Sentry Sample Sequencer 6 Instruction manual

Original Instructions
Installation, Operation
&Maintenance Manual
Sentry Sample Sequencer 6 Controller
Sample Conditioning Automation
S-SW-IOM-00375-0 04-23

2 Sentry Equipment Corp
Do not install, maintain, or operate this equipment without reading, understanding, and following the
appropriate Sentry Equipment Corp instructions. Otherwise, injury, damage, or both may result.
Copyright
© 2022 by Sentry Equipment Corp. All rights reserved. All product and company names are property of their respective
owners. This document contains proprietary information. No part of this document may be photocopied or reproduced
without the prior written consent of Sentry Equipment Corp.
Limit of Liability
Sentry Equipment Corp, its employees, agents, and the authors and contributors to this document specically disclaim
all liabilities and warranties, express or implied (including warranties of merchantability and tness for a particular
purpose), for the accuracy, currency, completeness, and/or reliability of the information contained herein and/or for the
tness for any particular use and/or for the performance of any material and/or equipment selected in whole or part
with the user of/or in reliance upon information contained herein. Selection of materials and/or equipment is at the
sole risk of the user of this publication.
Note
The information contained in this document is subject to change without notice.
Do not install, maintain, or operate this equipment without reading, understanding, and following the
appropriate Sentry Equipment Corp instructions. Otherwise, injury, damage, or both may result.
Copyright
© 2022 by Sentry Equipment Corp. All rights reserved. All product and company names are property of their respective
owners. This document contains proprietary information. No part of this document may be photocopied or reproduced
without the prior written consent of Sentry Equipment Corp.
Limit of Liability
Sentry Equipment Corp, its employees, agents, and the authors and contributors to this document specically disclaim
all liabilities and warranties, express or implied (including warranties of merchantability and tness for a particular
purpose), for the accuracy, currency, completeness, and/or reliability of the information contained herein and/or for the
tness for any particular use and/or for the performance of any material and/or equipment selected in whole or part
with the user of/or in reliance upon information contained herein. Selection of materials and/or equipment is at the
sole risk of the user of this publication.
Note
The information contained in this document is subject to change without notice.
Compliance Statement (Part 15.19)
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference that may cause undesired operation.
Warning (Part 15.21)
Changes or modications not expressly approved by the party responsible for compliance could void the user’s
authority to operate the equipment.
FCC Interference Statement (Part 15.105 (b))
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of
the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential
installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used
in accordance with the instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference
to radio or television reception, which can be determined by turning the equipment o and on, the user is encouraged
to try to correct the interference by one of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit dierent from that to which the receiver is connected.
• Consult the manufacturer for help.

Sentry Sample Sequencer 6 Controller 3
Table of Contents
Safety Information ..........................................................5
General Safety Precautions ..................................................5
Specications ...............................................................6
General Description .........................................................7
Modes of Operation ......................................................................8
Inputs and Outputs .......................................................................9
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Unpacking ...............................................................................11
Mounting ................................................................................11
Electrical Connections ....................................................................15
Operation...................................................................20
Operating LEDs...........................................................................20
Wireless ..................................................................................21
Operation Screen .........................................................................22
Conguration ............................................................................24
Operating Modes.........................................................................26
Valve Outputs ............................................................................29
Maintenance................................................................29
Cleaning .................................................................................29
Disassembly..............................................................................29
Troubleshooting Guide ......................................................30
List of Components..........................................................31
Accessories .................................................................31
Standard Warranty ..........................................................32
Customer Support...........................................................32

4 Sentry Equipment Corp
Appendix A: Recommended Piping...........................................33
Appendix B: Using a Sequencer with Hach Series 5000 Analyzer ...............33
Conguring the Series 5000 (Catalog No. 60000-XX through 60004-XX) for MARK END OF
MEASURE on Relay 4......................................................................34
Older Hach Series 5000 Analyzers .........................................................35
Model 5500X-XX Conguration ...........................................................35
Appendix C: Typical Continuous Analyzer Wiring,Reference
Orion 2117LL, 2117HL, 2118XP, 2111LL, and 2111XP ..........................37
Orion 2117LL, 2117HL, 2118XP, 2111LL, and 2111XP .......................................37
Thornton 2800Si Silica Analyzer ...........................................................38
Appendix D: Connection to PLC that requires a contact closure for point
identication ...............................................................39
Appendix E: Conguration of Web Panel......................................40
IP Parameters.............................................................................40
Web Parameters ..........................................................................41
Startup Parameters .......................................................................41
Save and Exit .............................................................................42

Sentry Sample Sequencer 6 Controller 5
Safety Information
Please read the entire manual before attempting to unpack, set up, or operate this product. Pay careful attention to all
Warnings, Cautions, and Notes. Failure to do so could result in serious personal injury and/or equipment damage.
Use of Hazard Information
If multiple hazards exist, the signal word corresponding to the greatest hazard shall be used.
Denitions
DANGER indicates a hazardous situation which, if not
avoided, will result in death or serious injury. WARNING indicates a hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION, used with the safety alert symbol, indicates a
hazardous situation which, if not avoided, could result in
minor or moderate injury.
NOTICE is used to address practices not related to personal
injury.
NOTE
Information that requires special emphasis.
TIP
Alternate techniques or clarifying information.
SHALL:
This word is understood to be mandatory.
SHOULD:
This word is understood to be advisory.
General Safety Precautions
Product Selection, Installation, and Use
Improper selection, installation, or use can cause personal injury or property damage. It is solely the responsibility of users,
through their own analysis and testing, to select products suitable for their specic application requirements, ensure they are
properly maintained, and limit their use to their intended purpose.
Follow proper local, state, and federal regulations for proper installation and operational requirements.
Always use caution and common sense when working with any chemical. Read the product label and Material Safety Data
Sheets (MSDS) carefully and follow the instructions exactly.
Potential Equipment Hazards
Hot surfaces! This equipment may have very hot surfaces. If an operator contacts a hot surface, injury may occur. Use
protective clothing to prevent injury. If other equipment comes in contact with a hot surface, damage to the equipment may
occur. Ensure the area around this equipment is kept clear to prevent damage from occurring.
High pressures! This equipment may contain uids at very high pressures. Prior to installing, removing or maintaining this
equipment, ensure that the equipment is isolated from all connecting piping, the equipment is depressurized, the contents
have been drained, and the equipment is cool.

6 Sentry Equipment Corp
Specications
Specications are subject to change without notice.
Applications: The Sentry® Sample Sequencer® 6 is designed to support sample stream switching in Batch or
Continuous modes.
*Solenoid voltage must match DC voltage
SPECIFICATIONS
Standard (7-06362C) (12V)
(7-06362K) (24V)
IO Extender 4 Point (7-06362D) (12V)
(7-06362P) (24V)
IO Extender 8 Point (7-06362E) (12V)
(7-06362N) (24V)
samples 848
analyzers 212
AC input
power (24V)
0.52A/115VAC, 0.26A/230VAC, 0.21A/277VAC Frequency: 50-60 Hz
AC input
power (12V)
1.5A/115VAC, 0.9A/230VAC, 0.75A/277VAC Frequency: 50-60 Hz
DC power
(24V)
1.9A/24VDC
DC power
(12V)
3.8A/12VDC
analyzer
inputs
Two (2) AI: 4-20mA (154.5 Ohm load)
+- 1% full scale (max)
6 Digital Inputs for end-of-analysis
indication and system alarm indication
max 28VDC
One (1) AI: 4-20mA (154.5 Ohm load)
+- 1% full scale (max)
3 Digital Inputs for end-of-analysis
indication and system alarm indication
max 28VDC
Two (2) AI: 4-20mA (154.5 Ohm load)
+- 1% full scale (max)
6 Digital Inputs for end-of-analysis
indication and system alarm indication
max 28VDC
outputs DO: 16
AO: 8 (4-20mA; 550 Ohm max load) +-
1% full scale (max)
SOL Out: 8 (12VDC) handling up to 2A*
DO: 8
AO: 4 (4-20mA; 550 Ohm max load) +-
1% full scale (max)
SOL Out: 4 (12VDC) handling up to 2A*
DO: 16
AO: 8 (4-20mA; 550 Ohm max load) +-
1% full scale (max)
SOL Out: 8 (12VDC) handling up to 2A*
resolution AI:5.37uA/LSB AO:5.306761 uA/LSB
user
interface
Wireless 802.11b/g/n, Ethernet N/A
enclosure IP 66
dimensions 8.88" H x 12.36" W x 4.72" D (226 mm H x 314 mm W x 196 mm D)
mounting surface
ambient
temperature
23 to 113 deg F (-5 -to 45 deg C), 5-95% Relative Humidity
shipping
weight
5 lbs (2.3 kg)
agency
approvals
(Approvals still pending) CE, FCC, Intertek, RoHS
SPECIFICATIONS
Touch Screen with Power Supply (7-06362M) Touch Screen without Power Supply (7-06362L)
AC input power 1.5A/115VAC, 0.9A/230VAC, 0.75A/277VAC Frequency:
50-60 Hz
-
DC power - 0.43A/24VDC
user interface 10.1" Touchscreen
enclosure IP 65
dimensions 14.3" H x 11.81" W x 5" D (363 mm H x 300 mm W x 127 mm D)
mounting surface
ambient temperature -20 to 70 deg C, 5-90% Relative Humidity
shipping weight 7.5 lbs (3.4 kg)
agency approvals CE, UL

Sentry Sample Sequencer 6 Controller 7
Environmental: Indoor use only
Operating Altitude: 0–2000 m (0–6562 ft) (Consult factory for other altitudes.)
Mains supply uctuation: -10% and +10%
Transient overvoltage: 1500 VAC
Pollution degree: 2
Shipping Weight: 5 lb (2.27 kg)
General Description
The Sentry Sample Sequencer 6 reduces the number of required analyzers while maintaining equivalent data accuracy.
Electronically switches up to eight sample streams, one at a time, to a shared analyzer and provides signals to
recorders.
If eight sample streams are not needed for one process, four to eight points may be assigned to one.
Track and hold outputs provide independent outputs for each sample point to remote data acquisition devices
(DCS, recorder, etc.).
Digital outputs provide channel number indication and/or replicate analyzer alarms for remote data.
Measures the analyzer analog output signal and converts the analog signal to digital data.

8 Sentry Equipment Corp
Modes of Operation
The Sample Sequencer 6 has two programmable modes of operation:
Continuous mode is used for analyzers that continuously produce an analytical result.
Batch mode is used for analyzers that take a sample“bite”(or batch), produce an analytical result, and then provide
a contact signal output, typically in 1-20 minutes.
Figure 1 illustrate typical system arrangements for sample stream switching with the Sample Sequencer 6. Customer-
supplied equipment is shown for clarication.
Figure 1. Sample Sequencer 6 with Continuous and Batch Analyzers

Sentry Sample Sequencer 6 Controller 9
ÂNOTE
See Appendix C on page 37 for interface protocols and wiring for a number of typical analyzers. If using the Sample
Sequencer 6 with a Hach analyzer, see Appendix B on page 33.
Inputs and Outputs
A variety of inputs and outputs are available on the Sample Sequencer 6. The Sample Sequencer 6 is a modular
design utilizing dedicated IO boards. Up to two IO boards can be tted into each enclosure. Up to eight IO boards
can be equipped with one Sample Sequencer 6. This solution would utilize 4 enclosures, one with the CPU and
communication board and 3 IO extender enclosures. In general one IO board corresponds to one four output manifold.
Analog Inputs
Up to two analog inputs are located in the sequencer enclosure for reading 4–20 mA signals from up to two dierent
analyzers.
The sequencer converts 4–20 mA signals into engineering units for display, and outputs the same signals for the
track and hold analog outputs.
Analyzer Alarm Inputs
Alarm inputs are typically used to indicate a loss of sample ow to the analyzer.
Two analyzer alarm inputs are located in the sequencer enclosure.
Alarm inputs cause the digital alarm output for that sequencer to be set.
In Continuous mode:
1. An alarm immediately closes the active sample valve and opens the next sample valve.
2. The next sample valve remains open until one of the following occurs:
a. the alarm clears
b. one minute elapses
c. the timer expires
3. After the sample valve closes, the sequencer opens the next sample valve.
4. The sequencer continues trying to establish ow until the analyzer alarm clears.
In Batch mode:
1. An alarm starts a timeout timer. (The operator can adjust the timeout timer for all samples simultaneously by
setting the sample timer.)
2. If the sequencer does not receive an end-of-batch indication from the analyzer, or if the alarm does not clear
before the timer expires, the sequencer closes the current sample valve and opens the next valve in the
sequence.
3. If the end-of-batch indication is received before the timer expires, the sequencer assumes the alarm is not a loss
of ow and continues with normal operation.

10 Sentry Equipment Corp
Calibration Hold and End-of-Batch Inputs
A calibration hold input indicates that the analyzer is performing a calibration.
Two calibration hold inputs are located in the sequencer enclosure.
In Continuous mode, when a calibration hold input is received:
– The sequencer waits at the active sample with the valve open.
– Once the calibration hold clears, the sequencer resumes operation, entering the Flush state for the sample that
was active before the calibration hold was received.
In Batch mode, calibration hold inputs do not perform a function.
Two end-of-batch inputs are located in the sequencer enclosure.
Without this input connected, the Sample Sequencer 6 does not advance to the next sample when the sequencer is
in Batch mode.
End-of-batch inputs are not used in Continuous mode.
Valve Outputs
Valve outputs supply 12 or 24 VDC to operate the sample valves.
Up to eight valve outputs are located in the sequencer enclosure.
The eight valves may be divided between two analyzers sequentially.
Track and Hold Outputs
Eight analog outputs (4–20 mA) are provided in the sequencer enclosure.
When a valid reading is taken from the analyzer for a given sample, the Sample Sequencer 6 tracks (updates) the
corresponding signal on the appropriate output.
When the sample is not currently being analyzed, the sequencer holds the corresponding signal on the appropriate
output at its last known value.
Digital Outputs
A total of eight digital outputs are available per IO board:
Four outputs are used to replicate the four analyzer alarm inputs
Four outputs are used for sample point indication.
An LED above the output indicates that the point is currently being sampled and the analyzer reading corresponds
to that sample point.
Ethernet Communications
The Sample Sequencer 6 is capable of connecting to a network for communicating to a PC or dedicate terminal
screen via a webpage.
Modbus communication protocol is available on request. Please contact Sentry for further information.

Sentry Sample Sequencer 6 Controller 11
For more information on these features, refer to Appendix D on page 39.
Optional Manifold Switching Valve
The Sample Sequencer 6 may be used to control most manifold valve arrangements; however, the Sentry Manifold
Switching Valve is a convenient optional accessory for obtaining representative samples.
The patented Sentry Manifold Switching Valve
incorporates block and bleed design
provides continuously owing samples
eliminates areas where cross contamination can occur
available in either four or eight sample line models
Contact Sentry Equipment for more information about this optional sampling valve.
Installation
This instrument should be installed by qualied technical personnel to ensure adherence to all applicable electrical codes.
To ensure the protection provided by this equipment is not impaired, this equipment must not be installed or used in any
manner other than that which is specied in this manual.
Unpacking
1. Carefully remove the Sample Sequencer 6 from its shipping carton.
2. Inspect it for any damage.
3. Report any damages to the shipper at once.
Mounting
Mount the Sample Sequencer 6 close to the analyzer to permit viewing of both devices at the same time.
The sample sequencer mounts to a at surface.
Mounting method must be capable of supporting at least 5lbs (2.27kg).
Figure 2. Enclosure Dimensions: Bottom View

12 Sentry Equipment Corp
Figure 3. Enclosure Dimensions: Side View
Mounting-Display Screen
1. The Display Screen mounts to a at surface
2. Mounting method must be capable of supporting at least 5 lbs.
Mounting-IO Extender
1. The IO Extender mounts to a at surface.
2. Mounting method must be capable of supporting at least 5 lbs.
Panel Mounting
1. Turn o all power to the Sample Sequencer 6 and disconnect all cables.
2. Use the dimension information shown in Figure 8 to mark positions on the panel for four mounting holes.
3. If replacing the Sequencer 5 with a Sequencer 6, the Sequencer 6 Display Screen must sit above the bottom edge
of the Sequencer 6 panel cut-out hole, as shown in 7.
4. Drill pilot holes in the marked positions for four #10 bolts.
5. Use four #10 bolts, lock washers, washer and nuts to secure the device to the panel, as shown in Figure 8.
6. Make electrical connections as described in the“Electrical Connections”section beginning on page 15.

Sentry Sample Sequencer 6 Controller 13
Figure 4. Panel Mount Dimensions
Figure 5. Replacing Sample Sequencer 5 with Sample Sequencer 6

14 Sentry Equipment Corp
Figure 6. Panel Mounting

Sentry Sample Sequencer 6 Controller 15
Electrical Connections
Electrical Shock Hazard – All input and output wiring should be installed in such a manner that access to bare copper is
prohibited during normal operation.
IcoTek KEL-DPZ 24/26 and KEL-ER 24/10 multi-membrane cable entry plates are provided on the bottom and right
side of the enclosure for up to 26 cables with 5 - 10.2mm cable diameter.
On the left side of the enclosure, cable entries are provided for the ethernet connection.
Figure 9 (board view) and Figure 10 (schematic view) show the power connections on the board.
Power Connection
The Sample Sequencer 6 uses a universal power supply that accepts an input voltage range of 85-305 VAC and
frequency of 47-440 Hz. A power supply lead of 12-24 AWG (4-0.25 mm2) copper wire is recommended.
Connect the sample sequencer to an electrical outlet or outlet strip with surge suppression and ltering. Using this
method, the plug serves as the disconnect device.
A 12-24 AWG, 60°C, VW-1, 600V, type SJT cord is recommended for non-detachable connections.
For permanent connections, use a exible SOW cable or better. Sizing must be in accordance to the CEC.
A suitable external over-current protection device such as a fuse (Class CC, 2A, Time delay) or circuit breaker (C trip
curve, 2 A) is recommended for both the hot (L) and neutral (N) leads.
The mains and output terminal are rated within the following condition:
Mains terminals rated up to 300V, 17.5A, 12-30 AWG, PA insulating material, Inammability class V0, Type: Plug, 8
position, and screw connection.
Output terminals rated up to 300V, 17.5A, 12-30 AWG, PA insulating material, Inammability class V0, Type: Plug, 8
position, and screw connection.
For permanently connected installations:
A switch or circuit breaker shall be included in the building installation.
It shall be in close proximity to the equipment and within easy reach of the operator.
It shall be marked as the disconnecting device for the equipment.
The cord and plug shall comply with the following standards:

16 Sentry Equipment Corp
Table 1. Power Cord Requirements
Location Cord Plug
United States ANSI/UL 817 ANSI/UL 498
Canada CSA C22.2 No. 21 or C22.2 No. 49 CSA C22.2 No. 21 and C22.2 No. 42
European Community IEC 60227 or 60245 Appropriate IEC standard
If desired, the Sample Sequencer 6 may also be hard wired to instrument-quality power using appropriate certied
ttings, and wiring.
A suitable external over-current protection device, such as a fuse (Class CC, 2A, Time delay) or circuit breaker (C trip
curve, 2 A), and disconnect device is recommended.
The over-current protection and disconnect devices shall be installed on both the hot (L) and neutral (N) leads.
The disconnect device shall be located near the equipment and marked with appropriate ON(|) OFF(O) markings as
specied by local codes.
For xed wiring methods, the ground conductor shall have insulation with a continuous outer nish that is either
green, or green with one or more yellow stripes.
During connection of the power wire, the grounding lead must be suitably routed and long enough so that it is the
last conductor to take the strain.
Installation should be performed by qualied personal in accordance with local codes and procedures.
Grounding and EMC Considerations
Power
To properly ground the Sample Sequencer 6:
• Use an insulated ground conductor that has a green or green with one or more yellow stripes and is the same size
as the mains conductor, and land the conductor on the“EGND”terminal on the power board.
I/O Cables
To maintain SGS and CE ratings for EMC:
All I/O cables must be shielded and grounded.
The shield (drain) wire for the analyzer inputs must be tied to an isolated ground located near the opposite cable
end (analyzer) from the sequencer, and not connected to the sequencer.
Sequencer Board Connections
Make sure all leads are lugged prior to wiring to the terminal blocks.This avoids the potential for miswiring and placing
transients on the analyzer(s) and sequencer. Be sure to ground the shield in the analyzer as detailed in“Grounding and EMC
Considerations”on page 16.

Sentry Sample Sequencer 6 Controller 17
Connections from the Analyzer(s)
Referencing Figure 10, analyzer connections include:
4–20mA output signals
analyzer alarm (typically a loss of ow or system failure alarm)
end-of-batch signal for batch analyzers
calibration signal for continuous analyzers
Use a 22 AWG multiple twisted pair shielded cable similar to Belden 9512 for each analyzer connected. This allows the
signals to be connected from each analyzer to the sequencer with one cable (seeFigure 9).
Analog Outputs from the Sequencer (Track and Hold)
Analog outputs (Track and Hold, 4–20 mA) are created for each sample being sequenced through the analyzer(s).
Outputs provide individual sample line data acquisition for local or remote indication.
Connections may be used with a recorder or in situations where analog I/O at the receiving device is limited.
Use a 22 AWG 4 pair twisted pair shielded cable similar to Belden 9514 for wiring to each removable terminal block.
Maximum load is 400 ohms. (See“Specications”on page 6 for additional details.)
Digital Outputs from the Sequencer
The Sample Sequencer 6 has 16 sinking digital outputs to provide point indication for sample being analyzed:
eight (8) per I/O board
four (4) replicated analyzer alarms
four (4) point indication
Use a 22 AWG multi-conductor shielded cable similar to Belden 9423 for wiring to each removable terminal block.
Valve Output Connections
The valve outputs of the Sample Sequencer 6 output 12 or 24VDC, and are intended for use with the Sentry Manifold
Sample Valve. Other manifold valves may be used; however, additional hardware such as interposing relays may be
required.
Use an 18 AWG multi-conductor shielded cable similar to Belden 9409 for wiring to each solenoid valve used. Typical
valve connections are shown in Figure 10.
Summary of Wiring Recommendations
Power cable Volex 19348 or equivalent
Connections from the analyzers Belden 1033A or equivalent
Valve output connections Belden 9316 or equivalent
Analog outputs from the Sample Sequencer 6 Belden 1033A or equivalent
Digital outputs from the Sample Sequencer 6 Belden 9316 or equivalent

18 Sentry Equipment Corp
Figure 7. Sequencer board connections

Sentry Sample Sequencer 6 Controller 19
Figure 8. Electrical Wiring Diagram
JUMPERS
Analys+
Analys-
BatCal+
BatCal-
Alarm+
Alarm-
FloSns+
FloSns-
V+
V-
SO1+
SO1-
SO2+
SO2-
SO3+
SO3-
SO4+
SO4-
DO1+
DO1-
DO2+
DO2-
DO3+
DO3-
DO4+
DO4-
DO5+
DO5-
DO6+
DO6-
DO7+
DO7-
DO8+
DO8-
AN1+
AN1-
AN2+
AN2-
AN3+
AN3-
AN4+
AN4-
V+ V-
AC / N
AC / L
EGND
CPU_P
CPU_N
GND
I/O_P
I/O_N
KC-100
SENTRY EQUIPMENT
Analys+
Analys-
BatCal+
BatCal-
Alarm+
Alarm-
FloSns+
FloSns-
V+
V-
SO1+
SO1-
SO2+
SO2-
SO3+
SO3-
SO4+
SO4-
DO1+
DO1-
DO2+
DO2-
DO3+
DO3-
DO4+
DO4-
DO5+
DO5-
DO6+
DO6-
DO7+
DO7-
DO8+
DO8-
AN1+
AN1-
AN2+
AN2-
AN3+
AN3-
AN4+
AN4-
ANALYZER 2
ANALYZER 1
SEQUENCER 6
CPU
HMI-100 (OPTIONAL)
IF1 IF2 IF3
IF4
-+
ETH ETH USB
DC
SAMPLE A
12
ANALYZER 1
INPUTS
4-20mA OUTPUTS
TO DCS
(SAMPLES A,B,C,D)
85-305 VAC
47-440 Hz
250 VA MAX
TO
SEQUENCER 6
EXTENDER
SAMPLE B
12
SAMPLE C
12
SAMPLE D
12
SOLENOID VALVE MANIFOLD 1
ANALYZER 2
INPUTS
4-20mA OUTPUTS
TO DCS
(SAMPLES E,F,G,H)
TO SOLENOID VALVE MANIFOLD 2 (NOT SHOWN)
(SAMPLES E,F,G,H)
TO
CUSTOMER
NETWORK
4-20mA
End of
Batch
Alarm
4-20mA
End of
Batch
Alarm

20 Sentry Equipment Corp
Jumpers and Addressing
All IO Boards in the Sequencer 6 have jumpers to indicate to the CPU which IO board they are. Each jumper position
presents 3 options to the board; connect top and middle pins, middle and bottom pins, or no connection. Together the
jumpers form an address for each IO board. It is important for all IO boards to be sequentially numbered. Analyzer 1
must be paired with IO board 1 which will contain the address 65. Figures below show the four jumpers on a sequencer
backplane and how they would be congured for the rst two IO boards.
Operation
To ensure the protection provided by this equipment is not impaired, this equipment must not be installed or used in any
manner other than that which is specied in this manual.
Operating LEDs
The front of the Sequencer 6 main unit houses several LEDs to indicate the status of the unit at a glance.
1. Indicates which output of the rst manifold is active.
2. Indicates which output of the second manifold is active.
NOTE: only the rst 8 points being analyzed will be indicated by the LEDs.
3. Indicates an alarm is present in the system.
4. Functions as a heartbeat to show the system is active. Will ash on and o.
5. Indicates whether the WiFi is active (ON), inactive (OFF), or in the process of becoming active (Blinking).
6. Button to turn the WiFi on and o.
This manual suits for next models
1
Table of contents
Other Sentry Controllers manuals