
Operating instructions— Type 549
Line Voltage Selection
115 or 230 volt line voltage is selected by means of an
internal switch SW602. Three voltage ranges, LOW 10 /
208, MED 115/230 and HIGH 127/25 are selected by the
LINE RANGE switch SW602 on the back panel. The instru
ment is normally shipped with these switches set for 115 V.
Line frequency is 50 to 60 Hz.
CAUTION
Before applying power to the instrument, check
that the 115V /2 3 0 V internal selector switch
(located at the rear of the vertical amplifier
chassis) is positioned correctly for the power
source voltage.
FUNCTION OF CONTROLS
AND CONNECTORS
NOTE
The Time Base A and Time Base B controls serve
identical functions with the exception of the
LENGTH control.
TRIGGERING
LEVEL. Selects the amplitude point on the triggering signal
where sweep triggering occurs.
STABILITY. Adjust the sweep triggering circuits for a
stable display. The STABILITY control can generally be set
to the PRESET position and left there. This PRESET position
with the LEVEL control midrange, provides stable triggering
for most displays.
MODE. AUTO position permits normal triggering on
signals with repetition rates higher than about 50 Hz. With
no trigger signal, or with a lower repetition rate, the time
base circuit free runs to provide a reference trace. TRIG
(triggered) position permits normal triggering on all trigger
ing signals. No trace occurs when the triggering signal is
removed.
SLOPE. Determines whether the time base is triggered on
the negative (—) or positive (+) going slope of the signal.
COUPLING. AC position blocks the DC component of the
triggering signal and allows triggering to occur only on the
changing portion of the signal. With frequencies below
about 30 Hz, use the DC position. AC LF REJ position attenu
ates trigger signal frequencies below about 17 kHz, allowing
the trigger circuits to respond only to higher frequencies.
DC position permits triggering on both high and low (to DC)
frequency signals.
SOURCE. INT NORM position uses a portion of the signal
applied to the vertical deflection plates of the CRT as a
trigger signal. INT PLUG IN position applies to multitrace
plug-in units that provide a single-channel trigger signal
through pin 5 of the interconnecting plug (e.g. Tektronix
Type 1A1 plug-in unit). LINE position uses a line-frequency
signal as a trigger. EXT position permits external triggering
on a signal applied to the TRIGGER INPUT connector.
Time Base Controls
TIME/CM. Selects the time-base sweep rate.
VARIABLE (Time Base A). Provides an uncalibrated sweep
rate adjustment which will slow the sweep rate to at least
2.5 times the setting of the TIME/CM switch. An UNCALI
BRATED lamp lights when the VARIABLE control is not in the
CAL position.
LENGTH (Time Base B). Controls the length of the B sweep.
HORIZONTAL DISPLAY. A.- Selects only Time Base A for
display on the CRT screen. B: Selects only Time Base B for
display on the CRT. B INTEN BY A: One of the delayed-
sweep functions. In this position a portion of Time Base B
is intensified during the time that Time Base A (the delayed
sweep) is in operation. A DLY'D: Another delayed-sweep
function. In this position, Time Base A is delayed at the
end of each delay period as determined by the B, TIME/CM
AND DELAY TIME switch setting and the DELAY-TIME M ULTI
PLIER 1:10 dial reading. EXT I X and 10X : Permits an
external signal to be applied to the horizontal deflection
circuit. Sensitivity is continuously variable with the VARIABLE
1-10 control.
READY Lamp. Lights when time-base circuit is ready for
triggering after being reset.
MAGNIFIER. Horizontally expands the portion of the sweep
in the center of the graticule by 5 times at any setting of
the TIME/CM switch.
NORM-SINGLE SWEEP (Sweep mode selector). Three-
position switch selects NORMAL or SINGLE SWEEP operation
for either time base. Switch is spring loaded at the RESET
position. In the SINGLE SWEEP position, selected time base
will sweep once upon receipt of trigger signal, then remain
inactive until the circuit has been reset with the RESET switch
or ERASE AND RESET button. Single sweep permits photo
graphing or storage of nonrepetitive waveforms.
DELAY-TIME MULTIPLIER 1-10. Works in conjunction with
the Time Base B TIME/CM AND DELAY TIME switch. Varies
sweep delay from 0 to 10 times the rate indicated by the
Time Base B TIME/CM AND DELAY TIM E switch.
HORIZONTAL POSITION and FINE. Positions the display
along the horizontal axis of the crt screen.
LOCATE button. When the Type 5 9 is operating as a
conventional oscilloscope, the LOCATE button provides a
trace finding mode. The gain of both the vertical and hori
zontal amplifiers is reduced so that regardless of the input
signal, an on-screen display is presented. With either of the
STORE buttons depressed the LOCATE button does not
affect the horizontal or vertical gain, but the Horizontal
Amplifier is DC offset and moves the beam to the left of the
CRT screen. This places the beam in the locate zone of the
target (a non-store area) and permits vertical positioning
of the beam.
In either mode, the intensity is brought to a preset level
when the LOCATE button is depressed.
Other Controls
AMPLITUDE CALIBRATOR. Determines the peak-to-peak
voltage available at the CAL OUT connector.
5 mA Current Strap. Provides a calibrated source of
square-wave current. The arrow indicates conventional cur
rent (i.e. positive to negative).
2-2 ®T