Automated Cleaning Instructions.
For Synthes Trauma Recon System.
Cleaning Chart
Synthes (USA)
1302 Wrights Lane East
West Chester, PA 19380
Telephone: (610) 719-5000
To order: (800) 523-0322
Fax: (610) 251-9056
Synthes (Canada) Ltd.
2566 Meadowpine Boulevard
Mississauga, Ontario L5N 6P9
Telephone: (905) 567-0440
To order: (800) 668-1119
Fax: (905) 567-3185
© 2009 Synthes, Inc. or its affiliates. All rights reserved. Synthes is a trademark of Synthes, Inc. or its affiliates. Printed in U.S.A. 5/11 J9539-B
www.synthes.com
7
Load washing machine basket
Place articles in the specially designed trays for machine
washing supplied by Synthes. Ensure that all cannulations
(handpiece and attachments), if applicable, are positioned
vertically, i.e., in an upright position as shown.
Note: Never immerse the handpiece, lid or attachments in
an ultrasonic bath or any other liquid as this could decrease
the service life of the system.
8
Mechanical cleaning cycle parameters
Minimum
Cycle time (minutes) Temperature/water Type of detergent
Rinse 2 Cold tap water N/ A
Prewash 1 Warm tap water (40ºC) Neutral pH enzymatic cleaner
Cleaning 2 Warm tap water (45ºC) Neutral pH enzymatic cleaner
Rinse 5 DI (Deionized) water N/ A
Thermal disinfection 5 Hot DI water (93ºC) N/ A
Dry 40 90ºC N/A
9
Inspect devices
Remove all devices from washing basket. Inspect the
cannulations, coupling sleeves, etc. for visible soil. If
necessary, repeat the manual preclean/mechanical cleaning
cycle. Please refer to Synthes Battery Powered Equipment
Lubrication Chart after cleaning and prior to sterilization.
Automated Cleaning Instructions with Manual Precleaning
Automated Cleaning Instructions. For Synthes Trauma Recon System.
3
Remove debris
Rinse device under running cold tap water for a
minimum of two minutes. Use a sponge, soft lint-free
cloth and/or soft-bristled brush to assist in the removal of
gross soil and debris. Clean all cannulations, handpiece
and attachments, with the Cleaning Brush (516.101).
Manipulate all moving parts such as the triggers, release
sleeves for attachments, mode switch, etc. under running
cold tap water to loosen and remove gross debris.
Note: Do not use pointed objects for cleaning.
Automated Cleaning Instructions with Manual Precleaning.
4
Spray with solution
Spray and wipe device using a neutral pH enzymatic
solution or foam spray for a minimum of two minutes.
Follow the enzymatic detergent manufacturer’s
instructions for use for correct temperature, water
quality and concentration/dilution.
6
Rinse with tap water
Rinse device thoroughly using running cool to lukewarm
tap water for a minimum of two minutes. Use a syringe,
pipette or water jet to flush lumens and channels. Actuate
joints, handles and other moveable device features in
order to rinse thoroughly under running water.
2
Return batteries to charger
Return batteries/power modules to charger after
each use.
Important: Do not sterilize batteries or power
modules! Batteries/power modules will no
longer function.
This information is provided in accordance with ISO 17664 and ANSI/AAMI ST81.
1
Disassemble
Disassemble device if applicable, and remove the battery
or power module from casing or handpiece.
Batteries/power modules and charger may be wiped
with a cloth. DO NOT disinfect, immerse, or sterilize
batteries/power modules.
Note: The screw coupling of the Sagittal Saw
Attachment, long, 05.001.224, completely
disassembles from saw for cleaning.
Visually inspect device. Repeat Steps 3–6until novisible soil remains.
Note: Manual precleaningas describedabovemust befollowedbythe
mechanical cleaningprocedure.
– Mechanical cleaning/disinfection is an additional
stress for power equipment, especially for seals and
bearings. Therefore, systems must be properly
lubricated and serviced regularly.
Note: Refer to back of chart for automatic washing instructions.
Important:
– Manual precleaning prior to mechanical cleaning/
disinfection is important to ensure cannulations and
other difficult to access sites are clean.
– Alternative cleaning/disinfection procedures other than
the procedure described below (including manual
precleaning) have not been validated by Synthes.
5
Clean with detergent
Clean device manually under running warm water using
a neutral pH enzymatic detergent for a minimum of five
minutes. Manipulate all moving parts under running
water. Use a soft-bristled brush and/or soft lint-free
cloth to remove all visible soil and debris. Follow the
enzymatic detergent manufacturer’s instructions for
use for correct temperature, water quality and
concentration/dilution.
Note: For Sagittal Saw Attachment, long, 05.001.224,
process the Screw Coupling in an ultrasonic cleaner
using a neutral pH enzymatic cleaner for a minimum
of 15 minutes.
Automated Cleaning Instructions.
For Synthes Trauma Recon System.
Cleaning Chart
Synthes (USA)
1302 Wrights Lane East
West Chester, PA 19380
Telephone: (610) 719-5000
To order: (800) 523-0322
Fax: (610) 251-9056
Synthes (Canada) Ltd.
2566 Meadowpine Boulevard
Mississauga, Ontario L5N 6P9
Telephone: (905) 567-0440
To order: (800) 668-1119
Fax: (905) 567-3185
© 2009 Synthes, Inc. or its affiliates. All rights reserved. Synthes is a trademark of Synthes, Inc. or its affiliates. Printed in U.S.A. 5/11 J9539-B
www.synthes.com
7
Load washing machine basket
Place articles in the specially designed trays for machine
washing supplied by Synthes. Ensure that all cannulations
(handpiece and attachments), if applicable, are positioned
vertically, i.e., in an upright position as shown.
Note: Never immerse the handpiece, lid or attachments in
an ultrasonic bath or any other liquid as this could decrease
the service life of the system.
8
Mechanical cleaning cycle parameters
Minimum
Cycle time (minutes) Temperature/water Type of detergent
Rinse 2 Cold tap water N/ A
Prewash 1 Warm tap water (40ºC) Neutral pH enzymatic cleaner
Cleaning 2 Warm tap water (45ºC) Neutral pH enzymatic cleaner
Rinse 5 DI (Deionized) water N/ A
Thermal disinfection 5 Hot DI water (93ºC) N/ A
Dry 40 90ºC N/A
9
Inspect devices
Remove all devices from washing basket. Inspect the
cannulations, coupling sleeves, etc. for visible soil. If
necessary, repeat the manual preclean/mechanical cleaning
cycle. Please refer to Synthes Battery Powered Equipment
Lubrication Chart after cleaning and prior to sterilization.
Automated Cleaning Instructions with Manual Precleaning

Other manuals for TRS

1

This manual suits for next models

1

Other Synthes Medical Equipment manuals

Synthes Hybrid Fixator User manual

Synthes

Synthes Hybrid Fixator User manual

Synthes PEAK User manual

Synthes

Synthes PEAK User manual

Synthes Colibri II User manual

Synthes

Synthes Colibri II User manual

Synthes LCP Distal Humerus Plates User manual

Synthes

Synthes LCP Distal Humerus Plates User manual

Synthes Pelvic C-Clamp User manual

Synthes

Synthes Pelvic C-Clamp User manual

Synthes MAXFRAME AUTOSTRUT User manual

Synthes

Synthes MAXFRAME AUTOSTRUT User manual

Synthes SPOTLIGHT User manual

Synthes

Synthes SPOTLIGHT User manual

Synthes ComPact Air Drive II User manual

Synthes

Synthes ComPact Air Drive II User manual

Synthes FMS 283573 User manual

Synthes

Synthes FMS 283573 User manual

Synthes Expert Tibial Nail User manual

Synthes

Synthes Expert Tibial Nail User manual

Synthes DFN User manual

Synthes

Synthes DFN User manual

Synthes Hybrid Fixator Proximal Tibia Frame User manual

Synthes

Synthes Hybrid Fixator Proximal Tibia Frame User manual

Synthes Delta Xtend User manual

Synthes

Synthes Delta Xtend User manual

Synthes Pelvic C-Clamp II User manual

Synthes

Synthes Pelvic C-Clamp II User manual

Synthes Spine Cable System Quick start guide

Synthes

Synthes Spine Cable System Quick start guide

Synthes Small Stature USS User manual

Synthes

Synthes Small Stature USS User manual

Synthes ACTIS DUOFIX User manual

Synthes

Synthes ACTIS DUOFIX User manual

Synthes EXPEDIUM VERSE Parts list manual

Synthes

Synthes EXPEDIUM VERSE Parts list manual

Synthes LCP User manual

Synthes

Synthes LCP User manual

Synthes VEPTR Quick start guide

Synthes

Synthes VEPTR Quick start guide

Synthes Maintenance Unit User manual

Synthes

Synthes Maintenance Unit User manual

Synthes USS II User manual

Synthes

Synthes USS II User manual

Synthes LUPINE BR Anchor User manual

Synthes

Synthes LUPINE BR Anchor User manual

Synthes Titanium Trochanteric FixationNail System User manual

Synthes

Synthes Titanium Trochanteric FixationNail System User manual

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.