Product name: MKF 2PW-MED
Part-no.: 1442721
Created on: 05.10.2022 Revision of data sheet: - D041837
Page
1 / 3
steute Technologies
GmbH & Co. KG
Brückenstraße 91
32584 Löhne, Germany
Phone +49 (0) 5731 745-0
Fax +49 (0) 5731 745-200
info@steute.com
www.steute.com
Specification
 neutral pedal in RAL 7035 Light grey
 two change-over contacts
 3 m shielded cable
Technical Data
Applied standards
IEC 60601-1 (Basic Safety)
IEC 60529 (Degrees of protection)
Note: see CSA certificate for
corresponding revision level.
Existing certifications CSA
Pedal shockproof thermoplastic, self-extinguishing / RAL 7035 Light grey
Pedal enclosure shockproof thermoplastic, self-extinguishing / black
Degree of protection IPX8 (1m / 35 Min.) according to DIN EN 60529
Contact element and material 2 micro switches / gold-plated
Connection type 3 m (6xAWG 24 UL/CSA RAL 7035)
Mechanical lifetime The expected operating life of the foot switch is limited to 10 years or 1,000,000 switching cycles of the
actuating units, whichever comes first.
Dimensions (W x L x H) 73 x 103 x 29 mm
Weight 0,3 kg
Product name: MKF 2PW-MED
Part-no.: 1442721
Created on: 05.10.2022 Revision of data sheet: - D041837
Page
2 / 3
steute Technologies
GmbH & Co. KG
Brückenstraße 91
32584 Löhne, Germany
Phone +49 (0) 5731 745-0
Fax +49 (0) 5731 745-200
info@steute.com
www.steute.com
Electrical Ratings
Switching voltage (UB): max. 25 V AC / 60 V DC
Switching current (IB): max. 1 A
Switching capacity (PB): max. 20 W
Environmental Conditions
Storage/Transport Operating
Temperature range -40 °C up to +70 °C -10 °C up to +60 °C
Relative humidity 10 % up to 100 % 10 % up to 100 %
Air pressure 500 hPa – 1120 hPa 800 hPa – 1060 hPa
Mounting/Wiring
Wire the foot switches according to the specified wire colours/terminal labelling. Contact symbols are shown for a not actuated switch.
Maintenance
Depending on the environmental/operating conditions we recommend routine maintenance as follows:
 Check enclosure and connecting cable for damage and destructive dirt
 Check functional elements and actuating elements for free operation
 For only manual cleaning, use just a cloth impregnated with water and a mild detergent.
 Do not use cleaning agents which may damage the plastic surfaces such as detergents, abrasive cleansers or solvent-based cleaners (such as
benzine, stain remover).
 During cleaning, avoid wiping with a cloth under the pedal, because this may cause the springs to move out of position.
Disposal
 Observe national, local and legal regulations concerning disposal.
 Recycle each material separately.
Product name: MKF 2PW-MED
Part-no.: 1442721
Created on: 05.10.2022 Revision of data sheet: - D041837
Page
3 / 3
steute Technologies
GmbH & Co. KG
Brückenstraße 91
32584 Löhne, Germany
Phone +49 (0) 5731 745-0
Fax +49 (0) 5731 745-200
info@steute.com
www.steute.com
General notes
The foot switch as a component of a medical device can be evaluated only together in the overall system of the customer. Therefore the conformity
evaluation including the classification of the overall system according to the Medial Device Directive 93/42/EEC respectively EU Medical Device
Regulation 2017/745 must be carried out at the customer. In addition, validation can also only be carried out by the customer in the overall system.
Furthermore, the intended use and the basic functions are verifiable only in interaction with the overall system. As a result, the customer must involve
the footswitch in his risk management, his usability analysis, the verification of the electromagnetic compatibility and, if applicable, the verification of the
biocompatibility.
Safety instructions
Despite careful production and inspection as well as redundant design of important components, an entirely safe function cannot be guaranteed at all
times (in case of malfunction, for example, a switch contact can turn off incorrectly, or a spring may break).
A functional test must be performed before every operation.
Risk of injury due to incorrect handling!
 The plug connections must not be subjected to mechanical loads!
 Switch off terminal before cleaning or maintenance work!
 Test functions when device is switched back on!
 Moving actuating components (e.g. carrying handle) can be a crushing hazard!
 Tripping hazard!
If components with reed contacts or hall sensors are used, strong magnetic fields can result in unwanted influences to the system. Additionally,
magnetically conducting materials are to be kept away from the direct foot switch area!
The electrical connection may only be carried out by authorized personnel.
Components, especially pedals or other actuating elements, may not be dissembled or dismembered under any circumstances! If this happens the
products need to be sent back immediately.
Errors and omissions excepted.

This manual suits for next models

1

Other steute Medical Equipment manuals

steute MKF 2 2S/2S-MED GP211 User manual

steute

steute MKF 2 2S/2S-MED GP211 User manual

steute MKF 3 2S SW2.4LE-MED GP311 User manual

steute

steute MKF 3 2S SW2.4LE-MED GP311 User manual

steute MKF 2 2S User manual

steute

steute MKF 2 2S User manual

steute RF 2PW-MED 2G User manual

steute

steute RF 2PW-MED 2G User manual

steute MKF 2 2PW/2PW-MED GP26 User manual

steute

steute MKF 2 2PW/2PW-MED GP26 User manual

steute MKFS 2PW-MED SH11 User manual

steute

steute MKFS 2PW-MED SH11 User manual

steute MKF 2 2S/2S SW2.4LE-MED GP212 User manual

steute

steute MKF 2 2S/2S SW2.4LE-MED GP212 User manual

steute KF 1PW-MED GP11 User manual

steute

steute KF 1PW-MED GP11 User manual

Popular Medical Equipment manuals by other brands

Castellini Puma Eli Series Instructions use and maintenance handbook

Castellini

Castellini Puma Eli Series Instructions use and maintenance handbook

Chattanooga Intelect HPL7 user manual

Chattanooga

Chattanooga Intelect HPL7 user manual

EKO DUO quick start guide

EKO

EKO DUO quick start guide

DERMAVIT Diamond Shield Crystal instruction manual

DERMAVIT

DERMAVIT Diamond Shield Crystal instruction manual

bort medical Dual-Tension Ankle Support manual

bort medical

bort medical Dual-Tension Ankle Support manual

Ametek Reichert Phoroptor VRx user guide

Ametek

Ametek Reichert Phoroptor VRx user guide

Olympic Medical AUTO-LENGTH 20 instruction manual

Olympic Medical

Olympic Medical AUTO-LENGTH 20 instruction manual

Ossur UNITY Instructions for use

Ossur

Ossur UNITY Instructions for use

OrthAlign KneeAlign 2 System TECHNIQUE MANUAL

OrthAlign

OrthAlign KneeAlign 2 System TECHNIQUE MANUAL

Winncare KALIN PO3P OPTION user manual

Winncare

Winncare KALIN PO3P OPTION user manual

Weinmann WM 24037 Instructions for use

Weinmann

Weinmann WM 24037 Instructions for use

Medela Invia Foam Dressing Kit Dressing Application Guide

Medela

Medela Invia Foam Dressing Kit Dressing Application Guide

GE Senographe 2000 D Operator's manual

GE

GE Senographe 2000 D Operator's manual

MEDAP MEDAP-BORA UP 2080 operating instructions

MEDAP

MEDAP MEDAP-BORA UP 2080 operating instructions

Timago yuwell YF-01 user manual

Timago

Timago yuwell YF-01 user manual

Aerogen Ultra AG-AS7500-EUR instruction manual

Aerogen

Aerogen Ultra AG-AS7500-EUR instruction manual

Innomax Perfection Series owner's manual

Innomax

Innomax Perfection Series owner's manual

Actron OBD II AutoScanner CP9175 instructions

Actron

Actron OBD II AutoScanner CP9175 instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.