Sealey TA302 User manual

DIGITAL AUTOMOTIVE ANALYSER 12
FUNCTION WITH IC
MODEL NO: TA302
Thank you for purchasing a Sealey product. Manufactured to a high standard, this product will, if used according to these
instructions, and properly maintained, give you years of trouble free performance.
IMPORTANT: PLEASE READ THESE INSTRUCTIONS CAREFULLY. NOTE THE SAFE OPERATIONAL REQUIREMENTS, WARNINGS & CAUTIONS. USE
THE PRODUCT CORRECTLY AND WITH CARE FOR THE PURPOSE FOR WHICH IT IS INTENDED. FAILURE TO DO SO MAY CAUSE DAMAGE AND/OR
PERSONAL INJURY AND WILL INVALIDATE THE WARRANTY. KEEP THESE INSTRUCTIONS SAFE FOR FUTURE USE.
1. SAFETY
1.1. PERSONAL PRECAUTIONS
9When using this multimeter, please observe all normal safety rules concerning:
Protection against the dangers of electrical current.
Protection of the meter against misuse.
9Full compliance with safety standards can only be guaranteed if used with the test probes supplied. If necessary, they must be replaced
with genuine Sealey leads with the same electronic ratings. Failure to do so will invalidate the warranty.
8DO NOT use leads if damaged or if the wire is bared in any way.
1.2. GENERAL SAFETY INSTRUCTIONS
9Familiarise yourself with the application and limitations of the multimeter as well as the potential hazards. IF IN ANY DOUBT CONSULT
A QUALIFIED ELECTRICIAN.
9Before commencing testing, follow instructions below and select the correct input sockets, function and range on the multimeter.
9When the meter is connected to a circuit, DO NOT touch any unused meter terminals.
9When the magnitude of the value to be measured is unknown beforehand, set the range selector to the highest value available.
9Before rotating the range selector to change functions, disconnect test probes from the circuit under test.
WARNING! Never perform resistance, transistor, diode or continuity measurements on live circuits.
9Always take care when working with voltages above 60V dc or 30V ac rms. These voltages are considered harmful.
WARNING! USE EXTREME CAUTION when working with high voltages.
9Always keep fingers behind the probe barriers whilst measuring and DO NOT use when hands are wet..
9Before attempting to insert transistors for testing, ensure that the test probes have been disconnected.
9Components should not be connected to the transistor socket when taking voltage measurements with the test probes.
8DO NOT test voltages above 600V ac or 1000V dc - the circuitry of the multimeter will be destroyed.
WARNING! NEVER connect the multimeter to a voltage source / live circuit when the rotary switch is set to any other function apart
from Voltage testing.
9ALWAYS discharge filter capacitors in power supplies and disconnect the power when making resistance or diode tests.
WARNING! Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the recessed
electrical contacts. Other means should be used to ensure that the terminals are not “live”.
8DO NOT use the multimeter in a potentially explosive atmosphere or where flammable material is present.
9ONLY operate the multimeter when the back cover is in place and fastened securely.
9If any abnormal readings are observed, the multimeter must be checked out by an authorised technician.
9ALWAYS turn off the multimeter and disconnect the test probes, before opening the back cover to replace the fuse or battery.
9When not in use, store the multimeter carefully in a safe, dry, childproof location out of direct sunlight. If storing for a long period of time,
remove the battery. Storage temperature range: -20°C to 60°C.
Note: The warnings, cautions and instructions referred to in this manual cannot cover all possible conditions and situations that may occur. It
must be understood that common sense and caution are factors which cannot be built into this product, but must be applied by the
operator.
2. INTRODUCTION
New generation, 12-function, auto-ranging automotive diagnostic multimeter. Combination digital/bar-graph display gives accurate
indication of component outputs. Large, easy to read high contrast display with bright, white backlight. Workshop-tough, durable
bi-composite case with integral stand. Reads standard automotive parameters including duty cycle. Features auto-ranging, and data hold
functions with overload protection on all ranges. Supplied with inductive coupler, test probes and thermocouple.
TA302 Issue 3 (HF) 02/10/18
Original Language Version
© Jack Sealey Limited
Refer to
instructions
Electrical shock
hazard
Warning

3. FEATURES
4. SPECIFICATION
Input Limits
Function Terminal Input Limit
ac volts V - Ω - rpm
600V ac rms
dc Volts 1000V dc
Frequency V - Ω - rpm 500V ac/dc
Ohm (resistance) V - Ω - rpm 250V ac/dc
Diode
ac/dc mA mA 320mA ac/dc
ac/dc 20A 20A *20A ac/dc
RPM
V - Ω - rpm 500V ac/dcDuty Cycle (%)
Dwell Angle
Contents/Layout:
1. LCD display
2. Alternate Function
3. Dwell
4. RPM Mode Button - Stroke
(4/2(DIS)
5. Rotary selector Switch
6. 20A Terminal
7. mA Terminal
8. Range Button
9. Backlight
10. Hold Button
11. Volts,
Ω, Hz, %, RPM,DWELL and
TEMP
Terminal
12. COM Terminal
13. Inductive Coupler
14. Thermocouple
15. Test Probes
g.1
WARNING! DO NOT make current
measurements between 1A and 20A
for longer than 30 seconds in every 15
minutes. Exceeding 30 seconds may
cause damage to the meter and test
probes.
WARNING! Ohms cannot be measured if
a voltage is present. Only measure in non
powered circuits.
DC Voltage (Auto Ranging)
Range Resolution Accuracy
32mV 0.1mV
± 1.2% of rdg +1 dgt
3.2V 1mV
32V 10mV
320V 100mV
1000V 1V
Input Impedance: 10MΩ.
Maximum Input: 1000Vdc or 750Vac rms
AC Voltage (Auto Ranging)
Range Resolution Accuracy
3.2V 0.1mV
± 2.0% of rdg +4 dgt
32V 1mV
320V 10mV
750V 100mV
Input Impedance: 10MΩ.
Frequency Range: 50 to 60Hz
Maximum Input: 1000Vdc or 750Vac rms
Duty Cycle
Range Resolution Accuracy
1.0%~90.0% 0.1% ± 2.0% of rdg +5 dgt
Pulse Width: >100us, <100ms
Overload Protection: 500Vdc or RMS ac
Dwell Angle
No. Cyl Range Resolution Accuracy
4 0~90°
0.1° ± 2.0% of rdg +5 dgt
5 0~72°
6 0~60°
8 0~45°
Overload Protection: 500Vdc or RMS ac
Temperature
Range Resolution Accuracy
-20~750°C 0.1°C ± 3.0% of rdg +2°C
-4~1400°F 0.1°F ± 3.0% of rdg +4°F
Sensor: Type K Thermocouple
Input Protection: 250Vdc or Vac rms
Current
Range Resolution Accuracy
32mA 0.1mA DC ± 2.0% of rdg +1dgt
AC ± 2.5% of rdg +4dgt
320mA 1mA DC ± 2.0% of rdg +1dgt
AC ± 2.5% of rdg +4dgt
10A 10mA DC ± 3.0% of rdg +3dgt
AC ± 3.5% of rdg +4dgt
Frequency Response: 50 to 60Hz
Input Protection: Fast acting ceramic fuse on mA and 10A input
Resistance (Auto Ranging)
Range Resolution Accuracy
320Ω0.1Ω
± 1.5% of rdg +3 dgt3.2kΩ1Ω
32kΩ10Ω
3.2MΩ1KΩ± 2.5% of rdg +3 dgt
32MΩ10KΩ± 5.0% of rdg +5 dgt
Overload Protection: 250Vdc or RMS ac
Frequency
Range Resolution Accuracy
320Hz 0.1Hz
± 1.0% of rdg +4 dgt
3200Hz 1Hz
32KHz 10Hz
Sensitivity: 3.5V rms min.
Overload Protection: 500Vdc or rms ac
Diode Test
Range Resolution Accuracy
0.6mA typical (Vf=0.6V) 1mA ± 10% of rdg +3 dgt
Open Circuit Voltage: 3Vdc Typical
Overload Protection: 250Vdc or rms ac
TA302 Issue 3 (HF) 02/10/18
Original Language Version
© Jack Sealey Limited

5. OPERATION
WARNING! Ensure that you read, understand and apply the safety and operational instructions before connecting the meter. Only
when you are sure that you understand the procedures is it safe to proceed with testing.
WARNING! Risk of electrocution. High voltage circuits, both ac and dc are very dangerous and should be measured with great care.
Operating temperature range 0°C to 40°C.
Remember to turn on meter before use and to turn it off when measurement is completed.
NOTE: IF “OL” appears in the display during a measurement, the value exceeds the range you have selected. Change to a higher range.
NOTE: On some low ac and dc ranges, with the test probes not connected to a device, the reading may show a random fluctuating reading.
This is normal and is caused by the high input sensitivity. The reading will stabilise and give a proper measurement when connected to a circuit.
5.1. ALTERNATE FUNCTION BUTTON (FIG.1.2)
5.1.1. Press the Alternate Function button to toggle between ac and dc in the voltage & current measurements. Press the Alternate
Function button to toggle between Continuity and Diode Test.
5.2. DWELL BUTTON (FIG.1.3)
5.2.1. To manually scroll through to the correct number of cylinders press the Dwell button.
5.3. DATA HOLD, (FIG.1.10)
5.3.1. The data hold function allows the meter to freeze a measurement reading for later reference.
5.3.2. Press the data hold button once to freeze the reading in the display. The indicator “hold” will appear in
the display.
5.3.3. Press the data hold button again to return to normal operation.
5.4. RANGE BUTTON (FIG.1.8)
5.4.1. The range is automatically selected by the meter.
5.4.2. To manually select a range within a function, press the range button.
5.4.3. To exit the range mode and return to autoranging, press and hold the range button for two seconds.
NOTE: If the range is too high, the meter will be less accurate.
If the range is too low, the meter displays ‘OL’ (Over Limit).
5.5. RPM MODE BUTTON (FIG.1.4)
5.5.1. Press rpm mode button to toggle between 4 stroke and 2 stroke / DIS (Distributerless Ignition
System).
5.5.2. Press and hold for two seconds to toggle between Negative (-) and Positive (+) Trigger slope.
5.5.3. Press the button repeatedly to adjust the trigger level if the meter reading is to high or unstable.
NOTE: The Trigger level has five steps and is different for each function combination.
5.6. BACKLIGHT BUTTON (FIG.1.9)
5.6.1. To turn the backlight on press the backlight button, to turn off press the backlight button again.
5.7. AC OR DC VOLTAGE MEASUREMENTS
5.7.1. Insert the black test probe into the negative “COM” terminal and the red test probe into the positive main positive terminal (fig.1.10)
terminal.
5.7.2. Turn the rotary switch to the ‘V’ position.
5.7.3. Press the Alternate Function button to select ac or dc voltage.
5.7.4. Touch the test probes to the circuit under test and read the voltage display.
g.2
Audible Continuity
Audible Threshold Resolution Test Current
<35Ω100mΩ<0.7mA
Input Protection: 250Vdc or Vac rms
RPM (Tach)
Range Resolution Accuracy
rpm 4
600-3200 1rpm
± 2.0% of rdg +4
dgt
6000~12000 10rpm
rpm 2/DIS 300~3200 1rpm
3000~6000 10rpm
Eective Reading: >600rpm
Overload Protection: 500Vdc or RMS ac
Display:..................3.5 digit (3200 counts) LCD display with function and unit sign annunciator.
Analogue Bar Graph:.......34 segments with measurements 12 times per second.
Polarity:..................Automatic, (-) negative polarity indication.
Over Range Indication:.....“OL” mark indication.
Low Battery Indication:.....The is displayed when the battery voltage drops below the operating level.
Measurement Rate:........2 times per second, nominal.
Operating Environment:....0°C to 50°C (32°F to 122°F) at <70% relative humidity.
Storage Environment:......-20°C to 60°C (-4°F to 140°F) at <80% relative humidity.
Temperature Coefficient:....0.2 x (specified accuracy) / °C (<18°C or >28°C).
Power:...................Single standard 9 Volt battery (PP3).
Battery Life:...............200 hours typical with alkaline battery.
Fuse:....................20A/250V, 5.5 x 32mm fast acting ceramic type.
....................0.5A/250V, 5 x 20mm fast acting ceramic type.
Dimensions:..............167 (H) x 79 (W) x 50 (D) mm.
Weight Approx:............366g.
Accuracy is given at 18°C to 28°C (65°F to 83°F) less than 70% relative humidity.
TA302 Issue 3 (HF) 02/10/18
Original Language Version
© Jack Sealey Limited

5.8. AC OR DC CURRENT MEASUREMENTS
WARNING! DO NOT make current measurements between 1A and 20A for longer than 30 seconds in every 15 minutes.
Exceeding 30 seconds may cause damage to the meter and test probes.
5.8.1. Insert the black test probe into the into the negative “COM” terminal and the red test probe into the:
a) Positive 20A terminal for currents to 20A (fig.1.6).
b) Positive mA terminal for currents to 320mA (fig.1.7).
NOTE: If you are unsure of the current draw select the 20A terminal.
5.8.2. Turn the rotary switch to the 20A or mA position.
5.8.3. Press the Alternate Function button to select ac or dc current.
5.8.4. Touch the test probes in series (fig.2) with the circuit under test and read the current on the display.
5.9. RESISTANCE
WARNING! To avoid electric shock, disconnect power to unit under test and discharge all capacitors before taking any
resistance measurements.
WARNING! Ohms cannot be measured if a voltage is present. Only measure in non powered circuits.
5.9.1. Insert the black test probe into the negative “COM” terminal and the red test probe into the positive main positive terminal (fig.1.10)
terminal.
5.9.2. Turn the rotary switch to the Ω position.
5.9.3. Connect the test probes to the two ends of the Resistance circuit to be measured.
5.9.4. Read the measured value from the display.
5.9.5. In resistance measurements, if greater accuracy is required press the Range button.
5.10. DIODE / CONTINUITY MEASUREMENTS
WARNING! To avoid electric shock, disconnect power to unit under test and discharge all capacitors before taking any
resistance measurements.
5.10.1. Insert the black test probe into the negative “COM” terminal and the red test probe into the positive main positive terminal (fig.1.10)
terminal.
5.10.2. Turn the rotary switch to the position.
5.10.3. Press the Alternate Function button (fig.1.2) to select Diode or Continuity.
5.10.4. Connect the test probes to the two ends of the Diode or Continuity circuit to be measured.
5.10.5. Read the measured value from the display.
5.10.6. When measuring the forward voltage across a good Diode, it will indicate 0.4V or 0.7V and the reverse voltage will indicate “OL”
(same as on open condition). For a short circuit diode, a value of 0mV will be displayed.
5.10.7. In continuity mode a complete circuit will beep continuously, if open circuit, there will be no beep.
5.11. FREQUENCY (HZ)
5.11.1. Insert the black test probe into the negative “COM” terminal and the red test probe into the positive main positive terminal (fig.1.10)
terminal.
5.11.2. Turn the rotary switch to either of the “Hz” positions (320Hz, 3200Hz, 32kHz).
5.11.3. Connect the negative “COM” test probe to ground.
5.11.4. Connect the positive main positive terminal (fig.1.10) test probe to the “signal out” wire of the sensor to be tested.
5.12. DUTY CYCLE (%)
5.12.1. Insert the black test probe into the negative “COM” terminal and the red test probe into the positive main positive terminal (fig.1.10)
terminal.
5.12.2. Select the %DUTY range with the rotary switch.
5.12.3. Connect the negative test probe to ground.
5.12.4. Connect the positive test probe to the signal wire circuit.
5.13. TEMPERATURE MEASUREMENTS
5.13.1. Insert the type K thermocouple plug into the negative “COM” terminal and the positive terminal ensuring the + symbol on the plug is
inserted into the positive + terminal and the negative symbol on the plug is inserted into the negative “COM” terminal.
5.13.2. Turn the rotary switch to the select °C or °F.
5.13.3. Read the temperature on the display.
5.14. RPM (TACH) MEASUREMENTS
5.14.1. Select the rpm range with the rotary switch.
5.14.2. Select the x10 rpm range with the rotary switch. Multiply the displayed reading times by 10 to get the actual rpm.
5.14.3. Press the rpm mode button (fig.1.4) to select through rpm 4 for 4-stroke or rpm 2 for 2-stroke DIS.
NOTE: RPM4 for RPM of 4-stroke engines which have 1 ignition on every 4 engine strokes.
RPM2 for RPM of DIS (Distributerless Ignition System) & 2-Stroke engines which have 1 ignition on every 2 engine strokes.
5.14.4. Insert the inductive pick-up leads into the meter. Black lead into the negative “COM” terminal and the red lead into the positive RPM
terminal.
5.14.5. Connect the inductive pick-up to a spark plug HT lead. If no reading is received, unhook the clamp, turn it over and connect again.
NOTE: Connect the pick-up as far away from the distributor and exhaust manifold as possible.
Position the pickup to within six inches of the spark plug or move it to another plug HT lead if no reading or an erratic reading is
obtained. The inductive pick-up has an adjustable sensitivity switch that may also be used to correct an unstable reading.
TA302 Issue 3 (HF) 02/10/18
Original Language Version
© Jack Sealey Limited

5.15. DWELL ANGLE MEASUREMENT
Dwell angle is the number of degrees through which the distributor cam rotates while the breaker points are closed.
5.15.1. Insert the black test probe into the negative “COM” terminal and the red test probe into the positive “V” terminal.
5.15.2. Turn the rotary switch to the DWELL position.
5.15.3. Set number of cylinders with the DWELL button (fig.1.3).
5.15.4. Connect the black test probe to the Ground terminal (-) on the vehicle battery and the red test probe to the contact breaker points or the
negative (-) terminal of the ignition coil.
5.15.5. When the engine is started the Dwell angle will be displayed.
NOTE: To reduce the dwell angle reading the points gap must be increased, to increase the dwell angle the points gap must be reduced.
Refer to your owners handbook for detailed procedures for dwell settings and adjustments.
6. MAINTENANCE
WARNING! DO NOT attempt to repair or service your meter unless you are qualified to do so and have the relevant calibration,
performance test, and service information. To avoid electrical shock or damage to the meter DO NOT get water inside the case.
6.1. REPLACING THE BATTERY
WARNING! To avoid electric shock, disconnect the test probes from any source of voltage before removing the battery door.
6.1.1. When the battery becomes exhausted or drops below the operating voltage, will appear in the right hand side of the display. Replace
the battery.
6.1.2. Disconnect the leads from the meter.
6.1.3. Open the battery door by loosening the screws using a Phillip’s screwdriver.
6.1.4. Remove the old battery and insert the new one, observing the correct polarity.
6.1.5. Replace the battery cover and secure with the two screws.
WARNING! To avoid electric shock, DO NOT operate the meter until the battery cover is secured in place.
6.2. REPLACING THE FUSES
WARNING! To avoid electric shock, disconnect the test probes from any source of voltage before accessing the fuses.
6.2.1. Open the battery door by loosening the screws using a Phillip’s screwdriver. Remove the battery.
6.2.2. Remove the six screws around the back of the meter using a Phillip’s screwdriver.
6.2.3. Gently part the casing then remove the blown fuse. Be careful not to damage the PCB.
6.2.4. Install the new fuse into its holder. Join the casing back together and secure with screws.
Note: Always use a fuse of the correct size and value.
0.5A/250V fast blow for the 320mA range.
20A/250V fast blow for the 20A range.
6.2.5. Replace the battery and cover and secure with the screws.
WARNING! To avoid electric shock, DO NOT use the meter until it has been fully re-assembled.
6.3. Periodically wipe the case with a damp cloth and mild detergent. DO NOT use solvents.
6.4. Turn the meter off when not in use and remove the battery if stored for a long period of time.
6.5. DO NOT store the meter in a place of high humidity or high temperature.
Sealey Group, Kempson Way, Suffolk Business Park, Bury St Edmunds, Suffolk. IP32 7AR
01284 757500 01284 703534 sales@sealey.co.uk www.sealey.co.uk
ENVIRONMENT PROTECTION
Recycle unwanted materials instead of disposing of them as waste. All tools, accessories and packaging should be sorted, taken to
a recycling centre and disposed of in a manner which is compatible with the environment. When the product becomes completely
unserviceable and requires disposal, drain any fluids (if applicable) into approved containers and dispose of the product and fluids
according to local regulations.
WEEE REGULATIONS
Dispose of this product at the end of its working life in compliance with the EU Directive on Waste Electrical and Electronic Equipment
(WEEE). When the product is no longer required, it must be disposed of in an environmentally protective way. Contact your local solid
waste authority for recycling information.
BATTERY REMOVAL
Under the Waste Batteries and Accumulators Regulations 2009, Jack Sealey Ltd are required to inform potential purchasers of products
containing batteries (as defined within these regulations), that they are registered with Valpak’s registered compliance scheme. Jack
Sealey Ltd Batteries Producer Registration Number (BPRN) is BPRN00705.
Note: It is our policy to continually improve products and as such we reserve the right to alter data, specifications and component parts without prior
notice.
Important: No Liability is accepted for incorrect use of this product.
Warranty: Guarantee is 12 months from purchase date, proof of which is required for any claim.
TA302 Issue 3 (HF) 02/10/18
Original Language Version
© Jack Sealey Limited
Table of contents
Other Sealey Measuring Instrument manuals

Sealey
Sealey TA303 User manual

Sealey
Sealey AK9636D.V2 User manual

Sealey
Sealey ak4565D User manual

Sealey
Sealey SA37/93.V3 User manual

Sealey
Sealey TA201 User manual

Sealey
Sealey VSE9540 User manual

Sealey
Sealey DBG508 User manual

Sealey
Sealey VS925.V2 User manual

Sealey
Sealey VSE5863 User manual

Sealey
Sealey MM404.261 User manual

Sealey
Sealey Professional AUTO SERVICE TA100 User manual

Sealey
Sealey TA101 User manual

Sealey
Sealey VS0563 User manual

Sealey
Sealey MM19.V3 User manual

Sealey
Sealey VS0567 User manual

Sealey
Sealey VS0052.V2 User manual

Sealey
Sealey TA070 User manual

Sealey
Sealey TA203.V2 User manual

Sealey
Sealey AK9637D.V2 User manual

Sealey
Sealey TM105 User manual