1This instruction manual............................................................................................5
1.1 Precautions ..................................................................................................5
1.2 Explanation of symbols and typeface...........................................................5
2Safety.......................................................................................................................6
2.1 Intended use.................................................................................................6
2.2 Predictable incorrect application..................................................................6
2.3 Safe handling ...............................................................................................6
2.4 Staff qualification..........................................................................................6
2.5 Calibration / adjustment ...............................................................................7
2.6 Modifications to the product .........................................................................7
2.7 Usage of spare parts and accessories.........................................................7
2.8 Liability information.......................................................................................7
3Product description..................................................................................................8
3.1 Scope of delivery..........................................................................................8
3.2 Properties and functions ..............................................................................8
3.3 Technical specifications ...............................................................................9
3.4 Approvals, tests and conformities..............................................................11
4Transport and storage ...........................................................................................11
5Commissioning......................................................................................................11
5.1 Using the IR printer ....................................................................................11
5.2 Using the Bluetooth Smart printer..............................................................12
6Operation...............................................................................................................14
6.1 Measuring mode.........................................................................................14
6.2 Changing the units .....................................................................................18
6.3 Setting time and date .................................................................................19
6.4 Setting the display......................................................................................19
6.5 Setting the key tone and the alarm tone ....................................................20
6.6 Displaying device information ....................................................................21
6.7 Generating QR-CODE................................................................................22
6.8 Data Logger function (option).....................................................................22
6.9 "Pressure Measurement" program.............................................................25
6.10 "Pressure Loss Measurement" program ....................................................26
6.11 "Pitot Measurement" program (option).......................................................28
6.12 "Leakage measurement" program (option)................................................30
6.13 "Leakage rate measurement" program (option).........................................32
7Memory mode and memory structure ...................................................................38
7.1 How to save................................................................................................38
7.2 Database Memory (option).........................................................................41
7.3 Entering the user address..........................................................................48

4 S4600-ST®
8Battery management.............................................................................................49
8.1 Battery operation / charging mode.............................................................49
8.2 Charging the batteries................................................................................49
9Maintenance..........................................................................................................51
10 Troubleshooting.....................................................................................................51
11 Disposal.................................................................................................................52
12 Warranty................................................................................................................52
13 Copyright ...............................................................................................................52
14 Customer satisfaction............................................................................................52
15 Addresses..............................................................................................................52
16 Appendix................................................................................................................53
16.1 DIN EN 50379 certificate............................................................................53

This instruction manual
S4600-ST® 5
1This instruction manual
This instruction manual is part of the product.
Read this manual before using the product.
Keep this manual during the entire service life of the product
and always have it readily available for reference.
Always hand this manual over to future owners or users of the
product.
1.1 Precautions
WARNING TERM
Type and source of the danger are shown here.
Precautions to take in order to avoid the danger are shown
here.
There are three different levels of warnings:
Warning term
Meaning
DANGER
Immediately imminent danger!
Failure to observe the information will result in
death or severe injuries.
WARNING
Possibly imminent danger!
Failure to observe the information may result in
death or severe injuries.
CAUTION
Dangerous situation!
Failure to observe the information may result in
minor or severe injuries as well as damage to
property.
1.2 Explanation of symbols and typeface
Symbol
Meaning
Prerequisite for an activity
Activity consisting of a single step
1.
Activity consisting of a several steps
Result of an activity
Bulleted list
Text
Indication on display
Highlighting
Highlighting

Safety
6 S4600-ST®
2Safety
2.1 Intended use
The pressure measuring instrument S4600-ST©may only be used to
measure pressure of gaseous media in non-hazardous areas. The
measuring instrument S4600-ST®must not be used for pressure
measurement at gas lines.
Any use other than the application explicitly permitted in this
instruction manual is not permitted.
2.2 Predictable incorrect application
The pressure measuring instrument S4600©must never be used in
the following cases:
Hazardous area (Ex)
If the device is operated in hazardous areas, sparks may cause
deflagrations, fires or explosions.
Applications involving persons and animals
2.3 Safe handling
This product represents state-of-the-art technology and is made
according to the pertinent safety regulations. Each device is
subjected to a function and safety test prior to shipping.
Operate the product only when it is in perfect condition. Always
observe the operating instructions, all pertinent local and
national directives and guidelines as well as the applicable
safety regulations and directives concerning the prevention of
accidents.
Extreme environmental conditions have negative effects on the
function of the product.
Protect the product from shocks.
Only use the product in rooms.
Protect the product from humidity.
2.4 Staff qualification
The product may only be mounted, commissioned, operated,
maintained, shut down and disposed of by qualified, specially trained
staff.
Electrical work may only be performed by trained electricians and in
compliance with all applicable local and national directives.

Safety
S4600-ST® 7
2.5 Calibration / adjustment
S4600-ST®must be calibrated on an annual basis. Calibration and
adjustment may only be performed by the manufacturer or by third
parties authorised by the manufacturer.
2.6 Modifications to the product
Changes or modifications made to the product by unauthorised
persons may lead to malfunctions and are prohibited for safety
reasons.
2.7 Usage of spare parts and accessories
Usage of unsuitable spare parts and accessories may cause
damage to the product.
Use only genuine spare parts and accessories of the
manufacturer.
2.8 Liability information
The manufacturer shall not be liable in any form whatsoever for
direct or consequential damage resulting from failure to observe the
technical instructions, guidelines and recommendations.
The manufacturer or the sales company shall not be liable for costs
or damages incurred by the user or by third parties in the usage or
application of this device, in particular in case of improper use of the
device, misuse or malfunction of the connection, malfunction of the
device or of connected devices. The manufacturer or the sales
company shall not be liable for damage whatsoever resulting from
any use other than the use explicitly permitted in this instruction
manual.
The manufacturer shall not be liable for misprints.

Product description
8 S4600-ST®
3Product description
3.1 Scope of delivery
Scope of delivery:
Measuring instrument S4600-ST®
USB mains adaptor
3.2 Properties and functions
1
Cancel ("Clear" key) / direct
access
2
Up key
3
On/off key
4
Protective sleeve
5
Down key
6
Menu / Enter key
7
Display
8
Pressure connections
9
Sensor
10
Charger/power supply unit
connection
11
IR printer connection
12
MicroSD card slot

Product description
S4600-ST® 9
3.3 Technical specifications
Table 1: Device description
Parameters
Value
General specifications
Dimensions of housing
with protective sleeve
(H x W x D)
143 mm x 66 mm x 37 mm
(5.6 in x 2.6 in x 1.5 in)
Weight (with protective
sleeve)
Approx. 220 g (7.76 oz)
Housing material
Polyamide (PA)
Display
High-resolution graphical 2.8" TFT display
(240 x 320).
Data communication
Infrared printer interface.
Bluetooth®Smart interface (Bluetooth®low
energy).
Printer
External infrared thermal printer
(EUROprinter)
Memory
MicroSD card with folder/file structure
Operating temperature range
Ambient
0 °C to +40 °C (+32 °F to +104 °F)
Medium
0 °C to +40 °C (+32 °F to +104 °F)
Storage
-20 °C to +50 °C (-4 °F to +122 °F)
Atmospheric pressure range
Ambient
750 hPa to +1100 hPa
Humidity range
Ambient
20 % rH to 80 % rH
Supply voltage
Battery operation
Lithium-ion battery 3.6 V / 1800 mAh
Mains operation
Mains adaptor (USB)
Electrical safety
Degree of protection
IP 40 EN 60529

Product description
10 S4600-ST®
Parameters
Value
Electromagnetic compatibility (EMC)
Interference
EN 55022 (VDE 0875-22)
Noise immunity
EN 61000 (VDE 0847-4-2)
ESD
EN 61000-4-2
Table 2: Device specifications (pressure measurement)
S4602-ST
S4601-ST
S4610-ST
S4650-ST
S4680-ST
S4699 ST
Measuring
range [mbar]
- nominal
- maximum
± 20
± 20
± 150
± 180
± 1,000
± 1,500
± 5,000
± 7,000
± 8,000
± 10,000
± 18.000
± 20.000
Max.
overpressure
[bar]
0.25
1.35
16.0
16.0
16.0
28.0
Resolution
[mbar]
0.001
0.01 (<99.99)
0.1 (>100.0)
0.1 (< 999.9)
1 (> 1000)
0.1 (< 999.9)
1 (> 1000)
0.1 (< 999.9)
1 (> 1000)
1
Accuracy
[% of
measured
value]
0.003 mbar
or
0.5 ± 1 digit
0.03 mbar
or
0.5 ± 1 digit
0.3 mbar
or
0.5 ± 1 digit
0.7 mbar
or
0.5 ± 1 digit
1.2 mbar
or
0.5 ± 1 digit
3 mbar
or
0.5 ± 1 Digit
Mechanical
connections
2 x standard
(8 mm)
2 x standard
(8 mm)
2 x standard
(8 mm)
2 x standard
(8 mm)
2 x standard
(8 mm)
2x Festo
(3 mm)
Table 3: Additional device specifications
Parameters
Value
Barometric pressure measurement
Measuring range
500 –1150 hPa
Max. deviation
± 1.5 hPa
Resolution
± 10 hPa
Sensor
Semiconductor sensor

Transport and storage
S4600-ST® 11
3.4 Approvals, tests and conformities
The S4600-ST®pressure measuring instrument is approved as per
the German Federal Immission Act (1. BImSchV) and EN 50379-2
and is TÜV-tested. It complies with the EMC Directive
(2004/108/EC).
4Transport and storage
CAUTION
Damage to the device due to improper transport.
Do not throw or drop the device.
CAUTION
Damage to the device due to improper storage.
Protect the device from shock when storing it.
Store the device in a clean and dry environment.
Only store the device within the permissible temperature
range.
5Commissioning
5.1 Using the IR printer
For data transmission from the measuring instrument to the
corresponding IR printer (EUROprinter), align the bottom of the
S4600-ST®and the printer as shown in the following illustration.
Switch on the printer. Start the data transmission via the menu of the
measuring instrument. The measurement log is printed. Keep a
minimum distance of approx. 25 cm! (Max. approx. 70 cm).
Fig. 1: Alignment of the two devices for printing
CAUTION
Transmission errors due to incorrect alignment
The transmission path must be straight and unobstructed.

Commissioning
12 S4600-ST®
5.2 Using the Bluetooth Smart printer
Measuring Data could be also transferred via Bluetooth Smart to the
“EUROprinter-BLE”. Activate the Bluetooth Smart in the S4600-ST®
Settings and on the printer. To activate the Bluetooth Smart on the
printer press the keys “OFF” and “ON” at once. A blue flashing LED
means activated Bluetooth Smart otherwise the IR modus is
activated. The Bluetooth Smart connection between device and
printer is described below:

Commissioning
S4600-ST® 13
Select the detected printer to activate the printer.
The IR modus could be set by deactivating the „Bluetooth Printer“ in
the settings menu. Set the printer to IR modus as well. Press the
keys „OFF“ and „ON“ at once and the blue flashing LED goes out.

Operation
14 S4600-ST®
6Operation
Switch on the device: Briefly press the "On/Off" key .
6.1 Measuring mode
Program selection
The program selection section displays the available programs as
symbols. Programs can be selected and started.
Fig. 2: Program selection in the start menu
(example: S4600-ST®Pressure Measurement)
Main display
1
Status line
2
Measured values
3
Information line
4
Optional information
line
5
Individual measured
value
Fig. 3: Display when measuring program is running

Operation
S4600-ST® 15
Status line
The status line provides information on the status of relevant
parameters data such as battery, Hold function, Pressure function,
Bluetooth®Smart function, MicroSD card, etc. The type of
information displayed depends on the operating mode and function-
specific criteria.
Information line
The information line provides information on the current measuring
program, time, date, etc.
Main menu
Display the main menu with the "Menu/Enter" key.
The main menu provides access to the main functions of the device.
All other functions and settings are accessed via submenus.

Operation
16 S4600-ST®
Optional information line
The optional information line provides additional information on the
corresponding measured value such as measuring velocity of the
pressure sensor, barometric pressure sensor during Pitot
measurement, etc.
Direct Access menu
Print measured value / terminate measurement.
Press the "Clear" key to display the Direct Access menu. In the
Direct Access Menu, you can print the measured values via the
"Menu/Enter" key or terminate the measurement and return to the
Start menu.
As soon as the print command is chosen, the measurement log is
printed in parallel with the measuring task (multitasking function),
i.e. the measurement mode remains active without limitations.
Edit mode
The Edit mode is activated if you select certain submenus so that
you can edit the appropriate values, such as the interval time for the
data logger function.
Editing the interval time of the data logger function
For example, if you want to change the interval time for the Data
Logger function, you must first select the appropriate line using the
arrow keys and then confirm with the "Menu/Enter" key. Once the
background colour of the line changes to blue, you are in Edit mode
and can change the corresponding value using the arrow keys. The
change must be confirmed with the "Menu/Enter" key. The Edit mode
is activated for the next number. Once you have changed and
confirmed all numbers, the Edit mode is disabled and the blue
background disappears.

Operation
S4600-ST® 17
Fig. 4: Edit mode, example: changing the interval time
Switching off the device, automatic Off function
Switching off the device: Briefly press the "On/Off" key and
confirm with the "Menu/Enter" key. The S4600-ST®measuring
instrument also features a time-dependent automatic Off function;
this function can be enabled or disabled.
The switch off time can be changed in Edit mode. The Edit mode is
described in chapter 6.1, page 16.

Operation
18 S4600-ST®
6.2 Changing the units
Displaying the main menu.
Various units are available for pressure measurements.

Operation
S4600-ST® 19
6.3 Setting time and date
Changing time and date
Time and date can be changed in Edit mode. The Edit mode is
described in chapter 6.1, page 16. The device automatically
considers leap years and daylight-saving time/wintertime; the time
and date formats are adjustable.
6.4 Setting the display
The intelligent power management of S4600-ST©allows for
optimisation of the battery life.
Three display settings are available: "Normal", "Automatic" and "Eco
Mode". The selected setting influences the battery life, see chapter
8.1, page 49.
Change the display setting.

Operation
20 S4600-ST®
6.5 Setting the key tone and the alarm tone
Setting the key tone and the alarm tone.
There are four sound levels for the key tone and the alarm tone:
Off
Low
Medium
Loud
Table of contents
Other SYSTRONIK Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

sparklike
sparklike Handheld instruction manual

Sartorius
Sartorius LMA200PM operating instructions

Burster
Burster DIGIFORCE 9310 manual

SPL-Laboratory
SPL-Laboratory Race Meter 2018 user manual

MetroPWR
MetroPWR FX773 user manual

Audio Precision
Audio Precision APx555 Installation Instructions and Specifications