
TB 9-6625-2005-35
14. Frequency Span Accuracy and Linearity
a. Performance Check
(1) Disconnect test oscillator (A9) output from
input and connect CALIBRATOR output to INPUT using
cable (B7).
(2) Position controls as listed in (a) through (f)
below: (a)DOT FREQUENCY control for 0.00
kHz. (b)RESOLUTION (Hz) control for 30
kHz. (c)FREQUENCY SPAN/DIV control for
MAX.(d)TIME/DIV control to AUTO.
(e)REFERENCE LEVEL control for -30
dBm. (f)LOG 10 dB/DIV pushbutton on.
(3) Aline second marker with second graticule
line and verify that each marker right from the second
graticule line is coincident with each graticule line. If 8th
marker is not within 5 percent (±5 division), perform b
below. (4) Disconnect CALIBRATOR output from
INPUT and connect oscilloscope calibrator (A8) marker
output to INPUT, using cable (B7).
(5) Position oscilloscope calibrator controls
for an output of 5 µs.
(6) Position controls as listed in (a) through
(e) below: (a)DOT FREQUENCY control for
4000.00 kHz. (b)FREQUENCY SPAN/DIV control for
200 kHz. (c)RESOLUTION (Hz) control for 10
kHz. (d)REFERENCE LEVEL control for
+10 dBm or as required.
(e)DOT MKR control to MAX CW (dot
to left edge of graticule lines.
(7) Repeat technique of (3) above and verify
8th marker coincident within +.4 division.
(8) Position FREQUENCY SPAN/DIV control
to 100 kHz and position oscilloscope calibrator controls
for an output of 10 ps. Set FREQUENCY SPAN/DIV
control to 100 kHz.
(9) Repeat technique of (3) above and verify
8th marker coincident within ±.5 division.
(10) Repeat technique of (3) above for settings
in table 7.
NOTE
Adjust RESOLUTION (Hz) control as
required to optimize signal
amplitude.
b. Adjustments
(1) Connect oscilloscope calibrator (A8)
marker output to vertical amplifier (A10) input (CH1),
using cable (B7).
(2) Position controls as listed in (a) through
(e) below: (a)TIME DIV control to 10 ms.
(b)FREQUENCY SPAN/DIV control 0.
(c)DIGITAL STORAGE pushbuttons
OFF. (d)TRIGGERING SOURCE INT
pushbutton ON.(e)MODE NORM pushbutton ON.
(3) Adjust oscilloscope calibrator controls for
10 ms markers output.
(4) Adjust triggering LEVEL control for a
triggered display on crt.
(5) Adjust HORIZ POSITION control and
R685 (fig. 6) to position one marker per division.
(6) Repeat technique of (4) and (5) above for
remaining TIME/DIV settings while adjusting
oscilloscope calibrator controls for appropriate time
marker output. Observe crt time markers will be within
±≥5 % of TIME/DIV control setting.
18