ATTENTION
Beforeuse,readcarefullytheusermanual,payingattentiontothe
warningsandcontraindicationslistedinit.
UseonlyTesmedelectrodesastheyareconstructedwithmaterialsabletoguarantee
maximumconductivityofthepulsesofourelectrostimulators.
Thelargerelectrodesarepositionedinthesamewayasthesmallerones.
Ifthecablesareabithardtoinsertintothedevice,youcantrytograduallyforcethe
insertion.
Incaseoflackofstimulationorothermalfunctions,checktheconditionofthe
electrodesandtheiradhesiontotheskin,thecorrectconnectionofallthepartsand
possiblythebatterycharge.
Youcancustomizetheuseofyourstimulator:
1.usinglesschannels;
2.usingTesmedelectrodesofdifferentsizes;
3.usingyourfavoriteprogramintheareayouwishtotreat,regardlessoftheindication
inthemanual.Forexample,theabdomentoningtreatmentcanwellbeperformedon
anotherarea,suchaslegs,pectorals,gluteus,etc…
Youcanuseyourfavoriteprogramonmultiplezones.Youcannotsettwodifferent
programsatthesametime.
REMEMBER
Programs21to99arepresetwithdifferentstimulationsteps.
Eachphasealternatesmomentsofactivitywithmomentsofrest.Itisnotpossibletoincrease
theintensityduringthemomentofrest.So,donotincreasetheintensityintherestphases
duringprogramexecution,butonlyduringactivephases.
Ifyoukeepalowintensity,forexampleatlevel1orlevel2,atthephasetransitionduring
execution,theprogrammaybereset.Inthiscase,restartwiththeprogramandkeepthe
intensityabithigher.
Youmayhavetheperceptionthatontheonesidetheimpulsesarelessintense.Thestimulator
emitsthesameimpulsesonallchannels,soprobablyyouhavenotwellidentifiedthemotor
pointsofthemuscle,orononesidethereismorefattissue,ortheelectrodesarewornout.Try
movingtheelectrodes,makingsuretheyarewelladherenttotheskin.
Thelevelofintensitytoachievethedesiredmusclecontractioncanvarydependingonmultiplefactors,
includingtheplacementoftheelectrodesandthepresenceofadiposetissue.
Therefore,thesamecurrentintensitycanprovidedifferentsensationsfrompersontoperson,
fromdaytoday,fromtheleftsidetotherightside.
Ifyouneedtechnicalassistancewriteto[email protected],orcall00390305050163
fromMondaytoFridayfrom09.00to16.30,excludingholidays.
Wesolve100%ofrequestsquicklybecauseourpriorityiscustomersatisfaction.
Beforeuse,readcarefullytheusermanual,payingattentiontothewarningsand
contraindicationslistedinit.
QUICKUSERGUIDEANDUSEFULTIPS
Batteryalreadyinserted,devicereadytouse.Ifthebatterysymbolflashesafterswitchingon,charge.
1.Connectthe4
bipolarcables
2.Inserttheelectrodesinthecablesand
placethemonthetreatmentarea
Youmayrefertotheelectrodes
placementphotosinthemanual
EQUIPMENTDESCRIPTION
LCDdisplayandkeys
1Tesmedlogo
2Programnumber
3Programsymbol
4Lowbatterysymbol
5Frequencysymbol
6Selectedoutputchannels(n.4)
7Relativeintensityvalueforeach
4output
8Functionskey(MODE)
9Decreasekey()
10Increasekey()
11ON/OFFandselectchannels
doublefunctionkey
12Intensitysymbol
13KeyboardblockandONblock
symbol
14Timersymbol
15Treatmenttimeandremaining
time
16Clocksymbol
17Outputchannels
18Batteryrechargersocket
19Batterydoor
20Removablebelthook

Thebatteryisalreadyinstalledinthedevice.
IftheflashingbatterysymbolappearsonthedisplayortheTesmeddoesnot
lightup,chargethedevice.Whenthesymbolstopsflashingordisappears,the
batteryischarged.IfthebatteryischargedandtheTesmedisconnectedto
thebatterycharger,theTesmedscreenwillnotlightupandnosignalwill
appearbecausethebatterychargerdetectsthatthereisnoneedforcharging
andthereforedoesnotactivate.
41
Setinitiallyalowlevelandthengraduallyincrease.Theintensitylevelmustbe
suchthatthetreatmentispleasant.
Symbolappearsonthedisplay(fig.F)andyoucannowadjusttheintensityusing
keys(from1to99).Theintensityrisesorfallsonalltheselectedchannelsat
thesametime.
Whenyousettheintensity,theclockbeginstoflashindicatingthattheequipmentis
providingenergy(fig.G).
Attention:Programs21to99arepresetwithdifferentstimulationsteps.
Eachphasealternatesmomentsofactivitywithmomentsofrest.Itisnotpossi
bletoincreasetheintensityduringthemomentofrest.Sodonotincreasethe
intensityintherestphasesduringprogramexecution,butonlyduringactive
phases.
Ifyoukeepalowintensity,forexampleatlevel1orlevel2,atthephasetransition
duringexecution,theprogrammaybereset.Inthiscase,restartwiththeprogram
andkeeptheintensityabithigher.
Youcan,however,accesseachchannelduringusetocontroltheintensityindependently:
simplypresskeytoselectthechannelwhoseintensityyouwanttochangeandthen
usekeystoincreaseordecreasethevalue.Fig.Hshowsthattheintensity
valuesforthe4channelscanbedifferent.Attheendofthetreatment,theTesmedclears
theintensitysettingforalloutputchannels;Youcannowswitchtheequipmentoffby
pressingkeyfor2secondsorsetnewparameterstobeginanewtreatment.Use
keystoincreaseordecreasethevalueofanyparameter,eachtimethesekeys
arepressed,theparameterisincreasedordecreasedinstepsofoneunit,keepthekeys
pressedforfasterincrease/decreaseoftheparameter.
3.Usemode
Switchingon
Pressthekeyforapproximately2seconds:thesystemrunsaselftestprocedure,during
whichthedisplayshowsalltheavailablefunctions(fig.A);
thenappearsonthedisplaywhatshowninfig.B.
"Displaylighting"function
Thedisplayilluminationturnsoffautomaticallyafterabout20secondsintheabsenceof
interactions.
Thedisplayislitagainwhenyoupressanykey.
"Selfswitchingoff"function
TheTesmedautomaticallyswitchesoffisnotusedformorethan5minutesinordertosave
thebattery.
Afterswitchingonthedevice,pressinsuccessionkeytoselectthenumberof
channelsyouwanttousetosuitthetreatmentandthenkeytoselecttheprogram
(fig.C).
Attention:TheprogramsintheTesmedaremodulatestoprovidevariousstages(different
frequenciesofwaves)ofmassagerangingfrom1toamaximumof5stages.
SomeprogramshaveoneormoreWAIMSSYSTEM®stages.
Duringtreatment(andespeciallywiththeWAIMSSYSTEM®)wealwaysrecommendthat
youadapttheintensityofeachchannel,asyoursensitivitywillvaryvariousfromonegroupof
musclestoanother.
Programs
Onlyinprogramsfrom01to20youcanchangetimeandfrequency
Inprogramsfrom01to20,onceselectedtheprogram,it'spossibletocustomizethe
treatmenttimeandfrequency.
Presskeyandtherelevantsymbolstartsflashingonthedisplay(fig.D).Use
keystoadjustthetreatmenttime:frommin1minutetomax60minutes.
Presskeyagaintoswitchtotheintensityadjustment,whichisadjustedwiththe2
keys.
Youcanadjustthefrequencyonlyaftergivingintensityandstartedtreatment.Presskey
for3secondstoenterthefrequencycontrolmode.Presskeysrepeatedlyto
increaseortodecreasethefrequency.Thesymbolflashes(atavaryingspeed)to
indicatefrequencycontrolmode(fig.E).Presskeyoncetoquitfrequencycontrolmode.
Thefrequencyisthenumberofpulsespersecond.Whenyouchangethefrequency,youfeel
anincreaseordecreaseinthespeedofstimulation.
Attention:ifyouareinfrequencycontrolmodeduringatreatmentandyouwanttoclearthe
intensity,presskeytwice.
Programs21to99:presetwithdifferentstimulationsteps
Afterhavingselectedaprogramfrom21to99,presskeytogostraighttotheintensity
adjustment,whichisadjustedwiththe2keys.
Fortheprogramschoice,refertothetabsinthemanualwiththeindicationof
treatments,program,electrodesphotos,duration,phasesanddays.
Errorwarning(electrodesdisconnected)
Inthecaseofadetachedelectrode,thedevicegoesintoerror,displayingtheword"Er",
flashingforabout5",inthechannelwheretheproblemoccurred(fig.I).
Allpulsesarestoppedimmediatelyand"00"appearsastheinsensityvalueonthe
display.Theerrormessagecanmeanthat:
•Therearenoelectrodesconnectedtothechannelinerrormode.
•Theelectrodesarenotproperlyadheredtotheskin:repositiontheelectrodesor
replacethemiftheyaretooold.
•Thecableusedtoconnecttheelectrodesisfaulty:checkbyinvertingtwocables.
Thesystemautomaticallyresetsitselfoncetheintensityvalueincreases.
"Keyboardblock"function
Keepingthekeypressed,presskeyandreleaseit.Symbolwillappearon
thedisplay(fig.L).Followthesameproceduretounlockthekeyboard.
"ONblock"function
Usefultoavoidarandomswitchon.
Withthedeviceon,keepingtheandkeyspressedsimultaneouslyfor3seconds,
thesymbolandtheTesmedlogowillappearonthedisplay(Fig.M).Followthe
sameproceduretoswitchtheequipmenton.
23
Intensitycontrol:levels
Treatmenttimeadjustment
Frequencyadjustement

Other manuals for MAX 830

2

This manual suits for next models

2

Other Tesmed Medical Equipment manuals

Tesmed MAX 830 User manual

Tesmed

Tesmed MAX 830 User manual

Tesmed MAX 830 User manual

Tesmed

Tesmed MAX 830 User manual

Popular Medical Equipment manuals by other brands

Optopol REVO nx user manual

Optopol

Optopol REVO nx user manual

BIOTRONIK Cardiac Airbag Technical manual

BIOTRONIK

BIOTRONIK Cardiac Airbag Technical manual

Getinge Arjohuntleigh AtmosAir 300 owner's guide

Getinge

Getinge Arjohuntleigh AtmosAir 300 owner's guide

hinor POCA06-5L manual

hinor

hinor POCA06-5L manual

Mindray M9 Service manual

Mindray

Mindray M9 Service manual

Dräger D-Vapor Instructions for use

Dräger

Dräger D-Vapor Instructions for use

ResMed AirFit P10 Fittings guide

ResMed

ResMed AirFit P10 Fittings guide

Accutome a-scan plus user guide

Accutome

Accutome a-scan plus user guide

iHealth PO3M Operation manual

iHealth

iHealth PO3M Operation manual

VITACON C5 Assembly instructions

VITACON

VITACON C5 Assembly instructions

Aesculap ELAN 4 Instructions for use/Technical description

Aesculap

Aesculap ELAN 4 Instructions for use/Technical description

Henry Schein 900-8275 Instructions for use

Henry Schein

Henry Schein 900-8275 Instructions for use

Activ8rlives PUFFClicker manual

Activ8rlives

Activ8rlives PUFFClicker manual

GEIGER TCU 150 Operation manual

GEIGER

GEIGER TCU 150 Operation manual

3RAY Electronics ECG-2301G Operation manual

3RAY Electronics

3RAY Electronics ECG-2301G Operation manual

Breas Vivo 45 LS Quick reference guide

Breas

Breas Vivo 45 LS Quick reference guide

orfit HP PRO Instructions for use

orfit

orfit HP PRO Instructions for use

Rossmax SB200 Instructions for use

Rossmax

Rossmax SB200 Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.