Agera®Quick Start Guide
A60-1018-921 ver. 2.0
SETTING UP THE INSTRUMENT
STANDARDIZATION
1. Congure UV Mode: The UV mode is
shown in the status bar. To change the
mode, click the UV Button and select a UV
mode. Then click APPLY.
2. Initiate Standardization: Press the
Standardization Button on the status bar.
The port plate size and UV mode is shown
in the standardization dialog. To change
the port plate, place the new port plate
on Agera and click BACK to update the
port plate information.
3. Read Black Glass: Place the Agera black
glass at the sensor port and press READ.
4. Read White Tile: Remove the black glass
and place the Agera white tile at the port.
Press READ to continue.
1. Create a New Workspace: Press
Workspace > New Workspace and enter
a Name for this Workspace.
A new job will be opened together with
this new Workspace.
CREATE A WORKSPACE FOR
YOUR APPLICATION
For more detailed instructions on using the instrument, please refer to the Agera®User’s Manual.
Visit support.hunterlab.com for more information on maintenance, operation and repair of your instrumentation.
A
B
READING SAMPLES
D
C
1. Unpack your Box: Place the Agera®on the
bench. Retain the packaging in case of
instrument return to HunterLab.
2. Connect to the Power: Insert the power
supply into the back of the instrument and
connect the other end to an electrical
outlet.
2. Congure the New Workspace:
A Workspace is a measurement template
for a series of sample readings. With this
new Workspace open, the following can
be changed and saved:
To change measurement scales, select
Workspace > Color Scales. This congures
the desired color scales, indices, and
differences.
To change measurement procedure,
select Workspace > Read Options.
To add tolerances, select Workspace >
Standards & Tolerances.
To select the view screens, go to
Workspace > Views. To congure each
view screen, click View Options
in the toolbar.
3. Turn Power On: Using the rocker switch on
the back side, turn the power on. Once
inside the Essentials software, the Color
Data Table (CDT) is automatically
displayed.
Note: The default setting for Agera is a port up
position. If port forward is needed, go to
Workspace Menu > Preferences and check
Reverse Screen Orientation, click Apply.
Then cycle power on the Agera to implement
the new screen orientation.
3. Start a New Job: A job is a place to store
sample and standard readings. To start a
new job for this product, press Workspace
button on status bar to load this congured
workspace into a new job. To open an
existing job, press JOB button on the status
bar.
5. Standardization Complete: Remove the
Agera White Tile and click OK.
Standardization is updated and reported
in the bottom status bar along with UV
Mode and Port Plate Size.
A60-1018-921 ver. 2.0
1. Read Samples: Use the Read
Measurement icon to read samples.
To preview an image of a sample prior
to measurement, go to WORKSPACE >
READ OPTIONS > IMAGE VIEWER. Check
the IMAGE VIEWER and press APPLY.
When this option is checked, the camera
will look on sample before taking each
measurement. You can just view the
sample’s area without any capture or
capture an image and save it together
with a measurement.
2. Output Data: Select the Jobs icon from the
upper right corner. Under Jobs, data can
be saved, sent to a printer, emailed
to the network or exported to a ash drive.
Note: Agera supports multiple modes allowing
for switching among different valid modes
after standardization.
Note: Agera includes a built-in 60-degree gloss
meter, located right below the sample port
plate, that provides measurements in
conformance with ASTM D523. Gloss is
automatically measured with each sample
and standard measurement. To show
Gloss value, please select Gloss index in
Workspace > Color Scales > Indices.
Note: The active workspace name is displayed
in the status bar on the lower right corner.
Note: Orient the tiles by aligning the white line to
the line on the port plate.
Job icon
Read Measurement icon
Workspace icon
View Option icon

Other manuals for Agera

2

Other HunterLab Measuring Instrument manuals

HunterLab UltraScan VIS Owner's manual

HunterLab

HunterLab UltraScan VIS Owner's manual

HunterLab ColorFlex EZ User manual

HunterLab

HunterLab ColorFlex EZ User manual

HunterLab UltraScan VIS Owner's manual

HunterLab

HunterLab UltraScan VIS Owner's manual

HunterLab ColorFlex EZ User manual

HunterLab

HunterLab ColorFlex EZ User manual

HunterLab MiniScan EZ User manual

HunterLab

HunterLab MiniScan EZ User manual

HunterLab D25 NC Manual

HunterLab

HunterLab D25 NC Manual

HunterLab Vista User manual

HunterLab

HunterLab Vista User manual

HunterLab ColorFlex EZ Tomato User manual

HunterLab

HunterLab ColorFlex EZ Tomato User manual

HunterLab Vista Owner's manual

HunterLab

HunterLab Vista Owner's manual

HunterLab ColorFlex 13-1 User manual

HunterLab

HunterLab ColorFlex 13-1 User manual

HunterLab ColorFlex EZ Tomato User manual

HunterLab

HunterLab ColorFlex EZ Tomato User manual

HunterLab MiniScan EZ Owner's manual

HunterLab

HunterLab MiniScan EZ Owner's manual

HunterLab ColorFlex EZ Citrus User manual

HunterLab

HunterLab ColorFlex EZ Citrus User manual

HunterLab EasyMatch User manual

HunterLab

HunterLab EasyMatch User manual

HunterLab ColorFlex EZ Citrus User manual

HunterLab

HunterLab ColorFlex EZ Citrus User manual

HunterLab Vista Owner's manual

HunterLab

HunterLab Vista Owner's manual

HunterLab LabScan XE User manual

HunterLab

HunterLab LabScan XE User manual

HunterLab ColorFlex EZ Owner's manual

HunterLab

HunterLab ColorFlex EZ Owner's manual

HunterLab LabScan XE Owner's manual

HunterLab

HunterLab LabScan XE Owner's manual

HunterLab Agera EasyMatch QC Owner's manual

HunterLab

HunterLab Agera EasyMatch QC Owner's manual

HunterLab UltraScan PRO Owner's manual

HunterLab

HunterLab UltraScan PRO Owner's manual

HunterLab Agera User manual

HunterLab

HunterLab Agera User manual

HunterLab SpectraTrend HT User manual

HunterLab

HunterLab SpectraTrend HT User manual

HunterLab Vista User manual

HunterLab

HunterLab Vista User manual

Popular Measuring Instrument manuals by other brands

BURG WATCHER ProfiScale CROSS operating instructions

BURG WATCHER

BURG WATCHER ProfiScale CROSS operating instructions

Sonotec SONAPHONE RD operating manual

Sonotec

Sonotec SONAPHONE RD operating manual

ETouch ETM-G01E quick start guide

ETouch

ETouch ETM-G01E quick start guide

Pro-dig Intelli-Tork owner's manual

Pro-dig

Pro-dig Intelli-Tork owner's manual

Tintometer Lovibond MD100 instruction manual

Tintometer

Tintometer Lovibond MD100 instruction manual

RadComm RC2PLUS user manual

RadComm

RadComm RC2PLUS user manual

Katronic KATflow 150 Quick start manual

Katronic

Katronic KATflow 150 Quick start manual

Konica Minolta CM-3700D brochure

Konica Minolta

Konica Minolta CM-3700D brochure

TESTO 552 instruction manual

TESTO

TESTO 552 instruction manual

Velleman CAMCOLBUL4 quick start guide

Velleman

Velleman CAMCOLBUL4 quick start guide

janitza UMG 96S Handbook

janitza

janitza UMG 96S Handbook

ECEFAST TDS-110 instruction manual

ECEFAST

ECEFAST TDS-110 instruction manual

OPTILAB Phoenix Series user manual

OPTILAB

OPTILAB Phoenix Series user manual

WATERGUARD V21I installation guide

WATERGUARD

WATERGUARD V21I installation guide

hager EC377M User instructions

hager

hager EC377M User instructions

Cirrus Research Trojan2 User's manual guide

Cirrus Research

Cirrus Research Trojan2 User's manual guide

Mercury 600.652 instruction manual

Mercury

Mercury 600.652 instruction manual

Chauvin Arnoux B102 user manual

Chauvin Arnoux

Chauvin Arnoux B102 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.