
16
4.2.2 Operation of the Detector with the MCP Front Side connected to a Reference Voltage
Follow this procedure for all later operation starts, when operating the detector with the front side of
the MCP stack connected to an external high voltage potential
• Finish the complete cabling as described in Chapter 3 and as described in the manual of the Single HV
Supply.
• Connect the external reference voltage to the “MCP-F” connector. DO NOT APPLY ANY HIGH VOLTAGE
yet.
• Turn on the TDC.
• Start your end-user software (e.g. GUI software) and within this software the rate meter (if available).
For details see your end-user software manual.
• Be sure, that the vacuum pressure at the detector is remarkably below 1E-6 mbar, otherwise the micro
channel plates might be damaged by a local discharging (in general: the lower the pressure, the longer
the lifetime of the MCPs).
• Turn o all sources for electrons, ions, light or X-rays that might hit the detector.
• Increase the detector voltage for “MCP-B” to the specied operation voltage stepwise within a couple
of minutes.
• Turn o all sources for electrons, ions, light or X-rays that might hit the detector.
• Now increase the external reference voltage until reaching the desired voltage for“MCP-F”. (Important:
Respect the maximum reference voltage for “MCP-F”. It is given within the specication sheet.) The
voltage increase should not exceed 400V per minute, when operating the detector with an external
high voltage applied to “MCP-F” for the rst time. Otherwise turn up the high voltage carefully and
stepwise within a couple of minutes to the operation voltage.
• The voltage increase should not exceed 400V per minute, when operating the detector with an external
high voltage applied to “MCP-F” for the rst time. Otherwise turn up the high voltage carefully and
stepwise within a couple of minutes to the operation voltage.
• Increase the external reference voltage until reaching the desired voltage for “MCP-F”.
• Now turn up the the high voltage carefully and stepwise within a couple of minutes to the operation
voltage. The starting operation voltage is specied in the specication sheet of the detector.
• A schematic sketch on how to ramp the voltages is given in Figure 5.
• Watch the vacuum pressure while increasing the high voltage; turn the voltages back, if an unusual
increase is observed in the pressure (indicator for high voltage sparking).
High voltage sparks may seriously damage the detector anode/meander or the MCPs.
Observe the chamber pressure carefully every time the high voltage is turned on. Switch
o the high voltage immediately in case of a temporary pressure rise by an order of
magnitude or more. This indicates high voltage sparking.
If sparking occurs, turn down the high voltage immediately and wait some time (up to 5
min.). Start the“Start-Up”procedure again with an increased ramp time. Turn o the high
voltage completely, stop the procedure and call your provider for further assistance, if is
it not possible to reach the operation voltage without sparking.
DLD8080 R4.30 & R4.31 Manual | Surface Concept GmbH