SIKA UC RTD.2 User manual

Please keep this operating manual for future reference.
If the device is resold, please provide the operating manual along with it.
© SIKA •Ba_UC-RTD.2_en •01/2019 .
Monofunction simulator
Type UC RTD.2
Operating manual
Operating manual
..................................................page 1 - 46

UC RTD.2
- 2 - © SIKA •Ba_UC-RTD.2_en •01/2019
Table of contents page
0About this operating manual......................................................................................4
1General.....................................................................................................................7
1.1 Introduction.............................................................................................................7
1.2 Parts........................................................................................................................7
1.3 Safety......................................................................................................................7
1.3.1 Compliance with safety standards......................................................................7
1.3.2 Environmental conditions....................................................................................8
1.3.3 Worn devices......................................................................................................8
1.3.4 Device destruction procedure.............................................................................8
1.3.5 Instructions .........................................................................................................8
1.3.6 Making measurements........................................................................................9
1.3.7 Defects and abnormal stresses ..........................................................................9
1.3.8 Definitions...........................................................................................................9
1.4 Maintenance .........................................................................................................10
2Using the instrument...............................................................................................11
2.1 Power-up...............................................................................................................11
2.1.1 Keyboard ..........................................................................................................11
2.1.2 Measuring and simulation terminals..................................................................12
2.1.3 USB connector..................................................................................................14
2.1.4 Screen ..............................................................................................................14
2.1.5 Getting started (after power-up)........................................................................15
2.1.6 Operating modes..............................................................................................16
3Mode Programming.................................................................................................17
3.1.1 Resistance or temperature measurement using resistive sensors....................17
3.1.2 Resistance or temperature simulation using resistive sensors .........................18
4Related Functions...................................................................................................23
4.1 Scaling (linearisation)............................................................................................23
4.2 Nulling (tare/offset)................................................................................................24
4.3 Calibrated sensors (point correction)....................................................................25
4.4 Configuration predefined set points ......................................................................27
4.5 Storage of acquisitions in progress.......................................................................28
4.5.1 Save measurement...........................................................................................28
4.5.2 Run...................................................................................................................29
4.5.3 Stop ..................................................................................................................29
4.5.4 Parameters.......................................................................................................29
4.5.5 Display burst.....................................................................................................29
4.5.6 Save burst.........................................................................................................32
4.5.7 Open a burst.....................................................................................................32
4.5.8 New free burst ..................................................................................................32
4.5.9 Burst management............................................................................................32
4.5.10 Statistics ...........................................................................................................32

UC RTD.2
Technical changes reserved - 3 -
4.6 Synthesiser configuration......................................................................................33
4.6.1 Configuration of synthesiser points...................................................................33
4.6.2 Configuration of synthesiser parameters...........................................................33
4.7 Configuration of the ramp generation....................................................................34
4.7.1 Incremental signal configuration........................................................................34
4.7.2 Single ramp signal configuration.......................................................................35
4.7.3 Cyclic ramp signal configuration........................................................................36
5Parameter settings ..................................................................................................38
5.1 Contrast adjustment..............................................................................................38
5.2 Date and time setting.............................................................................................38
5.3“Preferences” setting.............................................................................................39
5.3.1 Filtering setting..................................................................................................39
5.3.2 Display resolution setting ..................................................................................39
5.3.3 Lighting duration setting....................................................................................39
5.3.4 Key beeping setting...........................................................................................40
5.3.5 Language setting...............................................................................................40
5.3.6 Temperature unit setting ...................................................................................40
5.4 Maintenance menu................................................................................................40
5.5 About the instrument menu ...................................................................................41
6Technical specifications...........................................................................................42
6.1 Measurement Function..........................................................................................42
6.1.1 Resistance measurement..................................................................................42
6.1.2 Temperature by resistive sensors (measurement)............................................43
6.2 Simulation Function...............................................................................................43
6.2.1 Resistance simulation .......................................................................................43
6.2.2 Temperature by resistive sensors (simulation)..................................................44
6.3 Power supply - Autonomy......................................................................................44
Urheberschutzvermerk:
Weitergabe sowie Vervielfältigung dieser Betriebsanleitung, Verwertung und Mit
teilung seines Inhalts
sind verboten, soweit nicht ausdrücklich gestattet. Zuwiderhandlungen verpflichten zu Schadenersatz.
Alle Rechte für den Fall der Patent
-, Gebrauchsmuster- oder Geschmacksmustereintragung vorbehal-
ten.

About this operating manual UC RTD.2
- 4 - © SIKA •Ba_UC-RTD.2_en •01/2019
0 About this operating manual
•The operating manual is aimed at specialists and semi-skilled personnel.
•Before each step, read through the relevant advice carefully and keep to the specified
order.
•Thoroughly read and understand the information in the section “Safety”.
If you have any problems or questions, please contact your supplier or contact us directly at:
Dr. Siebert & Kühn GmbH & Co. KG
Struthweg 7-9 • D - 34260 Kaufungen
+49 5605 803-0 • +49 5605 803-555
info@sika.net • www.sika.net

UC RTD.2 About this operating manual
Technical changes reserved - 5 -
Thanks
We would like to take advantage of this opportunity to thank you for choosing this SIKA precision
measuring device, the result of our hundred-year experience in manufacturing top-quality precision
measuring devices.
Consequently, we can continue to pursue this constant innovation policy, which has benefited our us-
ers for over 100 years. SIKA encourages all comments and welcomes any suggestions you might
have in order to allow us to fine-tune our know-how and improve our future products.
LIMITED GUARANTEE and LIABILITY
SIKA guarantees that this product is free of any manufacturing defects under normal conditions of use
and care. The guarantee period is one year with effect from the date of despatch. Parts, product re-
pairs and services are guaranteed for a period of 90 days. This guarantee only applies to the original
purchaser or to the end user if the latter is a distributor authorised by SIKA, and it does not apply to
fuses, inter-changeable batteries or any product which, according to SIKA, has been tampered with,
modified, neglected or damaged by accident or subjected to abnormal conditions of use and handling.
SIKA guarantees that the software will operate for the most part in compliance with the functional
specifications for a period of 90 days and that said software was recorded on non-faulty media. SIKA
does not guarantee that the software does not contain any mistakes or that it operates continuously.
The distributors authorised by SIKA shall apply this guarantee to products sold new and which have
not been used, but they are not authorised to apply an extended or different guarantee on behalf of
SIKA. Guarantee support is offered if the product has been purchased through a sales outlet author-
ised by SIKA or if the purchaser has paid the applicable international price. SIKA reserves the right to
invoice the purchaser the import charges for repaired parts or spare parts if the product purchased in
a country is sent to another country for repair.
The guarantee obligation on the part of SIKA is limited, at the discretion of SIKA, to refunding the pur-
chase price, or to repairing/replacing free of charge a faulty product returned to a service centre au-
thorised by SIKA within the guarantee period.
To benefit from the guarantee service, contact your nearest SIKA agency or send the product, along
with a description of the problem, carriage and insurance paid (free to destination), to your nearest
SIKA service centre. SIKA will not be held liable for any damage incurred during transport. Once re-
paired under guarantee, the product will be returned to the purchaser, carriage paid in advance (free
to destination). If SIKA considers that the problem was caused by misuse, a modification, an accident
or abnormal operating or handling conditions, SIKA shall provide an estimate of the repair costs and
shall only begin the repair work once appropriate authorisation has been received. Once repaired, the
product will be returned to the purchaser, carriage paid in advance, and the repair and transport costs
will be invoiced to the latter.
THIS GUARANTEE IS EXCLUSIVE AND INCLUDES ALL OTHER GUARANTEES, WHETHER
EXPLICIT OR IMPLICIT, INCLUDING, BUT NOT ONLY, ANY IMPLICIT GUARANTEE AS TO THE
SUITABILITY OF THE PRODUCT FOR SALE OR FOR APPLICATION FOR A SPECIFIC END OR
USE. SIKA CANNOT BE HELD LIABLE FOR ANY PARTICULAR DAMAGE, WHETHER INDIRECT,
ACCIDENTAL OR ENSUING, NOR ANY DAMAGE OR LOSS OF DATA, WHETHER FOLLOWING
INFRINGEMENT TO THE GUARANTEE OBLIGATIONS, ON A CONTRACTUAL, NON-
CONTRACTUAL OR OTHER BASIS.
Since certain countries or states do not allow limitations to an implicit guarantee condition, or the ex-
clusion or limitation of accidental or ensuing damage, the limitations and exclusions of this guarantee
may not apply to every purchaser. If any provision set forth in this guarantee is deemed invalid or in-
applicable by a court with jurisdiction, any such decision will not affect in any way the validity or the
applicability of all other provisions.

About this operating manual UC RTD.2
- 6 - © SIKA •Ba_UC-RTD.2_en •01/2019
Contents checks
The UC RTD.2 was checked mechanically and electrically prior to despatch. The necessary precau-
tions have been taken to ensure it reaches the user without being damaged.
Nonetheless, it is wise to perform a rapid check to detect any deterioration which may have occurred
during transport. If this is the case, inform the carrier immediately thereof.
The standard accessories are the following:
•This user's manual
•Four 1.5V AA batteries.
•A wrist-strap
•A protective sheath
If the product needs to be returned, use the original packaging where possible and indicate as clearly
as possible the reasons for the return in a note accompanying the device.

UC RTD.2 General
Technical changes reserved - 7 -
1 General
1.1 Introduction
The UC RTD.2 is a portable resistive sensor temperature calibrator (compliant with EC standards). It
is especially designed for calibration and maintenance. It makes it possible to measure and to simu-
late temperatures both on site as well as in a laboratory.
The UC RTD.2 features a large number of related functions which extend its range of application, in-
cluding:
•Generation of default values, increases, single or cyclic ramps.
•Storage of acquisitions and display in the form of tables or trend curves.
•Use of calibrated sensors with their coefficients of correction
A range of improvements facilitates its operation:
•Rapid access to all functions.
•Intuitive user interface.
•160x160 graphic display
•Connection via 4 mm safety plugs or a circular connector (4 pin).
•Power supply via 4 AA batteries or rechargeable batteries with rapid internal charger (Option).
The device is fitted in an elastomer-sheathed ABS case.
1.2 Parts
General characteristics:
•Portable device powered by 4 AA batteries (pack of Ni-MH storage batteries, 1.7 Ah optional).
•Hand strap for carrying and use on-site
•Graphic liquid crystal display: 160 x 160 pixels.
•Choice of language used for messages and programming of functions, gauges and parame-
ters via 6-key keyboard + 1 navigator.
•Backlit display accessible via a keyboard key, with the possibility of automatic black-out after a
specific programmable period of inactivity.
•Appearance: ABS case (elastomer-sheathed).
•Dimensions : 157 mm x 85 mm x 45 mm (without coating).
•Weight: 306 g without coating.
•IP54 tightness in compliance with standard EN 60529
1.3 Safety
1.3.1 Compliance with safety standards
The device complies with the applicable standards in force on the subject of electrical safety (EN
61010) as well as on the electromagnetic compatibility of the electrical measuring instruments (EMC:
EN61326).
These instructions for use contain information and warnings which must be observed by the user to
protect the latter against the dangers of electricity, to ensure the safe operation of the device and to
protect it against any mishandling which could damage or compromise the safety of use of the device.

General UC RTD.2
- 8 - © SIKA •Ba_UC-RTD.2_en •01/2019
1.3.2 Environmental conditions
In accordance with publication CEI 359: operating category I.
Range of application of standards from 0 to 2200 m.
Reference temperature range: 23°C ± 5°C, relative humidity: 45 % to 75 %.
Nominal operating range: -10°C to +50°C, relative humidity: 20 % to 80 % non-condensing.
Operating range limit: -15°C to +55°C, relative humidity: 10 % to 80 % (70 % at 55°C).
Storage and transport temperature range limit: - 30°C to + 60°C (without the batteries).
(without AA batteries or rechargeable battery pack).
1.3.3 Worn devices
Worn electrical devices can pollute the environment. We recommend you refrain from disposing of
this device in an ordinary waste bin, but rather that you use the recycling circuits available locally. If
not, you can return the device to us, and we will take care of its disposal free of charge.
1.3.3.1 Waste generated by the device
List of waste classified according to the decree published in the Official French Gazette dated 20th
April 2002. Decree no. 2002-540.
•16.02.14: Waste originating from electronic equipment:
Printed circuit boards making up the device.
•16.06.02: Batteries and storage battery (dangerous)
Alkaline Batteries (or NI-MH batteries).
•15.01.02: Packaging
ABS plastic device casing.
Elastomer conduit.
1.3.4 Device destruction procedure
Opening the device: unscrew the screw on the battery compartment, followed by the 5 screws secur-
ing the 2 shells. Separate the 2 shells. Separate the PCB from the upper shell.
With regard to the batteries, you will find them in the battery compartment (see commissioning chap-
ter).
In the case of the pack of batteries, there are 2 contaminants: NI-MH (Nickel-Metal Hybride) batteries
and a PCB. Separate these 2 items.
1.3.5 Instructions
The device was designed to operate safely if the instructions provided in the accompanying docu-
ments are followed. Any other use may jeopardise the safety of the operator. Any use other than
those specified in the instructions is therefore dangerous and forbidden.

UC RTD.2 General
Technical changes reserved - 9 -
1.3.6 Making measurements
The measuring leads and wires must be in good condition and must be replaced if their insulation ap-
pears faulty (insulating material cut, burned, etc...).
When the instrument is connected to the measuring circuits the terminals can be dangerous, therefore
do not place your hands near a terminal, whether used or not.
Never exceed the protection value limits indicated in the specifications.
When the order of size of the measured value is not known, ensure that the starting measurement
range is has high as possible, or select automatic range change mode.
Before changing function, disconnect the measuring wires from the external circuit. When voltage
measurements are being made, even weak ones, keep in mind that the circuits may feature a dan-
gerous voltage for the operator compared to the ground.
Do not make any measurements when the device is linked up to another device using the USB link or
when the batteries are being charged (option).
1.3.7 Defects and abnormal stresses
Every time you believe the protection may have been compromised, switch off the device and prevent
it from being switched back on unexpectedly.
The protection may be impaired in the following cases, for example:
•The device is visibly worn.
•The device is no longer able to make precise measurements.
•The device was stored in unfavourable conditions.
•The device has undergone severe stresses during transport.
1.3.8 Definitions
1.3.8.1 Definition of the category and degree of pollution
CAT II 60V:
The notion of categories determines the maximum transient voltage that can be applied to the meas-
urement inputs (it is also called overvoltage category). For the SIKA UC RTD.2, the maximum per-
missible overvoltage is 60V (DC or AC)
POL 2:
The notion of pollution determines the clearance between circuits. Degree 2 authorises temporary
conductivity caused by condensation.

General UC RTD.2
- 10 - © SIKA •Ba_UC-RTD.2_en •01/2019
1.3.8.2 Table of symbols used
Symbol
Name
Attention: see the accompanying documents
Earth
Compliant with the European Union directives
Category II, Pollution 2.
Maximum common mode voltage compared with
the ground=60V
1.4 Maintenance
The device must always be repositioned in accordance with the instructions provided herein. Any in-
complete or incorrect assembly could compromise the safety of the operator.
The authority responsible must regularly ensure that all safety-related items are not worn and ensure
all the preventive maintenance operations required are performed.
Before opening the device for any maintenance operations, you must make sure that all the wires are
disconnected from the appliance.
All adjustments, maintenance and repair work on the open device must be avoided as much as pos-
sible and, when these are indispensable, they must be performed by qualified staff, who are well
aware of the risks involved.

UC RTD.2 Using the instrument
Technical changes reserved - 11 -
2 Using the instrument
In order to use the device in all the safety required, all operators must read the paragraph on safety
carefully, along with the paragraph below.
2.1 Power-up
The device is delivered with 4 AA batteries of 1.5V each. It is wise to place these batteries in the
compartment provided for this purpose. To open up the compartment, unscrew the screw on the back
of the box. Once the batteries are in place, screw the cover back on.
Observe the polarity: an incorrect battery positioning could damage the device. The correct polarity is
indicated inside the compartment.
The figure below illustrates how to open the battery compartment as well as the correct positioning of
each battery.
2.1.1 Keyboard
The keyboard features:
•2 function keys (F1 and F2) for the selection of the various menus displayed on the screen.
•The navigator, consisting of 4 arrows (up (
↑
), down (
↑
),right (
), left (
)
•A clear key (CLEAR).
•A device on/off and backlighting on/off key (ON/OFF).
Press briefly to start the device. During operation, press briefly to turn the lighting on or off.
Press it longer for 2 seconds to stop the device.
•A validation key (VAL).
•A HOLD key allows you to suspend a process temporarily (when pressed briefly). If you press
it longer, this key makes it possible to switch from measuring mode to emission mode and
vice-versa.
Battery compartment
cover screw

Using the instrument UC RTD.2
- 12 - © SIKA •Ba_UC-RTD.2_en •01/2019
2.1.2 Measuring and simulation terminals
The UC RTD.2 is fitted with 4 safety bushes (4 mm in diameter) ad a circular 4-point connector. This
wiring is used both in measurement and emission mode (non-simultaneous).
Simulation / Generation Output
HOLD
key
Function
key F1
CLEAR key
Navigation
key
ON/OFF key
Validation key
(VAL)
Function key
F2
Measurement:
Current terminal (positive)
Simulation:
Current terminal (negative)
4-wire voltage /
measurement
terminal
Measurement:
Current terminal
(negative)
Simulation:
Current terminal
(positive)
3-wire voltage /
measurement
terminal

UC RTD.2 Using the instrument
Technical changes reserved - 13 -
Measurement Input
2-wire measurement
3-wire measurement
4-wire measurement
NOTE:
An adjustment of 2, 3 or 4 wire measurement is not necessary. The numbers of wires will be detected
automatically.
NOTE:
When using the device with resistance measurement/simulators, you must respect the polarities.
Measurement:
(negative)
4-wire voltage /
measurment
terminal
3-wire voltage /
measurement
terminal
Measurement:
(positive)

Using the instrument UC RTD.2
- 14 - © SIKA •Ba_UC-RTD.2_en •01/2019
2.1.3 USB connector
The UC RTD.2 is fitted with a USB connector (mini B) intended for uploading new software versions,
device adjustment and general connection to PC.
2.1.4 Screen
The UC RTD.2 is fitted with a graphic LCD display with back-lighting. The display resolution is 160 x
160 pixels. In normal operating conditions, the display is divided up into seven horizontal fields:
•The 1st field indicates the operating mode (Measurement, emission).
•The 2nd field indicates the date, time and battery charge.
•The 3rd field is reserved for icons indicating the operating mode
(related functions: Scaling, filtering…etc).
•The 4th field indicates the operating mode, the range and certain related functions.
•The 5th field indicates the value of the measurement or of the emission.
These values are expressed in Ω, °C, °F or %.
•The 6th field indicates (in measurement mode) the min., average and max. values of the
measurement.
•The 7th field indicates the function of keys F1 and F2.
USB port
(mini B connector)
1st field indicating
the operating
mode
2nd field indicating
the Date, time and
battery charge
3rd field indicating
the configuration
by pictograms
4th field indicating
the measurement,
the range and the
main related
functions
5th field indicating
the value of the
measurement (in Ω,
°C, °F or %)
6th field indicating
the min, ave and
max. measurement
values
7th field indicating
the functions of
the F1 and F2 keys

UC RTD.2 Using the instrument
Technical changes reserved - 15 -
The table below provides a definition of each pictogram displayed on the screen:
Symbol
Description
Emission mode in increases
Emission mode in single ramp
Emission mode in cyclic ramp
Scaling
On hold
Filtering
%FS (full scale) function
Error (over-calibration in measurement or error on the
value emitted…)
Incremental mode using the arrows
Battery life indication
Acquisition in progress (the value on the right of the pic-
togram indicates the number of values recorded)
The table below provides a definition of each pictogram of the function keys
Symbol
Description
Tab key
Open a drop-down list
Close a drop-down list
Delete the selected item
Clear the selection
Add the item being edited
2.1.5 Getting started (after power-up)
On power-up (inserting the batteries or accu pack), the device is automatically turned on (loading the
software in the memory). At this stage, we advise against connecting the device to an external circuit.
To avoid any signal conflicts, the device switches to measurement mode.

Using the instrument UC RTD.2
- 16 - © SIKA •Ba_UC-RTD.2_en •01/2019
2.1.6 Operating modes
There are 2 different operating modes:
•Measurement of resistive sensors (displayed in Ω or °C or °F ),
•Simulation of resistive sensors (value displayed in Ω or °C or °F),
The functional and electrical characteristics not to be exceeded are described below:
2.1.6.1 Resistance/temperature measurement
The following ranges are available:
Range
400 Ω
(for PT100)
3600 Ω
(for PT1000)
Resolution (display)
10 mΩ or 0.01 °C or 0.01 °F
100 mΩ or 0.01 °C or 0.01 °F
Scope of range:
0 Ω up to 400 Ω
-220 °C up to 850 °C
-364 °F up to 1562 °F
0 Ω to 3600 Ω
-220 °C to 760 °C
-364 °F up to 1400 °F
Scaling
yes
yes
2.1.6.2 Resistance/temperature simulation
The following ranges are available (for a current of 0.1 mA to 1 mA):
Range
400 Ω
(for PT100)
3500 Ω
(for PT1000)
Resolution (display)
1 mΩ or 0.01 °C or 0.01 °F
10 mΩ or 0.01 °C or 0.01 °F
Scope of range:
0 Ω to 400 Ω
-220 °C to 850 °C
-364 °F up to 1562 °F
0 Ω to 3500 Ω
-220°C to 715 °C
-364 °F up to 1319 °F
Scaling
yes
yes
2.1.6.3 Electrical characteristics not to be exceeded.
Function
Range
max Vin
I measurement
Ωmeasurement
400 Ω/ 3600 Ω
60 V
Ω
simulation
400 Ω/ 3500 Ω
5 mA

UC RTD.2 Mode Programming
Technical changes reserved - 17 -
3 Mode Programming
3.1.1 Resistance or temperature measurement using resistive sensors
•The choice of measurement or emission mode is made using the F2 key (mode menu).
•Using the navigation keys (
↑
and
↓
), position the cursor in the Measurement field going down
the menu.
•Confirm your choice using the VAL key.
NOTE:
The measurement mode is the mode selected by default.
•The function type selection (Resistive sensor type) is made using the F1 key
(configuration menu).
•Using the navigation keys (
↑
and
↓
), position the cursor in the function field.
•Confirm the latter using the VAL key.
In the MEASUREMENT CONFIG. menu, position the cursor in the Probe field using the F1 key.
•Enter the Probe menu using the F2 key.
•Choose the type of sensor (PT50, PT100, PT200…), using the navigation keys (
↑
and
↓
).

Mode Programming UC RTD.2
- 18 - © SIKA •Ba_UC-RTD.2_en •01/2019
•Press VAL to confirm.
•Using the F1 key, define the Unit by positioning the cursor on it.
•Enter the menu by pressing F2.
•Using the navigation keys (
↑
and
↓
), choose the unit.
•Press VAL to confirm.
NOTE:
Attention, the choice of °C or °F is made in the Setup\Preferences\Display unit menu
•Press VAL (again) to confirm the desired function and go back to the measurement screen.
The Measure mode makes it possible to display the Min (bottom left), Average (bottom centre) and
Max values (bottom right) from the last min/max reset command.
•Access this command by pressing the F2 key.
•Using the navigation keys (
↑
and
↓
), position the cursor in the min/max reset field.
•Confirm the latter using the VAL key.
3.1.2 Resistance or temperature simulation using resistive sensors
To access the Transmission mode:
•The choice of Transmission mode is made using the F2 key (mode menu).
•Using the navigation keys (
↑
and
↓
), position the cursor in the Transmission field going down
the menu.
•Confirm your choice using the VAL key.
The function type selection (Resistive sensor type) and the choice of unit displayed is made using the
F1 key (configuration menu) as in the measurement mode (please refer to the previous chapter).

UC RTD.2 Mode Programming
Technical changes reserved - 19 -
However, in simulation mode, when using an alternating current transmitter, you must indicate this to
the device.
•In the CONFIGURATION/MEASURE menu, position the cursor in the Current field using the
F1 key.
•Enter the Current menu using the F2 key.
•Choose the type of current (direct or alternating), using the navigation keys (
↑
and
↓
).
Confirm using the VAL key.
Once you have confirmed the Emission mode and the type of sensor used, define the type of genera-
tion:
•Continuous (manual or arrows or default editing).
•Incremental (by step or default).
•Single ramp (only one ramp emitted).
•Cyclic ramp.
•Synthesiser ("automatic default”).
3.1.2.1 Manual editing
•Press the F2 key to display the edit menu.
•Using the navigation keys (
↑
and
↓
), choose the manual editing mode and confirm (VAL
key).
•Press VAL again and enter your value using the navigation keys:
o
↑
and
↓
to increase or decrease the value
o
and
to select the digit to modify
NOTE:
If the Scaling mode is ON, the value to edit is expressed as a %, otherwise this value is expressed in
Ω or °C or °F.

Mode Programming UC RTD.2
- 20 - © SIKA •Ba_UC-RTD.2_en •01/2019
3.1.2.2 Editing with arrow
•Press the F2 key to display the edit menu.
•Using the navigation keys (
↑
and
↓
), choose the arrows edit mode and confirm (VAL key).
•Use the navigation keys to enter the value:
o
↑
and
↓
to increase or decrease the value
o
and
to select the digit to modify
3.1.2.3 Incremental editing
•Press the F2 key to display the edit menu.
•Using the navigation keys (
↑
and
↓
), choose the INCREMENTS mode and confirm (VAL key).
The values emitted are those defined in the CONFIGURATION/RAMP menu
a) automatically
•Using the navigation key (
↑
), start the automatic increases phase (following the parameters
programmed in the (CONFIGURATION/RAMP menu).
•Using the navigation key (
↓
), you can decrease automatically starting from the max. pro-
grammed resistance (or temperature).
b) manually
•Using the navigation key (
), you can increase manually the resistance (or temperature) emit-
ted (following the parameters programmed in the (CONFIGURATION/RAMP menu).
•Using the navigation key (
), you can decrease manually the resistance (or temperature)
emitted starting from the max. programmed resistance.
Indicates that the
mode used is the
arrows mode
Digit selected
during editing
Indicates that
the emission is
incorrect
Table of contents
Other SIKA Switch manuals
Popular Switch manuals by other brands

Multilink
Multilink 010-505-10 Installation and operation manual

HYDACELECTRONIC
HYDACELECTRONIC EDS 3100 operating instructions

Planet
Planet GS-4210-8UP2S user manual

Ntron
Ntron 500 Series installation guide

Lake Shore
Lake Shore 3f Wye Operation & maintenance manual

Tripp Lite
Tripp Lite B170-002-R user manual