Sievers TOC 800 User manual

800 Series Total Organic Carbon Analyzer
Operation and Maintenance Manual
Minimum Firmware Version:
TOC Firmware Versions 3.52 (CAS) with or without
DataProAutosampler Software, or
TOC Firmware Versions 3.62 (CAS) with
DataGuardand with or without DataPro Autosampler Software, or
TOC Firmware Versions 3.15 (CMS), or
TOC Firmware Versions 2.12 (CTB) Turbo, or
TOC Firmware Version 2.11 (CBI)
Instrument Business Group
6060 Spine Road
Boulder, CO 80301 USA
phone 800.255.6964 •303.444.2009 •fax 303.444.9543 DLM 30007-08 Rev. A
www.IonicsInstruments.com Printed in USA 2004

Ionics Instruments 2004 Page i DLM 30007-08 Rev. A
Identification Records
Default Administrator Password: IIBG
Record the instrument’s serial number as it appears on the identification label located
on the left side panel of the instrument:
Instrument’s serial number
Record the date of receipt and installation of the instrument; this is the warranty start
date:
Date of receipt/installation of the instrument ______________

Ionics Instruments 2004 Page ii DLM 30007-08 Rev. A
Revision History
New editions are complete revisions of the manual and incorporate all previous update pages and
write-in instructions. This manual will be revised as necessary. Revisions can be in the form of new
editions, update pages, or write-in instructions.
Revision A and B On-Line Software Versions 0.10 – 1.17......................................... July 1992
Revision C On-Line Software Version 2.0 and updates; Autosampler
Firmware Versions 20 – 30...................................................... July 1993
Revision D and E On-Line Software Version 2.0 and updates; Autosampler
Firmware Versions 20 – 30 ...........................................September 1994
Revision F Firmware Version 3.00 (AS or MS) and updates...........................May 1995
Revision G Firmware Version 3.06 (AS or MS) and updates; RAM Card
PC Software Version 3.08RC and updates; Autosampler
PC Software Version 3.07PC and updates.......................... August 1995
Revision H Appendices B and C updates...................................................... March 1996
Revision J Appendices B and C updates.................................................... January 1997
DLM 30007-01 Rev. A TUV rating; Firmware 3.13CAS, 3.12CMS, 2.0CTB,
3.11PRA .................................................................................. July 1998
DLM 30007-02 Rev. A Year 2000 Readiness................................................................ January 1999
DLM 30007-03 Rev. A New DI Loop Maintenance............................................................June 1999
DLM 30007-04 Rev. A Firm/Software Versions 3.20CAS, 2.11CTB and 3.20PRA .........April 2000
DLM 30007-05 Rev. A Firm/Software Versions 3.30CAS, 2.12CTB 2.11CBI ...............March 2001
DLM 30007-06 Rev. A Firm/Software Versions 3.50CAS, 3.60CAS.................................May 2003
DLM 30007-06 Rev. B Formatting Adjustments................................................................. July 2003
DLM 30007-07 Rev. A Firm/Software Version 3.51CAS ................................................... July 2003
DLM 30007-08 Rev. A Firm/Software Version 3.52CAS, 3.62CAS ............................ October 2004

Ionics Instruments 2004 Page iii DLM 30007-08 Rev. A
Trademarks and Patents
Sieversis a registered trademark of Ionics Instrument Business Group; DataProand DataGuardare
trademarks of Ionics Instrument Business Group.
Teflonis a registered trademark of E.I. du Pont de Nemours and Company; Tygonis a registered
trademark of Norton Performance Plastics Corporation; Swagelokis a registered trademark of the
Swagelok Company. Mininertis a registered trademark of Valco Instruments Co. Inc. Windowsis
a registered trademark of Microsoft Corporation. Lotus 1-2-3is a registered trademark of IBM-Lotus
Corporation.
The analyzer described in this manual is covered by one or more patents issued to and owned or
pending by Ionics Instrument Business Group.
Confidentiality
The information contained in this manual may be confidential and proprietary and is the property of
Ionics Instrument Business Group. Information disclosed herein shall not be used to manufacture,
construct, or otherwise reproduce the goods disclosed herein. The information disclosed herein shall
not be disclosed to others or made public in any manner without the expressed written consent of the
Ionics Instrument Business Group.

Ionics Instruments 2004 Page iv DLM 30007-08 Rev. A
Standard Limited Warranty
Ionics Instrument Business Group warrants its products (Sievers™, Ionics™, and Leakwise™) for defects in materials and
workmanship. Ionics Instrument Business Group will, at its option, repair or replace instrument components that prove to be
defective with new or remanufactured components (i.e., equivalent to new). The warranty set forth is exclusive and no other
warranty, whether written or oral, is expressed or implied.
Warranty Term
The Ionics Instrument Business Group warranty term is thirteen (13) months ex-works, or twelve (12) months from
installation or start up by Ionics Instruments certified service personnel. In no event shall the standard limited warranty
coverage extend beyond thirteen (13) months from original shipment date.
Warranty Service
Warranty Service is provided to customers through telephone support (1-800-255-6964 or 1-888-245-2595), Monday -
Friday, from 8:00 a.m. to 5:00 p.m. (Mountain Time), excluding all company and legal holidays. Telephone support is
provided for troubleshooting and determination of parts to be shipped from Ionics Instrument Business Group to the customer
in order to return the product to operation. If telephone support is not effective, the product may be returned to Ionics
Instrument for repair or replacement. In some instances, suitable instruments may be available for short duration loan or
lease.
Ionics Instrument Business Group warrants that any labor services provided shall conform to the reasonable standards of
technical competency and performance effective at the time of delivery. All service interventions are to be reviewed and
authorized as correct and complete at the completion of the service by a customer representative, or designate. Ionics
Instrument Business Group warrants these services for 30 days after the authorization and will correct any qualifying
deficiency in labor provided that the labor service deficiency is exactly related to the originating event. No other remedy,
other than the provision of labor services, may be applicable.
Repair components (parts and materials), but not consumables, provided in the course of a repair, or purchased individually,
are warranted for 90 days ex-works for materials and workmanship. In no event will the incorporation of a warranted repair
component into an instrument extend the whole instrument’s warranty beyond its original term.
Consumables (e.g., dilution standards, verification solutions, reagents, and UV lamps, etc.) are warranted to the extent of
their stated shelf life, provided these items are maintained within the stated environmental limitations. Warranty claims for
consumables, reagents, and verification standards are limited to the replacement of the defective items, prorated from the
time of claim to the expiration of shelf life.
Shipping
A Repair Authorization Number (RA) must be obtained from the Technical Support Group before any product can be
returned to the factory. Ionics Instrument Business Group will pay freight charges, exclusive of any taxes and duties, for
replacement or repaired products shipped to the customer site. Customers shall pay freight charges, including all taxes and
duties, for all products returning to Ionics Instruments. Any product returned to the factory without an RA number will be
returned to the customer.
Limiting Conditions
The foregoing warranty shall not apply to defects resulting from improper or inadequate installation, maintenance,
adjustment, calibration, or operation by customer. Installation, maintenance, adjustment, calibration, or operation
must be performed in accordance with instructions stated in the Operation and Maintenance Manual. Usage of non-
recommended maintenance materials may void a warranty claim.
Limitation of Remedies and Liability
The remedies provided herein are the customer's sole and exclusive remedies. In no event shall Ionics Instrument Business
Group be liable for direct, indirect, special, incidental or consequential damages (including loss of profits) whether based on
contract, tort, or any other legal theory. The Operation and Maintenance Manual is believed to be accurate at the time of
publication and no responsibility is taken for any errors that may be present. In no event shall Ionics Instrument Business
Group be liable for incidental or consequential damages in connection with or arising from the use of the manual and its
accompanying related materials. Warranty is valid only for the original purchaser on the Ionics Instruments quote or invoice.
This Limited Warranty is not transferable from the original purchaser to any other party without the express written consent
from Ionics Instrument Business Group. Ionics Instrument Business Group specifically disclaims the implied warranties
of merchantability and fitness for a particular purpose.

Ionics Instruments 2004 Page v DLM 30007-08 Rev. A
Declaration of Conformity

Ionics Instruments 2004 Page vi DLM 30007-08 Rev. A
TABLE OF CONTENTS
IDENTIFICATION RECORDS.....................................................................................I
REVISION HISTORY .................................................................................................II
TRADEMARKS AND PATENTS ..............................................................................III
CONFIDENTIALITY..................................................................................................III
STANDARD LIMITED WARRANTY ........................................................................ IV
TABLE OF CONTENTS........................................................................................... VI
1INTRODUCTION.............................................................................................. 1-1
Warnings – English ............................................................................................. 1-2
Warnings – Español ............................................................................................ 1-3
Warnings – Francais ........................................................................................... 1-4
Warnings – Deutsch............................................................................................ 1-5
Warnings – Italiano ............................................................................................. 1-6
Warnings – Dutch................................................................................................ 1-7
2SYSTEM SPECIFICATIONS ........................................................................... 2-2
3SYSTEM DESCRIPTION ................................................................................. 3-1
Sample Inlet Systems and Sample Pump ........................................................... 3-3
On-Line Sampling from Pressurized Source................................................................. 3-3
Grab Sampling .............................................................................................................. 3-3
Calibration Valve........................................................................................................... 3-4
Chemical Reagent Reservoirs............................................................................. 3-4
Chemical Reagent Syringe Pumps ..................................................................... 3-5
Stream Splitter .................................................................................................... 3-5
Oxidation Reactor ............................................................................................... 3-6
CO2Sensors ....................................................................................................... 3-6
DI Water Loop............................................................................................................... 3-7
Membrane Module ........................................................................................................ 3-7
Conductivity and Temperature Measurement Cell........................................................ 3-7
Microprocessor Controller and Electronics.......................................................... 3-8
On-board Data Storage....................................................................................... 3-8

Ionics Instruments 2004 Page vii DLM 30007-08 Rev. A
Data Outputs ....................................................................................................... 3-9
CO2Sensor Measurement Cycle........................................................................ 3-9
Major Accessories and Configurations Available for the Analyzer .................... 3-10
IC Removal Module .................................................................................................... 3-10
Autosampler................................................................................................................ 3-10
Multistream ................................................................................................................. 3-11
Turbo...........................................................................................................................3-12
Flow Switch and Binary Input Module......................................................................... 3-12
Ozone Destruct Kit...................................................................................................... 3-12
Environmental Enclosure ............................................................................................ 3-12
Front Panel Displays and Controls.................................................................... 3-13
4INSTALLATION............................................................................................... 4-1
Unpacking and Inspecting ................................................................................... 4-1
Collecting Additional Equipment.......................................................................... 4-2
Site Selection/Location........................................................................................ 4-3
Power Requirements........................................................................................... 4-3
Environmental Consideration .............................................................................. 4-3
Installing Sample Port ......................................................................................... 4-3
Plumbing ....................................................................................................................... 4-4
Installing On-Line Sample Inlet System – ............................................................................................. 4-4
Installing Grab Sample Inlet System –................................................................................................... 4-5
Electrical.............................................................................................................. 4-8
Installing the Recorder, Alarm, and Printer Cables - Wiring ......................................... 4-8
Recorder and Alarm Outputs ........................................................................................ 4-9
Printer Cable ............................................................................................................... 4-11
Installing Reagent Cartridges ............................................................................ 4-11
Selecting the Sample Mode .............................................................................. 4-14
Entering the Date and Time .............................................................................. 4-17
Enter the Installation Dates for Reagents.......................................................... 4-17
Oxidizer Reagent ........................................................................................................ 4-18
Acid Reagent .............................................................................................................. 4-20
Selecting Oxidizer and Acid Flow Rates............................................................ 4-21
Preset Flow Rates....................................................................................................... 4-24
Manual/Grab Flow Rates ............................................................................................ 4-26
Setting Up a Password...................................................................................... 4-29
Entering Printer Setup....................................................................................... 4-30
Print Interval................................................................................................................ 4-32
Print Constants.................................................................................................. 4-33
Reagent Flush................................................................................................... 4-33
5ANALOG AND DIGITAL OUTPUTS................................................................ 5-1
Baud Rate ........................................................................................................... 5-1
RS-232 Data Out................................................................................................. 5-2
Serial Commands ......................................................................................................... 5-5
Alarms................................................................................................................. 5-7
Analog Outputs ................................................................................................... 5-9

Ionics Instruments 2004 Page viii DLM 30007-08 Rev. A
Display .............................................................................................................. 5-11
Warning 42........................................................................................................ 5-12
6OPERATION AND DISPLAY MENUS............................................................. 6-1
Menu Map ........................................................................................................... 6-2
Main Menu .......................................................................................................... 6-3
Power-Up and Initialization (On-Line Sampling).................................................. 6-3
Measurement Display.......................................................................................... 6-5
RunTime Menu.................................................................................................... 6-6
History........................................................................................................................... 6-7
Errors ............................................................................................................................ 6-9
Stop TOC .................................................................................................................... 6-10
Viewing Consumables Status............................................................................ 6-10
UV Lamp Status.......................................................................................................... 6-11
Reagent Status ........................................................................................................... 6-12
Fatal Error Display ............................................................................................ 6-14
Turning Off Power to the Analyzer .................................................................... 6-14
Grab Sample Measurements ............................................................................ 6-15
Setting the Sample Mode to Grab Sampling............................................................... 6-16
Changing from On-line to Grab Sampling................................................................... 6-17
Setting Up the Grab Sample Mode ............................................................................. 6-17
Running Grab Samples............................................................................................... 6-22
Changing to or Returning to On-line Measurements .................................................. 6-24
Changing the Password .................................................................................... 6-26
Disabling Password Protection.......................................................................... 6-29
Service Menu .................................................................................................... 6-29
7DATAGUARD OPERATION............................................................................ 7-1
Menu Map ........................................................................................................... 7-2
Main Menu .......................................................................................................... 7-3
RunTime Menu.................................................................................................... 7-3
Stop TOC Option .......................................................................................................... 7-4
Setup................................................................................................................... 7-5
Admin.................................................................................................................. 7-6
Password ...................................................................................................................... 7-6
Entering a Password –............................................................................................................................ 7-8
Setting a Password – .............................................................................................................................. 7-9
Disabling a Password – ........................................................................................................................ 7-11
Login Timeout ............................................................................................................. 7-11
Audit Trail .......................................................................................................... 7-12
View Audit Trail ........................................................................................................... 7-14
Print Audit Trail ........................................................................................................... 7-14
Export Audit Trail ........................................................................................................ 7-15
Reset Audit Trail ......................................................................................................... 7-16
Logging Out....................................................................................................... 7-17
8MAINTENANCE............................................................................................... 8-1

Ionics Instruments 2004 Page ix DLM 30007-08 Rev. A
Replacing the In-line Filter................................................................................... 8-2
Replacing the Reagents ...................................................................................... 8-4
Replacing the UV Lamp ...................................................................................... 8-8
Replacing the Pump Tubing .............................................................................. 8-12
Entering the Installation Dates for Reagents, UV Lamp, and Pump Tubing ...... 8-15
Oxidizer Reagent ........................................................................................................ 8-16
Acid Reagent .............................................................................................................. 8-18
Lamp ...........................................................................................................................8-19
Pump Tubing............................................................................................................... 8-20
Checking and Refilling the DI Water Reservoir ................................................. 8-22
Cleaning the Analyzer ....................................................................................... 8-23
9CALIBRATION ................................................................................................ 9-1
Supplies for TC and IC Standards....................................................................... 9-2
TC Zero Offset .................................................................................................... 9-3
TC Calibration Constant ...................................................................................... 9-8
IC Calibration Constant ..................................................................................... 9-15
TC and IC Calibration Verification ..................................................................... 9-19
10 TROUBLESHOOTING................................................................................ 10-1
Checking the Error Stack .................................................................................. 10-1
Corrective Actions ............................................................................................. 10-4
Printing the Error Stack ..................................................................................... 10-9
No Power to the Analyzer.................................................................................. 10-2
No Flow Through the Analyzer.......................................................................... 10-3
On-line Sampling ........................................................................................................ 10-4
Grab Sampling ............................................................................................................ 10-4
pH Of Sample Stream Is Too High.................................................................... 10-5
Measure the pH (non-deionized water only) ............................................................... 10-5
Oxidizer Flow Rate Too High ............................................................................ 10-6
Oxidizer Flow Rate Too Low ............................................................................. 10-6
DI Reservoir Too Low........................................................................................ 10-7
Gas Bubbles in Reagent Lines or Syringes....................................................... 10-7
Negative or Zero TOC Measurements .............................................................. 10-8
Diagnostic Testing........................................................................................... 10-11
Syringe Pump.................................................................................................. 10-12
Returning the Analyzer to Ionics Instruments.................................................. 10-12
Preparing the Analyzer for Return Shipment ............................................................ 10-13
Repackaging the Analyzer for Shipment................................................................... 10-16
11 RAM CARDS, CARD READER, AND RAMREADER SOFTWARE.......... 11-2
Installing the Card Reader Hardware ................................................................ 11-2
Installing RAMReader Software ........................................................................ 11-2
Removing the RAM Card .................................................................................. 11-2
Placing the RAM Card in the Card Reader Device............................................ 11-3
Installing the RAM Card in the Analyzer............................................................ 11-4

Ionics Instruments 2004 Page x DLM 30007-08 Rev. A
RAM Card Battery Maintenance........................................................................ 11-5
APPENDIX A – 810/810AS ANALYZER OPERATING FREQUENCY .................A-1
Selecting the Operating Frequency..................................................................... A-1
Calibration ........................................................................................................... A-2
APPENDIX B – 820/820AS ANALYZER OPERATING FREQUENCY .................B-1
Checking the Operating Frequency..................................................................... B-1
CE Mark Compliance .......................................................................................... B-2
Printer Cable ....................................................................................................... B-2
APPENDIX C - TURBO TOC ANALYZERS ..........................................................C-1
Introduction .........................................................................................................C-1
Setting Turbo Sampling Model ............................................................................C-2
Setting Reagent Flow Rates................................................................................C-3
Setting Alarms.....................................................................................................C-5
Calibration in Turbo Sampling Mode ...................................................................C-5
Turbo Mode TC Zero Offset......................................................................................... C-6
Turbo Mode TC Channel Calibration ........................................................................... C-7
Turbo Mode IC Channel Calibration ............................................................................ C-8
Alarm Settings.....................................................................................................C-9
Accuracy Specifications ......................................................................................C-9
APPENDIX D – BINARY INPUT MODULE – FLOW SWITCH..............................D-1
Mechanical Installation ........................................................................................D-2
Firmware Installation ...........................................................................................D-4
Replacing the EEPROM Chip (IC chip) ....................................................................... D-4
Replacing the IC Chip from an Zero Insertion Force-type Socket –..................................................... D-4
Replacing the IC Chip from a Low Profile-type Socket – .................................................................... D-6
Software Configuration........................................................................................D-8
APPENDIX E – CALIBRATION STANDARD PREPARATION............................. E-1
Preparing the TC Calibration Standard ............................................................... E-1
Part 1 – Measure TC of Water Used for Preparing Standards .....................................E-1
Part 2 – Prepare a Stock TC Standard .........................................................................E-4
Part 3 – Prepare a Dilute TC Standard ........................................................................E-5
Preparation of the IC Calibration Standard.......................................................... E-5
Part 1 – Prepare a Stock IC Standard: .........................................................................E-5
Part 2 – Prepare a Dilute IC Standard: ........................................................................E-7
APPENDIX F – CONSUMABLES AND ACCESSORIES ...................................... F-1

Ionics Instruments 2004 Page xi DLM 30007-08 Rev. A
TABLES
Table 1. Analog Outputs on Terminal Strip ...........................................4-10
Table 2. Recommended Reagent Flow Rates........................................4-23
Table 3. Preset Acid and Oxidizer Flow Rates ......................................4-24
Table 4. RS-232 Data Output Format ........................................................5-3
Table 5. Operating Messages....................................................................6-6
Table 6. User IDs.........................................................................................7-7
Table 7. RS-232 Audit Trail Output Format............................................7-16
Table 8. Replacement/Maintenance Schedule.........................................8-1
Table 9. Available RAM Cards and Storage Capacity.............................8-2
Table 10. Warnings and Error Messages ...............................................10-3

Ionics Instruments 2004 Page 1-1 DLM 30007-08 Rev. A
1 INTRODUCTION
The Sievers 800 Series Total Organic Carbon (TOC) Analyzer from Ionics
Instruments (U.S. Patent No. 5,132,094) is a high-sensitivity analyzer used to
measure the concentration of total organic carbon (TOC), total inorganic carbon
(TIC), and total carbon (TC = TOC+ TIC) in all water samples. The 800
designation is for 120V systems, 810 is for 100 V systems and 820 designation
is for 230V systems (generically referred to as a Sievers 800 TOC Analyzer).
The analyzer is based on the oxidation of organic compounds to form carbon
dioxide using UV radiation and a chemical oxidizing agent (ammonium
persulfate). Carbon dioxide is measured using a sensitive and selective
membrane-based conductometric detection technique as described by Carlson
(R. M. Carlson, "Method and Apparatus for the Determination of Volatile
Electrolytes", U.S. Patent No. 4,209,299, licensed by Ionics Instruments) For
each TOC measurement, the concentration of inorganic carbon species (CO2,
HCO3-, and CO3-2) is determined and, after oxidation of the organic compounds,
the total carbon content of the sample is measured. The concentration of the
organic compounds is then calculated from the difference between total carbon
and total inorganic carbon concentrations
(TOC=TC–TIC).
The analyzer can be used to monitor water samples ranging from high-purity
water containing < 0.5 parts per billion carbon to water samples containing up to
50 parts per million TOC. The analyzer is easy to operate, with extremely low
maintenance, and no special training or chemical knowledge required for
preparing reagents. The analyzer is calibrated at the factory, with the calibration
stable for approximately one year. Recalibration and validation can be easily
performed at the user’s site.
NOTE
Throughout this manual, the term Sievers 800 TOC Analyzer refers to all models
(800/810/820) of the analyzer.

Ionics Instruments 2004 Page 1-2 DLM 30007-08 Rev. A
Warnings – English
WARNING
Any operation requiring access to the inside of the equipment could result in
injury. To avoid potentially dangerous shock, disconnect from power supply
before opening the equipment.
WARNING
For continued protection against fire hazard, replace fuse with same type and
rating.
WARNING
This symbol, on the instrument indicates that the user should refer to the
manual for operating instructions.
WARNING
This is a Safety Class I product. It must be wired to a mains supply with a
protective earthing ground incorporated into the power cord. Any interruption of
the protective conductor, inside or outside the equipment, is likely to make the
instrument dangerous. Intentional interruption is prohibited.
WARNING
If this instrument is used in a manner not specified by Ionics Instruments Business
Group USA, the protection provided by the instrument may be impaired.
WARNING
Disposal of RAM card Lithium batteries must follow local environmental
regulations.

Ionics Instruments 2004 Page 1-3 DLM 30007-08 Rev. A
Warnings – Español
ATENCION
Cualquier operación que requiera acceso al interior del equipo, puede causar una
lesión. Para evitar peligros potenciales, desconectarlo de la alimentación a red
antes de abrir el equipo.
ATENCION
Para protección contínua contra el peligro de fuego, sustituir el fusible por uno del
mismo tipo y características.
ATENCION
Este símbolo, en el instrumento indica que el usuario debería referirse al manual
para instrucciones de funcionamiento.
ATENCION
Esto es un producto con clase I de seguridad. Debe conectarse a una red que
disponga de tierra protectora en el cable de red. Cualquier interrupción del
conductor protector, dentro o fuera del equipo, puede ser peligroso. Se prohibe la
interrupción intencionada.
ATENCION
Si este instrumento se usa de una forma no especificada por Ionics Instruments
Business Group, USA, puede desactivarse la protección suministrada por el
instrumento.
ATENCION
Las pilas de litio de la RAM card deshechadas deben seguir las regulaciones
medioambientales locales.

Ionics Instruments 2004 Page 1-4 DLM 30007-08 Rev. A
Warnings – Francais
ATTENTION
Chaque opération à l’intérieur de l’appareil, peut causer du préjudice. Afin
d’éviter un shock qui pourrait être dangereux, disconnectez l’appareil du réseau
avant de l’ouvrir.
ATTENTION
Afin de protéger l’appareil continuellement contre l’incendie, échangez le fusible
par un fusible du même type et valeur.
ATTENTION
Le symbol, indique que l’utilisateur doit consulter le manuel d’instructions.
ATTENTION
Ceci est un produit de Classe de sécurité I. L’instrument doit être branché sur
l’alimentation secteur par un fil de secteur prévu d’une prise de masse. Chaque
interruption du conducteur protégeant, à l’intérieur ou á l’extérieur de l’appareil
peut rendre l’instrument dangereux. Interruption intentionnelle est interdite.
ATTENTION
Si l’instrument n’est pas utilisé suivant les instructions de Ionics Instruments
Business Group, USA, les dispositions de sécurité de l’appareil ne sont plus
valables.
ATTENTION
Les batteries RAM card Lithium doivent être déposés suivant les régulations
d’environnement locales.

Ionics Instruments 2004 Page 1-5 DLM 30007-08 Rev. A
Warnings – Deutsch
WARNHINWEIS
Vor dem Öffnen des Gerätes Netzstecker ziehen!
WARNHINWEIS
Für kontinuierlichen Schutz gegen Brandgefahr dürfen bei Sicherungswechsel nur
Sicherungen der gleichen Stärke verwendet werden!
WARNHINWEIS
Dieses Symbol auf dem Gerät weist darauf hin, dass der Anwender zuerst das
entsprechende Kapitel in der Bedienungsanleitung lesen sollte.
WARNHINWEIS
Dies ist ein Gerät der Sicherheitsklasse I und darf nur mit einem Netzkabel mit
Schutzleiter betrieben werden. Jede Unterbrechung des Schutzleiters auβerhalb
oder innerhalb des Gerätes kann das Gerät elektrisch gefährlich machen.
Absichtliches Unterbrechen des Schutzleiters ist ausdrücklich verboten.
WARNHINWEIS
Wenn das Gerät nicht wie durch die Firma Ionics Instruments Business Group,
USA, vorgeschrieben und im Handbuch beschrieben betrieben wird, können die im
Gerät eingebauten Schutzvorrichtungen beeinträchtigt werden.
WARNHINWEIS
Die Entsorgung der Lithium-Batterie in der RAM-Karte darf nur nach den
geltenden Umweltschutzregeln erfolgen.

Ionics Instruments 2004 Page 1-6 DLM 30007-08 Rev. A
Warnings – Italiano
ATTENZIONE
Qualsiasi intervento debba essere effettuato sullo strumento può essere
potenzialmente pericoloso a causa della corrente elettrica. Il cavo di alimentazione
deve essere staccato dallo strumento prima della sua apertura.
ATTENZIONE
Per la protezione da rischi da incendio in seguito a corto circuito, sostituire I
fusibili di protezione con quelli dello stesso tipo e caratteristiche.
ATTENZIONE
Il simbolo, sullo strumento avverte l’utilizzatore di consultare il Manuale di
Istruzioni alla sezione specifica.
ATTENZIONE
Questo strumento è conforme alle specifiche per I prodotti in Classe I - Il cavo di
alimentazione dalla rete deve essere munito di “terra”. Qualsiasi interruzione del
cavo di terra all’interno ed all’esterno dello strumento potrebbe risultare
pericolòsa. Sono proibite interruzioni intenzionali.
ATTENZIONE
Se questo strumento viene utilizzato in maniera non conforme alle specifiche di
Ionics Instruments Business Group USA, le protezioni di cui esso è dotato
potrebbero essere alterate.
ATTENZION
Le batterie al Litio sulla RAM card, quando sono esaurite, devono essere gettate
secondo le regolamentazioni vigenti localmente.

Ionics Instruments 2004 Page 1-7 DLM 30007-08 Rev. A
Warnings – Dutch
OPGELET
Iedere handeling binnenin het toestel kan beschadiging veroorzaken. Om iedere
mogelijk gevaarlijke shock te vermijden moet de aansluiting met het net verbroken
worden, vóór het openen van het toestel.
OPGELET
Voor een continue bescherming tegen brandgevaar, vervang de zekering door een
zekering van hetzelfde type en waarde.
OPGELET
Het symbool, geeft aan dat de gebruiker de instructies in de handleiding moet
raadplegen.
OPGELET
Dit is een produkt van veiligheidsklasse I. Het toestel moet aangesloten worden op
het net via een geaard netsnoer. Bij onderbreking van de beschermende geleider,
aan de binnenzijde of aan de buitenzijde van het toestel, kan gebruik het toestel
gevaarlijk maken. Opzettelijke onderbreking is verboden.
OPGELET
Indien het toestel niet gebruikt wordt volgens de richtlijnen van Ionics Instruments
Business Group, USA gelden de veiligheidsvoorzieningen niet meer.
OPGELET
RAM kaart Lithium batterijen dienen volgens de lokale afvalwetgeving verwijderd
te worden.

Ionics Instruments 2004 Page 1-8 DLM 30007-08 Rev. A
WARNING
For on-line applications, an in-line particle filter must be installed on the sample
inlet line to prevent damage to the instrument. If the water sample to be monitored
contains high levels of suspended solids, then the in-line filter must be replaced on
a regular basis.
WARNING
Do not run the instrument with the sample inlet line capped off as this can result in
false TOC readings and possible damage to the instrument.
WARNING
Hazardous chemical reagents (ammonium persulfate and phosphoric acid) are used
in the instrument. The waste stream from the instrument is acidic and must be
disposed of properly. Consult your state and local regulations.
WARNING
Do not adjust the pressure regulator on the sample inlet system. The pressure is
preset to 5 psi. Changing the inlet pressure can damage the analyzer and void the
warranty. Do not let the inlet pressure exceed 5 psi.
WARNING
Always stop TOC measurements before turning off or unplugging the instrument.
Turning off the instrument while it is writing to the RAM card may damage the
card.
WARNING
Make sure the DI Reservoir is full, particularly when running samples with high
TOC or high salt concentrations. Always “clean-up” the instrument by running
low TOC DI water after running high TOC or salt samples.
WARNING
This is a Class A product. In a domestic environment, this product may cause
electromagnetic interference in which case the user may be required to take
adequate measures to correct the interference.
This manual suits for next models
2
Table of contents
Popular Measuring Instrument manuals by other brands

YSI
YSI EcoSense DO200 Operation manual

Chauvin Arnoux
Chauvin Arnoux AEMC PowerPad 8436 quick start guide

Tractel
Tractel HF36/1/LPT Operation and maintenance manual

Vega
Vega VEGAMIP R62 operating instructions

ELGO Electronic
ELGO Electronic EMIX23 series operating manual

Hanna Instruments
Hanna Instruments HI 99161 instruction manual

YOKOGAWA
YOKOGAWA PH450G user manual

Lutron Electronics
Lutron Electronics AM-4201 Operation manual

RTW
RTW BroadcastMonitor X Series operating manual

ROOTECH
ROOTECH ACCURA 3500 user guide

Somogyi Elektronic
Somogyi Elektronic home DM 4070 instruction manual

PCB Piezotronics
PCB Piezotronics J353B33 Installation and operating manual