Operating Instructions—7S1
1
10 kll. The signal is not affected by the Oscilloscope Verti-
cal Mode switch. The signal is changed by the Units/Div
switch, the DC OFFSET controls, and the NORMAL/
SMOOTH switches. The INVERT switch does not invert
the signal. Output voltage swing is limited to 2.5 volts peak
or about 6divisions above and below the graticule center
depending on the actual position. The Units/Div switch is
active but the output voltage is always proportional to the
CRT display (0.2 V/Div) up to 12 divisions maximum peak
to peak. Since the signal at the Vertical Output jack is an
equivalent time signal, it is useful for pen recorder applica-
tions when the sampling sweep unit external input is driven
by the recorder.
The OFFSET OUT monitor jack permits accurate
measurement of the DC voltage set by the DC OFFSET
controls. The output voltage has arange of -^10 to —10
volts. Unless otherwise specified on the sampling head front
panel, the monitor jack output voltage is ten times the
internal DC offset voltage. The output resistance is 10 kT2.
BASIC APPLICATIONS
Vertical Deflection Measurements
Vertical displacement of the trace on the CRT is directly
proportional to the signal at the sampling head input con-
nector installed in the 7S11. The amount of displacement
for agiven signal can be selected with the Units/Div switch.
To provide sufficient deflection for best resolution, set the
Units/Div switch so the display spans alarge portion of the
graticule. Also, when measuring between points on the dis-
play, be sure to measure consistently from either the
bottom, middle, or top of the trace. This prevents the
width of the trace from affecting the measurements.
To make avertical deflection difference measurement
between two points on the display, proceed as follows:
1. Note the vertical deflection, in graticule divisions,
between the two points on the display. Make sure the
VARIABLE control is in the CAL position.
NOTE
The vertical deflection factor is determined by the
Units/Div switch and the value of the units of
measure as stated on the sampling head front panel.
The ratio of the input signal to the resultant deflec-
tion is called the deflection factor. For example, with
the sampling head unit values stated as mV/Div, and
the Units/Div switch set to 20, the vertical deflection
factor will be 20 mV/division.
2. Multiply the divisions of vertical deflection by the
deflection factor, and the external attenuator or probe
attenuation factor (if any). The product is the voltage
difference between the two points measured.
For example, suppose you measure 4.4 divisions of de-
flection between two points on the display and the Units/
Div switch is set for 20 mV/Div. Multiply 20 millivolts/
division by 4.4 divisions, the product is 88 millivolts. This is
the voltage difference at the input connector between the
two points on the display. Now assume there is a10X
external attenuator (probe) between the input connector
and the signal source. To determine the actual signal voltage
at the source, multiply 10 (the attenuation factor of the
probe) by 88 millivolts; this product (880 millivolts or 0.88
volts) is the actual voltage at the signal source.
It is also possible to measure an instantaneous (or DC)
voltage to ground from the display. This measurement is
accomplished in the same manner, except that with no sig-
nal applied, you must first establish aground-reference
point on the CRT.
NOTE
To establish the ground reference point be sure the
input connector is either terminated by a50 El resis-
tor or coaxial cable connected to the Input.
To do this, trigger the sampling sweep unit with the
signal source and display atrace. Then, position the trace so
it is exactly aligned with one of the graticule lines. The
actual graticule line you select will be largely determined by
the polarity and amplitude of the applied signal. After
establishing the ground reference, make no further adjust-
ments with the DC OFFSET control.
Apply the signal and measure the voltage in the manner
previously described. Make ail measurements from the
established ground reference point. Accuracy of this meas-
urement is within ±3%.
If the applied signal has a relatively high DC level, the
ground-reference point and the actual signal may be so far
apart that neither will appear on the CRT. In this case, refer
to the following discussion on "Voltage Measurements
Using the DC Offset Control."
Voltage Measurements Using the DC OFFSET
Control. Unless otherwise stated on the sampling head
front panel, the DC offset voltage cancels the effect of an
applied DC voltage of up to ±1 volt at the sampling head
input. Also, accurate slideback amplitude measurements of
the applied signal can be obtained by positioning the dis-
play at various points and measuring the amount of voltage
change at the OFFSET OUT monitor jack.
2-9