TA Instruments DHR Series User manual

Revision H Issued November 2013
Getting Started Guide
DHR Series and AR Series
Asphalt Submersion Cell

Page 2 DHR/AR Series Asphalt Submersion Cell Getting Started Guide
Notice
The material contained in this manual, and in the online help for the software used to support this instru-
ment, is believed adequate for the intended use of the instrument. If the instrument or procedures are used
for purposes other than those specified herein, confirmation of their suitability must be obtained from TA
Instruments. Otherwise, TA Instruments does not guarantee any results and assumes no obligation or liabil-
ity. TA Instruments also reserves the right to revise this document and to make changes without notice.
TA Instruments may have patents, patent applications, trademarks, copyrights, or other intellectual prop-
erty covering subject matter in this document. Except as expressly provided in written license agreement
from TA Instruments, the furnishing of this document does not give you any license to these patents, trade-
marks, copyrights, or other intellectual property.
TA Instruments Operating Software, as well as Module, Data Analysis, and Utility Software and their asso-
ciated manuals and online help, are proprietary and copyrighted by TA Instruments. Purchasers are granted
a license to use these software programs on the module and controller with which they were purchased.
These programs may not be duplicated by the purchaser without the prior written consent of TA Instru-
ments. Each licensed program shall remain the exclusive property of TA Instruments, and no rights or
licenses are granted to the purchaser other than as specified above.
©2013 by
TA Instruments — Waters LLC
159 Lukens Drive
New Castle, DE 19720

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 3
Introduction
Important: TA Instruments Manual Supplement
Please click the TA Manual Supplement link to access the following important information
supplemental to this Getting Started Guide:
• TA Instruments Trademarks
• TA Instruments Patents
• Other Trademarks
• TA Instruments End-User License Agreement
• TA Instruments Offices

Page 4 DHR/AR Series Asphalt Submersion Cell Getting Started Guide
Notes, Cautions, and Warnings
This manual uses NOTES, CAUTIONS, and WARNINGS to emphasize important and critical
instructions. In the body of the manual these may be found in the shaded box on the outside of the page.
NOTE: A NOTE highlights important information about equipment or procedures.
CAUTION: A CAUTION emphasizes a procedure that may damage equipment or cause loss of data
if not followed correctly.
MISE EN GARDE: UNE MISE EN GARDE met l'accent sur une procédure susceptible d'endom-
mager l'équipement ou de causer la perte des données si elle n'est pas correctement suivie.
Regulatory Compliance
Safety Standards
For Canada
CAN/CSA-C22.2 No. 61010-1 Safety requirements for electrical equipment for measurement, control, and
laboratory use, Part 1: General Requirements.
CAN/CSA-C22.2 No. 61010-2-010 Particular requirements for laboratory equipment for the heating of
materials.
For European Economic Area
(In accordance with Council Directive 2006/95/EC of 12 December 2006 on the harmonization of the laws
of Member States relating to electrical equipment designed for use within certain voltage limits.)
EN 61010-1:2001 Safety requirements for electrical equipment for measurement, control, and laboratory
use, Part 1: General Requirements + Amendments.
EN 61010-2-010:2003 Particular requirements for laboratory equipment for the heating of materials +
Amendments.
For United States
UL61010-1:2004 Electrical Equipment for Laboratory Use; Part 1: General Requirements.
UL61010A-2-010:2002 Particular requirements for laboratory equipment for the heating of materials +
Amendments.
A WARNING indicates a procedure that may be hazardous to the operator or
to the environment if not followed correctly.
Un AVERTISSEMENT indique une procédure qui peut être dangereuse pour
l'opérateur ou l'environnement si elle n'est pas correctement suivie.

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 5
Electromagnetic Compatibility Standards
For Australia and New Zealand
AS/NZS CISPR11:2004 Limits and methods of measurement of electronic disturbance characteristics of
industrial, scientific and medical (ISM) radio frequency equipment.
For Canada
ICES-001 Issue 4 June 2006 Interference-Causing Equipment Standard: Industrial, Scientific, and Medical
Radio Frequency Generators.
For the European Economic Area
(In accordance with Council Directive 2004/108/EC of 15 December 2004 on the approximation of the
laws of the Member States relating to electromagnetic compatibility.)
EN61326-1:2006 Electrical equipment for measurement, control, and laboratory use-EMC requirements-
Part 1: General Requirements. Emissions: Meets Class A requirements per CISPR 11. Immunity: Per
Table 1 - Basic immunity test requirements.
For the United States
CFR Title 47 Telecommunication Chapter I Federal Communications Commission, Part 15 Radio
frequency devices (FCC regulation pertaining to radio frequency emissions).

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 6
Safety
Do not attempt to service this instrument, as it contains no user-serviceable components.
Required Equipment
While operating this accessory, you must wear eye protection that either meets or exceeds ANSI Z87.1
standards. Additionally, wear protective clothing that has been approved for protection against the
materials under test and the test temperatures.
Instrument Symbols
The following label is displayed on the accessory for your protection:
Please heed the warning labels and take the necessary precautions when dealing with these areas. This
Getting Started Guide contains cautions and warnings that must be followed for your own safety.
Symbol Explanation
This symbol indicates that a hot surface may be present. Take care not to touch
this area or allow any material that may melt or burn come in contact with this
hot surface.
Ce symbole indique la présence possible d'une surface chaude. Prenez soin de
ne pas toucher cette zone ou de laisser un matériau susceptible de fondre ou de
brûler entrer en contact avec cette surface chaude.

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 7
Table of Contents
Introduction ......................................................................................................................................... 3
Important: TA Instruments Manual Supplement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Notes, Cautions, and Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Regulatory Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Safety Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Electromagnetic Compatibility Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Instrument Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Chapter 1:
Introducing the Asphalt Submersion Cell .................................................................... 9
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Techniques Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Chapter 2:
Installing the System ..................................................................................................... 10
Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Important Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
System Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Chapter 3:
Preparing and Maintaining the Rheometer ................................................................ 14
Filling and Draining the Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Selecting Geometries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Filters and Plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Installing and Removing Plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Chapter 4:
Troubleshooting ............................................................................................................. 17
Solving Common Asphalt Submersion Cell Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Flow Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Temperature Control Cell Overflows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Poor Temperature Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Leaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Appendix A:
Julabo Circulator (HE Style) .................................................................................... 19
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Setting Up the CCFC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Using the Membrane Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 8
Setting the Excess Temperature Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Appendix A:
REVISED Julabo Circulator (HE Style) ................................................................. 23
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Startup Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Index ................................................................................................................................................... 25

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 9
Chapter 1:
Introducing the Asphalt Submersion Cell
Overview
The TA Instruments Asphalt Submersion Cell is a Smart Swap™ accessory designed to be used with the
DHR Series and AR Series Rheometers. For the general operation of those instruments, including use of
the Smart Swap system, see the appropriate operator's manual.
The Submersion Cell is specifically designed to satisfy the latest Federal Highway Administration
(FHWA) requirements for asphalt binder testing. When used with the cell, the DHR Series and AR Series
Rheometers meet the requirements of Dynamic Shear Rheometer (DSR) as defined by the FHWA.
The cell is designed to be used with a Julabo Computer Controlled Fluid Circulator (CCFC), normally
model FP35-HE. The operation of the cell and CCFC is directly controlled through TRIOS Software. For
full details of the operation of the Julabo and Safety information, see the manufacturers operating manual.
A summary is provided in Appendix A of this manual.
Historically, asphalt characterization was based on measurements of penetration, capillary viscosity and
ductility. While these measurements were repeatable, they could not be easily correlated to end-use perfor-
mance characteristics such as rutting or cracking. Recently an exhaustive multiyear study (SHRP) by
FHWA resulted in the adoption of alternative measurements that solve this correlation problem. These new
measurements provide a set of alternative criteria for selecting asphalt binders and aggregate mixes to meet
specific end-use conditions.
Techniques Used
For asphalt binders, the new measurements are based on the Bending Beam Rheometer, the Direct Tension
Tester, and the Dynamic Shear Rheometer (DSR). The first two techniques evaluate low temperature per-
formance, whereas the DSR evaluates performance over a broader temperature range (5 to 85°C).
Specifically, the DSR test measurements include:
• AASHTO T315 & ASTM D7175: Determining the Rheological Properties of Asphalt Binder Using a
Dynamic Shear Rheometer (DSR).
• AASHTO R29 & M320: Grading or Verifying the Performance Grade of an Asphalt Binder.
• AASHTO TP70 & ASTM D7405: Multiple Stress Creep and Recovery (MSCR) of Asphalt Binder
Using a Dynamic Shear Rheometer
• AASHTO XXXX: High stress repeated creep
• AASHTO XXXX: Determining the mixing and compaction temperature of asphalt binder
This manual should be read in conjunction with these Standard Methods. The cell can be used with TRIOS
Software, but for full automated testing using the standards listed above, Rheology Navigator and
Rheology Advantage are required. More information about the operation of this software can be found in
CSA II Asphalt Rheometerwith Rheology Navigator Software PDF.

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 10
Chapter 2:
Installing the System
Components
The Asphalt Submersion Cell consists of the following items:
• Smart Swap Asphalt Submersion Cell
• 8-mm parallel plate set
• 25-mm parallel plate set
• Lower plate removal tool
• In-line water filter housing unit
• Filter material
• Mold (1 pair)
• Solenoid box
• Interconnecting cables
• Interconnecting water tubing and adaptors
• Manual tap
Important Instructions
The following items are important to the installation and use of the Asphalt Submersion Cell:
• For general instructions on the use of the Julabo CCFC, see the manufacturer's operating manual. For
settings particular to the operation of Asphalt Submersion Cell, see Appendix A of this manual.
• Position the CCFC close to the rheometer to keep the tubing run as sort as possible. However, locate
the circulator slightly lower than the rheometer, otherwise water can siphon into the Submersion Cell
after the system has been turned off.
• Position the CCFC so that its top reservoir fill line opening is lower than the Asphalt Submersion Cell
Overflow.
• For the standard testing temperatures, use deionized water as the circulating fluid. Below 5°C, it may
be necessary to use a glycol/water mixture.

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 11
System Connections
A schematic of the Asphalt Submersion Cell is shown below.
Figure 1 Schematic of the Asphalt Submersion Cell.
For the AR Series Rheometers, mount the solenoid box on the rear of the rheometer using the four 3-mm
panhead Pozidrive screws supplied with the instrument, as shown below.
Figure 2 Solenoid box mounted on rear of rheometer (DHR shown).
For the DHR Series Rheometers, fit the adapter bracket to the back of the instrument before mounting the
solenoid box.
Lower drain
Event
Upper drain
Water out
Event B
Inline filter
Not used with
AsphaltSubmersion Cell
Adapter bracket

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 12
Mount the Asphalt Submersion Cell on the rheometer using the Smart Swap connector, as shown below.
Figure 3 Asphalt Submersion Cell mounted on rheometer (DHR shown).
Make the following connections, referring to the diagrams for guidance. All tubing should be cut to length.
• The lower drain port on the Submersion Cell to the lower drain port on the solenoid box using the
6-mm tubing (blue) provided.
• The upper drain connector on the Submersion Cell to the upper drain port on the solenoid box using
the 6-mm tubing (blue) provided. Insert the in-line filter into the line, as shown in Figure 4.
• The water inlet port on the Submersion Cell to the fluid outlet port on the CCFC using the 4-mm tub-
ing (white) provided. It may be necessary to replace the connector on the CCFC with that provided.
The manual tap should be inserted into the line. This can be used to cut off the supply of water to the
cell in the event of overflow, but should otherwise be in the fully open position.
• The water outlet port on solenoid box to the fluid Inlet connector on the CCFC using the 6-mm tubing
(blue). It may be necessary to replace the connector on the CCFC with that provided.
• The overflow port on the Submersion Cell to a suitable container/drain using the 6-mm tubing (blue).
•TheEvent connection on the Solenoid box to the Event B connection on the rheometer rear using the
9-pin to Lemo cable provided.
•TheRS232 auxiliary port on the Rheometer Electronics box to the SERIAL port on the CCFC using
the 9 to 9-pin RS232 cable provided.
• The Lemo connector on the Submersion Cell Pt100 cable to the Pt100 connection on the CCFC.
In-line filter
Upper drain
Lower drain
Water inlet
Overflow
Pt 100 cable

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 13
Figure 4 Connections for Julabo CCFC with HE style head.
Pt 100
Serial port
Chiller

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 14
Chapter 3:
Preparing and Maintaining the Rheometer
Filling and Draining the Cell
The Environmental section of the Instrument Control panel in TRIOS software contains icons for filling
and draining the cell. These icons are only present when the cell is connected.
Selecting Geometries
You may need to change the geometry set installed, depending upon the desired use:
• To test Original Binder and RTFO residue, install the 25-mm geometry set.
• To test for PAV residue, install the 8-mm set.
The components are shown below.
Figure 5

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 15
Routine Maintenance
It is recommended that the following procedures be carried out on a regular basis:
•Circulation fluid: This should be changed once per week, or if the water becomes contaminated.
•Inline water filter: The filter should be removed, and cleaned or replaced, when visibly contaminated.
See the next section for details.
•Asphalt Temperature Control Cell filters: These should be changed when visibly contaminated. See
the next section for details
•TheFlow Deflector Turret, shown below, should be unscrewed and cleaned.
Figure 6 Flow Deflector Turret.
Filters and Plates
Before the Asphalt Submersion Cell is used, you will need to perform several functions:
• Check the drain and inline filters on the cell. The filters should be clean and free of debris.
aTo replace the lower drain filter, which is a 6-mm single hole punched from the filter sheet, use
tweezers to place the filter in the drain. Then secure the filter with a quarter turn.
bTo replace the inline filter, disassemble the filter-housing unit and remove the tubing. Cut a filter
from the sheet provided—it should be approximately 25 mm long and 10 mm wide. Then use a pair
of tweezers to replace the filter.
CAUTION: Do not pack the filter material too tightly as this may result in an overflow situation.
MISE EN GARDE: Ne serrez pas trop l'emballage du matériel du filtre car cela risque de provoquer
un débordement.
• Clean the plates used in the cell regularly. To remove or install the plates, see the next section. The
plates are manufactured from stainless steel so they can be cleaned in a solvent bath.

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 16
Installing and Removing Plates
To install or remove the cell plates, follow these instructions:
Upper Plate
The upper plate is simply pushed onto the end of the shaft, and fixed in position by screwing up the draw-
rod until finger tight. To remove it, reverse the procedure.
Lower Plate
The lower plate is screwed into the bottom of the asphalt bath. Use the removal tool provided to tighten
and loosen this plate (see the figure below).
Figure 7 Lower plate removal/installation.
Removal tool
Lower plate

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 17
Chapter 4:
Troubleshooting
Solving Common Asphalt Submersion Cell Problems
This chapter provides answers to some common problems that can occur while using the Asphalt Submer-
sion Cell. If you cannot solve the problem using one of the suggestions on the following pages, contact
your TA Instruments Service Representative for assistance.
Flow Problems
The flow in and out of the temperature control cell is balanced by the size of the tubing and the efficiency
of the pump and suction. The temperature control parameters are optimized for this flow rate. Temperature
stability is generally related to flow rates. The on/off valve should not be used to reduce the flow rate, other
than during sample loading, if splashing of the lower plate is a problem.
Temperature Control Cell Overflows
This is caused by inefficiency in the suction side of the system. If you use the on/off valve to reduce the
flow rate and stop the overflow, poor temperature control is likely to result.
Try to isolate the problem area. Check the following:
• Remove all filter material.
• Disconnect and reconnect all pipes.
• Consider replacing pipes if they are visibly contaminated.
• Remove and clean solenoid valve located behind column back panel. Replace the valve, if it appears
worn. Turn off power to the instrument before removing the back panel.
Poor Temperature Control
Try the following solutions to poor temperature control:
• Make sure that the on/off valve is fully open.
• Make sure that the configuration parameters in the CCFC have not been tampered with.
• Remove the flow deflection turret and check for blockages.
• Disconnect the pipe from the inlet to the cell and measure the flow rate at the height of the cell. If the
flow rate is less than 450ml/min try replacing the pipe, or flush the CCFC with solvent.

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 18
Leaks
If you find leaks occurring, first locate the pipe that is leaking. Turn off the fluid circulator’s main power
switch before disconnecting any lines. Disconnect and reconnect the problem pipe. If problem persists, use
a sharp straight tool to trim approximately 3 mm from the end of the pipe (do not use scissors, as this will
pinch a hose flat). Then reconnect the pipe.

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 19
Appendix A:
Julabo Circulator (HE Style)
Introduction
This appendix covers those aspects of the operation of the Julabo Computer Controlled Fluid Circulator
(CCFC) with HE style head, relevant to the Asphalt Submersion Cell. The head incorporates a VFD (vac-
uum fluorescent display), and membrane keypad through which parameters settings must be entered.
Setting Up the CCFC
1Fill the CCFC to the level recommended by manufacturer. For the standard testing temperatures, use
deionized water as the circulating fluid. Below 5°C it may be necessary to use a glycol/water mixture.
A small amount of algaecide can be added at this point (TA part number 952377.901).
2Set up the connections between the rheometer, the Submersion Cell, and CCFC.
3Switch on the rheometer and the CCFC head and pump. The CCFC will perform a self-test at this point.
When this test has finished, leave the head switched on, but switch off the pump and proceed to the next
section.
Using the Membrane Keypad
The following list describes the functioning of the keypad shown below:
• MENU Key: Used to enter the various menus.
• ENTER Key: Used to save entered parameters.
• ESC Key: Used to proceed to the next menu.
• UP/DOWN Arrow Keys: Used to select values.
• LEFT/RIGHT Arrow Keys: Used to change the value of a number. Press the LEFT or RIGHT arrow
keys until the desired digit is activated then press the UP or DOWN arrow keys until the desired num-
ber is reached.
WARNING: Use caution when working with the CCFC at high temperatures. Read
and study the safety information contained in the manufacturers operating man-
ual thoroughly prior to use.
AVERTISSEMENT: Soyez prudent lorsque vous travaillez avec du CCFC à des
températures élevées. Lisez et étudiez parfaitement les consignes de sécurité
contenues dans le manuel d'utilisation du fabricant avant d'utiliser l'instrument.

DHR/AR Series Asphalt Submersion Cell Getting Started Guide Page 20
NOTE: Please note that if you enter the values as prescribed below and then for any reason go back and
change any one parameter, all the previous PID settings will be altered. This applies to parameters in the
Control Section only.
Menu Settings
Select the following menus settings using the keys as instructed above:
Pump: Adjustable Pump Capacity
Pump Speed: 2
Control: Intelligent Cascade Control Parameters
Control: External (Note: must have pt100 cable connected between bath and circulator. If not display
will default to Internal Control.)
Config: Configuration of the Circulator
Set: Serial
A-start: off
Off Mode: Pump off
Reset: no (to restore defaults select yes and then enter)
Actvar: control
Time/Date: Setting time and date
110v/120v 220/240v
Self-Tuning Off Off
Cospeed 0 0
Control PID values PID values
XP 0.6 0.5
TN 70 90
TV 7 24
XPU 4.5 Default 4.5 Default
Safety temperature
Pump Start/Stop
Enter
Other manuals for DHR Series
3
This manual suits for next models
1
Table of contents
Other TA Instruments Measuring Instrument manuals

TA Instruments
TA Instruments Nano ITC Series User manual

TA Instruments
TA Instruments TMA 2940 User manual

TA Instruments
TA Instruments TMA 2940 User manual

TA Instruments
TA Instruments DSC Q Series User manual

TA Instruments
TA Instruments AR 2000 User manual

TA Instruments
TA Instruments DHR Series User manual

TA Instruments
TA Instruments DISCOVERY XENON FLASH User manual

TA Instruments
TA Instruments TGA5500 User manual

TA Instruments
TA Instruments TMA Q400 User manual

TA Instruments
TA Instruments TGA 2050 User manual