Surface Concept HVPS User manual

19“ HVPS Detector Supply
1 Channel
(Release R020)
Manual

2
User Manual for the
19” HVPS Detector Supply 1 Channel
Release: R020
Manual Version 2.0
Printed on 2020-11-13
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
Surface Concept GmbH
Am Sägewerk 23a
55124 Mainz
Germany
phone: +49 6131 62716 0
fax: +49 6131 62716 29
email: info@surface-concept.de
web: www.surface-concept.de
All rights reserved. No part of this manual may
be reproduced without the prior permission
of Surface Concept GmbH.

3
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
1 Table of Contents
1 Table of Contents ......................................................................................................................3
2 Introduction................................................................................................................................5
2.1 General Information..........................................................................................................5
2.2 Safety Instructions .............................................................................................................5
2.3 General Overview...............................................................................................................6
3 Introduction................................................................................................................................7
3.1 Initial Inspection.................................................................................................................7
3.2 Installation............................................................................................................................7
4 Device Layout & Operation ...................................................................................................9
4.1 Device Layout.....................................................................................................................9
4.2 General Device Operation........................................................................................... 10
4.3 Operation of a MCP/CEM Detector.......................................................................... 11
4.4 Schematic Layout of the 19”HVPS Detector Supply R020.............................. 12
5 List of Figure ........................................................................................................................... 13

419” HVPS Detector Supply R020 Manual | Surface Concept GmbH
This side has been left blank on purpose.

5
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
2 Introduction
2.1 General Information
This manual is intended to assist users in the installation, operation and maintenance of Release Version
020 of the 19” HVPS Detector Supply. It is divided into 5 chapters.
2.2 Safety Instructions
Please read this manual carefully before performing any electrical or electronic operations and strictly
follow the safety rules given within this manual.
The following symbols may appear throughout the manual:
The“caution symbol”marks warnings, which are given to prevent an accidental damaging
of the device. Do NOT ignore these warnings and follow them strictly. Otherwise no
guarantee is given for arose damages.
Note
The “note symbol” marks text passages that contain important information/hints about
the operation of the device. Follow this information to ensure a proper operation of the
device.
The “high voltage symbol” marks warnings, given in context with the description of the
operation/use of high voltage supplies and/or high voltage carrying parts.
Hazardous voltages are present that can cause serious or fatal injuries. Therefore only
persons with the appropriate training are allowed to carry out the installation, adjustment
and repair work.

619” HVPS Detector Supply R020 Manual | Surface Concept GmbH
2.3 General Overview
The Surface Concept 19” HVPS Detector Supply R020 is a 1 channel high voltage module especially
designed for the Surface Concept 19”Basic Unit, a modular supply system.
The 19” HVPS Detector Supply R020 is laid out for the operation of e.g. MCP based detectors with metal
anode or CEM detectors. It provides one single high voltage with a positive polarity. It also provides the
capability to oat on an external reference potential.
The 19” HVPS Detector Supply R020 comes with a maximum output voltage of +3000V and can oat on
voltages of up to +/- 1000V. Check the specication sheet of your specic 19”HVPS Detector Supply R020
for further specic information.
The device can produce lethal high voltages of up to several kV. Hazardous voltages are
present, therefore only persons with the appropriate training are allowed to carry out
the installation, adjustment and repair work.
Do not open the power supply, while it is in operation. Hazardous voltages are present.
In case that the device must be opened, turn o the device rst AND pull out the power
plug.

7
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
3 Introduction
3.1 Initial Inspection
Visual inspection of the system is required to ensure that no damage has occurred during shipping. If
there are any signs of damage, please contact SURFACE CONCEPT immediately. Please check the delivery
according to the packing list (see Table 1) for completeness.
• 1x 19“ HVPS-HV Module R019
• 1x SHV Termination Plug
Table1: Packing list for the 19“ Basic Unit R019.
3.2 Installation
The general connection scheme of the 19”HVPS-HC Module R020 is as follows:
Figure 1: Connection scheme of the 19” HVPS Detector Supply R020 (the detector ange is given for illustration only.
Please check the manual of your MCP/CEM detector for the correct ange layout and feedthrough assignment).

819” HVPS Detector Supply R020 Manual | Surface Concept GmbH
Finish the complete cabling of the 19” HVPS Detector Supply R020 before switching on
the 19” Basic Unit and switch o the device rst before performing any changes to the
cabling.
• Install the 19”HVPS Detector Supply R020 into a free slot of the 19”Basic Unit (if not already installed).
• High voltage output is provided to the SHV socket named“MCP-B”.
• The socket named“MCP-F”connects the external reference voltage which is given to the device via the
“HV Ref. In” input to the front side of the MCP stack/ CEM of the detector
• Use SHV cables to connect the outputs of the 19”HVPS Detector Supply R020 to the corresponding HV
input/s of the MCP/ CEM detector (see your MCP/ CEM detector manual for more details).
• Use the SHV termination plug to terminate the “HV Ref. In” input when not working with an external
reference voltage.
Note
In cases that no reference voltage is applied to the device, the termination plug must be
used to ground the reference input of the 19”HVPS Detector Supply R020. With a missing
reference potential the device is not providing any output voltage.

9
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
4 Device Layout & Operation
4.1 Device Layout
The layout of the 19” HVPS-HC Module R020 is given below.
1. SHV connector for output of operation voltage for a
MCP/CEM detector.
2. SHV connector for output of reference potential to
the front side of a MCP/CEM detector.
3. SHV connector for input of an external reference
potential (max. +/- 1,000V, check with your MCP
or CEM detector manual for details on maximum
reference potentials).
Figure 2: Layout of the 19”HVPS Detector Supply R020.
The 19” HVPS Detector Supply R020 is specied for the operation with an external
reference voltage of maximum +/- 1,000 V (see no. 3 in Figure 2).
Please also respect the corresponding specications for the maximum voltage for the
MCP front potential of your MCP/CEM detector (see the manual and the specication
sheet of your MCP/CEM detector for further details).

10 19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
4.2 General Device Operation
After switching on the 19” Basic Unit, its display shows the “Surface Concept” animated logo, while the
device is scanning for the 19” HVPS Detector Supply R020 and its specic settings. This can take up to
several seconds. If the 19” Basic Unit is ready for operation, it switches into the standby mode and shows
an empty mask for the voltage adjustment (see Figure 3).
Push the“Start/Standby” button in the lower left
corner of the display to switch on the high
voltage.
Alternatively one can press the“Channel”control
knob.
After switching on the high voltage the device
is in the so called operation mode. In operation
mode the display shows the name of the selected
channel in the top line (in this case“HV 1”) as well
as the output voltage of that channel.
The dierent HV channels can be selected by
turning the “Channel” control knob.
Figure 3: Standby mode.
Figure 4: Operation mode.
Figure 5: Operation mode – voltage adjustment.
The “Adjust” control knob is used to adjust the
output voltage. Turn the “Adjust” control knob
clockwise/counterclockwise to increase/decrease
the value of the output voltage in a step width as
dened in the line “edit step”.
The line “set value” displays the nominal value
for the output voltage as adjusted by the user.
Voltage adjustment can only be made in this line.
The line “actual” displays the actual value for the
output voltage on the output connector as measured by the device.The device always regulates the actual
value of the output voltage to t to the nominal value as set by the user. Hereby the voltage measurement
is always a relative measurement between the two HV outputs of one channel. Additional reference
voltages (e.g. in oating operation) are not measured and therefore are also not displayed (see the manual
of the specic module for further details). Turn the“Adjust”control knob clockwise/counterclockwise while
pushing it to increase/decrease the step width in the line “edit step”.
Push the„Start/Standby“-button in the lower left corner of the display again to switch back to
the “Standby” mode.

11
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
4.3 Operation of a MCP/CEM Detector
In some cases a MCP detector is operated with the front side of the MCP stack terminated to ground. But
for many other applications it is necessary to apply some external reference voltage to the front side. In
this case the oating capability of the 19” HVPS Detector Supply R020 allows a more insusceptible supply
of the correct detector operation voltage because the external reference voltage must not be taken into
account for the MCP operation voltage. Additionally this wiring also saves the detector intrinsically from
any over-voltage in case of a sudden drop of the external reference voltage. In general there are three
dierent application cases concerning reference potential to the front side of the detector’s MCP stack.
For example: Assume an operation voltage of a detector of +1,900V.
Application case 1: The detector should be operated with the MCP front side connected to ground. In
this case the SHV termination plug is connected to the “HV Ref. In” input. “MCP-B” is set to the detectors
operation voltage. The display shows a value of MCP-B = 1,900 V. The output voltage in respect to the
ground potential is MCP-B = +1,900 V.
Application case 2: The detector should be operated with the MCP front side connected to +1000V. In this
case the external reference voltage is connected to the “HV Ref. In” input. First set the 19” HVPS Detector
Supply R020 to the detectors operation voltage. Then increase the external reference voltage to the
+1,000V. The display shows a value of MCP-B = 1,900V, but the output voltage in respect to the ground
potential is MCP-B = +2,900V.
Application case 3: The detector should be operated with the MCP front side connected to -500V. In this
case the external reference voltage is connected to the “HV Ref. In” input. First set the 19” HVPS Detector
Supply R020 to the detectors operation voltage. Then increase the external reference voltage to the -500V.
The display shows a value of MCP-B = 1,900V, but the output voltage in respect to the ground potential is
MCP-B = +1,400V.

12 19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
Figure 6: Schematic layout of the 19”HVPS Detector Supply R020 showing also the internal load
and measuring resistors.
4.4 Schematic Layout of the 19” HVPS Detector Supply R020
Figure 6 shows the schematic layout of the 19” HVPS Detector Supply R020 and especially the layout of
the HV outputs. An internal controller measures the output voltage and regulates it to the nominal value
entered by the user or set as default value within the device. Hereby the voltage measurement is always a
relative measurement. The output polarity is xed dened (see the specic specication sheet for details
on the max. output voltage and the polarity). The device comes with an integrated 2-fold fan-out for the
output voltage of each channel. Figure 6 also shows the internal load and measuring resistors.
Note
The 19”HVPS Detector Supply R020 is not producing any output voltage if the reference
input “HV Ref. In” is not terminated (either to ground or to an external HV potential),
because the HV module always need to be connected to a reference potential. A SHV
termination plug is part of the delivery to terminate the “HV Ref. In”input to ground.

13
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
Figure 1: Connection scheme of the 19” HVPS Detector Supply R020 (the detector ange is given for illustration
only. Please check the manual of your MCP/CEM detector for the correct ange layout and feedthrough
assignment).....................................................................................................................................................................................................7
Figure 2: Layout of the 19”HVPS Detector Supply R020.................................................................................................................9
Figure 3: Standby mode........................................................................................................................................................................... 10
Figure 4: Operation mode....................................................................................................................................................................... 10
Figure 5: Operation mode – voltage adjustment............................................................................................................................ 10
Figure 6: Schematic layout of the 19”HVPS Detector Supply R020 showing also the internal load and measuring
resistors.......................................................................................................................................................................................................... 12
5 List of Figure

14
Surface Concept GmbH
Am Sägewerk 23a
55124 Mainz
Germany
phone: +49 6131 62716 0
fax: +49 6131 62716 29
email: info@surface-concept.de
web: www.surface-concept.de
Manufacturer
Product
Model
High Voltage Power Supply
19” HVPS Detector Supply
The above named products comply with the following European directive:
89/336/EEC Electromagnetic Compability Directive, amended by 91/263/ EEC
and 92/31/ EEC and 93/68/EEC
73/23/EEC Low Voltage Equipment Directive, amended by 93/68/EEC
The compliance of the above named product to which this declaration relates is in conformity with the
following standards or other normative documents where relevant:
EN 61000-6-2:2005+AC:2005 Electromagnetic compatibility (EMC):
Generic standards - Immunity for industrial environments
EN 61000-6-4:2007+A1:2011 Electromagnetic compatibility (EMC):
Generic standards - Emission standard for industrial environments
EN 61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use
This declaration does not represent a commitment to features or capabilities of the instrument.The safety notes and regulations
given in the product related documentation must be observed at all times.
For and on behalf of Surface Concept GmbH
Mainz,……01.10.2019………. Legal Signature…………………………………………
(Date) (Dr. Andreas Oelsner)
EC Declaration of Conformity
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
Other manuals for HVPS
1
Table of contents
Other Surface Concept Security Sensor manuals

Surface Concept
Surface Concept Delayline DLD 8080 User manual

Surface Concept
Surface Concept DLD6060-8S User manual

Surface Concept
Surface Concept MCPD25 User manual

Surface Concept
Surface Concept DLD4040 User manual

Surface Concept
Surface Concept MCPD40 User manual

Surface Concept
Surface Concept CEM 4230 User manual

Surface Concept
Surface Concept Delayline DLD 8080 User manual

Surface Concept
Surface Concept MCPD18 User manual