Produktname: MKF 2 2S/2S SW2.4LE-MED GP212
Material-Nr.: 1423899
Created on: 31.03.2020 Revision of data sheet: E D022557
Page
2 / 3
steute Technologies
GmbH & Co. KG
Brückenstraße 91
32584 Löhne, Germany
Phone +49 (0) 5731 745-0
Fax +49 (0) 5731 745-200
info@steute.com
www.steute.com
Radio specifications
Performance
Distance 10 m -30 m (with internal antenna, depends on spatial conditions and the surrounding)
Output rating 3 dBm (max. 7 dBm optional)
Input sensitivity -93 dBm
Antenna Internal antenna with U.FL-plug
Radio technology
Frequency range 2402 to 2480 MHz
Chanel spacing 2 MHz
Modulation GFSK, adaptive frequency hopping of 40 channels
Applied radio standards
Europe EN 300 328 V2.1.1, EN 301 489-1 V2.2.0, EN 301 489-17 V3.2.0
USA , Part 15C , single Modular, Identifier: XK5-SW24LE
Canada RSS-247 Issue 1, Identifier: 5158A-SW24LE
Japan ARIB STD-T66, Identifier: 204-650001
Please note the instruction manual for the SW2.4LE system (D023336) (Document on request)
Environmental conditions
Storage/Transport (without batteries) Operating Battery charging
-20 °C up to +60 °C 0 °C up to +45 °C 0 °C up to +40 °C
10 % up to 100 % 10 % up to 100 % 10 % up to 100 %
500 hPa – 1060 hPa 700 hPa – 1060 hPa 800 hPa – 1060 hPa
General notes
as a component of a medical device can be evaluated only together in the overall system of the customer. Therefore the
conformity evaluation including the classification of the overall system according to the Medial Device Directive 93/42/EEC respectively EU
Medical Device Regulation 2017/745 must be carried out at the customer. In addition, validation can also only be carried out by the customer
in the overall system.
Furthermore, the intended use and the basic functions are verifiable only in interaction with the overall system. As a result, the customer
must involve the footswitch in his risk management, his usability analysis, the verification of the electromagnetic compatibility and, if
applicable, the verification of the biocompatibility.
Please refer to the Technical Description (D031288) for Medical synthetic footswitch SW2.4LE-MED/ Powered by rechargeable battery
Maintenance/cleaning
During cleaning, avoid wiping with a cloth under the pedal, because this may cause the springs to move out of position.
Safety instructions