Tesla TSR3DNM User manual

Manual also available on www.tesla.cz
EN
TERA Radon Program
TSR3DNM USB Radon Probe with Display and Big Memory
Technical Specifications & Operation Manual
v.1 –2017
Table of Contents
1 Introduction.............................................................................................................................................................2
2 Description and Utilization ......................................................................................................................................3
3 Scope of Delivery.....................................................................................................................................................4
4 Product Specification...............................................................................................................................................5
5 Operating Instructions.............................................................................................................................................6
6 Display Operation....................................................................................................................................................6
7 Basic Maintenance ..................................................................................................................................................8
8 EC Declaration of Conformity ..................................................................................................................................9
9 Repairs.....................................................................................................................................................................9
10 Warranty..................................................................................................................................................................9
11 Accessories ............................................................................................................................................................10
12 Alternatives ...........................................................................................................................................................10
13 Revision History.....................................................................................................................................................10

2/10
COPYRIGHT NOTICE
No part of this document may be reproduced, republished, translated or digitalized in any form or by
any means, without prior written permission of TESLA.
Information contained in this manual relates exclusively to the TERA system component specified on the
title page. New versions and modifications may be developed without prior notice to current users. TESLA
has made every attempt to provide you with complete, error-free and accurate information in this manual.
TESLA is not liable for errors or omissions contained in this document, or for any damages however resulted
from using or relying on any information contained herein. TESLA's liability for errors shall be strictly limited
to correcting such errors and providing advisory services as described below.
Users should be familiar with operation basis of used product. If you experience any problems with your
product, please contact us at:
TESLA
Podebradska 56/186
180 66 Prague 9
www.tesla.cz
1Introduction
This document describes technical specifications and user operation of the TSR3DNM USB Radon Probe
with Display and Big Memory.
Product was developed and manufactured in the Czech Republic. All rights reserved TESLA. Offer or
delivery of products or services related to the product does not include transfer of ownership rights.
Before using the product, please read this manual carefully and understand all operating and safety
precautions. Compliance with operational and safety precaution can prevent from damage to equipment or
injuries to personnel. Operating and safety instructions in the document are marked as follows:
Attention! This formatted text indicates the operating and safety instructions.
The product may only be used in the specified manner and for its intended purpose. The product may
be provided to third persons along with this documentation only.

3/10
2Description and Utilization
TSR3D is designed for continuous measuring of radon concentrations in buildings and results are also
shown on display.
Portable probe basis is a measuring chamber with a semiconductor detector. Radon enters the chamber
by diffusion through the input filter on the bottom of probe. The probe measures in autonomous and time
continuous way. It processes results every 4-minute intervals and from this counts short-term moving
average of radon concentration ( 1 hour moving average - average of 15 4-minute process intervals). It also
counts long-term moving average of radon concentration ( 24 hours moving average). The probe saves time
records of these radon concentration values including values of humidity and temperature within its
internal memory (typically at an interval of 1 hour). Next saved value is time record of measuring energy
spectrum (typically at an interval of 12 hours). All results are shown on display as well see paragraph
´Display Operation´.
The probe is random for location in measured place, but generally it is put on the bottom of the probe.
Bottom of the probe cannot be covered. The probe can be switched on/off by switch. LEDs „STAT“ a
„CHRG“ indicate current status of probe see ´Operation manual´below.
The resulting values can be downloaded continuously during measurement or at once at the end of
measurement. Data from the radon probe is downloaded to a PC directly via USB interface. Capacity of
internal memory is almost unlimited.

4/10
TSR4M Radon Probe can be operated by these ways:
A) Standalone probe - Thanks to its independent battery power, portable radon measuring probe
supports flexible placing options within monitored structures. Accumulator will last for more than
1 year after full charging. After switching on TSR4M immediately starts measuring and saving
results into internal memory. The resulting values are downloaded after end of the measurement
by B) way.
B) Probe connected via USB –Using usbRADONview app and USB cable is possible to download
results to PC continuously during measurement or at once at the end of measurement.
usbRADONview application, drivers and user manual is free downloaded on website:
http://www.tesla.cz/en/ke-stazeni/.
3Scope of Delivery
TSR3DNM USB Radon Probe with display and big memory
Power adapter 230VAC/5VDC
USB cable A-B
Probe holder
Operation Manual

5/10
4Product Specification
Product
TSR3D Wireless and USB Radon Probe with display
Type symbol
042 127 211 000
Average measurement sensitivity
0,25 count/hour/Bq.m-3
(method RaA+RaC; 15°C ÷ 30°C; rel. hum. 20% ÷ 40%)
Measuring range
MDA –100 000 Bq/m3;
MDA = 100 Bq/m3per 1 hour or 20 Bq/m3per 24 hours
Measurement uncertainty
< 13% at 300 Bq/m3per 1 hour;
< 3% at 300 Bq/m3per 24 hour
Measuring chamber capacity
0,176 dm3
Response rate
< 30 minutes (RaA); < 3 hours (RaA + RaC)
Radon records
calculated from RaA (quicker, less sensitive)
calculated form RaA + RaC (slower, more sensitive)
Measuring relative humidity range
10 –90 %
Measuring temperature range
-20 to + 60 °C
Records saving interval
1- 255 minutes, default 1 hour
Results memory capacity
> 1000 years
Powering
internal rechargeble accumulator; charging via USB
Accu life after full charging
>1 year
Built-in display
graphical, 128 x 64 pixels, orange
Current radon concentration results
short-term (0,5 hour running average from RaA)
long-term (24 hours running average from RaA + RaC)
Dimension
Ø 80 x 175 mm

6/10
5Operating Instructions
Switching on and off:
The probe measures radon concentration autonomously and communicates in wireless network only if
the switch is in position „I“( switch on). The switching on is signalized by LED diode „STAT“ according chart
below.
If the switch is in position „0“ (switch off) the probe doesn´t measure radon concentration and doesn´t
communicate in wireless network. In switching off mode the probe only keeps running real time for correct
date and time of records in case of switch it on again. By switching off the probe doesn´t lose previous
records of measurement. The switching on is signalized by LED diode „STAT“ according chart below.
Download data from probe over USB is possible only if switch is in position „I“( switch on).
LED diode „STAT“:
It signalizes status radon probe according to following chart:
Color
Description
Green blink 3x
Radon probe has just been turned on.
Green blink after 5s
Radon probe measures and works correctly
Yellow blink 3x
Radon probe has just been turned off.
Green / Yellow
blink after 5s
Radon probe measures but troubles are occur. –especially low capacity of
accumulator. Warnings and errors are shown in PC application.
No light, No blinking
Radon probe doesn´t measure or accumulator is empty or device is damaged.
Charging process of accumulator is described in chapter „Basic Maintenance/
Accumulator charging“
Power supply:
According to operation method the radon probe can be supplied:
1) By internal accumulator for portable use –Radon probe includes internal accumulator which is
able to ensure autonomous operation of probe for more than 12 months without charging.
Depends on climatic condition of probe use. Accumulator is charged with USB port and provided
USB cable. The USB cable is possible to connect to PC or to delivered power adapter. Status of
accumulator and charging process are described in paragraph ´Basic Maintenance/Accumulator
charging´
2) By mains power supply 230V/50Hz for stationary use –Radon probe is permanently supplied by
delivered power adapter. Power adapter is connected to probe via provided USB cable. In case of
blackout internal accumulator ensures UPS function.
Configuration:
Setting and configuration are realized by usbRADONview application. UsbRADONview application,
drivers and user manual with detail configuration description are free downloaded on website:
http://www.tesla.cz/en/ke-stazeni/.
6Display Operation
Built-in display can continuous show measured result and basic parameters of probe. By reason of
power saving the display isnt switched on permanently but it is switched on by pushing of button below
display. By next pushing of button is changed screen on another in cyclic order. Approximately after 1
minute without pushing of the button the display is automatically switched off.

7/10
After first pushing of button from switched off display status is shown intro screen. There is shown basic
information about probe as is version of firmware, serial number and status measurement running ON/OFF.
Go on basic data screen is possible by three ways:
-by next pushing of button on intro screen
-automatically after 5 seconds from intro screen
-directly from switched off display by pushing of button for longer than 1 second
Changing of three data screens is cyclic without intro screen and is possible to make by pushing of
button. Data screens are on following pictures.
Basic data screen firstly shows 1 hour moving average of radon concentration in Bq/m3. Time behaviour
of radon concentration is graphical shows in array of 128 x 24 pixels. Every columm means single data

8/10
record. So for example if data record time interval is 30 minutes the graph shows time behaviour of radon
concentration for last 64 hours.
Second graph shows current energy spectrum in array of 96 x 24 pixels and its parameters. Energy
spectrum is divide into 96 interval and radon concentration is calculated from intervals d1-d2 (RaA) or from
d1-d3 (RaA+RaC). Discriminations d1, d2, d3 are shows below graphs as well.
Numbers of impulses in discrimination intervals are counted during spectra record time interval (Trec)
or till then if one of particular intervals reaches 255 impulses.
Parameters on third data screen are especially service data but record time intervals are useful.
Notice: By display and button isnt possible to setup of probe. For setting of probe parameters is
neccessary to used the PC software application usbRADONview.
7Basic Maintenance
Accumulator charging:
During portable use of radon probe is essential to monitor state of internal accumulator and recharge it if
necessary. If accumulator is discharged the probe automatically turns off. The probe is switched on again
powering USB port.
Current state of accumulator can be determined in three ways:
1) By LED diode ´STAT´ - If LED starts blinking in green-yellow color it indicates that system is working
incorrectly and one of main case is low voltage of accumulator. (see paragraph “Operation Manual /
LED diode “STAT””)
2) On wireless Central Unit display –Symbol "#" on left edge of display second line means that
accumulator voltage is low.
3) In usbRADONview application - where you can check current accumulator voltage. Voltage should
not fall below 3.5 V, in limit conditions falls below 3.3V.
Accumulator is charged via USB port using supplied USB cable. USB cable can be connected to PC or to
supplied power adapter. Connect USB cable with power to USB port of probe. LED diode ´CHRG´ next to
USB port of probe indicates charging status according to following chart:

9/10
LED diode ´CHRG´
Color
Description
Green
Accumulator is fully charged
Yellow
Accumulator is being charged
Green - Yellow alternate blinking
Accumulator is damaged, contact Service Center
No light, No blinking
It is not connected to an external power supply or device is damaged.
Accumulator is fully charged when LED diode ´CHRG´ is green. You can disconnect USB cable.
8EC Declaration of Conformity
EC Declaration of Conformity will be delivered by Tesla producer on request. If interested, please use
contacts on the web www.tesla.cz .
9Repairs
Any repairs and non basic maintenance must be performed exclusively by TESLA manufacturer.
TESLA
Podebradska 56/186
180 66 Prague 9
www.tesla.cz
10 Warranty
-This product is covered by warranty of 24 months from purchase date.
-In case of warranty claim, please contact our Service Department.
-Warranty covers any defects in materials or workmanship and excludes any damage resulting from
or caused by transport or handling or by any misuse.
-Warranty ceases if product has been used improperly or its seal is broken.
-In case of warranty claim, warranty period is prolonged by number of days product was
undergoing warranty repairs.
-After the end of its life, product must be handled as e-waste.

10/10
11 Accessories
Radon Probe accessories are available at producer www.tesla.cz or at distributor.
Probe holder (is included in package)
12 Alternatives
TSR3DNM probe is easy and quick to change in manufacturer to new type of these probes:
TSR3D –WLAN Wireless and USB Radon Probe with Display
TSR3DS- SIGFOX Wireless and USB Radon Probe with Display
13 Revision History
Revision
Date
Comments
Rev.1:
31.1.2017
Initial release
Table of contents