6. While measuring, mix the sample thoroughly using the magnetic stirrer. Observe
the correct sample temperature (temperature indication on the display)
7. To start the measurement, press the START key.
8. Write down the measured value.
Please note: The device comes with an internal data logger. In the default set-
tings all measuring values are automatically saved within the used profile. To ex-
port measured values use the software SITA-ProcessLog. The measured values can
be displayed on the device by pressing the read key.
9. Rinse off the immersion probe with water (e.g. using a spray bottle). Continue
measuring with the next sample.
Cleaning
After finishing the measurements, carefully
rinse off the outside of the immersion probe
using water. Unscrew the limiter of the immer-
sion probe. Carefully clean the visible part of
the glass light guide using cotton sticks soaked
in water or alcohol. Please avoid damaging
the light guide.
You can pre-clean the immersion probe with a mild cleaning solution. To do so,
please prepare the measuring arrangement with screwed-on limiter and using the
cleaning solution instead of a sample.
Use demineralized water to rinse off cleaning agents and other residues after clean-
ing using water. Check the result (see ‘Checking the device’) and repeat the cleaning
process if necessary.
Please clean the stainless steel beaker after use.
Further information on cleaning can be found in the operating instructions.