SLS SLS4600 User manual

USER MANUAL
SLS Lab Basics 12 Place Microcentrifuge
United Kingdom:
International:
Northern Ireland:
Republic of Ireland:
High Speed Mini Centrifuge
Before using centrifuge, please carefully read this user manual for its efficient operation and safety.
12301570
Ver
.20211215
Before using centrifuge, please carefully read this user manual for its efficient operation and safety.
Code: SLS4600

SLS Lab Basics 12 Place Microcentrifuge User Manual
Safety Reminder............................................................................................................................................... 1
1. Specifications............................................................................................................................................... 3
2. Declaration of Conformity............................................................................................................................ 3
3. Required Operational Condition.................................................................................................................. 4
4. Installation....................................................................................................................................................
5
4.1 Location...................................................................................................................................................5
4.2 Connection of the power cord and grounding....................................................................................... 5
5. Structure....................................................................................................................................................... 6
6. Operation panel............................................................................................................................................ 7
7. Rotor Preparation......................................................................................................................................... 8
7.1 Prepare the samples............................................................................................................................. 8
7.2 Inject the samples into tubes............................................................................................................... 8
7.3 Keep the tubes balanced........................................................................................................................8
7.4 Inspect the rotor...................................................................................................................................
.
8
7.5 Symmetrically load centrifuge tubes in rotor....................................................................................... 8
8. Operation...................................................................................................................................................... 9
8.1 Normal Operation................................................................................................................................. 9
8.2 RCF Operation.................................................................................................................................... 12
8.3 Pulse Operation................................................................................................................................... 12
9. Maintenance............................................................................................................................................... 12
10. Troubleshooting........................................................................................................................................ 13
10.1 Frequent problems list...................................................................................................................... 13
10.2 How to open the door........................................................................................................................ 14
11. Instructions of rotor and tube....................................................................................................................15
11.1 The rotor instructions........................................................................................................................ 15
11.2 Tubes.................................................................................................................................................. 16
12. Calculating Relative Centrifuge Force (RCF)............................................................................................18
13. Warranty.....................................................................................................................................................18
13.1 Warranty of the centrifuge..................................................................................................................18
13.2 Warranty of the rotor...........................................................................................................................18
14. After-sales Service.....................................................................................................................................18
Contents

SLS Lab Basics 12 Place Microcentrifuge User Manual
Safety Reminder
Common safety precautions
Carefully read the following safety precautions for a thorough understanding.
• Follow the instructions and procedures described in this manual to operate this centrifuge safely
• Carefully read all safety messages in this manual and the safety instructions on the instrument
• Safety messages are labelled as indicated below. They are in combination with signal words of
“WARNING” and “CAUTION” with the safety alert symbol to call your attention to items or
operations that could be dangerous to you or other persons using this instrument. The definitions of
signal words are as follows:
WARNING: Personal Danger
Warning notes indicate any condition or practice, which if not strictly observed, could result in
personal injury or possible death.
CAUTION: Possible damage to instrument
Caution notes indicate any condition or practice, which if not strictly observed or remedied,
could result in damage or destruction of the instrument.
NOTE: Notes indicate an area or subject of special merit, emphasising either the product’s capability or
common errors in operation or maintenance.
• Do not operate this centrifuge in any manner not described in this User manual. When in doubt or
facing any troubles with this centrifuge, ASK FOR HELP
• The precautions described in this User manual are carefully developed in an attempt to cover all the
possible risks. However, it is also important that you are alert for unexpected incidents. Be carefully
operating this centrifuge
WARNING:
• This centrifuge is not explosion-proof. Never use explosive or flammable samples
• Do not install the centrifuge in or near places where inflammable gases are generated or chemicals
are stored
1

SLS Lab Basics 12 Place Microcentrifuge User Manual
• Do not place dangerous material within 30cm around the centrifuge
• It is your responsibility to make sure to prepare necessary safety measures before using samples that
are toxic, radioactive or contaminated with pathogenic micro-organisms
• If the instrument, rotor and/or accessories that has been contaminated by solutions with toxic,
radioactive or pathogenic materials, clean it according to the decontamination procedure that you are
specified
• If you require services at site, please sterilize and decontaminate it in advance, and then notify the SLS
Service Centre involved in the details of the particular materials
• Do not handle the power cord or turn on or off the POWER switch with wet hands to avoid electrical shocks
• For safety purposes, do not enter within 30cm around this centrifuge while it is in operation.
• While the rotor is rotating, never forcedly release the door lock
• Unauthorised repairs, disassembly, and other services to the centrifuge except by the SLS Service
Centre are strictly prohibited
CAUTION
• This centrifuge must be located on one firm and level table
• Make sure the centrifuge is horizontal before running
• Make sure the angle between the door and cover is greater than 70 degrees when the door is open
• Be careful not put your fingers or hands between the door and cover when the door is closing
• Do not move or relocate this centrifuge while it is running
• If fluid spills in the rotor chamber, please promptly clean and dry with a dry cloth to avoid sample
contamination
• Ensure to remove any objects and fragments of the tubes dropped inside the rotor chamber before
running this centrifuge
• Cautions be careful of the rotor
(1) Always check for corrosion and damages on the rotor surface before using it. Do not use the rotor if an
abnormality is found.
(2) Do not set the centrifuge speed beyond the allowable maximum speed of the rotorkits (rotor or
adapters). Make sure to run it below the allowable maximum speed.
(3) Do not exceed the allowable imbalance.
(4) Use the rotor and tubes within their actual capacities.
(5) If the rotor is attached with a lid, ensure it is tightened before operation.
• If any abnormal condition occurs during operation, please stop it immediately and contact the SLS
Service Centre. Notify the SLS Service Centre if a warning code if displayed
• Vibrations are likely to damage the centrifuge, contact the SLS Service Centre if abnormality observed
2

SLS Lab Basics 12 Place Microcentrifuge User Manual
3
1. Specifications
Maximum speed 15000rpm increment: 100rpm
Maximum RCF 15100 × g, increment: 100 × g
Maximum capacity 2ml × 12
Timer 30 seconds - 99 minutes HOLD continuous operation
Driving Motor Brushless DC motor
Safety devices Door interlock, over speed detector, over temperature detector, Error code
runtime display
Power requirements Single phase, 110V-240V, 50 Hz /60Hz, 3A
Dimensions (mm) (L) 255 × (W) 245 × (H )140
Weight 6Kgs
Additional features Speed/RCF switch, Pulse operation, LCD display of runtime status, buzzer
notification & alert
2. Declaration of Conformity
Construction in accordance with the following safety standards:
EN 61010-1
Construction in accordance with the following EMC standards:
EN 61326-1/ FCC Part 15 Subpart B/ ICES 001
Associated EU guidelines:
EMC-guidelines: 2004/108/EC
Instrument guidelines: 2006/95/EC
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
SLS Lab Basics 12 Place Microcentrifuge:
The centrifuge:
Construction in accordance with the following safety standards:
EN 61010-1
EN 61010 -2-10
Construction in accordance with the following EMC standards:
EN 61326-1/ FCC Part 15 Subpart B/ IECS 001
Associated EU guidelines:
EMC-guidelines: 2004/108/EC
Instrument guidelines: 2006/95/EC
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada

SLS Lab Basics 12 Place Microcentrifuge User Manual
4
Changes or modifications not expressly approved by the party responsible for compliance could void the
user’s authority to operate the equipment.
NOTE:
This centrifuge has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the centrifuge is operated in a commercial environment. The centrifuge
generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with
the user manual, may cause harmful interference to radio communications. Operation of centrifuge in a
residential area is likely to cause harmful interference in which case the user will be required to correct
the interference.
3. Required Operational Condition
3.1 Basic operational Conditions
(1) Power 110V-240 V, 50Hz /60 Hz, 3A
(2) Ambient temperature : 2ºC - 40ºC
(3) Relative humidity : ≤80%
(4) No vibration and airflow around
(5) No electric dust, explosive and corrosive gases around
3.2 Transport and storage condition
(1) Storage temperature : -40ºC - 55ºC
(2) Relative humidity : ≤93%

SLS Lab Basics 12 Place Microcentrifuge User Manual
5
This section describes the instructions that you should abide when installing the centrifuge to ensure your
safety and the optimum performance. Before moving the centrifuge, the rotor must be removed.
4. Installation
• Improper power supply may damage centrifuge
• Make sure the power source conforms to the required power supply before connecting
WARNING:
4.1 Location
(1) Place the centrifuge on a firm, flat and level table, ensure the four feet of this centrifuge stand on the
table firmly. Avoid installing on the slippery surface or surface prone to vibration.
(2) Ideal ambient temperature is 20ºC ±5ºC, avoid placing the centrifuge in direct sunlight if temperature
exceeds 30ºC.
(3) Keep clear of the centrifuge at least 10cm on both sides and at least 30cm behind it to guarantee the
cooling efficiency.
(4) Keep away from heat or water to avoid sample temperature issues or centrifuge failures.
4.2 Connection of the power cord and grounding
A minimum 10A outlet providing a sufficient ground is required, and this must meet with local safety
requirements.
• To avoid electrical shocks, ensure your hands are dry when touching the power cord
• This centrifuge must be grounded properly
WARNING:
SLS Lab Basics 12 Place Microcentrifuge User Manual

5. Structure
Rotor
LCD Display
Door Lock Hook
Foot
Operation Panel
Door
View Window
Air Vents
Power Inlet
Power Switch
Figure 5.1 Front view of centrifuge
Figure 5.2 Rear view of centrifuge
6
Door release hole
SLS Lab Basics 12 Place Microcentrifuge User Manual

6. Operation panel
7
Figure 6.1 Operation Panel
12 3 4 5
Item Symbol Name Function
1 Select button Press the button to choose the program which you want
to modify
2 Pulse button The speed can be accelerated and held at the speed
when pressing Pulse on
3 Parameter button Clockwise rotate to increase program values. Rotate
anti-clockwise to decrease parameter values.
Press the button, shift between speed and RCF display
4 Open/Lock button Press the button to open the door. The button is not
available when the centrifuge is running
5 Run/Stop button Press the button to start running. The centrifuge will
brake to stop running if pressed during centrifugation
Select
Pulse
Door
Run/Stop
High Speed Mini Micro-Centrifuge
7
6. Operation panel
Figure 6-1 Operation Panel
Item
Symbol
Name
Function
1
Select button
Press the button to choose the program which you want to
modify.
2
Pulse button
The speed can be accelerated and held at the speed when
pressing Pulse on.
3
Parameter button
Clockwise rotate to increase program values. Rotate
anti-clockwise to decrease parameter values.
Press the button, shift between speed and RCF display.
4
Open/ lock button
Press the button to open the door The button is not available
when the centrifuge is running.
5
Start/ Stop button
Press the button to start running. The centrifuge will brake to
stop running if pressed during centrifugation.
Figure 6-2 the main interface
Main interface is as figure 6-2. The speed is set to be 15000 rpm, running time is 12 minutes. When speed
symbol is rotating, indicating the centrifuge is running. Time display symbol display the ratio of
working to time setting. The total time setting is divided into 10 sections.
Speed area Lock status Time area
150 12 min
×
×
100rpm
Speed area Lock status Time area
Figure 6.2 The main interface
Main interface is as figure 6.2. The speed is set to be 15000 rpm, running time is 12 minutes. When speed
symbol is rotating, indicating the centrifuge is running. Time display symbol displays the ratio of
working to time setting. The total time setting is divided into 10 sections.
High Speed Mini Micro-Centrifuge
7
6. Operation panel
Figure 6-1 Operation Panel
Item
Symbol
Name
Function
1
Select button
Press the button to choose the program which you want to
modify.
2
Pulse button
The speed can be accelerated and held at the speed when
pressing Pulse on.
3
Parameter button
Clockwise rotate to increase program values. Rotate
anti-clockwise to decrease parameter values.
Press the button, shift between speed and RCF display.
4
Open/ lock button
Press the button to open the door The button is not available
when the centrifuge is running.
5
Start/ Stop button
Press the button to start running. The centrifuge will brake to
stop running if pressed during centrifugation.
Figure 6-2 the main interface
Main interface is as figure 6-2. The speed is set to be 15000 rpm, running time is 12 minutes. When speed
symbol is rotating, indicating the centrifuge is running. Time display symbol display the ratio of
working to time setting. The total time setting is divided into 10 sections.
Speed area Lock status Time area
150 12 min
××100rpm
High Speed Mini Micro-Centrifuge
7
6. Operation panel
Figure 6-1 Operation Panel
Item
Symbol
Name
Function
1
Select button
Press the button to choose the program which you want to
modify.
2
Pulse button
The speed can be accelerated and held at the speed when
pressing Pulse on.
3
Parameter button
Clockwise rotate to increase program values. Rotate
anti-clockwise to decrease parameter values.
Press the button, shift between speed and RCF display.
4
Open/ lock button
Press the button to open the door The button is not available
when the centrifuge is running.
5
Start/ Stop button
Press the button to start running. The centrifuge will brake to
stop running if pressed during centrifugation.
Figure 6-2 the main interface
Main interface is as figure 6-2. The speed is set to be 15000 rpm, running time is 12 minutes. When speed
symbol is rotating, indicating the centrifuge is running. Time display symbol display the ratio of
working to time setting. The total time setting is divided into 10 sections.
Speed area Lock status Time area
150 12 min
××100rpm
SLS Lab Basics 12 Place Microcentrifuge User Manual

7. Rotor Preparation
7.1 Prepare the samples
7.2 Inject the samples into tubes
CAUTION:
• Do not overload samples into the centrifuge which will cause leaking
• Do not exceed the actual capacity allowed in the user manual
7.3 Keep the tubes balanced
• Although the centrifuge can accept sample balancing by eye, we recommend that you keep this
centrifuge in a well balanced condition to extend its life expectancy
• Never intentionally run the centrifuge under unbalanced condition even though the allowable
imbalance is not exceeded
7.4 Inspect the rotor
Check the rotor for corrosion or scratches before using.
7.5 Symmetrically load centrifuge tubes in rotor
CAUTION:
• If any abnormality such as corrosions or scratches are found, stop using the rotor and contact the
SLS Service Centre
• Only manufacturers rotors must be used with the unit
CAUTION:
• Make sure the rotor lid is securely fixed on the rotor, and check that the rotor and shaft are tightened.
Otherwise, the rotor may be moved off while rotating and cause damage of the centrifuge and rotor
8
SLS Lab Basics 12 Place Microcentrifuge User Manual

8. Operation
CAUTION:
• Do not push or lean against the centrifuge while it is running
• Do not run the centrifuge when fragments or sample solutions are left in the centrifuge chamber
• Always keep the centrifugal chamber clean
• If the centrifuge makes strange noise during operation, stop it immediately and contact the SLS
Service Centre. Notify them of the warning code if displayed
8.1 Normal Operation
Turn on the power switch, centrifuge will display the running interface last time after passing the self-
diagnostic checks, see figure 8.1 below:
High Speed Mini Micro-Centrifuge
7
6. Operation panel
Figure 6-1 Operation Panel
Item
Symbol
Name
Function
1
Select button
Press the button to choose the program which you want to
modify.
2
Pulse button
The speed can be accelerated and held at the speed when
pressing Pulse on.
3
Parameter button
Clockwise rotate to increase program values. Rotate
anti-clockwise to decrease parameter values.
Press the button, shift between speed and RCF display.
4
Open/ lock button
Press the button to open the door The button is not available
when the centrifuge is running.
5
Start/ Stop button
Press the button to start running. The centrifuge will brake to
stop running if pressed during centrifugation.
Figure 6-2 the main interface
Main interface is as figure 6-2. The speed is set to be 15000 rpm, running time is 12 minutes. When speed
symbol is rotating, indicating the centrifuge is running. Time display symbol display the ratio of
working to time setting. The total time setting is divided into 10 sections.
Speed area Lock status Time area
150 12 min
×
×
100rpm
Figure 8.1 The last running interface
• Speed: 15000 rpm Running time: 12 minutes
• The door lock is released
8.1.1 Rotor loading and removal
CAUTION:
• Attach the rotor to the rotor shaft. Ensure the rotor is in position and connected with the shaft,
tightening the locking nut to secure the rotor with shaft, to prevent the rotor damaging the centrifuge
• Ensure the rotor lid is firmly tightened to the rotor
9
SLS Lab Basics 12 Place Microcentrifuge User Manual

SLS Lab Basics 12 Place Microcentrifuge User Manual
10
High Speed Mini Micro-Centrifuge
10
Figure 8-2 the rotor installation
⚫Load the rotor to the shaft to ensure rotor is in position until it connected well with the shaft.
⚫Rotate the rotor slightly by your fingers to check, if the rotor vibrates, if so attach the rotor again.
⚫Hold on the rotor with one hand, tighten clockwise the nut with the other hand, and make sure tighten
firmly.
⚫Put the rotor lid to the rotor, press the fastener down and make the fastener attach the locking nut
tightly.
⚫Close the door and start running.
⚫To release the rotor, firstly pull the fastener up to remove the lid, and then turn the locking nut
anti-clockwise.
CAUTION
⚫Check the rotor is firmly tightened before running.
8.1.2 Set the operation parameters
Press the button to select required parameters. The parameter can be modified when the parameter is
flashing. Clockwise rotate the parameter button to increase parameter value; counter-clockwise rotate the
parameter button to decrease parameter value. Parameter button rotate faster, parameter value
increase faster. The minimum speed increment is 100 rpm, the minimum time increment is 1 second.
(1)Set the speed
⚫Press the select button until the speed rpm is displayed.
⚫When the speed button is selected, the speed symbol will flash the speed value.
⚫The minimum speed value you can set 500rpm, the minimum increment is 100rpm.
⚫Rotate program button clockwise to increase speed value, rotate the program button anti-clockwise
to decrease speed value.
⚫You can speed-up set the speed value by rotating program button quickly.
⚫There is a circulating function to increase/decrease the speed values. Rotate the program button
Fastener
Figure 8.2 The rotor installation
• Load the rotor to the shaft to ensure rotor is in position until it is connected well with the shaft
• Rotate the rotor slightly by your fingers to check, if the rotor vibrates, if so attach the rotor again
• Hold on the rotor with one hand, tighten clockwise the nut with the other hand and make sure to
tighten firmly
• Put the rotor lid to the rotor, press the fastener down and make the fastener attach the locking nut tightly
• Close the door and start running
• To release the rotor, firstly pull the fastener up to remove the lid, and then turn the locking nut
anti clockwise
CAUTION:
• Check the rotor is firmly tightened before running
8.1.2 Set the operation parameters
Press the button to select required parameters. The parameter can be modified when the parameter
is flashing. Clockwise rotate the parameter button to increase parameter value, counter clockwise
rotation of the parameter button will decrease the parameter value. When the Parameter button is rotated
faster – the parameter value increases faster. The minimum speed increment is 100 rpm, the minimum
time increment is 1 second.
Select
(1) Set the speed
• Press the select button until the speed rpm is displayed
• When the speed button is selected, the speed symbol will flash the speed value
• The minimum speed value you can set 500rpm, the minimum increment is 100rpm
• Rotate program button clockwise to increase speed value, rotate the program button
anti-clockwise to decrease speed value
• You can speed up set the speed value by rotating program button quickly
Select

11
• There is a circulating function to increase decrease the speed values. Rotate the program button
clockwise to change settings from small → large → maximum → minimum. Rotate the program
button anti-clockwise to change settings from large → small → minimum → maximum
(2) Set the time
• Press button, time value flashes in the time setting mode
• Rotate the program button to set running time from 30 seconds to 99 minutes
• When time displays HD this is a continuous running mode
Select
SLS Lab Basics 12 Place Microcentrifuge User Manual
8.1.3 Start the operation
(1) Press running button to start running
• Timer will operate once the speed setting value is reached, the screen displays the remaining run time
(2) View and modify the operation programs
• Pressing button, returns the display to the program interface and displays settings programs.
Press the select button to the desired program. When flashing, rotate parameter button
to modify values. Release the button after 5 seconds, and the centrifuge will return to normal
operation mode and run according to the new value
• If the set time value has been modified the operation time is not affected and will continue
(3) Warning display
• If an error occurs during the operation, the centrifuge will brake to stop automatically, and display the
error code on the time/display area. The error code can be checked in the table 10.1, and corrective
actions can be applied accordingly
8.1.4 End the operation
(1) The centrifuge will brake when it reaches the setting time or button is pressed.
• When the rotor stops rotating, centrifuge will start beeping to alert the operation has finished
(2) Open the door
• The door can be released automatically when the operation has stopped
• With the door closed, you are able to press the button to open it
• After ending the operation, the program will store the setting parameters of this operation, and will
recall these parameters when restart the program
(3) Open the door and take out the rotor and samples.
Run/Stop
Select
Select
Run/Stop
Door

8.2 RCF Operation
(1) Turn on the power switch.
(2) Set an RCF (Relative Centrifugal Force) value.
• Press the select button and choose speed unit × g, the speed symbol will flash into RCF value
input status
• If no button is pressed after the speed value has flashed for 5 seconds, the input mode will be shut down
• Rotate program button to input a RCF value, RCF increment is100 ×g
(3) Set operating conditions
• The other operation, please refer to the section 8.1
8.3 Pulse Operation
This function is used to remove the residual samples adhered on the interior of the tubes.
Note: The button works only when the rotor stopped and the door is locked.
(1) Turn on the power switch and load the rotor to the shaft, tighten the rotor lid and make sure it is in
secured position, and then close the door.
(2) The centrifuge gets into preparation mode and displays last running values. The values can be reset.
(3) Press knob and hold, the centrifuge will speed up to the setting speed. While releasing
knob during acceleration, the centrifuge will start to decelerate and stop.
12
SLS Lab Basics 12 Place Microcentrifuge User Manual
Select
Pulse Pulse
9. Maintenance
CAUTION:
• If you do not follow the recommended instructions for cleaning or disinfecting, this may damage the
centrifuge
(1) Centrifuge
• If the centrifuge is exposed to ultraviolet rays for a long time, the color of the doors may change or the
label may come off. After using, cover the centrifuge with a piece of cloth to protect it from
direct exposure
• If the centrifuge needs cleaning, clean it with a cloth or sponge moistened with a neutral detergent
solution

13
SLS Lab Basics 12 Place Microcentrifuge User Manual
• Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution
(2) Rotor chamber
• If the rotor chamber needs cleaning, clean with cloth or sponge moistened with a neutral detergent
solution. Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution
(3) Drive shaft
• We recommend regular maintenance for drive shaft. You can wipe the drive shaft with soft cloth, and
then apply a thin coat of silicon grease
(4) Door
• Clean and sterilize the door using the same method as the step (1) above
(5) Rotor
• To prevent corrosion, remove the rotor from rotor chamber. If not in use for a long term, then detach
the rotor lid and turn upside down to dry the tube holes and keep clean
• For sample leaks in the rotor, rinse the rotor with water. Apply a thin coat of silicon grease to the rotor
when it is completely dry
• The rotor should be regular maintained, we recommend cleaning it every 3 months to ensure tube and
rotor holes keep clean, and then apply a thin coat of silicon grease
CAUTION:
• Do not directly pour water, neutral detergent or disinfectant solution into the rotor chamber.
Otherwise fluids may leak into the drive units and cause corrosion or deterioration to the bearings
This centrifuge has a self diagnostic function. If a problem occurs, an error/warning code will be displayed
on the time display screen and the operator can determine the malfunction with the warning code below.
Symptom Causes Solutions
Nothing appears on the screen
when the POWER is turned on
Building power circuit breaker
trips
The fuse was blown out
Remove the trouble and turn on the
POWER
Replace the fuse
Abnormal vibration
Rotor does not match with spindle
Samples are imbalance
Install again the rotor
Weighting scales, install symmetrically
10. Troubleshooting
10.1 Frequent problems list

14
SLS Lab Basics 12 Place Microcentrifuge User Manual
Rotor lid loose Tighten the rotor lid firmly
E–02 Door fault The door opened in running
The Run/Stop button is pressed
while the door opening
Close the door immediately
Close the door and then start
to operate.
E-06
Set wrong speed
The setting speed exceed the
allowable range
Modify the speed value
E-10–86 Read the service manual Contact the SLS Service Centre
Alarm code
appeared on the
time display screen
Table 10.1 1 Frequent problems and solutions
• Warning code E-1–9 are related to wrong operating. You can continue running the centrifuge after the
cause removed
10.2 How to open the door
10.2.1 In the case of power on
CAUTION:
• The door can be opened while the power is on and the rotor stops rotating
(1) Turn on the POWER switch the door lock will release automatically.
(2) The door lock will release automatically once the operation finished.
(3) It is available to release the door by pressing button once the rotor stops.
Door
10.2.2 In the case of power outage
The door cannot be opened automatically if there is a power outage. It is available to be opened manually.
(1) Ensure if the rotor has stopped rotating.
• Listen carefully to ensure no rotating sound can be heard
(2) Insert a screw driver into the hole to open the door.
• Hole is located on the top right side of the unit
• Insert a screw driver into the hole and push forward to release the door

CAUTION:
• Read the instructions tho roughly, correct use rotor
• Do not exceed the allowable maximum speed of rotor, tube and adapters etc. Take care that the
allowable maximum speed of some adapters are lower than the rotors maximum speed
15
SLS Lab Basics 12 Place Microcentrifuge User Manual
11. Instructions of rotor and tube
11.1 The rotor instructions
11.1.1 Rotor structure
High Speed Mini Micro-Centrifuge
15
11. Instructions of rotor and tube
CAUTION:
⚫Read the instructions thoroughly, correct use rotor.
⚫Do not exceed the allowable maximum speed of rotor、tube and adapters etc., be care that the allowable
maximum speed of some adapters are lower than the rotor’s maximum speed.
11.1 The rotor instructions
11.1.1 Rotor structure
Figure11-1 the rotor profile
11.1.2 Available rotors and adapters
Table 11.1 Rotors and adapters
Model
Rotor
type
ID
code
Tubes
Adapters
Maximum
speed(rpm)
Maximum
RCF(×g)
Allowable imbalance(*)
imbalance
Vo l u m e
imbalance
(**)
High Speed
Mini
Centrifuge plus
A12-2P
02
1.5/2.0ml
tube
15000
15100
2.0g/ tube
5mm/ tube
0.2ml PCR
A05P2
15000
11700
Locking nut Fastener Lid
Rotor block Spindle hole Tube
Locking Nut Fastener Lid
Rotor Block Spindle Hole Tube
Figure 11.1 The rotor profile
11.1.2 Available rotors and adapters
Table 11.1 Rotors and adapters
High Speed Mini Micro-Centrifuge
15
11. Instructions of rotor and tube
CAUTION:
⚫Read the instructions thoroughly, correct use rotor.
⚫Do not exceed the allowable maximum speed of rotor、tube and adapters etc., be care that the allowable
maximum speed of some adapters are lower than the rotor’s maximum speed.
11.1 The rotor instructions
11.1.1 Rotor structure
Figure11-1 the rotor profile
11.1.2 Available rotors and adapters
Table 11.1 Rotors and adapters
Model Rotor
type
ID
code Tubes Adapters
Maximum
speed(rpm)
Maximum
RCF(×g)
Allowable imbalance(*)
imbalance
Vo l u m e
imbalance
(**)
High Speed
Mini
Centrifuge plus
A12-2P 02
1.5/2.0ml
tube 15000 15100 2.0g/ tube 5mm/ tube
0.2ml PCR
A05P2
15000
11700
Locking nut Fastener Lid
Rotor block Spindle hole Tube
Model Rotor
Type
ID
Code Tubes Adapters Maximum
Speed
(RPM)
Maximum
RCF (X g)
Imbalance Volume
Imbalance
(**)
Allowable Imbalance
(*)
SLS Lab Basics
12 Place
Microcentrifuge
A12-2P 02
0.2mL PCR A05P2
15000
15000
15100
11700
2.0g/tube 5mm/tube
1.5/2.0mL
Tube

High Speed Mini Micro-Centrifuge
16
tube
0.5ml
micro tube A02P2 15000 12780
11.1.3 Notice
⚫The centrifuge rotor can separate sample which density lower than 2.0g/ml. if the separated samples
density is over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))1/2
11.1.4 Autoclaving
A12-2P rotor is made of plastic, cannot be high-pressure sterilization and UV irradiation, only ordinary
sterilization can be used.
CAUTION:
⚫The lid of the rotor is made of plastics, can not be high-pressure sterilization, only ordinary sterilization
can be used.
11.2 Tubes
11.2.1 Cleaning and sterilizing tubes
Table 11.2 Cleaning and sterilizing conditions for tubes
O:Applicable X:Inapplicable
Condition Material
PA
PC
PP
Cleaning
Cleaning fluids
Acidic(pH5 or lower)
X
X
X
Acidic(higher than pH5 )
O
O
O
Alkaline(higher than pH9 )
O
X
O
Alkaline(pH9 or lower)
O
O
O
Neutral(pH7)
O
O
O
Warm water(up to 70℃)
O
O
O
Ultrasonic cleaning
Neutral detergent(pH7)
O
O
O
Sterilization
Autoclaving
115℃(0.7kg/cm2)30minutes
O
O
O
121℃(1.0kg/cm2)20 minutes
X
O
O
126℃(1.4kg/cm2)15 minutes
X
X
X
Boiling
15 to 30 minutes
O
O
O
Ultraviolet sterilization
200-300nm
X
X
X
Gas sterilization
Ethylene oxide
O
X
O
11.1.3 Notice
• The centrifuge rotor can separate samples with density lower than 2.0g/ml. If the separated samples
density is over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm) = Maximum speed×(2.0(g/ml)/Sample density (g/ml)) 1/2
11.1.4 Autoclaving
A12-2P rotor is made of plastic, cannot take high-pressure sterilisation and UV irradiation, only ordinary
sterilisation can be used.
CAUTION:
• The lid of the rotor is made of plastics and cannot be subjected to high pressure sterilisation, only
ordinary sterilisation can be used
11.2 Tubes
11.2.1 Cleaning and sterilising tubes
Table 11.2 Cleaning and sterilising conditions for tubes
Condition
Material PA PC PP
Acidic (pH5 or lower) X X X
Acidic (higher than pH5) 0 0 0
Cleaning Cleaning Fluids
Alkaline (higher than pH9) 0 X 0
Alkaline (pH9 or lower) 0 0 0
Neutral (pH7) 0 0 0
Warm water (up to 70ºC) 0 0 0
Ultrasonic Cleaning
Neutral detergent (pH7) 0 0 0
115ºC (0.7kg/cm2) 30 minutes 0 0 0
Autoclaving 121ºC (10kg/cm2) 20 minutes X 0 0
Sterilisation 126ºC (1.4kg/cm2) 15 minutes X X X
Boiling 15 to 30 minutes 0 0 0
Ultraviolet sterilisation
200 - 300nm X X X
Gas sterilization Ethylene oxide 0 X 0
SLS Lab Basics 12 Place Microcentrifuge User Manual
16
Tube
0.5mL
Microtube A02P2 15000 12780

17
SLS Lab Basics 12 Place Microcentrifuge User Manual
11.2.2 Cleaning PC tubes
PC materials are low in chemical resistance against alkaline solutions. Avoid using neutral detergents
with pH higher than 9. Note that pH of some neutral detergents are still higher than 9 even if diluted
according to the instruction in the makers catalog. Use detergent with a pH between 7 and 9.
11.2.3 Autoclaving PA, PC and PP tubes
PA begins softening at about 120ºC, PC and PP at about 130ºC. Autoclave PA tubes at 115.0ºC (0.7kg/
cm2)for 30 minutes and PC and PP tubes at 121ºC (0.1kg/cm2) for 20 minutes. If a certain temperature is
exceeded, the tubes may be deformed.
When using a sterilising chamber, please operate as follows:
(1) Place tubes in vertical position, mouths upward. If tubes are placed sideways, they may deform
into an oval shape due to gravity.
(2) Remove screw nuts and inner covers to prevent from deformation or rupture.
(3) Wait until the sterilising chamber cools down to the room temperature before the tubes are removed.
11.2.4 Condition and life expectancy of tubes
The life expectancy of plastic tubes depends on the characteristics of samples, speed of the rotor used,
and temperature applied and so on. When the plastic tubes are used for centrifuge of ordinary aqueous
samples (pH between 5 and 9), their life expectancies are defined as follows.
Operated at the maximum speed:
High quality tubes (PA, PC, PP): 30-50 operations
Ordinary tubes (PA, PC, PP): around 10 operations (Using in low speed can extend the tube life)
Life expectancy of tubes also depends on the pretreatment conditions such as cleaning and sterilisation,
lifetime can be cut down.
Notice: Do not use damaged or cracked tubes.
Formaldehyde 0 0 0
PA: Polyallomer PC: Polycarbonate PP: Polypropylene

SLS Lab Basics 12 Place Microcentrifuge User Manual
18
12. Calculation Relative Centrifuge Force (RCF)
Relative Centrifuge Force (RCF) can be determined with the following calculation formula.
RCF= 1.118 × r × n 2 × 10-5
R– rotating radius, unit: cm; n-rotating speed, unit: rpm
13. Warranty
This centrifuge is guaranteed for one year from the date of delivery provided that it has been operated and
maintained properly.
13.1 Warranty of the centrifuge
13.2 Warranty of the rotor
The rotor is guaranteed for 1 year from the date of delivery upon manufacture. Please pay attention, do
not use the rotor once it has corrosion or fatigue damage. We do not guarantee this centrifuge and the
rotor under the following conditions even if within the guarantee period:
(1) Failures caused by incorrect installation
(2) Failures caused by rough or improper handling
(3) Failures caused by conveyance or relocation after installation
(4)Failures caused by unauthorized disassembly or modification
(5) Failures caused by using parts of the other companies, such as rotors and adapters
(6) Failures caused by natural disasters including fire, earthquakes and so on
(7) Consumables and parts have a limited guarantee period
14. After-sales Service
In order to ensure to operate centrifuge safely and efficiently, it is necessary for regular maintenance. If
your centrifuge has problems, do not attempt to repair it by yourself.
Contact our sales or the SLS Service Centre.
This manual suits for next models
1
Table of contents
Other SLS Laboratory Equipment manuals