series Training manual

Kurth Electronic
Kurth Electronic CCTS-03 operating manual

SMART
SMART KANAAD SBT XTREME 3G Series user manual

Agilent Technologies
Agilent Technologies BERT Serial Getting started